K150960 · Pensar Medical, LLC · OMP · Dec 1, 2016 · General, Plastic Surgery
Device Facts
Record ID
K150960
Device Name
WoundPro Negative Pressure Wound Therapy System
Applicant
Pensar Medical, LLC
Product Code
OMP · General, Plastic Surgery
Decision Date
Dec 1, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The WoundPro Negative Pressure Wound Therapy System may promote wound healing, through the drainage and removal of infectious material and other fluids from the wound site, with continuous and/or, intermittent negative pressure. Patients with chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic or pressure), flaps and grafts may benefit from the use of this device. The WoundPro is intended for use in a healthcare facility only.
Device Story
WoundPro is a powered suction pump providing continuous or intermittent negative pressure (20-200mmHg) to wound sites to promote healing via drainage of infectious material and fluids. System includes control unit with brushless DC motor, rechargeable battery or AC power, and electronic display. Operates with 300cc or 800cc single-use collection canisters containing integrated hydrophobic/bacterial filters. Dressing assembly consists of gauze/foam, polyurethane drape, and flange/tubing. Used in healthcare facilities; operated by clinicians. Device monitored by software with audio/visual alarms for pressure regulation. Mounts to IV poles, bedside rails, or countertops. Benefits patients by facilitating wound closure and fluid removal.
Clinical Evidence
Bench testing only. Evidence includes electrical safety and electromagnetic compatibility (EMC) testing, software verification and validation, performance verification testing, and biocompatibility assessment per ISO 10993-1.
Technological Characteristics
Powered suction pump; brushless DC motor; 20-200mmHg vacuum range; 300cc/800cc collection canisters with integrated hydrophobic/bacterial filters; injection-molded plastic housing; rechargeable Li-ion battery or AC power (90-264VAC); software-controlled; dimensions 6.75 x 8.5 x 3.5 in; weight 3.8 lbs. Biocompatibility per ISO 10993-1. Gamma sterilization for foam kits.
Indications for Use
Indicated for patients with chronic, acute, traumatic, subacute, and dehisced wounds, partial-thickness burns, diabetic/neuropathic ulcers, pressure ulcers, flaps, and grafts. Intended for use in healthcare facilities only.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, with flowing lines suggesting hair or movement. The profiles are stacked slightly overlapping each other. Encircling the image is the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 1, 2016
Pensar Medical, LLC % Elaine Duncan, Ms.ME, RAC President Paladin Medical, Inc. PO Box 560 Stillwater, Minnesota 55082
Re: K150960
Trade/Device Name: WoundPro Negative Pressure Wound Therapy System Regulation Number: 21 CFR 878.4780 Regulation Name: Powered suction pump Regulatory Class: Class II Product Code: OMP Dated: October 28, 2016 Received: October 31, 2016
Dear Ms. Duncan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## David Krause -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
#### 510(k) Number (if known) K150960
#### Device Name
PENSAR WoundPro Negative Pressure Wound Therapy System
#### Indications for Use (Describe)
The WoundPro Negative Pressure Wound Therapy System may promote wound healing, through the drainage and removal of infectious material and other fluids from the wound site, with continuous and/or, intermittent negative pressure. Patients with chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic or pressure), flaps and grafts may benefit from the use of this device. The WoundPro is intended for use in a healthcare facility only.
Additional Wounds Indicated are:
- · Diabetic/Neuropathic ulcers
- · Pressure ulcers
- · Chronic wounds
- · Acute wounds
- · Dehisced wounds
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
# 510(K) SUMMARY
#### SUBMITTER:
Submitted on behalf of:
| Company Name:<br>Address:<br>Telephone:<br>Fax: | Pensar Medical LLC<br>42225 Remington Rd, Unit A3<br>Temecula, CA 92590<br>(800) 669-4757<br>NA |
|-----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| by:<br>Telephone:<br>Fax: | Elaine Duncan, M.S.M.E., RAC<br>President, Paladin Medical, Inc.<br>PO Box 560<br>Stillwater, MN 55082<br>715-549-6035<br>715-549-5380 |
| CONTACT PERSON: | Elaine Duncan |
| DATE PREPARED: | December 1, 2016 |
| TRADE NAME:<br>COMMON NAME:<br>CLASSIFICATION NAME: | Pensar WoundPro Negative Pressure Wound Therapy<br>System<br>Negative Pressure Wound Therapy System<br>Negative Pressure Wound Therapy Suction Pump (and<br>components) |
| PRO CODE: | OMP (Class II, 21 CFR 878.4780) |
#### SUBSTANTIALLY EQUIVALENT TO:
PREDICATE: WoundPro Apex Negative Pressure Wound Therapy Suction Pump, Accuro Medical Products, LLC -K100823 and also features of Medela INVIA Liberty K080357
#### DESCRIPTION of the DEVICE:
The WoundPro is part of a family of products offered by Pensar Medical LLC. The WoundPro is a powered suction pump that uses controlled negative pressure to a dressing and flange assembly through tubing.
The control unit incorporates a maintenance free brushless DC motor and is powered by a rechargeable battery power source. Optionally the unit can be connected to mains power using the included Medical Grade Switching power converter. The device is monitored by software and contains an alarm function with audio and visual prompts.
The WoundPro has a negative pressure setting range of 20mm to 200mmHg which is electronically monitored and controlled. The pump includes push button user interface controls and an electronic display with audible and visual alarm indicators.
The device can accommodate 300cc and 800cc canister sizes.
The unit can be operated on a countertop and has provisions for connection to IV poles, bedside rails, or footboard mounting.
{4}------------------------------------------------
#### INDICATIONS FOR USE:
The WoundPro Negative Pressure Wound Therapy System may promote wound healing, through the drainage and removal of infectious material and other fluids from the wound site, with continuous and/or, intermittent negative pressure. Patients with chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic or pressure), flaps, grafts may benefit from the use of this device. The WoundPro is intended for use in a healthcare facility only.
Additional Wounds Indicated are:
- Diabetic/Neuropathic ulcers ●
- Pressure ulcers ●
- . Chronic wounds
- Acute wounds
- Dehisced wounds
#### SUMMARY OF THE TECHNICAL CHARACTERISTICS OF THE DEVICE COMPARED TO PREDICATE
The primary predicate is the Accuro APEX WoundPro K100823 with the same indication for use and intended use.
| | PREDICATE | SUBJECT | SIMILARITY or DIFFERENCE |
|--------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------|
| | Accuro APEX WoundPro<br>K100823 | Pensar Medical<br>WoundPro | |
| Manufacturer | Accuro<br>(acquired by Pensar) | Pensar Medical | Pensar acquired stock and<br>refurbished units |
| Product Type | Negative Pressure Wound<br>Therapy System | Negative Pressure Wound Therapy<br>System | SAME |
| Product Code | OMP (Class 2, 21 CFR 878.4780) | OMP (Class 2, 21 CFR 878.4780) | SAME |
| Indication For Use | The WoundPro Negative<br>Pressure Wound Therapy<br>System may promote wound<br>healing, through the drainage<br>and removal of infectious<br>material and other fluids from<br>the wound site. The WoundPro<br>can provide continuous,<br>intermittent and variable<br>intermittent negative pressure<br>therapies to achieve this goal.<br>Patients with chronic, acute,<br>traumatic, subacute and<br>dehisced wounds, partial-<br>thickness burns, ulcers (such as<br>diabetic or pressure), flaps and<br>grafts may benefit from the use<br>of this device. At no time should<br>the WoundPro, or any NPWT<br>device, be used without an<br>order from a physician.<br>Types of Wounds Indicated are:<br>· Diabetic/Neuropathic ulcers<br>· Pressure ulcers<br>· Chronic wounds<br>· Acute wounds<br>· Dehisced wounds | The WoundPro Negative Pressure<br>Wound Therapy System may<br>promote wound healing, through<br>the drainage and removal of<br>infectious material and other fluids<br>from the wound site, with<br>continuous and/or, intermittent<br>negative pressure. Patients with<br>chronic, acute, traumatic, subacute<br>and dehisced wounds, partial-<br>thickness burns, ulcers (such as<br>diabetic or pressure), flaps, grafts<br>may benefit from the use of this<br>device. The WoundPro is intended<br>for use in a healthcare facility only.<br>Additional Wounds Indicated are:<br>● Diabetic/Neuropathic<br>ulcers<br>● Pressure ulcers<br>● Chronic wounds<br>● Acute wounds<br>● Dehisced wounds | SAME except for minor<br>editorial difference |
| Therapy Unit | Software controlled pump for<br>delivery of negative pressure<br>wound therapy | Software controlled pump for delivery of<br>negative pressure wound therapy | SAME |
{5}------------------------------------------------
## 510(k) Summary-Continued
| | PREDICATE | SUBJECT | SIMILARITY or DIFFERENCE |
|------------------------------------------|--------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|--------------------------------------------------|
| | Accuro APEX WoundPro<br>K100823 | Pensar Medical<br>WoundPro | |
| Housing | Injection molded plastic | Injection molded plastic | SAME |
| Modes | Continuous or Intermittent Modes | Continuous or Intermittent Modes | SAME |
| Vacuum Range | 20-150 mgHg | 20-200mg Hg | Same as Medela, see chart<br>below |
| Weight (control unit) | 3.8 lbs (1.7kg) | 3.8 lbs (1.7kg) | SAME |
| Dimensions | 6.75 x 8.5 x 3.5 in<br>(17 x 21 x 9 cm) | 6.75 x 8.5 x 3.5 in<br>(17 x 21 x 9 cm) | SAME |
| External Power Supply | Input: 90-264VAC, 0.8-0.4A, 47-63Hz;Output:18VDC 1.67A | Input: 90-264VAC, 0.8-0.4A, 47-63Hz;Output:18VDC 1.67A | SAME |
| Electrical Battery Powered | 18VDC 25 Watts | 18VDC 25 Watts | SAME |
| Power Source | Lithium Ion Battery or AC | Lithium Ion Battery or AC | SAME |
| Expected Useful Life | 3 years | 3 years | SAME |
| Collection Canisters | 300 c and 800c | 300 c and 800c | SAME |
| Filter | Hydrophobic overflow<br>protection/bacterial filter<br>integrated into the single-use<br>collection canister | Hydrophobic overflow<br>protection/bacterial filter integrated<br>into the single-use collection canister | SAME |
| Dressing System | Gauze dressing with<br>polyurethane drape. | Gauze and foam dressing with<br>polyurethane drape. | SAME |
| Gauze Dressing Kit Assembly | Off the shelf components<br>purchased sterilized. | Off the shelf components purchased<br>sterilized. | SAME |
| Drape | Polyurethane with adhesive | Polyurethane with adhesive | SAME |
| Dome/exudates line assembly | Flange and Tubing | Flange and Tubing | SAME |
| Method of Sterilization | (gauze dressing purchased<br>sterile) | gamma sterilization black foam kit | Qualified by vendor;<br>documented in submission |
| Single-Use | Single-use in Tyvek pouch | Single-use in Tyvek pouch | SAME |
| Disposable<br>Packaging for<br>Dressings | | | |
| Biocompatibility | ISO10993-1 | ISO10993-1 | SAME |
| Accessories | EX clamp system for<br>head/footboard or I.V. pole<br>-WoundPro Carrying Case | EX clamp system for head/footboard or<br>I.V. pole<br>-WoundPro Carrying Case | SAME |
The Pensar WOUNDPRO is also compared to the Medela predicate for pressure range only.
| PERFORMANCE | MEDELA INVIA Liberty | WOUNDPRO PENSAR- | DIFFERENCE/ impact of |
|---------------------|-------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|---------------------------|
| PARAMETER | K080357 | subject of this 510(k) | difference |
| weight | <2.2 lbs | 3.8lbs (1.7kg) | Pensar a bit heavier |
| dimensions | 3.74" x 6.69 "x5.91 " | 6 ¾" (w) x 8 ½" (h) x 3 ½" (d) (17 x<br>21 x 9 cm) | Pensar is similar in size |
| Power source | Lithium Ion Battery<br>AC Power | Lithium Ion Battery<br>AC Power | No difference |
| Electrical voltage | 110-240V AC<br>47-63 Hz | 90-260V AC<br>47-63Hz | No difference |
| Electrical amperage | 0.8-4A | .8-.4A | No difference |
| Housing | Injection Molded Plastic | Injection Molded Plastic | No difference |
| Vacuum range | 60-200mmHg | 20-200 mm Hg | Similar upper limit |
| Suction capacity | 5 L/min | 5L/min | No difference |
| Canister volumes | 0.3L/ 0.8L | 0.3L/0.8L | No difference |
| Modes | Continuous or<br>Variable/Intermittent Modes | Continuous or<br>Variable/Intermittent Modes | No difference |
| Filters | Hydrophobic overflow<br>protection/bacteria filter<br>integrated into the single use<br>collection canister | Hydrophobic overflow<br>protection/bacteria filter<br>integrated into the single use<br>collection canister | No difference |
{6}------------------------------------------------
### 510(k) Summary-Continued
This comparison shows no differences in technological characteristics and functions between the Pensar WoundPro Negative Pressure Wound Therapy System and the predicate devices.
#### SUMMARY OF PERFORMANCE TESTING:
The following performance data was provided in support of the substantial equivalence determination.
- . Electrical Safety and Electromagnetic Compatibility (EMC)
- Software Verification and Validation Testing ●
- . Verification and Validation Testing
- . Biocompatibility Assessment
#### SUMMARY OF CONCLUSIONS
Substantial Equivalence for the Pensar WoundPro Negative Pressure Wound Therapy System is based on the same indications, intended use, technological features and functions as the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.