K150951 · Bryan Medical, Inc. · KTI · Jun 25, 2015 · Ear, Nose, Throat
Device Facts
Record ID
K150951
Device Name
aeris Balloon Dialation Catheter
Applicant
Bryan Medical, Inc.
Product Code
KTI · Ear, Nose, Throat
Decision Date
Jun 25, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4680
Device Class
Class 2
Attributes
Therapeutic, Pediatric, 3rd-Party Reviewed
Indications for Use
The aeris Balloon Dilation Catheter is intended for use in adult and pediations to dilate strictures of the airway.
Device Story
Single lumen balloon dilation catheter; high-pressure balloon at distal tip; includes stylet for advancement; luer lock for inflation/stylet placement; radiopaque markers for fluoroscopic placement. Used in clinical settings under endoscopic visualization; physician-operated. Device advanced to stricture site; stylet removed; balloon inflated with sterile water to dilate airway; monitored via endoscopy; deflated via vacuum for removal. Benefits include mechanical dilation of airway strictures to improve patency.
Clinical Evidence
Bench testing only. Testing included balloon working length, inflation/deflation times, fatigue (10 cycles), burst pressure, hub measurement, insertability, removability, luer fitting, torque, and bond strength. Transit and 1-year accelerated aging tests performed. Biocompatibility testing (cytotoxicity, sensitization, systemic toxicity, irritation) confirmed safety.
Technological Characteristics
Materials: Nylon balloon, Pebax catheter. Energy: Hydraulic (sterile water inflation). Dimensions: 5-16mm diameter, 30-40mm length, 55cm catheter length. Connectivity: None. Sterilization: Ethylene Oxide. Standards: ISO 594-1/2 (luer), ISO 10555 (bond strength). Single lumen design.
Indications for Use
Indicated for adult and pediatric patients requiring dilation of airway strictures.
Regulatory Classification
Identification
A bronchoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the bronchoscope and is intended to examine or treat the larynx and tracheobronchial tree. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel or flexible plastic. This generic type of device includes the rigid ventilating bronchoscope, rigid nonventilating bronchoscope, nonrigid bronchoscope, laryngeal-bronchial telescope, flexible foreign body claw, bronchoscope tubing, flexible biopsy forceps, rigid biopsy curette, flexible biopsy brush, rigid biopsy forceps, flexible biopsy curette, and rigid bronchoscope aspirating tube, but excludes the fiberoptic light source and carrier.
Predicate Devices
Acclarent Inspira AIR Balloon Dilation System (K090660)
Boston Scientific CRE Pulmonary Balloon Dilation Catheter (K023337)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three human profiles facing right, with flowing lines extending from their faces. The profiles are arranged in a way that suggests unity and connection. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the image.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 25, 2015
Bryan Medical, Inc. c/o Mr. Mark Job Regulatory Technology Services LLC 1394 25th Street, NW Buffalo, MN 55313
Re: K150951
Trade/Device Name: Aeris Balloon Dilation Catheter Regulation Number: 21 CFR 874.4680 Regulation Name: Bronchoscope (Flexible or Rigid) and Accessories Regulatory Class: Class II Product Code: KTI Dated: June 9, 2015 Received: June 10, 2015
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Mark Job
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Eric A. Mann -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use Statement
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
#### Device Name
aeris Balloon Dilation Catheter
Indications for Use (Describe)
The aeris Balloon Dilation Catheter is intended for use in adult and pediations to dilate strictures of the airway.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14)
Page 1 of 1
PSC Publishing Services (200) 483-6740
{3}------------------------------------------------
## 510(k) Summary
Bryan Medical, aeris Balloon Dilation Catheter
The assigned 510(k) number is: K150951 510(k) Owner: Andrew Georgilis Bryan Medical, Inc. 3906 Oak St. Cincinnati, OH 45227 Phone: (513) 272-1600 Fax: (513) 272-1610 Contact Person: Andy Black Medical Murray, Inc. 400 North Rand Road North Barrington, IL 60010 Phone: (847) 620-7990 Fax: (847) 620-7995 January 30, 2015 Date Prepared: Device Name and Classification Classification Name: Bronchoscope (Flexible or Rigid) and Accessories Common Name: Balloon Dilation Catheter
Name: aeris Balloon Dilation Catheter
Device Classification: Class II
Number: 21 CFR Ref. § 874.4680
Product Code: KTI
Review Panel: Ear, Nose, & Throat
Predicate Device: Acclarent Inspira AIR Balloon Dilation System (K090660)
> Boston Scientific CRE Pulmonary Balloon Dilation Catheter (K023337)
{4}------------------------------------------------
### Device Description
The aeris Balloon Dilation Catheter is comprised of a sinqle lumen catheter with a high pressure balloon near the distal tip. A stylet is provided to facilitate advancement of the balloon dilation catheter to the desired location. The stylet must be removed before inflation of the high pressure balloon. A luer lock at the proximal end is used for placement of the stylet and injecting sterile water into the balloon. Two radiopaque markers, located on the catheter, inside the balloon, can be used to confirm balloon placement under Fluoroscopy.
The aeris Balloon Dilation Catheter includes inflation balloons of diameters 5, 7, 8, 9, 10, 12, 14, and 16 mm with complementing characteristics as shown in Table 1.
| aeris Balloon<br>Dilation<br>Catheter Size | Product<br>Number | Balloon<br>Ø (mm) | Balloon<br>Working<br>Length (mm) | Recommended<br>Use Pressure<br>(ATM) | Catheter<br>OD (mm) | Catheter<br>Length<br>(cm) |
|--------------------------------------------|-------------------|-------------------|-----------------------------------|--------------------------------------|---------------------|----------------------------|
| 5 x 30 mm | KG0530 | 5 | 30 | 17 | 1.55 | 55 |
| 7 x 30 mm | KG0730 | 7 | 30 | 17 | 1.55 | 55 |
| 8 x 30 mm | KG0830 | 8 | 30 | 17 | 1.55 | 55 |
| 9 x 30 mm | KG0930 | 9 | 30 | 17 | 1.55 | 55 |
| 10 x 30 mm | KG1030 | 10 | 30 | 17 | 2.34 | 55 |
| 12 x 40 mm | KG1240 | 12 | 40 | 10 | 2.34 | 55 |
| 14 x 40 mm | KG1440 | 14 | 40 | 10 | 2.34 | 55 |
| 16 x 40 mm | KG1640 | 16 | 40 | 10 | 2.34 | 55 |
Table 1: aeris Balloon Dilation Catheter sizes and corresponding characteristics.
The aeris Balloon Dilation Catheter is provided in sterile packaging and with a protective sheath covering the balloon. After removing the catheter from the packaging and protective sheath, it is wiped down with a gauze pad soaked in sterile water. The catheter should then be gently advanced into the airway, to the site of the stricture under endoscopic visualization. The balloon portion of the device is then centered across the restriction and the stylet is removed while holding the catheter securely in place. The balloon dilation catheter luer is attached to the Inflation Device and the balloon is inflated to desired pressure with sterile water. Monitor the balloon via endoscopy during inflation assessing the diameter, shape, and position of the balloon ensuring that the proximal end of the balloon remains proximal to the stricture.
The balloon can be completely deflated while maintaining an endoscopic view of the balloon using a vacuum. Once visual confirmation of the fully deflated balloon is achieved, the balloon may be removed from the airway.
## Indications for Use
The aeris Balloon Dilation Catheter is intended to dilate strictures of the airway.
{5}------------------------------------------------
# Substantial Equivalence Claim
Based on comparison of device features, materials, intended use, and performance, the Bryan Medical, aeris Balloon Dilation Catheter is substantially equivalent to the commercially available predicate device Acclarent Inspira AIR approved by the FDA under 510k number K090660 and Boston Scientific CRE approved by the FDA under 510k number K023337. Table 2 shows a substantial equivalence comparison summary of the aeris Balloon Dilation Catheter compared to the predicate devices.
| Table 2. Comparison table of aeris Balloon Dilation Catheter characteristics to predicate devices | | |
|----------------------------------------------------------------------------------------------------|--|--|
| | | |
| Attribute | Acclarent Inspira AIR | Boston Scientific CRE | Bryan Medical aeris |
|--------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| A. Intended Use | | | |
| A.1 Indications for<br>Use Statement | The Airway Balloon<br>Catheter is an instrument<br>intended to dilate strictures<br>of the airway tree. | The CRE Pulmonary Balloon<br>Dilatation Catheter is intended<br>to be used endoscopically to<br>dilate strictures of the airway<br>tree. | The aeris Balloon Dilation<br>Catheter is intended to dilate<br>strictures of the airway. |
| A.2 Labeling | Equivalent in content (label<br>included in Section N) | Equivalent in content (label<br>included in Section N) | Equivalent content (label<br>included in Section N) |
| A.3 Target<br>Population | All patients meeting<br>intended use and labeling<br>requirements. | All patients meeting intended use<br>and labeling requirements. | All patients meeting intended<br>use and labeling<br>requirements. |
| A.4 Product code<br>& 21 CFR<br>regulation # | KTI, 874.4680 | KTI, 874.4680 | KTI, 874.4680 |
| A.5<br>Contraindications | • Balloon dilation is<br>contraindicated in any<br>patient whose degree of<br>respiratory failure would<br>not allow the patient to<br>tolerate the manipulation<br>required to accomplish<br>balloon dilation.<br>• Balloon dilation is<br>contraindicated in the<br>presence of:<br>o significant active<br>bleeding from the site<br>of the proposed<br>dilation<br>o and/or presence of a<br>known perforation at<br>the site of proposed<br>dilation<br>o and/or presence of a<br>known fistula between<br>the tracheobronchial<br>tree and esophagus,<br>mediastinum or<br>pleural space. | • Balloon dilation is<br>contraindicated in any patient<br>whose general medical<br>condition and degree of<br>respiratory failure would not<br>allow the patient to tolerate<br>bronchoscopy (rigid or<br>flexible) and/or the<br>manipulation required to<br>accomplish balloon dilation.<br>• Balloon dilation is<br>contraindicated in the<br>presence of:<br>significant active bleeding<br>from the site of the proposed<br>dilatation,<br>and/or presence of a known<br>perforation at the site of<br>proposed dilatation,<br>and/or presence of a known<br>fistula between the<br>tracheobronchial tree and<br>esophagus, mediastinum or<br>pleural space unless the<br>dilatation was being<br>performed in preparation for<br>the placement of a stent to<br>treat the perforation or fistula. | • Balloon dilation of the<br>airway is contraindicated in<br>any patient whose degree<br>of respiratory failure would<br>not allow the patient to<br>tolerate the manipulation<br>required to accomplish<br>balloon dilation.<br>• Balloon dilation is<br>contraindicated in the<br>presence of:<br>o significant active<br>bleeding from the site of<br>the proposed dilation<br>o and/or presence of a<br>known perforation at the<br>site of proposed dilation<br>o and/or presence of a<br>known fistula between<br>the tracheobronchial tree<br>and esophagus,<br>mediastinum or pleural<br>space |
| Table 2 cont. | | | |
| Attribute | Acclarent Inspira AIR | Boston Scientific CRE | Bryan Medical aeris |
| B. Technology | | | |
| B.1 Balloon Shape<br>at < 2 ATM / Max.<br>Inflation Pressure | Straight body and conical<br>ends / straight body and<br>spherical ends | Straight body and conical<br>ends / straight body and<br>spherical ends | Dog-bone body and conical<br>ends / straight body with<br>spherical ends |
| B.2 Used with<br>Stylet | Optional | Yes (Guidewire) | Optional |
| B.3 Radiopaque<br>Markers | No | Yes | Yes |
| B.4 Patient<br>Contacting<br>Materials | Unknown | Unknown | Nylon balloon, Pebax catheter |
| B.5 Balloon<br>Diameters | 5 mm<br>7 mm<br>8.5 mm<br>10 mm<br>12 mm<br>14 mm<br>16 mm | 8 – 9 – 10 mm<br>10 – 11 – 12 mm<br>12 – 13.5 – 15 mm<br>15 – 16.5 – 18 mm<br>18 – 19 – 20 mm | 5 mm<br>7 mm<br>8 mm<br>9 mm<br>10 mm<br>12 mm<br>14 mm<br>16 mm |
| B.6 Balloon<br>Working Lengths<br>(mm) | 24 for diameters 5, 7, & 8.5<br>40 for diameters 10, 12, 14,<br>& 16 | 30 for diameters 8 - 12<br>55 for diameters 12 – 18 per<br>manufacturer's website | 30 for diameters 5 - 10<br>40 for diameters 12 - 16 |
| B.7 Catheter<br>Outside Diameter<br>(mm) | 2.1 | 2.5 | 1.55 for diameters 5 - 9<br>2.34 for diameters 10 - 16 |
| B.8 Catheter<br>Length (cm) | 45 | 75 | 55 |
| B.9 Deflation Time<br>(Seconds) | ≤ 15 & 25 per clearance<br>letter for K110218 | Unknown | ≤ 15 |
| B.10 Maximum<br>Inflation Pressure | 8 – 16 ATM | 6 - 9 ATM | 10 & 17 ATM |
| B.11 Flexible | Yes | Yes | Yes |
| B.12 Shaft Design | Coaxial lumen | Coaxial lumen | Single lumen |
| B.13 Single-Use | Yes | Yes | Yes |
| B.14 Sterilization | Ethylene Oxide | Ethylene Oxide | Ethylene Oxide |
| C. Performance Specifications | | | |
| C.1 Inflation Time | Equivalent | Equivalent | ≤ 10 seconds |
| C.2 Insertability | Equivalent | Equivalent | Through equal ID tubing with 10°<br>bend around 4.5" radius |
| C.3 Removability | Equivalent | Equivalent | Through airway ≤ 60% of<br>balloon diameter |
{6}------------------------------------------------
{7}------------------------------------------------
Table 2 cont.
| Attribute | Acclarent Inspira AIR | Boston Scientific CRE | Bryan Medical aeris |
|------------------------------|-----------------------|-----------------------|------------------------------------------------------------------|
| C.4 Proximal Luer<br>Fitting | Yes | Yes | Yes |
| C.5 Torqueable | Equivalent | Equivalent | Min. 2 turns |
| C.6 Bond Force | Unknown | Unknown | Min. 10 N for diameters 5 - 9<br>Min. 15 N for diameters 10 - 16 |
The intended use of the aeris Balloon Dilation Catheter is substantially equivalent to the predicate devices in intended use, contraindications, and content of labeling. All devices meet the description set forth by the FDA requlation number 21 CFR Ref. § 874.4680.
The technology incorporated into the aeris Balloon Dilation Catheter is substantially equivalent to the technology for the predicate devices. The major components of these devices include a distally located inflation balloon, a flexible catheter shaft, and proximal luer to which an inflation device may be connected. The catheter shaft can be stiffened with a stylet (guidewire) running axially in an interior lumen. Attached to the catheter shaft under each end of the inflation balloon are radiopaque marker bands similar to the Boston Scientific CRE predicate device. All materials of the aeris Balloon Dilation Catheter were tested for biocompatibility for their intended use.
The inflation balloon sizes of the aeris Balloon Dilation Catheter are equivalent to those of the predicate devices. The predicate device's catalog of sizes includes a 5 mm through 18 mm diameter balloon at inflation pressure with intermediate balloon sizes approximately every 2 millimeters.
The aeris Balloon Dilation Catheter differs from the predicate devices in the design of the balloon. The aeris Balloon Dilation Catheter has hubs proximally and distally located on the balloon which begin to take shape upon inflation creating a dog bone balloon shape. As the pressure approaches the inflation pressure, each balloon end begins to take a spherical end shape. The predicate devices' balloon ends have a conical radius shape at pressures approximately equal to or less than two atmospheres. Above two atmospheres, a spherical end shape, similar to the aeris Balloon Dilation Catheter device takes shape.
The aeris Balloon Dilation Catheter is a sinqle lumen catheter design opposed to a dual lumen design of the predicates. The predicate devices have one lumen for the stylet (quidewire) to run through and a separate inflation lumen. The single aeris Balloon Dilation Catheter lumen is able to serve both of these functions without creating any additional risk.
The performance of the aeris Balloon Dilation Catheter was compared to the predicate devices in the insertability and removability tests. A sample device of each the Acclarent Inspira AIR and Boston Scientific CRE were tested for equivalency. Each predicate device passed the tests, showing equivalency between the aeris Balloon Dilation Catheter and each of the predicates were
{8}------------------------------------------------
applicable. The aeris Balloon Dilation Catheter also shows equivalency to performance specifications of the predicate devices in the proximal luer fitting design. Equivalency of the remaining performance specifications for both predicate devices is unknown at this time due to the specifications being unspecified by the manufactures. The aeris Balloon Dilation Catheter performance specification values confirm there are no increased risks or safety concerns introduced to the patient.
## Summary of Testing
The conclusions drawn from the nonclinical tests demonstrate that the device is as safe, as effective, and performs at least as safely and effectively as the legally marketed predicate devices identified in this submission.
| Test | Study Endpoint | Requirement | Acceptance<br>Requirement | Result |
|--------------------------------------|-------------------------------------------|--------------------------------------------------------------------------|--------------------------------------------------------------------------|--------|
| Balloon<br>Working Length | Balloon pressure<br>at 1 ATM | Balloon working<br>length within<br>tolerance | 30 ± 1 mm | Pass |
| Inflation Time | Balloon pressure<br>at 17 ATM | Balloon able to<br>inflate | < 10 seconds | Pass |
| Deflation Time | Balloon Pressure<br>at 0 ATM | Balloon able to<br>deflate | ≤ 15 seconds | Pass |
| Catheter<br>Fatigue | Balloon inflated<br>10 times to 17<br>ATM | 10 cycles to<br>recommended use<br>pressure | No evidence of<br>balloon leakage | Pass |
| Maximum<br>Inflation<br>Pressure | Balloon ruptures | Balloon failure<br>pressure ><br>recommended use<br>pressure | Balloon failure<br>pressure > 17 ATM | Pass |
| Balloon Hub<br>Measurement | Balloon pressure<br>at 17 ATM | Measure diameter of<br>hub and working<br>length | Hub diameter minus<br>working length<br>diameter < 1 mm | Pass |
| Balloon<br>Diameter | Balloon pressure<br>at 17 ATM | Balloon diameter<br>within tolerance | 5 ± .4 mm | Pass |
| Insertability | Balloon located | Distal end inserted<br>through 10° bend in<br>tube with 5 mm<br>diameter | No visual signs of<br>kinking or damage<br>done to catheter or<br>stylet | Pass |
| Removability | Catheter<br>retracted | Distal end withdrawn<br>through 3 mm tube | No visual signs of<br>damage | Pass |
| Proximal Luer<br>Fitting | Tested | Per ISO 594-1 &<br>594-2 | Meet acceptance<br>criteria of ISO 594-1 &<br>594-2 | Pass |
| Torqueable | Catheter rotated<br>2 times | Catheter turned at<br>least 2 times | No visual signs of<br>damage | Pass |
| Luer/Catheter<br>Shaft<br>Connection | Connection<br>failure | Bond tensile force<br>per ISO 10555 | ≥ 10 N | Pass |
| | | Table 3. Summary of baseline bench testing for 5x30mm aeris Balloon Dilation Catheter |
|--|--|---------------------------------------------------------------------------------------|
| | | |
{9}------------------------------------------------
| Test | Study Endpoint | Requirement | Acceptance<br>Requirement | Result |
|--------------------------------------|-------------------------------------------|--------------------------------------------------------------------------|--------------------------------------------------------------------------|--------|
| Balloon<br>Working Length | Balloon pressure<br>at 1 ATM | Balloon working<br>length within<br>tolerance | 30 ± .6 mm | Pass |
| Inflation Time | Balloon pressure<br>at 17 ATM | Balloon able to<br>inflate | < 10 seconds | Pass |
| Deflation Time | Balloon Pressure<br>at 0 ATM | Balloon able to<br>deflate | ≤ 15 seconds | Pass |
| Catheter<br>Fatique | Balloon inflated<br>10 times to 17<br>ATM | 10 cycles to<br>recommended use<br>pressure | No evidence of<br>balloon leakage | Pass |
| Maximum<br>Inflation<br>Pressure | Balloon ruptures | Balloon failure<br>pressure ><br>recommended use<br>pressure | Balloon failure<br>pressure > 17 ATM | Pass |
| Balloon Hu…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.