K150912 · National Biological Corporation · FTC · Jul 10, 2015 · General, Plastic Surgery
Device Facts
Record ID
K150912
Device Name
Houva-NET Control System
Applicant
National Biological Corporation
Product Code
FTC · General, Plastic Surgery
Decision Date
Jul 10, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4630
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Houva-NET Control System is a therapeutic product designed for individuals who require specific Ultraviolet radiation therapy for the treatment of psoriasis and vitiligo.
Device Story
The Houva-NET Control System is a software-based controller integrated into existing phototherapy devices (e.g., Houva V, Houva III, Daavlin 3 Series) to enable centralized monitoring and management of multiple units from a single computer. It utilizes a custom interface board (5PCB-040) installed in each phototherapy unit to communicate with the central computer. The system maintains the same UVA/UVB radiation delivery, digital timer control, and clinical workflow as the predicate Houva III system. Used primarily in clinical settings by healthcare professionals, the system provides a centralized interface for managing treatment parameters, alarms, and lockouts across multiple phototherapy machines. It benefits patients by allowing clinicians to efficiently oversee multiple treatment sessions, ensuring consistent delivery of prescribed UV therapy for dermatologic conditions.
Clinical Evidence
No clinical data. Substantial equivalence established through bench testing, including system verification and software validation. Performance data demonstrates conformance with IEC 60601-1 Ed.2.
Technological Characteristics
System consists of a central computer and custom interface board (5PCB-040) retrofitted into phototherapy units. Uses UVA (320-400 nm) and UVB (311-313 nm or 265-320 nm) lamps. Control is via a crystal-controlled digital timer. Features LCD touch screen, user-programmable options, alarms, and lockouts. Complies with IEC 60601-1 electrical safety and electromagnetic compatibility standards.
Indications for Use
Indicated for individuals requiring specific Ultraviolet radiation therapy for the treatment of psoriasis and vitiligo.
Regulatory Classification
Identification
An ultraviolet lamp for dermatologic disorders is a device (including a fixture) intended to provide ultraviolet radiation of the body to photoactivate a drug in the treatment of a dermatologic disorder if the labeling of the drug intended for use with the device bears adequate directions for the device's use with that drug.
Predicate Devices
Houva III Phototherapy System with PhotoSense II™ (K041212)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around it. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 10, 2015
National Biological Corporation Ms. Jennifer Cartledge REU Associates Incorporated 409 Woodridge Drive Seneca, South Carolina 29672
Re: K150912
Trade/Device Name: HOUVA-Net Control System Regulation Number: 21 CFR 878.4630 Regulation Name: Ultraviolet lamp for dermatologic disorders Regulatory Class: Class II Product Code: FTC Dated: June 15, 2015 Received: June 16, 2015
Dear Ms. Cartledge:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Jennifer R. Stevenson -S
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K150912
Device Name
HOUVA-Net Control System
Indications for Use (Describe)
The Houva-NET Control System is a therapeutic product designed for individuals who require specific Ultraviolet radiation therapy for the treatment of psoriasis and vitiligo.
Type of Use (Select one or both, as applicable)
区 Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for National Biological Corporation. The logo features the company name in a bold, sans-serif font, with the words "NATIONAL" stacked above "BIOLOGICAL". To the left of the text is a blue square with a white sine wave running through it. Below the company name is the tagline "The Phototherapy Experts!" in a smaller, italicized font.
# 5. 510(k) Summary
Type of submission: Traditional 510(k); this is the first submission for this device.
Date of Submission: March 28, 2015
Preparation of Submission: This submission has been prepared by J. Cartledge on behalf of National Biological Corporation.
## Name and Address of Manufacturer and 510(k) Owner:
National Biological Corporation 23700 Mercantile Road Beachwood, Ohio 44122 (216) 831-0600 X227 (216) 765-0271 FAX Establishment registration number: 1521608 Contact: Lynn Keller, Director of Regulatory Affairs, Quality and Engineering
#### US contact person:
Jennifer Cartledge REU Associates Inc. 409 Woodridge Drive Seneca, SC 29672 Tel.: (864) 500-0097 Email: jcartledge@reuassociates.com
## Device Identification
| Trade Name: | <i>Houva V</i> with <i>Houva-NET</i> Control System |
|----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------|
| Model number: | <i>Houva V</i> |
| Common names: | Ultraviolet lamp for dermatologic disorders |
| Classification(s) of the device: | Light, Ultraviolet, Dermatological, 21CFR 878.4630<br>Product Code: FTC<br>Classification Panel: General and Plastic Surgery<br>Class II |
## Equivalent legally marketed devices:
- Houva III Phototherapy System with PhotoSense II™ (K041212)
Houva V with Houva-NET Control System – Traditional 510(k) 510(k) Summary Section 5: Page 1 of 5
{4}------------------------------------------------
The Daavlin Smart Touch Multi-Machine Software (K110912) provides an example of a cleared device considering a single control board managing multiple phototherapy devices via a single control board. Smart Touch UV Therapy Multiple Machine. Under this pre-market clearance, the Smart Touch UV Therapy Multiple Machine utilizes a single control board to manage operation of multiple phototherapy devices.
# Device Description:
The Houva V with Houva-NET Control System is simply the integration of the new Houva-NET Control System software, into the predicate device, the Houva III Phototherapy System with PhotoSense II™. The Houva -NET Control System software is utilized to monitor multiple phototherapy devices from a single computer, the Houva-NET Central Computer. The central computer may be used to monitor mulitiple phototherapy devices from one central location. A custom built interface board. 5PCB-040, supplied by National Biological Corporation is installed in each phototherapy device. An "off-the shelf" replacement is not possible and a replacement may only be obtained from National Biological Corporation. Schematics are located in the Device Description of this 510(k) submission, Section 11.
The integration of the new Houva-NET Control System software into the predicate device, the Houva III Phototherapy System with PhotoSense II™, results in the modified system or the Houva V with Houva-NET Control System. The modification, as compared to the predicate, is limited to the replacement of the PhotoSense II Controller with the new Houva-NET Control. The Houva V with Houva-NET Control System software is designed to provide Ultraviolet A (UVA) and/or Ultraviolet B (UVB) radiation treatments in the same manner as the predicate device.
The Houva-NET Control System may be retrofitted into any Houva Phototherapy Systems, constructed in the same design configuration as the predicate device. Houva III Phototherapy System with PhotoSense II™ (K041212) and utilizing identical energy source (UV Lamps) and materials of identical composition. The Houva V with Houva-NET Control System's only variation from the predicate device is that the control system hardware and software have been updated to control multiple phototherapy units. The intended use, general and specific indications for use, spectral output, mode of operation, labeling, treatment area, and general operating principals of the Houva V with Houva-NET Control System are the same as those of the predicate. Additionally, the use of control system software and interface board to manage multiple phototherapy devices have already received FDA market clearance under K110912 integrated into the Daavlin Smart Touch Multi-Machine Software.
# Intended Use:
●
The Houva-NET Control System may also be integrated into the following HOUVA models:
- Houva II UVA (K885025)
- Houva II UVB (K885026)
- Houva II UVA/UVB (K885029) Houva V with Houva-NET Control System – Traditional 510(k) 510(k) Summary Section 5: Page 2 of 5
{5}------------------------------------------------
- Houva III Phototherapy System with PhotoSense II™ (K041212)
- Houva III/IV Phototherapy System with PhotoSense II™ ●
NOTE- the Houva III and Houva III/IV Phototherapy System are identical in every way apart from a Marketing Name change; therefore, the clearance for Houva III also applies to the Houva III/V system.
Additionally, the Houva -NET Control System is intended as a retro-fit kit for the following Daavlin Distributing Company models:
- . 3 Series SP Models (K063621):
- 3 Series PC Models (K063621): ●
- 3 Series X Models (K042502): ●
- Spectra (K828654): ●
The integration of the Houva -NET Control System as a retro-fit kit for the above noted models has been assessed through extensive verification and validation testing as documented in Section 18 of this submission, Section 18C, HN-502 Daavlin System Test. This testing confirms that retrofitting the above noted models with the Houva -NET Control System results in an equivalent system as compared to the Houva III (III/V) Phototherapy System with PhotoSense II™ predicate.
The Indication for Use for the Houva V with Houva -NET Control System is identical to that of the predicate. the Houva III Phototherapy System with PhotoSense II™ and is as follows:
The Houva V with Houva -NET Control System is a therapeutic product designed for individuals who require specific Ultraviolet radiation therapy for diagnosed skin disorders.
The Houva-NET Control System will be applicable to additional phototherapy systems once the complete verification and validation are completed for each individual system.
## Comparison to Predicate Device
The modified Houva V with Houva-NET Control System is substantially equivalent to the cleared Houva 3 Phototherapy System with PhotoSense II™ in terms of intended use, technology, and performance. Table 5-1 compares the proposed Houva -NET Control System with the predicate.
{6}------------------------------------------------
| Table 5-1: Substantial Equivalence Table | |
|------------------------------------------|--|
|------------------------------------------|--|
| Characteristic | Houva V with Houva-NET<br>Control System | Houva 3 Phototherapy<br>System with PhotoSense<br>IITM | Comparison |
|---------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Intended Use<br>Target population<br>Indication for Use | "The Houva-NET Control<br>System is a therapeutic<br>product designed for<br>individuals who require<br>specific Ultraviolet<br>radiation therapy for the<br>treatment of psoriasis and<br>vitiligo." | "The Houva 3<br>Phototherapy System with<br>PhotoSense II is a<br>therapeutic product<br>designed for individuals<br>who require specific<br>Ultraviolet radiation<br>therapy for the treatment of<br>psoriasis and vitiligo." | Identical |
| Software | The controller board<br>manages the function for<br>the applied system, but is<br>designed to interface with<br>multiple systems; however,<br>the control boards use on a<br>single unit continues to<br>manage the functions for<br>that applied system in an<br>equivalent fashion | The controller board<br>manages the function for<br>the applied system | Equivalent |
| Lamp Type(s) and<br>Wavelengths | UVA-F72T12/BL/HO<br>(PUVA) @ 320-400 nm<br>UVB-NB-TL100W/01<br>FS72 @ 311-313 nm | UVA-F72T12/BL/HO<br>(PUVA) @ 320-400 nm<br>UVB-BB-<br>FSX72T12/UVB-HO @<br>265-320 nm<br>UVB-NB-TL100W/01<br>FS72 @ 311-313 nm | Identical |
| Control (Timer)<br>Type | Digital Timer Crystal<br>controlled | Digital Timer Crystal<br>controlled | Identical |
| Electrical Safety | Devices is certified to meet<br>current electrical safety<br>standards | Designed and is certified to<br>current UL and CSA<br>standards | Identical |
| Compatibility with<br>environment and | Device meets current<br>electromagnetic | Device meets current<br>electromagnetic | Identical |
| other devices | compatibility standards | compatibility standards | |
| Where Used | Primarily used in a clinical<br>settings | Primarily used in clinical<br>settings | Identical |
| Mechanical Safety | Physical characteristics are<br>identical | Physical characteristics are<br>identical | Identical |
| Human Factors | LCD Touch screen, user<br>programmable options,<br>alarms and lockouts | LCD Touch screen, user<br>programmable options,<br>alarms and lockouts | Identical |
Houva V with Houva-NET Control System – Traditional 510(k)
{7}------------------------------------------------
# Summary of technological characteristics / performance data:
This modified device has the same indications for use and technological characteristics as the predicate devices. Comparisons of the following technological characteristics, and non-clinical performance data (indicated by *), were assessed, and the results demonstrate the substantial equivalence to the predicates:
# Description of Testing
The safe and effective performance of the Houva V with Houva-NET Control System has been clearly demonstrated by bench tests:
- System Verification Section 18 A-M
- Software Validation Section 16 A-L ●
Performance data demonstrates continued conformance with IEC 60601-1 Ed.2
Clinical testing was not required to establish equivalency of the device.
The comparison of technological characteristics, non-clinical performance data, safety testing, and software validation, demonstrates that the device is as safe, as effective, and performs as well or better than the predicate devices.
## Conclusion:
National Biological Corporation concludes that the device is substantially equivalent to the currently legally marketed predicate devices. The Houva V with Houva -NET Control System does not introduce new indications for use or intended use, has identical or equivalent technological characteristics, provides images of equivalent diagnostic capability, and does not introduce new potential hazards or safety risks. The device is as safe, as effective, and performs as well or better than the predicate devices.
> Houva V with Houva-NET Control System – Traditional 510(k) 510(k) Summary Section 5: Page 5 of 5
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.