← Product Code [OUR](/productcode/OUR) · K150875

# iFuse Implant System (K150875)

_SI-BONE, Inc. · OUR · Jul 22, 2015 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K150875

## Device Facts

- **Applicant:** SI-BONE, Inc.
- **Product Code:** [OUR](/productcode/OUR.md)
- **Decision Date:** Jul 22, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3040
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The iFuse Implant System is intended for sacroiliac fusion for conditions including sacroiliac joint dysfunction that is a direct result of sacroiliac joint disruption and degenerative sacroiliitis. This includes conditions whose symptoms began during pregnancy or in the peripartum period and have persisted postpartum for more than 6 months.

## Device Story

iFuse Implant System consists of porous plasma spray coated titanium implants; used for sacroiliac fusion to achieve biological fixation and stabilization. Implants range 30-90mm length, 4-7mm diameter. System includes associated surgical instruments. Used by surgeons in clinical settings for patients with sacroiliac joint dysfunction. Device provides mechanical stabilization of joint to facilitate fusion; benefits patient by addressing chronic pain associated with joint disruption/degeneration.

## Clinical Evidence

No clinical data or performance testing was required for this submission as it only involves a clarification of existing labeling.

## Technological Characteristics

Metallic (titanium) rods with porous plasma spray coating. Dimensions: 30-90mm length, 4-7mm diameter. Intended for surgical implantation to create biological fixation and stabilization. No changes to technological characteristics.

## Regulatory Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- iFuse Implant System ([K080398](/device/K080398.md))
- iFuse Implant System ([K092375](/device/K092375.md))
- iFuse Implant System ([K110838](/device/K110838.md))
- iFuse Implant System ([K122074](/device/K122074.md))
- iFuse Implant System ([K123850](/device/K123850.md))
- iFuse Implant System ([K131405](/device/K131405.md))
- iFuse Implant System ([K141049](/device/K141049.md))

## Reference Devices

- iFuse Implant System ([K150714](/device/K150714.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

July 22, 2015

SI-BONE, Incorporated Ms. Roxanne Dubois Vice President, Regulatory Affairs and Quality Assurance 3055 Olin Avenue, Suite 2200 San Jose, California 95128

Re: K150875

Trade/Device Name: iFuse Implant System® Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation Regulatory Class: Class II Product Code: OUR Dated: April 22, 2015 Received: April 23, 2015

Dear Ms. Dubois:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

Page 2 - Ms. Roxanne Dubois

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

# Lori A. Wiggins -S

for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known): K150875

Device Name: SI-BONE iFuse Implant System

Indications for Use:

The iFuse Implant System is intended for sacroiliac fusion for conditions including sacroiliac joint dysfunction that is a direct result of sacroiliac joint disruption and degenerative sacroiliitis. This includes conditions whose symptoms began during pregnancy or in the peripartum period and have persisted postpartum for more than 6 months

Prescription Use ___X__ (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

{3}------------------------------------------------

# 510(k) SUMMARY - iFuse Implant System®

#### 510(k) Owner's Name, Address, and Telephone Number

SI-BONE, Inc., 3055 Olin Avenue, Suite 2200, San Jose, CA 95128; (408) 207-0700

#### Contact Person

Roxanne Dubois, VP, Regulatory and Quality, SI-BONE, Inc. Email: rdubois@si-bone.com 408-828-5019; Office: 408-207-0700 x2236; Facsimile: 408-557-8312 Mobile:

Date Prepared: July 21, 2015

Trade Name of Device: iFuse Implant System®

Common or Usual Name: Orthopedic Rod

Classification Name: 21 CFR 888.3040 - Smooth or threaded metallic bone fastener; Product Code: OUR

Predicate Devices: iFuse Implant System by SI-BONE, Inc. (K080398, K092375, K110838. K122074. K123850. K131405 and K141049)

#### Intended Use

The iFuse Implant System is intended for sacroiliac fusion for conditions including sacroiliac joint dysfunction that is a direct result of sacroiliac joint disruption and degenerative sacroiliitis. This includes conditions whose symptoms began during pregnancy or in the peripartum period and have persisted postpartum for more than 6 months.

## Device Description

The iFuse Implant System consists of porous plasma spray coated titanium implants and associated surgical instruments. The iFuse Implant lengths range from 30-90mm with a diameter of 4-7mm. The fusion rods are implanted using instrumentation previously described in K080398, K092375, K110838, K122074, K123850, K131405 and K141049 and K150714.

## Technological Characteristics

The iFuse Implant System® consists of a series of metallic (titanium), porous plasma spray coated rods, intended for surgical implant within the bone to create biological fixation, immediate stabilization and fusion. This 510(k) does not involve any changes to the technological characteristics of the device.

#### Performance Data

No performance testing was required to support the clarification of existing labeling that is the subject of this 510(k).

## Substantial Equivalence and Conclusion

The iFuse Implant System is substantially equivalent to the predicate device (iFuse Implant System).

---

**Source:** [https://fda.innolitics.com/device/K150875](https://fda.innolitics.com/device/K150875)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
