Checkme Pro Health Monitor

K150869 · Viatom Technology Co., Ltd. · MWI · Dec 10, 2015 · Cardiovascular

Device Facts

Record IDK150869
Device NameCheckme Pro Health Monitor
ApplicantViatom Technology Co., Ltd.
Product CodeMWI · Cardiovascular
Decision DateDec 10, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.2300
Device ClassClass 2
AttributesPediatric

Indications for Use

The Checkme Pro Health Monitor is intended to be used for measuring, displaying, reviewing and storing of ECG (adults only), oxygen saturation and pulse rate (adults only for continuous data collection and recording, adults and pediatrics for spot checking) and temperature in the home or in healthcare facilities. This device is not intended to substitute for a hospital diagnostic ECG device and also not to be used on patients with implanted cardiac devices, such as pacemakers and/or implanted cardio-defibrillators (ICDs).

Device Story

Checkme Pro Health Monitor is a portable, handheld device for home or clinical use. Inputs: ECG (via dry electrodes or external cable), SpO2/PR (via integrated or external sensor), and temperature (via infrared thermopile sensor). Device also includes a 3D smart sensor for pedometer functions. Operation: User initiates measurements; device processes signals to display waveforms and vital signs on an integrated LCD. Data is stored locally for review. Healthcare providers use the output for spot-checking or continuous monitoring (SpO2/PR) to assist in clinical assessment. Benefits include convenient, multi-parameter monitoring for patients and clinicians. Device is not for diagnostic ECG use.

Clinical Evidence

Bench testing only. Compliance demonstrated with IEC 60601-2-47 (ECG), ISO 80601-2-61 (SpO2), ISO 80601-2-56 (thermometer), and IEC 60601-1-11 (home healthcare). Biocompatibility testing per ISO 10993-1. Electrical safety and EMC testing per IEC 60601-1 and EN 60601-1-2. Battery safety per IEC 62133. System-level testing confirmed device meets all performance specifications.

Technological Characteristics

Handheld monitor with integrated LCD. Materials: Surface-contacting, biocompatible per ISO 10993-1. Sensing: Dry conduction electrodes (ECG), infrared thermopile (temperature), pulse oximetry sensor. Power: 560mAh rechargeable Li-ion polymer battery. Connectivity: Standalone storage. Software: Moderate level of concern. Standards: IEC 60601-1, IEC 60601-2-47, ISO 80601-2-61, ISO 80601-2-56, IEC 60601-1-11, IEC 62133.

Indications for Use

Indicated for adults (ECG, continuous SpO2/PR) and pediatrics (spot-check SpO2/PR, temperature) for measuring, reviewing, and storing physiological parameters in home or healthcare settings. Contraindicated for patients with implanted cardiac devices (pacemakers/ICDs). Not for use as a substitute for hospital diagnostic ECG devices.

Regulatory Classification

Identification

A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized graphic of three human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, creating a silhouette effect. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the graphic. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 December 10, 2015 Viatom Technology Co., Ltd. % Donna-Bea Tillman Senior Consultant Biologics Consulting Group, Inc. 400 N. Washington St. Suite 100 Alexandria, Virginia 22314 Re: K150869 Trade/Device Name: CheckMe Pro Health Monitor Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer and Rate Alarm) Regulatory Class: Class II Product Code: MWI Dated: November 6, 2015 Received: November 10, 2015 Dear Donna-Bea Tillman: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply {1}------------------------------------------------ with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours. Mitchell Stein for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K150869 Device Name Checkme Pro Health Monitor #### Indications for Use (Describe) The Checkme Pro Health Monitor is intended to be used for measuring, reviewing and storing of ECG (adults only), oxygen saturation and pulse rate (adults only for continuous data collection and recording, adults and pediatrics for spot checking) and temperature in the home or in healthcare facilities. This device is not intended to subspital diagnostic ECG device and also not to be used on patients with implanted cardiac devices, such as pacemakers and/or implanted cardio-defibrillators (ICDs). | Type of Use (Select one or both, as applicable) | | |------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------| | <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary | Device Common Name: | Multiparameter Monitor | |----------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Trade Name: | Checkme Pro Health Monitor | | Applicant: | Viatom Technology Co., Ltd.<br>C607,Languang Park, No. 7 Xinxi Road, Hi-Tech<br>Park(north) Nanshan, Shenzhen, 518057, P.R.China<br>www.viatomtech.com | | Contact: | Zhou Saixin<br>General Manager<br>Viatom Technology Co., Ltd.<br>C607,Languang Park, No. 7 Xinxi Road, Hi-Tech<br>Park(north) Nanshan, Shenzhen, 518057, P.R.China<br>Phone: +86 755 86638929<br>Fax: +86 755 22649904<br>Email: zhousaixin@viatomtech.com | | Prepared by: | Donna-Bea Tillman, Ph.D.<br>Senior Consultant, Devices<br>Biologics Consulting Group, Inc.<br>Phone: 410-531-6542<br>Email: dtillman@bcg-usa.com | | Date Prepared: | November 6, 2015 | | Classification Regulation: | 21 CFR 870.2300 – Cardiac Monitor | | Panel: | Cardiovascular | | Product Code: | MWI, Physiological Patient Monitor | | Secondary Product Codes: | DQA (21 CFR 870.2700) | | | DPS (21 CFR 870.2340) | | | FLL (21 CFR 880.2910) | | | DRT (21 CFR 870.2300) | | Predicate Device: | K131762, M800 Handheld Monitor | | Secondary Predicates: | K062894, InstantCheck RTD-ECG Monitor | | | K100953, Microlife Non-Contact Infrared Forehead | {4}------------------------------------------------ ### Indications for Use: The Checkme Pro Health Monitor is intended to be used for measuring, displaying, reviewing and storing of ECG (adults only), oxygen saturation and pulse rate (adults only for continuous data collection and recording, adults and pediatrics for spot checking) and temperature in the home or in healthcare facilities. This device is not intended to substitute for a hospital diagnostic ECG device and also not to be used on patients with implanted cardiac devices, such as pacemakers and/or implanted cardiodefibrillators (ICDs). ### Device Description: The Checkme Pro Health Monitor is a lightweight, portable health monitor for use in the home or health care facility. The Checkme Pro Health Monitor is used for measuring, displaying, reviewing and storing of multiple physiological parameters including ECG, pulse oxygen saturation (SpO2), pulse rate and temperature. This device is not intended to substitute for a hospital diagnostic ECG device. The device is capable of measuring and storing data for up to twelve users. | Functions | Measured parameters | Features | |------------------------------------------------------------|------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------| | ECG | ECG waveform<br>Heart Rate (HR) | Cable or cable-free, four modes, 30s waveform<br>replay, voice memo, 100 measurements storage | | Pulse Oximeter:<br>Spot-check | Pleth waveform<br>SpO2, Pulse Rate (PR) | Cable or cable-free, 20 second measurement, 100<br>measurements storage | | Pulse Oximeter:<br>Continuous<br>Recording<br>(adult only) | Total duration, <90% STAT,<br>Average saturation, Lowest<br>saturation | Recording for up to 10 hrs | | Thermometer | Temperature | Integrated infrared sensor, 3 second measurement,<br>or °F configurable | | Pedometer | Steps, distance, calories, fat<br>burned, speed | Built-in high accuracy 3D smart sensor, algorithm<br>to calculate distance | #### Measurements Provided by Checkme Pro # Performance Testing: The performance testing provided for the Checkme Pro Health Monitor consists of the following: # ECG Testing The ECG algorithm used in the Checkme Pro Health Monitor has been tested according to the requirements of IEC 60601-2-47. Medical electrical equipment Part 2: Particular requirements for the basic safety and essential performance of ambulatory electrocardiographic systems. {5}------------------------------------------------ # SpO2 Testing The Checkme Pro Health Monitor has been tested and shown to comply with the requirements of ISO 80601-2-61. Medical electrical equipment - Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment # Thermometer Testing The Checkme Pro has been tested and shown to comply with the requirements of the FDArecognized standard ISO 80601-2-56:2009: Medical electrical equipment - Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement. # Home Healthcare Testing The Checkme Pro has been tested and shown to comply with the following FDA-recognized standard: IEC 60601-1-11 MEDICAL ELECTRICAL EQUIPMENT - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment. ### System Level Testing Viatom conducted system level performance testing to demonstrate that the device meets its specifications. The device passed all tests and performed according to specifications. #### Biocompatibility Testing: It is a surface contacting device with less than 24 hours contact duration. The patient contacting materials were testing in accordance with ISO 10993-1: 2009/AC 2010. # Software Verification and Validation: Similar to the predicate device, the software level of concern for the Checkme Pro Health Monitor is MODERATE. Per FDA Guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices - Guidance for Industry and FDA Staff", the required software documentation for a MODERATE Level of Concern device has been provided. # EMC and Electrical Safety: The following EMC and Electrical Safety tests were performed on the device: - . IEC 60601-1: 2005 + Corrigendum 1 + Corrigendum 2 + AM1 (2012) IEC 60601-1 Medical electrical equipment Part 1: General requirements for basic safety and essential performance - EN 60601-1-2: 2007 + AC: 2010 medical electrical equipment part 1-2: general . requirements for basic safety and essential performance - collateral standard: electromagnetic compatibility - requirements and tests. (General II (ES/EMC)) The results of these tests demonstrate the electrical safety and electromagnetic compatibility of the Checkme Pro device. {6}------------------------------------------------ The Checkme Pro Health Monitor is provided with a rechargeable lithium ion polymer battery. Therefore, Viatom has also conducted battery safety testing in accordance with the FDArecognized standard IEC 62133:2012 Secondary cells and batteries containing alkaline or other non-acid electrolytes – Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications. The battery passed all required tests, and therefore has been shown to be safe for the proposed intended use. The device comparison table below compares the primary features of the proposed Checkme device and the predicate M800 device. | | Subject Device | Predicate Device | |---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 510(k) Number | K150869 | K131762 | | Applicant | Viatom Technology Co. Ltd. | GUANGDONG BIOLIGHT<br>MEDITECH CO., LTD | | Device Name | Checkme Pro Health Monitor | M800 Handheld Monitor | | Classification<br>Regulation | 870.2300 - Physiological Patient<br>Monitor | 870.2300 – Physiological Patient<br>Monitor | | Product Code<br>(Primary) | MWI | MWI | | Indications for Use | The Checkme Pro Health<br>Monitor is intended to be used for<br>measuring, displaying, reviewing<br>and storing of ECG (adults only),<br>oxygen saturation and pulse rate<br>(adults only for continuous data<br>collection and recording, adults<br>and pediatrics for spot checking)<br>and temperature in the home or in<br>healthcare facilities.<br><br>This device is not intended to<br>substitute for a hospital<br>diagnostic ECG device and also<br>not to be used on patients with<br>implanted cardiac devices, such<br>as pacemakers and/or implanted<br>cardio-defibrillators (ICDs). | M800 handheld monitor is intended for<br>continuously monitoring or spot<br>checking of SpO2, PR, ECG and HR of<br>adult, pediatric and neonatal patients in<br>hospital, hospital type facilities as well<br>as in the home care environment. | | Physiological<br>Parameters Monitored | ECG, HR, SpO2, Temperature | ECG, HR, SpO2 | | Location | Home and Hospital | Home and Hospital | | Rx or OTC | Rx | Rx | | Power Supply | 560mAh rechargeable lithium-ion<br>polymer battery | Rechargeable Lithium Battery or AA<br>alkaline or lithium batteries | #### Device Comparison Table {7}------------------------------------------------ The comparison of ECG Measurements table below provides a more detailed comparison of the Checkme device with the additional predicate for ECG monitoring, the DailyCare InstantCheck RTD-ECG Monitor (K062894) #### Comparison of ECG Measurements | | Subject Device | Predicate Device (Additional) | | |---------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------| | 510(k) Number | K150869 | K062894 | | | Applicant | Viatom Technology Co. Ltd. | Daily Care BioMedical, Inc. | | | Device Name | Checkme Pro Health Monitor | InstantCheck RTD-ECG Monitor | | | Classification<br>Regulation | 870.2300 – Physiological Patient<br>Monitor | 870.2340 - Electrocardiograph | | | Measured ECG<br>Parameter | ECG waveform, Heart Rate (HR) | ECG waveform, Heart Rate (HR), ST<br>segment and QRS interval | | | ECG Rhythm<br>Classification | Not included | Regular Heart Rate<br>Fast Heart Rate<br>Slow Heart Rate<br>High ST Value<br>Low ST Value<br>High QRS Value<br>Irregular Heart Rate<br>Impossible to Analyze | | | ECG Lead Type | External ECG<br>Cable and<br>Electrodes | Integrated<br>ECG<br>Electrodes<br>External and Integrated ECG Electrodes | | | Input impedance | > 10 M Ohm | > 20 M Ohm | | | Input dynamic range | +/- 3 mV | +/- 3 mV | | | Bandwidth | 0.05 – 40 Hz | 0.1 - 40 Hz | | | A/D conversion | 16 bit | 12 bit | | | Sampling Rate | 500 samples/sec | 250 samples/sec | | | Measurement Time | 30 seconds | 30 seconds | | | Display | 400*240 Dot-matrix LCD display | 240 X 128 Dot-matrix LCD display | | | Input | Dry conduction electrodes and/or<br>external auxiliary electrodes | Dry conduction electrodes and/or external<br>auxiliary electrodes | | | Heart rate Range | 30~250 bpm, ±2 bpm or ±2%<br>whichever is larger | 45 to 180 bpm | | | | Subject Device | Predicate Device | | | 510(k) Number | K150869 | K131762 | | | Device Name | Checkme Pro Health Monitor | M800 Handheld Monitor | | | Classification<br>Regulation | 870.2300 - Physiological Patient<br>Monitor | 870.2300 - Physiological Patient<br>Monitor | | | Display Data | SpO2, Pulse Rate (PR) | SpO2, Pulse Rate (PR) | | | Mode | Continuous Recording and Spot-Check | Continuous Recording and Spot-Check | | | Sensor Types | Integrated and external | External | | | Population | Adult and Pediatric (continuous<br>recording only for adult) | Adult, Pediatric and Neonate | | | SpO2 Display Range | 0 - 100% | 0 - 100% | | | Accuracy | Integrated<br>Sensor<br>70%-100%:<br>$\pm$ 3%<br>(Arms:2.11)<br>70-80%: $\pm$ 3%<br>80-90%: $\pm$ 3%<br>90-100%: $\pm$ 2%<br>0% - 69% not<br>defined | External Sensor<br>70%-100%: $\pm$ 2%<br>(Arms:1.88)<br>70-80%: $\pm$ 3%<br>80-90%: $\pm$ 2%<br>90-100%: $\pm$ 2%<br>0% - 69% not<br>defined | 70% - 100%: $\pm$ 2%<br>0% - 69% not defined | | Resolution | 1% | 1% | | | PR range | 30-250 BPM | 25-250 BPM | | | PR Accuracy | 30-250 BPM, $\pm$ 2 bpm or $\pm$ 2%<br>whichever is larger | 25-250 BPM, $\pm$ 1 bpm or $\pm$ 2%<br>whichever is larger | | | Resolution | 1bpm | 1bpm | | | | Subject Device | Predicate Device (Additional) | | | 510(k) Number | K150869 | K100953 | | | Applicant | Viatom Technology Co. Ltd. | Microlife Intellectual Property, GMBH | | | Device Name | Checkme Pro Health Monitor | Microlife Non-Contact Infrared<br>Forehead Thermometer | | | Classification<br>Regulation | 870.2300 - Physiological Patient Monitor | 880.2910 - Thermometer, Electronic,<br>Clinical | | | Measurement<br>technique | infrared sensor (thermopile) | infrared sensor (thermopile) | | | Measurement<br>site | Forehead | Forehead | | | Unit | °F | °C/°F | | | Body mode<br>measuring<br>range | 93.2°F to 108.0°F | 34°C to 42.2°C (93.2°F to 108.0°F) | | | Body mode<br>Accuracy | ±0.4°F | 36°C to 39°C (96.8°F to 102.2°F): +/-<br>0.2°C /0.4°F<br>34.0°C to 35.9°C (93.2°F to 96.6°F),<br>39.1 °C to 42.2°C (102.4°F to 108.0°F): | | {8}------------------------------------------------ The Comparison of Pulse Oximeter Measurement and Analysis table below provides a more detailed comparison of the pulse oximetry feature of the Checkme device with the primary predicate, theM800 Handheld Monitor. Comparison of Pulse Oximeter Measurement and Analysis The Comparison of Temperature Measurement and Analysis table below provides a more detailed comparison of the Checkme and the additional predicate for temperature measurement, the Microlife Non-Contact Infrared Forehead Thermometer (K100953). {9}------------------------------------------------ #### Comparison of Temperature Measurement and Analysis #### Substantial Equivalence Conclusion: The primary predicate device for the Checkme Pro Health Monitor is M800 Handheld Monitor cleared in K131762. The M800 is also a handheld device for monitoring physiological parameters in the home or health care environment. The Checkme Pro Health Monitor has three primary medical device functions, and each of these has been compared to a predicate device. While there are some minor differences in technological characteristics, this do not raise different questions of safety and effectiveness. Each function has been tested in accordance with the appropriate standards, and the results of testing demonstrate that the Checkme Pro Health Monitor complies with the standards and has equivalent performance to the reference predicate devices. Therefore, the Checkme Pro Health Monitor has been shown to be substantially equivalent to the M800 Handheld Monitor.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...