K150867 · Biomet, Inc. · HSB · Jun 10, 2015 · Orthopedic
Device Facts
Record ID
K150867
Device Name
Affixus Tibial Nailing System
Applicant
Biomet, Inc.
Product Code
HSB · Orthopedic
Decision Date
Jun 10, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3020
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Indicated for alignment, stabilization of fractures caused by trauma or disease, and the fixation of long bones that have been surgically prepared (osteotomy) for correction of deformity and for arthrodesis. This includes transverse fractures, oblique fractures, spiral fractures, segmental fractures, open fractures, open fractures, pathologic fractures, metaphyseal fractures with bone loss, prophylactic fixation of impending fractures, pseudoarthrosis of the tibial shaft, bone transport, nonunions, and corrective osteotomies.
Device Story
Intramedullary nailing system for tibial fracture fixation; includes titanium alloy nails, 4.0/5.0mm non-locking screws, and end caps. Used by orthopedic surgeons in clinical/surgical settings to stabilize long bone fractures or deformities. Implanted via surgical instrumentation to provide mechanical support for bone healing. Benefits include anatomical design for insertion, versatile proximal/distal locking options for fracture stability, and support for complex fracture patterns.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by non-clinical performance testing, including static and fatigue construct strength testing of the nail/screw combination, which showed superior load-bearing capacity compared to the predicate.
Technological Characteristics
Fabricated from Ti6Al4V titanium alloy. Intramedullary fixation rod design with proximal (4 holes) and distal (5 holes) oblique screw positions. Sterile configuration. Mechanical fixation via 4.0mm and 5.0mm non-locking screws.
Indications for Use
Indicated for patients requiring alignment, stabilization, and fixation of tibial fractures (transverse, oblique, spiral, segmental, open, pathologic, metaphyseal), bone loss, impending fractures, pseudoarthrosis, nonunions, malunions, or corrective osteotomies.
Regulatory Classification
Identification
An intramedullary fixation rod is a device intended to be implanted that consists of a rod made of alloys such as cobalt-chromium-molybdenum and stainless steel. It is inserted into the medullary (bone marrow) canal of long bones for the fixation of fractures.
Predicate Devices
Biomet Phoenix Tibial Locking Nail System (K063570)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three heads in profile, facing right. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 10, 2015
Biomet Incorporated Ms. Suzana Otano RA Manager, Trauma 56 East Bell Drive Warsaw, Indiana 46581
Re: K150867
Trade/Device Name: Affixus Tibial Nailing System Regulation Number: 21 CFR 888.3020 Regulation Name: Intramedullary fixation rod Regulatory Class: Class II Product Code: HSB, JDS Dated: March 27, 2015 Received: April 1, 2015
Dear Ms. Otano:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{2}------------------------------------------------
Page 2 - Ms. Suzana Otano
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known)
K150867
Device Name
Affixus Tibial Nailing System
Indications for Use (Describe)
Indicated for alignment, stabilization of fractures caused by trauma or disease, and the fixation of long bones that have been surgically prepared (osteotomy) for correction of deformity and for arthrodesis. This includes transverse fractures, oblique fractures, spiral fractures, segmental fractures, open fractures, open fractures, pathologic fractures, metaphyseal fractures with bone loss, prophylactic fixation of impending fractures, pseudoarthrosis of the tibial shaft, bone transport, nonunions, and corrective osteotomies.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| <div> <span> <b> ☑ </b> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "BIOMET" in all caps. The letters are bold and black, and the background is white. The letters are evenly spaced and the word is centered.
## 510(k) Summary
In accordance with 21 CFR §807.92 and the Safe Medical Devices Act of 1990, the following information is provided for the AFFIXUS TIBIAL NAILING SYSTEM 510(k) premarket notification. The submission was prepared in accordance with the FDA guidance document, 'Format for Traditional and Abbreviated 510(k)s', issued on August 12, 2005.
| Sponsor: | Biomet Inc.<br>56 East Bell Drive<br>PO Box 587<br>Warsaw, IN 46581<br>Establishment Registration Number: 1825034 |
|-----------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Suzana Otaño<br>Regulatory Manager, Trauma |
| Date: | 27 March 2015 |
| Subject Device: | Trade Name: AFFIXUS TIBIAL NAILING SYSTEM<br>Common Name: INTRAMEDULLARY FIXATION ROD<br>Classification Name:<br>• HSB- ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES (21<br>CFR 888.3020) |
#### Legally marketed devices to which substantial equivalence is claimed:
- Biomet Phoenix Tibial Locking Nail System [K063570] ●
- . VersaNail Tibial Nail [K032097]
- TriGen Meta-Nail Tibial Nails [K061019] ●
#### Device Description
The new Affixus Tibial Nailing System is part of a long bone nailing system that offers an implant design to treat a range of tibial fractures from simple to complex, with versatile locking options. These intramedullary nails offer specific anatomical features designed to aid in insertion, while also offering options to address both proximal and distal fractures. The proximal end of the nail features four screw holes while the distal end contains five. Both the proximal and distal ends offer oblique screw positions to provide maximum fracture stability and/or facilitate capture of distal fragments, as needed. The nails are intended to be used with 4.0 and 5.0mm non-locking screws and the system offers optional end caps that can be inserted into the top of the nail following implantation. The implants shall be offered in a sterile configuration. The system is to be used with a series of instrumentation to aid during the preparation and implantation of the nail, screws and end caps.
#### Intended Use and Indications for Use
Indicated for alignment, stabilization, and fixation of fractures caused by trauma or disease, and the fixation of long bones that have been surgically prepared (osteotomy) for correction of deformity and for arthrodesis. This includes transverse fractures, oblique fractures, spiral fractures, segmental fractures, comminuted fractures, open fractures, pathologic fractures,
{5}------------------------------------------------
metaphyseal fractures and fractures with bone loss, prophylactic fixation of impending fractures, pseudoarthrosis of the tibial shaft, bone transport, nonunions, malunions, and corrective osteotomies.
#### Summary of Technological Characteristics
The rationale for substantial equivalence is based on consideration of the following characteristics:
- . Intended Use: The Affixus Tibial Nailing System and all of the predicate devices are intended for intramedullary fixation of the tibia.
- Indications for Use: The Affixus Tibial Nailing System has the same indications for use as the predicate devices.
- Materials: The Affixus Tibial Nailing System is fabricated from Ti6Al4V Titanium alloy, the same material used by the predicate devices.
- Design Features: The Affixus Tibial Nailing System is similar in overall design, diameters and lengths as the predicate devices.
- . Sterilization: The sterilization configuration of the Affixus Tibial Nailing System is the same as the predicate devices.
#### Summary of Performance Data (Nonclinical and/or Clinical)
- . Non-Clinical Tests
- Non-clinical performance testing including static and fatigue construct o (nail/screw combination) strength testing was performed and demonstrated that the new Affixus Tibial Nail construct withstood greater static load as well as fatigue load compared to the predicate device.
- . Clinical Tests
- o None provided as a basis for substantial equivalence.
#### Substantial Equivalence Conclusion
The Affixus Tibial Nailing System is substantially equivalent to the commercially available Biomet Phoenix Tibial Nailing System, the Biomet VersaNail Tibial Nailing System and the S&N TriGen Meta-Nail cleared via K063570, K032097 and K061019, respectively.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.