iHealth Align Gluco-Monitoring System The iHealth Align Gluco-Monitoring System consists of the iHealth Align Glucose meter (BG1), iHealth Blood Glucose Test Strips (AGS-1000I), and the iHealth Gluco-Smart App mobile application as the display component of the iHealth Align Gluco-Monitoring System. The iHealth Align Gluco-Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip, palm, forearm, upper arm, calf, or thigh. The iHealth Align Gluco-Monitoring System is intended to be used by a single person and should not be shared. The iHealth Align Gluco-Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The iHealth Align Gluco-Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly). iHealth BG5 wireless Smart Gluco-Monitoring System The iHealth BG5 wireless Smart Gluco-Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip, palm, forearm, upper arm, calf or thigh. The iHealth BG5 wireless Smart Gluco-Monitoring System is intended to be used by a single person and should not be shared. The iHealth BG5 wireless Smart Gluco-Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The iHealth BG5 wireless Smart Gluco-Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly). The AGS1000I test strips are for use with the iHealth BG5 meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, upper arm, calf or thigh. iHealth BG5L wireless Smart Gluco-Monitoring System The iHealth BG5L wireless Smart Gluco-Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip, palm, forearm, upper arm, calf or thigh. The iHealth BG5L wireless Smart Gluco-Monitoring System is intended to be used by a single person and should not be shared. The iHealth BG5L wireless Smart Gluco-Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The iHealth BG5L wireless Smart Gluco-Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly). The AGS1000I test strips are for use with the iHealth BG5L meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, upper arm, calf or thigh.
Device Story
System measures capillary whole blood glucose via electrochemical biosensor (glucose oxidase) and capillary action; test strips draw blood into chamber; results displayed in 5 seconds. Devices (BG1, BG5, BG5L) connect to iOS mobile devices (iPhone, iPod, iPad) via 3.5mm jack (BG1) or Bluetooth (BG5/BG5L). iHealth Gluco-Smart App acts as display, command/control interface, and data manager. Users perform self-testing at home; results aid diabetes management. App uses camera to scan QR codes on test strips for calibration. Data transmission accuracy verified between meters and mobile platforms. Single-patient use; disinfection validated for meter materials.
Clinical Evidence
No clinical data. Analytical performance (precision, linearity, specificity, method comparison) and clinical supportive data were established in previous submissions (k110017, k133790). Usability study (n=40) confirmed ease of use on representative Apple platforms. Data transmission accuracy verified.
Technological Characteristics
Glucose oxidase enzyme-based electrochemical sensing. Measuring range: 20-600 mg/dL. Hematocrit range: 20-60%. Connectivity: 3.5mm auxiliary jack (BG1), Bluetooth 3.0 (BG5), Bluetooth 4.0 (BG5L). Power: battery-operated. Form factor: handheld meters. Software: iOS-based mobile application.
Indications for Use
Indicated for people with diabetes for self-testing of capillary whole blood glucose (fingertip, palm, forearm, upper arm, calf, or thigh) to monitor diabetes control effectiveness. Contraindicated for neonates, arterial blood, serum, plasma, critically ill patients, or patients with dehydration, hypertension, hypotension, or shock. Not for diagnosis or screening of diabetes.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
Predicate Devices
iHealth Align Mini Gluco-Monitoring System (k133790)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION MEMORANDUM
## ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
k150833
B. Purpose for Submission:
Modified devices to add compatibility with additional mobile platforms; see device comparison table in section J, below.
C. Measurand:
Capillary Whole Blood Glucose from the fingertip palm, forearm, upper arm, calf or thigh
D. Type of Test:
Quantitative, amperometric method, glucose oxidase
E. Applicant:
Andon Health Co., Ltd
F. Proprietary and Established Names:
iHealth Align Gluco-Monitoring System
iHealth BG5 wireless Smart Gluco-Monitoring System
iHealth BG5L wireless Smart Gluco-Monitoring System
G. Regulatory Information:
1. Regulation section:
21 CFR 862.1345 - Glucose test system
21CFR 862.2100 - Calculator/data processing module for clinical use
2. Classification:
Class II
Class I, exempt
3. Product code:
NBW - system, test, blood glucose, over the counter
{1}
CGA - Glucose Oxidase
JQP - calculator/data processing module, for clinical use
4. Panel:
Clinical Chemistry (75)
H. Intended Use:
1. Intended use(s):
See indications for use, below.
2. Indication(s) for use:
iHealth Align Gluco-Monitoring System
The iHealth Align Gluco-Monitoring System consists of the iHealth Align Glucose meter (BG1), iHealth Blood Glucose Test Strips (AGS-1000I), and the iHealth Gluco-Smart App mobile application as the display component of the iHealth Align Gluco-Monitoring System. The iHealth Align Gluco-Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip, palm, forearm, upper arm, calf, or thigh. The iHealth Align Gluco-Monitoring System is intended to be used by a single person and should not be shared.
The iHealth Align Gluco-Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The iHealth Align Gluco-Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly).
iHealth BG5 wireless Smart Gluco-Monitoring System
The iHealth BG5 wireless Smart Gluco-Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip, palm, forearm, upper arm, calf or thigh. The iHealth BG5 wireless Smart Gluco-Monitoring System is intended to be used by a single person and should not be shared.
The iHealth BG5 wireless Smart Gluco-Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The iHealth BG5 wireless Smart Gluco-Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly).
The AGS1000I test strips are for use with the iHealth BG5 meter to quantitatively
2
{2}
measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, upper arm, calf or thigh.
## iHealth BG5L wireless Smart Gluco-Monitoring System
The iHealth BG5L wireless Smart Gluco-Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertip, palm, forearm, upper arm, calf or thigh. The iHealth BG5L wireless Smart Gluco-Monitoring System is intended to be used by a single person and should not be shared.
The iHealth BG5L wireless Smart Gluco-Monitoring System is intended for self testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. The iHealth BG5L wireless Smart Gluco-Monitoring System should not be used for the diagnosis of or screening of diabetes or for neonatal use. Alternative site testing should be done only during steady - state times (when glucose is not changing rapidly).
The AGS1000I test strips are for use with the iHealth BG5L meter to quantitatively measure glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips, palm, forearm, upper arm, calf or thigh.
## 3. Special conditions for use statement(s):
For the iHealth Align, iHealth BG5 wireless Smart, and iHealth BG5L wireless Smart Gluco-Monitoring Systems:
- Not intended for use on neonates.
- Not intended for use on artery blood, serum, and plasma.
- For single patient use.
- Should not be used for the diagnosis of or screening for diabetes
- Not for use for patients in a hyperglycemic-hyperosmolar state, with or without ketosis.
- Not for use on critically ill patients.
- Not for patients who are dehydrated, hypertensive, hypotensive, or in shock.
- Very low or very high red blood cell count (hematocrit) can lead to incorrect test results.
- AST should be used only during steady-state times when blood glucose levels are not changing rapidly.
- Alternate Site Testing should not be used to calibrate continuous glucose monitoring systems (CGMs).
- Results from Alternate Site Testing should not be used in insulin dose calculations.
## 4. Special instrument requirements:
iHealth Align Glucose meter (BG1)
iHealth BG5 meter
{3}
iHealth BG5L meter
# I. Device Description:
The iHealth Align Gluco-Monitoring System consists of a blood glucose meter, test strips, iHealth Gluco-Smart App, sterile lancets, lancing device and AGS1000I Control Solutions (Level I, Level II and Level III). The iHealth Align Gluco-Monitoring System cannot display test results and must be used with an iPhone or iPod touch via an $3.5\mathrm{mm}$ auxiliary jack.
The iHealth BG5 wireless Smart and iHealth BG5L wireless Smart Gluco-Monitoring Systems consist of the BG5 and BG5L wireless Smart blood glucose meters, respectively, AGS 1000I Test Strips, sterile lancets, lancing device and the iHealth control solutions control solutions. (Control solutions provided are for Level 1, II, and III). iHealth BG5L uses Bluetooth 4.0 wireless radio technology; while iHealth BG5 uses Bluetooth 3.0 wireless radio technology. The iHealth BG5 and BG5L meters can display the test results and the test results can also be transmitted to an iPhone, iPod touch or iPad through blue tooth.
iHealth Gluco-Smart App is iOS- based software for use with the iHealth Align Glucose meter (BG1), iHealth BG5 meter, and iHealth BG5L meter. When used with these meters, Apps-AG01 iHealth App acts as a display and allows command and control of the meter. The App can transfer data from the device's memory, manage, and share the data.
# J. Substantial Equivalence Information:
1. Predicate device name(s):
iHealth Align Mini Gluco-Monitoring System
2. Predicate 510(k) number(s):
k133790
3. Comparison with predicate:
| Item | Predicate device | Candidate Devices | |
| --- | --- | --- | --- |
| Device name | iHealth Align Mini Gluco-Monitoring System | iHealth BG5L wireless Smart Gluco-Monitoring System and iHealth BG5L wireless Smart Gluco-Monitoring System | iHealth Align Gluco-Monitoring system |
{4}
| Item | Predicate device | Candidate Devices | |
| --- | --- | --- | --- |
| IFU | Is intended to be used for quantitative measurement of glucose in fresh capillary whole blood as an aid to monitor the effectiveness of diabetes control in people with diabetes. | Same | Same |
| Model | BG1 | BG5 (Bluetooth)BG5L (Bluetooth low energy) | Same |
| Enzyme | Glucose oxidase | Same | Same |
| Measuring range | 20 – 600 mg/dL | Same | Same |
| Hematocrit range | 20-60% | Same | Same |
| Connectivity to Meter | Earphone jack | Bluetooth and Bluetooth low energy | Earphone jack |
| Display | Connect to Apple platform | Connect to Apple platform and LED meter display | Same |
| Test Strip Calibration | QR code scan | Same | Same |
| Dimensions | 52 mm×30 mm×9.5 mm | 9 mm × 34.5mm ×19mm | Same |
| Mobile App name | iHealth Gluco-Smart App | Same | Same |
| Mobile App | V2.3 | V3.8.5 | V3.8.5 |
| Compatible iOS version | 5,6,7 | 7,8,9 | 7,8,9 |
{5}
| Item | Predicate device | Candidate Devices | |
| --- | --- | --- | --- |
| Phone Platform | iPhone 5s
iPhone 5c
iPhone 5
iPhone 4S
iPhone 4
iPhone 3GS
iPod touch (4th generation)
iPod touch (5th generation) | iPhone 5s
iPhone 5c
iPhone 5
iPhone 4S
iPhone 4
iPhone 3GS
iPod touch (4th generation)
iPod touch (5th generation)
iPhone 6S
iPhone 6S Plus
iPhone 6
iPhone 6 PLUS
iPad2
iPad3
iPad4
iPad Mini
iPad mini 2
iPad Air
iPad Air 2 | iPhone 5s
iPhone 5c
iPhone 5
iPhone 4S
iPhone 4
iPhone 3GS
iPod touch (4th generation)
iPod touch (5th generation)
iPhone 6S
iPhone 6S Plus
iPhone 6
iPhone 6 PLUS |
K. Standard/Guidance Document Referenced (if applicable):
None
L. Test Principle:
The iHealth Align Gluco-Monitoring System, iHealth BG5 Wireless Smart Gluco-Monitoring System, and BG5L Wireless Smart Gluco-Monitoring Systems are based on an electrochemical biosensor technology (electrochemical) and the principle of capillary action. Capillary action at the end of the test strip draws the blood into the action chamber and the blood glucose result is displayed in 5 seconds.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
As established in k110017
b. Linearity/assay reportable range:
As established in k110017
{6}
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
Traceability:
The system is traceable to the NIST SRM 917 glucose reference material.
Traceability was evaluated in k110017.
Test Strip Stability:
The AGS-1000I blood glucose test strips are identical to the test strips in the predicate device, iHealth Align Mini Gluco-Monitoring System (BG1 - k133790). Stability protocols and acceptance criteria were reviewed in k110017 and support the sponsors claimed shelf life of 24 month shelf-life when stored at 39 to 86°F (2-30°C) with relative humidity of <80% and a 90 day stability after opening when stored at 39 to 86°F with relative humidity of <80%.
Control Solution Stability:
The iHealth control solutions are identical to the control solutions in the predicate device (k133790). Value assignment and stability protocols and acceptance criteria were reviewed in k110017 and support the sponsors claimed shelf life of 24 month shelf-life when stored at 39 to 86°F (2-30°C) with relative humidity of <80% and a 90 day stability after opening when stored at 39 to 86°F with relative humidity of <80%.
d. Detection limit:
The reportable range is 20 to 600 mg/dL based on the linearity/assay reportable range study above (section M.1.).
e. Analytical specificity:
As established in k110017
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
As established in k110017
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
{7}
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
As established in k110017
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
| Time of day | People without diabetes |
| --- | --- |
| Fasting and before meals | <100 mg/dL |
| 2 hours after meals | <140 mg/dL |
(American Diabetes Association: Diagnosis and Classification of Diabetes Mellitus (Position Statement). Diabetes Care 36 (Supp. 1) S71, 2013.
N. Instrument Name:
iHealth Align Gluco-Monitoring System
iHealth BG5 wireless Smart Gluco-Monitoring System
iHealth BG5L wireless Smart Gluco-Monitoring System
O. System Descriptions:
1. Modes of Operation:
Does the applicant’s device contain the ability to transmit data to a computer, webserver, or mobile device?
Yes ☐ X ☐ or No ☐
Does the applicant’s device transmit data to a computer, webserver, or mobile device using wireless transmission?
Yes ☐ X ☐ or No ☐
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for
{8}
this line of product types:
Yes ☐ X ☐ or No ☐
3. Specimen Identification:
Samples are applied directly to the test strip as they are collected.
4. Specimen Sampling and Handling:
This device is intended to be used with capillary whole blood from AST (Alternative site testing, including palm, forearm, upper arm, calf, or thigh) and finger.
5. Calibration:
The user scans a QR code to code the test strips. The App controls the platform's camera. The QR scan code App function was successfully validated on representative Apple platforms for the BGL, BG5 and BG5L meters.
6. Quality Control:
The iHealth Control solution are used as a quality control checks to make sure that the iHealth Align Gluco-Monitoring System (BG1) and iHealth BG5/5L Wireless Smart Gluco-Monitoring Systems and the AGS-1000I blood glucose test strips are working correctly. The labeling provides instructions on when quality control testing should be performed.
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
1. Software: Software documentation was reviewed and demonstrated that the device was developed under appropriate software lifecycle processes.
2. Electromagnetic compatibility (EMC) (radiated emissions and immunity) testing: As established in k120813
3. Usability Questionnaire: Using representative Apple platforms, a 40 participant study was conducted to verify users with average education and computer skills and no prior knowledge of the system can use the new device without direct instruction or training. Study results demonstrated that participants rated the ease of use of the devices as normal or above.
4. Data transmission accuracy: Demonstrated, using representative Apple platforms and 5 of each meter (iHealth Align, iHealth BG5 and BG5L), that sent and received data matched.
{9}
10
5. Infection Control and Robustness Studies:
The iHealth Align Gluco-Monitoring System, iHealth BG5 Wireless Smart Gluco-Monitoring System, and BG5L Wireless Smart Gluco-Monitoring System are intended for single-patient use. The sponsor stated that the disinfection efficacy for the meters was established in k110017 (BG5 and BG5L) and in k123935 (Align Mini) and that the materials comprising the candidate meters are identical. CaviWipes Disinfecting Towelettes (EPA registration #46781-8) were previously validated through disinfection efficacy studies demonstrating complete inactivation of hepatitis B (HBV) virus using materials comprising the meter.
Robustness testing was previously established in k133790 (Align Mini) and k123935 (BG5 and BG5L) demonstrating that there was no change in performance or in the external materials of the meter after 11,000 cleaning and disinfection cycles (one cycle includes one cleaning wipe plus one disinfecting wipe) to simulate 5 years of single-patient use. Labeling was reviewed for adequate instructions for the validated cleaning and disinfection procedures.
The sponsor provides a disposable sleeve for use with the iPod or iPhone to be used with each test strip, not to be reused and to be disposed of as bio-hazard, therefore no disinfection studies were required of the iPod or iPhone. The sleeve is intended to be used with the iHealth Align Gluco-Monitoring System.
6. Hematocrit: As established in k110017
7. Altitude: As established in k110017
8. Temperature and Humidity operating conditions: As established in k110017
9. Sample volume: As established in k110017
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.