The Nutra Face LIFT Model PE8050 device is intended for facial stimulation and indicated for over-the-counter cosmetic use.
Device Story
Hand-held facial toning device; applies low-level electrical impulses to face via two chrome-plated spherical probes; requires conductive derma gel. User-operated; powered by internal Ni-MH rechargeable battery. User adjusts intensity (0-400 microamps) via up/down buttons. Device alternates polarity between probes. Intended for home use to improve facial skin tone. Benefits include non-invasive cosmetic facial stimulation.
Clinical Evidence
Usability and Product Self-Selection Study conducted with 80 participants. Study goals: verify intended user population, confirm ability to self-select via labeling, test knowledge of instructions, and demonstrate correct operation/care. All study goals were met. No clinical efficacy data provided; device relies on bench testing for safety and performance verification.
Indicated for over-the-counter cosmetic facial stimulation.
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference. Inside the circle is a stylized emblem featuring three human profiles facing to the right, stacked on top of each other, with flowing lines suggesting movement or connection.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 10, 2016
Nutra Luxe MD, LLC Gloria Avendano Regulatory Affairs Manager 12801 Commonwealth Dr. Units 2-6 Fort Myers, FL 33913
Re: K150826 Trade/Device Name: Nutra Face LIFT Model PE8050 Regulation Number: 21 CFR 882.5890 Regulation Name: Transcutaneous electrical nerve stimulator for pain relief Regulatory Class: Class II Product Code: NFO Dated: January 4, 2016 Received: January 13, 2016
Dear Gloria Avendano:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Michael J. Hoffmann -A
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K150826
Device Name Nutra Face LIFT Model PE8050
Indications for Use (Describe)
The Nutra Face LIFT Model PE8050 device is intended for facial stimulation and indicated for over-the-counter cosmetic use.
Type of Use (*Select **one** or **both**, as applicable*)Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Nutra Luxe MD. The words "nutra" and "luxe MD" are written in different fonts and colors. A black flower is placed between the two words.
## FDA 510K Summary
## The Nutra Face LIFT
#### 1. General Information
Submitter:
Nutra Luxe MD, LLC 12801 Commonwealth Dr. Units 2-6 Fort Myers, FL 33913
| Contact Person: | Gloria Avendano |
|-----------------|----------------------------------|
| | Nutra Luxe MD |
| | 12801 Commonwealth Dr. Units 2-6 |
| | Fort Myers, FL 33913 |
| | gloria@nutraluxemd.com |
Summary Preparation Date:
February 9, 2016.
## 2. Device Name
| Trade Name: | Nutra Face Lift Model PE8050 |
|----------------------|------------------------------------------------------------|
| Common Name: | Facial Toning Device |
| Classification Name: | Transcutaneous electrical nerve stimulator 21 CFR 882.5890 |
| Product Code: | NFO |
#### 3. Predicate Device
The Nutra Face Lift is substantially equivalent to the NuFace Facial Toning Device K072260
#### 4. Device Description
The Nutra Face Lift is a hand-held device intended to apply low level electrical impulses to strategic locations on the face. The Nutra Face Lift probes are designed for optimal contact with the face. The device continually alternates between the positive and negative probes and allows the user to adjust the settings from 0 to 400 microamps for personalized comfort level by pressing the up/down button. The Nutra Face Lift requires the use of a conductive derma gel.
The Nutra Face Lift unit contains a power supply and internal Ni-MH rechargeable battery. The enclosure is made of medical grade biocompatility plastics and the output contacts (Probes) consist of chrome-plated spheres. The on/off button on the unit directly allows the user to immediately remove all power from the unit.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the words "nutra luxe MD" with a flower in between the words "nutra" and "luxe". The word "nutra" is in black, the flower is in black, and the words "luxe MD" are in pink. The words and flower are reflected below.
#### 5. Intended Use and Indications:
The Nutra Face Lift device is ntended for facial stimulation and is indicated for-over-the counter cosmetic use.
#### Substantial Equivalency & Comparison of Technological Similarities and Differences 6.
The Nutra Face Lift is as safe and effective as the predicate device. The Nutra Face Lift has the same indications, as the predicate device.
The Nutra Face Lift is substantially equivalent in technological characteristics to the predicate device, including the positive and negative probes. The device powered by a 5V-500Ma AC/DC charger, produces microcurrent that is discharged through the two fixed, smooth probes.
The Nutra Face Lift has the following similarities to the NuFace
- 1. Has the same indicated use
- Uses the same operating principle 2.
- 3. Incorporates similar materials.
- Utilizes a same treatment duration (i.e. 5 seconds per step, repeat each step 3 times) 4.
- Utilizes a similar treatment regimen (i.e. 5 times a week for 90 days) న్.
One of the differences between the Nutra Face Lift and its predicate device is the predicate device uses a thumbwheel to adjust the desired intensity while the Nutra Face Lift uses an up/down button.
Any differences between The Nutra Face Lift and the predicates device are not significant to its safety or effectiveness for its intended use.
#### Section 1: Device Descriptions
| Elements of Comparison | Nutra Face Lift | Predicate Device |
|----------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K150826 | K072260 |
| Regulation Number | 21 C.F.R 882.5890 | 21 C.F.R 882.5890 |
| Regulation Name | Transcutaneous Electrical Nerve<br>Stimulator | Transcutaneous Electrical Nerve<br>Stimulator |
| Regulatory Class | Class II | Class II |
| Product Code | NFO | NFO |
| Intended use | Stimulate the face; skin toning | Stimulate the face; skin toning |
| Indications for Use | Intended for facial stimulation and is<br>indicated for-over-the counter<br>cosmetic use. | Intended for facial stimulation<br>and is indicated for-over-the counter<br>cosmetic use. |
| Technological Characteristics | The Nutra Face Lift is a hand-held<br>device intended to apply low level<br>electrical impulses to strategic locations<br>on the face. The Nutra Face Lift probes<br>are designed for optimal contact with<br>the face. | NuFace® is a Facial Toning Device<br>intended for facial stimulation. The<br>device measures 7" L x 2.5" W x 1"<br>D. Its outer case is injection molded of<br>thermoplastic resin, ABS UL 94 HB,<br>and the output contacts (probes)<br>consist of chrome-plated spheres. The |
| | The device continually alternates<br>between the positive and negative<br>probes and allows the user to adjust the<br>settings from 0 to 400 microamps for<br>personalized comfort level by pressing<br>the up/down button. The Nutra Face<br>Lift requires the use of a conductive<br>derma gel. | device, powered by a 9-volt battery,<br>produces microcurrent that is<br>Discharged through the two fixed,<br>smooth spherical probes. To turn the<br>device on, the thumbwheel is pushed<br>upwards. A Green LED light will<br>Then illuminate, indicating the unit is<br>ready for use. Users then follow the<br>instructions for use. The two probes<br>gently glide over the skin to deliver<br>low-level electrical impulses to<br>strategic locations on the face. |
| Power Source | One 3.6 V battery | One 8V Battery |
| Number of output modules | 1 | 1 |
| Number of output channels | 1 output channel | 1 output channel |
| Regulated current or regulated<br>Voltage? | Both | Both |
| Software/Firmware/Microprocessor<br>Control? | Yes | Yes |
| Automatic Overload Trip? | No | No |
| Automatic Shut off? | No | No |
| Patient override control? | Yes | Yes |
| Indicator Display | Yes | No |
| Timer range | No Timer | No timer |
| Type of protection | Type BF | Type BF |
| Number of output modules | 1 | 1 |
| Number of output channels | 1 | 1 |
| Software/Firmware | No | No |
| On/off status | No | No |
| Standards Compliance | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-10<br>60601-1-11 | IEC 60601-1<br>IEC 60601-1-2<br>IEC 60601-2-10 |
| Biocompatibility | ISO 10993-5, -10 | ISO 10993-5, -10 |
| Waveform | Pulsed Monophasic | Pulsed Monophasic |
| Shape | Modulated Square | Rectangular Pulses |
| Maximum output voltage | 50Vpp (No load condition)<br>8Vpp (10kΩ) | 158 Mv @ 500Ω<br>780 Mv @ 2kΩ<br>2.6 V @ 10kΩ |
| Maximum output current | +/- 400μA @ (10kΩ) | 223 μA @ 500 Ω |
| Output tolerance | +/- 10% | +/- 10% |
| Pulse Width | 60ms | 112 ms |
| Frequency (Hz) | 8.34 Hz | 8.39 Hz |
| Symmetrical phases | Not Multiphasic | Not Multiphasic |
| Phase duration | Not determined | Not determined |
| Net Charge (µC per pulsetrain) | 0 µC per pulsetrain | 0 µC per pulsetrain |
| Maximum Phase Charge (µC) | 18.13 µC @ 500 Ω | 18.13 µC @ 500 Ω |
| Maximum current Density<br>(Ma/Cm²) | 0.341mA/cm² @ 500 Ω | 0.341mA/cm² @ 500 Ω |
| Maximum Power Density<br>(μW/cm²) | 3.02μW/cm² @ 500 Ω | 3.02μW/cm² @ 500 Ω |
| Pulses per burst | 20 | 20 |
| Burst duration | 2.4 | 2.3 |
| ON Time (seconds) | Constant | Constant |
| OFF Time (Seconds) | None | None |
#### Nutra Face Lift and NuFace Original Comparison Table
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the words "nutra luxe MD" with a flower in between the words "nutra" and "luxe". The word "nutra" is in black, the flower is in black, and the words "luxe MD" are in pink. The words and flower have a shadow effect.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for "nutra luxe MD". The word "nutra" is in black, followed by a black hibiscus flower. The words "luxe MD" are in pink. The logo has a reflection underneath.
## Section 2: Basic Characteristics
| 510(k) Number | K150826 | K072260 |
|--------------------|-----------------|--------------------|
| Device Name, Model | Nutra Face Lift | NuFace® |
| Manufacturer | NutraLuxe MD | Carol Cole Company |
| Power Source(s) | 2.4 | 2.4 |
| ON Time (seconds) | Constant | Constant |
| OFF Time (Seconds) | None | None |
#### 7. Clinical Performance Data
A Usability and Product Self Selection Study was conducted utilizing 80 participants: with the following four goals:
(1) To attract participants that represented the "intended users" of the device;
- (2) To determine if consumers could correctly self-select using the Packaging labeling only:
- (3) To test consumer knowledge of the Packaging labeling and Instruction Manual;
- (4) To have consumers demonstrate their ability to adhere to what they had read in the Instruction Manual and actually operate and care for the device correctly. All four goals of the Study were met.
#### Nonclinical Performance Data 8.
Based on the Risk Analysis, the verification and validation tests that were performed and the acceptance criteria applied for each are listed in Section 17.
The Nutra Face Lift device was tested and found to be in compliance with FDA's performance standards set forth in 21 C.F.R. §898.
The Nutra Face Lift was also tested and complies with Electrical Safety and EMC testing, which include the requirements of IEC 60601-1:2012 3rd Edition "Medical Electrical Equipment Part 1 - General Requirements for Safety" IEC 60601-1-2 "Medical Electrical Equipment Part 1-2, General Requirements for Safety - Collateral Standard Electromagnetic Compatibility Requirements and Tests IEC 60601-1-11 Home healthcare medical equipment and IEC 60601-2-10:2012 2nd Edition "Medical Electrical equipment -- part 2-10: particular requirements for the basic safety and essential performance of nerve and muscle stimulators. (Neurology). In addition, testing and analysis have demonstrated compliance of the plastic within ISO 10993 (Biocompatibility).
#### 9. Regulatory Requirements
NutraLuxe MD manufactures under strict quality assurance guidelines and US FDA Good Manufacturing Practice (GMP)
Nutra Luxe MD is fully compliant with 21 CFR Part 820, US FDA Quality Systems Regulations (QSR), Risk Analysis and Risk Management files (RMF) conforms to ISO 14971,
Conclusion: Nutra Luxe MD, Inc. found The Nutra Face Lift device to be substantially equivalent to the legally marketed predicate devices.