K150825 · Collagen Matrix, Inc. · GXQ · Nov 20, 2015 · Neurology
Device Facts
Record ID
K150825
Device Name
Collagen Dural Regeneration Matrix
Applicant
Collagen Matrix, Inc.
Product Code
GXQ · Neurology
Decision Date
Nov 20, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5910
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Collagen Dural Regeneration Matrix is intended for use as a dura substitute for the repair of dura mater.
Device Story
Collagen Dural Regeneration Matrix is a white, non-friable, resorbable, porous type I collagen matrix derived from bovine Achilles tendon. It is designed to conform to the contours of a dural defect site. The device is supplied sterile and non-pyrogenic for single-use application. It functions as a physical barrier to facilitate dural repair and undergoes gradual resorption in vivo. It is intended for use by surgeons during neurosurgical procedures to repair the dura mater. The device provides a scaffold for tissue repair, helping to prevent cerebrospinal fluid (CSF) leakage and support healing at the surgical site.
Clinical Evidence
No clinical trials were performed. Evidence consists of non-clinical bench testing (dimensions, pH, tensile strength, conformability, hydrothermal transition, liquid permeability, burst strength) and biocompatibility testing (ISO 10993 series: cytotoxicity, sensitization, intracutaneous reactivity, systemic toxicity, genotoxicity, pyrogenicity, muscle/subcutaneous implantation). An animal efficacy study in a rabbit dural defect model compared the device to the predicate for repair and resorption. Clinical history of the predicate device was referenced.
Technological Characteristics
Material: Type I collagen from bovine Achilles tendon. Form: Porous, sponge-like matrix with one smooth surface. Properties: Non-friable, conformable, resorbable. Sterilization: Sterile (SAL 10^-6). Standards: ISO 22442-1/2/3 (animal tissues), ISO 10993-3/5/6/10/11 (biocompatibility).
Indications for Use
Indicated for the repair of dura mater in patients requiring a dura substitute.
Regulatory Classification
Identification
A dura substitute is a sheet or material that is used to repair the dura mater (the membrane surrounding the brain).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure, with three overlapping profiles facing to the right. The profiles are rendered in black and appear to be connected, forming a single, flowing shape. The logo is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 20, 2015
Collagen Matrix, Inc. Ms. Gloria Zuclich Senior Manager of Regulatory Affairs 15 Thornton Road Oakland, New Jersey 07436
Re: K150825
Trade/Device Name: Collagen Dural Regeneration Matrix Regulation Number: 21 CFR 882.5910 Regulation Name: Dura Substitute Regulatory Class: Class II Product Code: GXQ Dated: November 13, 2015 Received: November 16, 2015
Dear Ms. Zuclich:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Carlos L. Pena -S FD/△
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K150825
Device Name Collagen Dural Regeneration Matrix
Indications for Use (Describe)
Collagen Dural Regeneration Matrix is intended for use as a dura substitute for the repair of dura mater.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="font-family: Arial;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|-----------------------------------------------------------------------------------------|
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
### 1. Applicant Information
| Applicant Name: | Collagen Matrix, Inc. |
|-----------------|--------------------------------------------------------|
| Address: | 15 Thornton Road<br>Oakland, New Jersey 07436 USA |
| Telephone: | (201) 405-1477 Ext. 317 |
| Fax: | (201) 405-1355 |
| Contact Person: | Gloria Zuclich<br>Senior Manager of Regulatory Affairs |
| Date Prepared: | November 19, 2015 |
#### 2. Name of the Device
| Device Trade Name: | Collagen Dural Regeneration Matrix |
|-----------------------------|-------------------------------------------------------------------|
| Device Common Name: | Collagen Dura Substitute |
| Device Classification Name: | Dura Substitute |
| | Regulation Number 882.5910<br>Product Code GXQ<br>Device Class II |
#### 3. Leqally Marketed Devices to Which Substantial Equivalence is Claimed
DuraGen Plus® Dural Regeneration Matrix Predicate Device(s): K032693
#### 4. Description of the Device
Collagen Dural Regeneration Matrix is a white, non-friable, resorbable and biocompatible type I collagen matrix made from purified bovine Achilles tendon. Collagen Dural Regeneration Matrix is a porous, sponge-like collagen matrix with one smooth surface that conforms to the contours of the defect site. It is supplied sterile, non-pyrogenic, in various sizes, and for single use only.
#### 5. Intended Use
Collagen Dural Regeneration Matrix is intended for use as a dura substitute for the repair of dura mater.
#### 6. Summary/Comparison of Technical Characteristics
Collagen Dural Regeneration Matrix has been determined to be substantially equivalent to the predicate device having similar technological characteristics as follows:
{4}------------------------------------------------
| Parameter | Collagen Dural Regeneration<br>Matrix (This submission) | DuraGen Plus® Dural Graft<br>Matrix |
|---------------------|---------------------------------------------------------------------------|---------------------------------------------------------------------------|
| 510(k) | K150825 | K032693 |
| Indications for Use | Intended for use as a dura<br>substitute for the repair of dura<br>mater. | Intended for use as a dura<br>substitute for the repair of dura<br>mater. |
| Collagen Source | Bovine Achilles tendon | Bovine Achilles tendon |
| Form | Porous Collagen Matrix | Porous Collagen Matrix |
| Color | White to off-white | White to off-white |
| Physical Integrity | Non-friable | Non-friable |
| Sizes | Variety of sizes | Variety of sizes |
| Conformability | Conformable | Conformable |
| Biocompatibility | Biocompatible | Biocompatible |
| In Vivo Stability | Gradual resorption | Gradual resorption |
| Sterility | Sterile, SAL 10-6 | Sterile, SAL 10-6 |
| Pyrogenicity | Non-pyrogenic | Non-pyrogenic |
| Single Use/Reuse | Single use only | Single use only |
| Packaging | Double blister | Double blister |
# 7. Discussion of Non-clinical Testing
The substantial equivalence of Collagen Dural Regeneration Matrix and its predicate was demonstrated based on in vitro characterization studies, biocompatibility studies, and an animal efficacy study.
Non-clinical testing was performed in accordance with FDA recognized consensus standards and FDA guidelines as follows:
FDA Guidance Document entitled, "Guidance document for Dura Substitute Device: Guidance for Industry", issued on November 9, 2000
ISO 22442-1 Animal Tissues and Their Derivatives Utilized in the Manufacture of Medical Devices – Part 1 Analysis and Risk Management
ISO 22442-2 Animal Tissues and Their Derivatives Utilized in the Manufacture of Medical Devices - Part 2 Controls on Sourcing, Collection, and Handling
ISO 22442-3 Animal Tissues and Their Derivatives Utilized in the Manufacture of Medical Devices – Part 3 Validation of the Elimination and/or Inactivation of Viruses and Transmissible Agents
ISO 10993-3:2009 Biological Evaluation of Medical Devices- Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicity
ISO 10993-5:2009 Biological Evaluation of Medical Devices- Part 5: Tests for in vitro cytotoxicity
{5}------------------------------------------------
ISO 10993-6:2009 Biological Evaluation of Medical Devices- Part 6: Test for local effects after implantation
ISO 10993-10:2009 Biological Evaluation of Medical Devices- Part 10 Test for local effects after implantation
ISO 10993-11:2009 Biological Evaluation of Medical devices - Part 11 Tests for systemic toxicity
## Non-clinical Testing Conducted
In vitro product characterization testing was performed to demonstrate substantial equivalence of the subject device to its predicate device. A series of bench tests were conducted to evaluate material properties, biological properties, chemical and physical properties. The comparative bench testing is summarized in the table below.
| Test | Test Method | Results |
|-------------------------------------------|--------------------------------------------------------------------|---------------------------------------------------------------------|
| Dimensions | Measurements | Dimensions similar to predicate device |
| pH | Internal test method using<br>pH meter | pH similar to predicate device |
| Tensile<br>strength | Internal test method using<br>mechanical test apparatus | Tensile strength similar to predicate device |
| Conformability | Internal test method to<br>measure drape angle | Conformability similar to predicate device |
| Hydrothermal<br>transition<br>temperature | Internal test method using<br>differential scanning<br>calorimeter | Hydrothermal transition temperature<br>similar to predicate device. |
| Liquid<br>Permeability | Internal test method to<br>measure permeability to<br>liquid | Minimally permeable; similar to predicate |
| Burst strength | Internal test method to<br>measure burst strength | Adequate for cerebrospinal fluid (CSF)<br>pressure |
A series of in vitro and in vivo biocompatibility testing was performed to assess safety of the Collagen Dural Regeneration Matrix as an implantable material. The biocompatibility testing performed is summarized in the table below.
| Test | Test Method/ Model | Results |
|------------------------------|---------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Cytotoxicity | Agarose Overlay, ISO 10993-5 | Non-cytotoxic |
| Sensitization | Guinea Pig Maximization, ISO<br>10993-10 | Under the conditions of this protocol,<br>the test article did not elicit a<br>sensitization response. |
| Intracutaneous<br>Reactivity | Intracutaneous Reactivity in<br>Rabbit, ISO 10993-10 | Polar Extract<br>Under the conditions of the study,<br>there was no erythema and no edema<br>from the test extract injected<br>intracutaneously into the rabbits. |
| Test | Test Method/ Model | Results |
| | | Non-polar Extract<br>Under the conditions of the study,<br>there was no to very slight erythema or<br>edema from the extract injected<br>intracutaneously into rabbits. |
| Acute Systemic<br>Toxicity | Acute Systemic Toxicity in<br>Mice, ISO 10993-11 | No mortality or evidence of systemic<br>toxicity. |
| Genotoxicity | Bacterial Reverse Mutagenic<br>Study, ISO 10993-3 | Non-mutagenic to Salmonella<br><i>typhimurium</i> (Test Strains: TA98,<br>TA1535, and TA1537) and to<br><i>Escherichia coli</i> (Test Strain WP2uvrA) |
| Genotoxicity | Mouse Lymphoma Assay, ISO<br>10993-3 | The test article extracts did not cause<br>a two-fold or greater increase in the<br>mean mutant frequency of the<br>L5178Y/TK+/- cell line either in the<br>presence or absence of metabolic<br>activation. The test article is<br>considered non-mutagenic. |
| Pyrogenicity | USP (151) Pyrogen Study –<br>Material Mediated | The test article was judged as<br>nonpyrogenic. |
| Muscle<br>Implantation | Muscle Implantation Study in<br>Rabbits, 2 weeks, ISO 10993-6 | The macroscopic reaction was not<br>significant as compared to the sponsor<br>provided control article or to the<br>negative control article.<br>Microscopically, the test article was<br>classified as a non-irritant as<br>compared to the sponsor provided<br>control article and as a slight irritant as<br>compared to the negative control<br>article. |
| Subchronic<br>Toxicity | Subcutaneous Implantation in<br>Rabbits, ISO 10993-11 | Under the conditions of the 13-week<br>study, there was no evidence of<br>systemic toxicity or adverse findings<br>attributed to the test article when<br>compared with the predicate control. |
| Chronic<br>Toxicity | Subcutaneous Implantation in<br>Rabbits, ISO 10993-11 | Under the conditions of the 26-week<br>subcutaneous implant toxicity study,<br>there was no evidence of systemic<br>toxicity or adverse findings attributed to<br>the test article. Based upon the<br>differences between the Test Article<br>Group Average Irritation Scores, the<br>test article was considered non-irritant<br>when compared to the predicate<br>control at 26 weeks. |
{6}------------------------------------------------
An animal efficacy study utilizing a rabbit dural defect repair model was conducted to evaluate the device as compared to its predicate device with regards to dura repair and resorption. No clinical tests were performed on the product; however clinical history of the predicate device was referenced in the submission.
{7}------------------------------------------------
A Viral inactivation study was performed to ensure the viral safety of the product.
# 8. Conclusion of Non-clinical Studies
The predicate device was cleared based on the results of non-clinical data. The subject device demonstrates equivalence to the predicate device.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.