← Product Code [DQX](/productcode/DQX) · K150802

# Heavy Double Flexible Tipped Wire Guide (K150802)

_Cook Incorporated · DQX · May 28, 2015 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K150802

## Device Facts

- **Applicant:** Cook Incorporated
- **Product Code:** [DQX](/productcode/DQX.md)
- **Decision Date:** May 28, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1330
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The Heavy Double Flexible Tipped Wire Guide is intended to facilitate placement of devices used during diagnostic and interventional procedures.

## Device Story

The Heavy Double Flexible Tipped Wire Guide is a 0.032-inch diameter, 60 cm long stainless steel wire guide; features stainless steel coils surrounding a stainless steel mandril; includes two flexible tips and four sets of centimeter markings etched at 10 cm intervals. Used by clinicians during diagnostic and interventional procedures to facilitate the placement of other medical devices. The device is supplied sterile in a Tyvek peel pouch. It functions as a mechanical guide wire to navigate vasculature or anatomical pathways, allowing the physician to track catheters or other interventional tools to a target site.

## Clinical Evidence

Bench testing only. Testing included fracture testing, tensile testing (per ISO 11070:1999), corrosion resistance, flex testing, and acute performance evaluation. Packaging integrity was verified via bubble leak and seal strength testing. Biocompatibility testing was conducted per ISO 10993-1:2009 for external communicating devices in contact with circulating blood (≤24 hours), including cytotoxicity, sensitization, irritation/intracutaneous reactivity, systemic toxicity, and hemocompatibility. All predetermined acceptance criteria were met.

## Technological Characteristics

Materials: stainless steel coils and mandril. Dimensions: 0.032 inch diameter, 60 cm length. Features: two flexible tips, etched centimeter markings. Sterilization: sterile. Biocompatibility: ISO 10993-1:2009 compliant. Mechanical testing: ISO 11070:1999 compliant.

## Regulatory Identification

A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.

## Special Controls

*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.

## Predicate Devices

- Cardiovascular Spring Guides ([K770977](/device/K770977.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol. The logo is black and white.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 28, 2015

Cook, Inc. Ms. Julia Ferguson Regulatory Affairs Specialist 750 Daniels Way Bloomington, IN 47404

Re: K150802

Trade/Device Name: Heavy Double Flexible Tipped Wire Guide Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DQX Dated: March 25, 2015 Received: March 26, 2015

Dear Ms. Ferguson:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in

{1}------------------------------------------------

the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Kenneth J. Cavanaugh -S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K150802

Device Name

Heavy Double Flexible Tipped Wire Guide

Indications for Use (Describe)

The Heavy Double Flexible Tipped Wire Guide is intended to facilitate placement of devices used during diagnostic and interventional procedures.

Type of Use (Select one or both, as applicable)

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Cook Medical. The logo consists of the word "COOK" in large, white, sans-serif font on a red background. Below the word "COOK" is the word "MEDICAL" in a smaller, white, sans-serif font. The logo is simple and clean, and the red and white color scheme is eye-catching.

## Heavy Double Flexible Tipped Wire Guide 21 CFR §870.1310 Date Prepared: May 26, 2015

### Submitted By:

| Applicant:            | Cook Incorporated                                             |
|-----------------------|---------------------------------------------------------------|
| Contact:              | Julia Ferguson                                                |
| Applicant Address:    | Cook Incorporated<br>750 Daniels Way<br>Bloomington, IN 47404 |
| Contact Phone Number: | (812) 335-3575 x104554                                        |
| Contact Fax Number:   | (812) 332-0281                                                |

### Device Information:

| Trade Name:          | Heavy Double Flexible Tipped Wire Guide |
|----------------------|-----------------------------------------|
| Common Name:         | Catheter Wire Guide                     |
| Classification Name: | Wire, Guide, Catheter                   |
| Regulation:          | 21 CFR §870.1330                        |
| Product Code:        | DQX                                     |

## Predicate Device:

The predicate device is the Cardiovascular Spring Guides (Lake Region Mfg., Inc., K770977).

### Device Description:

The Heavy Double Flexible Tipped Wire Guide is constructed of stainless steel coils surrounding a stainless steel mandril. It is a 0.032 inch diameter, 60 cm long wire guide with two flexible tips and four sets of centimeter markings etched at 10 cm intervals along the wire shaft. The device is supplied sterile, packaged within a Tyvek peel pouch.

## Intended Use:

The Heavy Double Flexible Tipped Wire Guide is intended to facilitate placement of devices used during diagnostic and interventional procedures.

### Comparison to Predicates:

Cook Incorporated's Heavy Double Flexible Tipped Wire Guide is substantially equivalent to the predicate device, the Cardiovascular Spring Guides (K770977), in that these devices have identical designs, methods of construction and operation, and indications for use.

{4}------------------------------------------------

# Technological Characteristics:

The following tests have been conducted to ensure reliable design and performance under the specified design requirements.

- Fracture Testing: Testing verified that there will be no signs of fracture after . manipulation, excluding the region of fixation and the first turn. The predetermined acceptance criterion was met.
- Tensile Testing: Testing verified that there will be no sign of loosening of any unions . after an appropriate level of force defined by ISO 11070:1999 is applied. The predetermined acceptance criterion was met.
- Corrosion Resistance: Testing verified that there will be no signs of corrosion that will . affect the functional performance or biocompatibility of the wire guide. The predetermined acceptance criterion was met.
- Flex Testing: Testing verified that there will be no signs of defects or damage when the . wire guide is subjected to repeated flexing. The predetermined acceptance criterion was met.
- . Acute Performance Evaluation: The testing confirmed that the wire guide is rated as acceptable for normal clinical practice for the performance parameters evaluated. The predetermined acceptance criterion was met.
- Bubble Leak (Packaging): The package shall show no signs of pin holes or imperfect . seals indicated by bubbles created from the pressurized air leaking into the surrounding solution. The predetermined acceptance criterion was met.
- Seal Strength (Packaging): All specimens must have a mean seal force that exceeds 0.50 . lbf. The predetermined acceptance criterion was met.
- Biocompatibility Testing: Per ISO 10993-1:2009, the subject devices are categorized as . external communicating devices in contact with circulating blood for a limited (≤24 hours) duration. The following tests were completed and the biocompatibility was deemed acceptable: cytotoxicity, sensitization, irritation/intracutaneous reactivity, systemic toxicity, and hemocompatibility.

# Conclusion:

The results of these tests support a conclusion that the Cook Incorporated Heavy Double Flexible Tipped Wire Guide met the design input requirements based on its intended use. The results also support the conclusion that these devices do not raise new questions of safety or effectiveness and are substantially equivalent to the predicate device, the Cardiovascular Spring Guides (K770977).

---

**Source:** [https://fda.innolitics.com/device/K150802](https://fda.innolitics.com/device/K150802)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
