Puracyn(R) Plus Duo-Care(TM) Antimicrobial Wound & Skin Hydrogel, Puracyn(R) Plus Antimicrobial Hydrogel Professional Formula

K150799 · Innovacyn, Inc. · FRO · Nov 19, 2015 · SU

Device Facts

Record IDK150799
Device NamePuracyn(R) Plus Duo-Care(TM) Antimicrobial Wound & Skin Hydrogel, Puracyn(R) Plus Antimicrobial Hydrogel Professional Formula
ApplicantInnovacyn, Inc.
Product CodeFRO · SU
Decision DateNov 19, 2015
DecisionSESE
Submission TypeTraditional
Device ClassClass U
AttributesTherapeutic

Indications for Use

For Over-the-Counter Use: Puracyn® Plus Duo-Care™ Antimicrobial Wound & Skin Hydrogel is intended for OTC use to relieve itch and pain from minor skin irritations, minor cuts, exit sites, minor lacerations, minor abrasions and minor burns, including sunburns. Hydrogel is also intended to moisten and lubricate absorbent wound dressings and moisten the wound bed. A moist wound and skin environment facilitates autolytic debridement and is beneficial to wound management and the healing process. For Professional Use: Puracyn® Plus Antimicrobial Hydrogel Professional Formula is intended for use by healthcare professionals to moisten the wound bed and facilitate autolytic debridement of acute and chronic dermal lesions, as indicated below. Puracyn® Plus Antimicrobial Hydrogel Professional Formula relieves itch and pain associated with dermal irritation, sores, injuries and ulcers of dermal tissue in addition to moistening and lubricating absorbent wound dressings. Puracyn® Plus Antimicrobial Hydrogel Professional Formula is indicated for the management of partial or full thickness wounds such as 1st and 2nd degree burns, stage I – IV pressure ulcers, diabetic and stasis ulcers, abrasions and skin irritations, surgical wounds (donor and graft sites, incisions), trauma wounds, and various dermatoses including atopic dermatitis.

Device Story

Topical hydrogel dressing; aqueous-based; pH balanced. Contains hypochlorous acid and sodium hypochlorite as antimicrobial preservatives. Applied topically to skin and wound areas by patients (OTC) or healthcare professionals. Maintains moist wound environment; facilitates autolytic debridement; relieves itch and pain. Inhibits microbial growth within the hydrogel. Supplied in various packaging configurations.

Clinical Evidence

Bench testing only. Evidence includes stability studies, biocompatibility testing per ISO 10993-1, and antimicrobial/preservative effectiveness testing per modified USP <51>. No clinical data presented.

Technological Characteristics

Aqueous-based topical hydrogel; pH balanced. Ingredients include hypochlorous acid, sodium hypochlorite, and magnesium lithium sodium silicate (nanoclay) as a viscosity-increasing agent. Antimicrobial preservative effectiveness demonstrated against various pathogens (e.g., MRSA, VRE, E. coli, P. aeruginosa). Biocompatible per ISO 10993-1.

Indications for Use

Indicated for OTC use for minor skin irritations, cuts, exit sites, lacerations, abrasions, and burns (including sunburn). Professional use indicated for acute and chronic dermal lesions, including 1st/2nd degree burns, stage I-IV pressure ulcers, diabetic/stasis ulcers, surgical wounds, trauma wounds, and dermatoses (e.g., atopic dermatitis).

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure, with three profiles overlapping to create a sense of depth and connection. The profiles are rendered in black and white. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the figure. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 November 19, 2015 Innovacyn Incorporated % Ms. Roshana Ahmed, MA, RAC OptumInsight Incorporated 4 Innovative Drive Dundas. Ontario L9H 7P3 Canada Re: K150799 Trade/Device Name: Puracyn® Plus Duo-Care™ Antimicrobial Wound & Skin Hydrogel Puracyn® Plus Antimicrobial Hydrogel Professional Formula Regulatory Class: Unclassified Product Code: FRO Dated: October 21, 2015 Received: October 21, 2015 Dear Ms. Ahmed: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {1}------------------------------------------------ device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # David Krause -S - Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K150799 #### Device Name Puracyn® Plus Duo-Care™ Antimicrobial Wound & Skin Hydrogel Puracyn® Plus Antimicrobial Hydrogel Professional Formula #### Indications for Use (Describe) For Over-the-Counter Use: Puracyn® Plus Duo-Care™ Antimicrobial Wound & Skin Hydrogel is intended for OTC use to relieve itch and pain from minor skin irritations, minor lacerations, minor abrasions and minor burns, including sunburns. Hydrogel is also intended to moisten and lubricate absorbent wound dressings and moisten the wound bed. A moist wound and skin environment facilitates autolytic debridement and is beneficial to wound management and the healing process. For Professional Use: Puracyn® Plus Antimicrobial Hydrogel Professional Formula is intended for use by healthcare professionals to moisten the wound bed and facilitate autolytic debridement of acute and chronic dermal lesions, as indicated below. Puracyn® Plus Antimicrobial Hydrogel Professional Formula relieves itch and pain associated with dermal irritation, sores, injuries and ulcers of dermal tissue in addition to moistening absorbent wound dressings. Puracyn® Plus Antimicrobial Hydrogel Professional Formula is indicated for the management of partial or full thickness wounds such as 1st and 2nd degree burns, stage I - IV pressure ulcers, diabetic and stasions and skin irritations, surgical wounds (donor and graft sites, incisions), trauma wounds, and various dermatoses including atopic dermatitis. Type of Use (Select one or both, as applicable) | <div> <span> </span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | |---------------------------------------------------------------------------------------| | <div> <span> </span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> | #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 3. 510(k) Summary | Manufacturer Name: | Innovacyn, Inc. | |--------------------|----------------------------------------------| | Address: | 3546 N. Riverside Avenue<br>Rialto, CA 92377 | | Contact Name: | Victor Torcat | | Title: | Vice President, Regulatory and Compliance | | Phone Number: | 866-318-3116 | | Fax Number: | 909-428-1947 | | Date Prepared: | October 15, 2015 | | Device Proprietary Names: | Puracyn® Plus Duo-Care™ Antimicrobial Wound & Skin<br>Hydrogel;<br>Puracyn® Plus Antimicrobial Hydrogel Professional<br>Formula | |---------------------------|---------------------------------------------------------------------------------------------------------------------------------| | Common or Usual Name: | Hydrogel Wound Dressing | | Classification Name: | Dressing, Wound, Drug | | Classification Code: | FRO | | Regulation Number: | N/A | | Device Classification | Unclassified | #### Predicate Devices: Substantial equivalence is claimed to the following devices as related to intended use, design, and material characteristics: - Microcyn™ Skin and Wound Hydrogel, Oculus Innovative Sciences, Inc., . K093585 - Anasept™ Antimicrobial Skin and Wound Gel, Anacapa Technologies, Inc., . K073547 - Puricore Wound Hydrogel Spray Dressing, Puricore Inc., K141863 ● - Epicyn™ HydroGel, Oculus Innovative Sciences, Inc., K102945 . Innovacyn's Puracyn Plus Skin and Wound Care Solutions (K133542) is used as a reference device. #### Description of the Device Puracyn® Plus Duo-Care™ Antimicrobial Wound & Skin Hydrogel and the Puracyn® Plus Antimicrobial Hydrogel Professional Formula are opaque and pH balanced hydrogel dressings that are topically applied to skin and wound areas. The gel provides relief from {4}------------------------------------------------ itching and pain, and helps maintain a moist wound environment to assist with the wound healing process. The hydrogel dressings will be supplied in various packaging configurations. Puracyn® Plus Duo-Care™ Antimicrobial Wound & Skin Hydrogel and the Puracyn® Plus Antimicrobial Hydrogel Professional Formula contain hypochlorous acid, a known antimicrobial, which serves as a preservative to inhibit the growth of microorganisms in the hydrogel. ## Intended Use/Indications for Use For Over-the-Counter Use: Puracyn® Plus Duo-Care™ Antimicrobial Wound & Skin Hydrogel is intended for OTC use to relieve itch and pain from minor skin irritations. minor cuts, exit sites, minor lacerations, minor abrasions and minor burns, including sunburns. Hydrogel is also intended to moisten and lubricate absorbent wound dressings and moisten the wound bed. A moist wound and skin environment facilitates autolytic debridement and is beneficial to wound management and the healing process. For Professional Use: Puracyn® Plus Antimicrobial Hydrogel Professional Formula is intended for use by healthcare professionals to moisten the wound bed and facilitate autolytic debridement of acute and chronic dermal lesions, as indicated below. Puracyn® Plus Antimicrobial Hydrogel Professional Formula relieves itch and pain associated with dermal irritation, sores, injuries and ulcers of dermal tissue in addition to moistening and lubricating absorbent wound dressings. Puracyn® Plus Antimicrobial Hydrogel Professional Formula is indicated for the management of partial or full thickness wounds such as 1st and 2nd degree burns, stage I – IV pressure ulcers, diabetic and stasis ulcers, abrasions and skin irritations, surgical wounds (donor and graft sites, incisions), trauma wounds, and various dermatoses including atopic dermatitis. #### Technological Characteristics Puracyn® Plus Duo-Care™ Antimicrobial Wound & Skin Hydrogel and the Puracyn® Plus Antimicrobial Hydrogel Professional Formula are opaque and pH balanced hydrogel dressings that are topically applied to skin and wound areas. The gel provides relief from itching and pain, and helps maintain a moist wound environment to assist with the wound healing process. Puracyn® Plus hydrogels contain hypochlorous acid, a known antimicrobial preservative, which is shown to inhibit the growth of microorganisms such as Candida albicans, Aspergillus brasiliensis, acinetobacter albumin, Vancomycinresistant Enterococcus (VRE), Escherichia coli, Pseudomonas aeruginosa, Staphylococcus epidermidis, and Methicillin-resistant Staphylococcus aureus (MRSA) in the hydrogel. A comparison of technological characteristics between the subject and predicate devices is provided in the table below. {5}------------------------------------------------ | | Puracyn® Plus Duo-<br>Care™ Antimicrobial<br>Wound & Skin<br>Hydrogel;<br>Puracyn® Plus<br>Antimicrobial<br>Hydrogel<br>Professional Formula | Microcyn™ Skin and<br>Wound HydroGel<br>(K093585) | Anasept™<br>Antimicrobial Skin<br>and Wound Gel<br>(K073547) | Puricore Wound<br>Hydrogel<br>(K141863) | Epicyn™ HydroGel<br>(K102945) | |---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------|--------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------| | Technology | Aqueous based topical<br>hydrogel maintaining a<br>moist wound<br>environment and<br>encouraging autolytic<br>debridement | Topical Hydrogel<br>encouraging autolytic<br>debridement | Topical Hydrogel | Aqueous based topical<br>hydrogel maintaining a<br>moist wound<br>environment and<br>encouraging autolytic<br>debridement | Aqueous based topical<br>hydrogel maintaining a<br>moist wound<br>environment and<br>encouraging autolytic<br>debridement | | Preservatives | Hypochlorous acid<br>Sodium hypochlorite | Hypochlorous acid<br>Sodium hypochlorite | Sodium hypochlorite | Hypochlorous acid<br>Sodium hypochlorite | Hypochlorous acid<br>Sodium hypochlorite | | Antimicrobial<br>Effectiveness | Yes | Yes | Yes | Yes | Yes | | Viscosity<br>Increasing<br>Ingredient | Magnesium lithium<br>sodium silicate<br>(nanoclay) | Magnesium lithium<br>fluorosilicate<br>(nanoclay) | Sodium magnesium<br>silicate (nanoclay) | Sodium magnesium<br>fluorosilicate<br>(nanoclay) | Magnesium lithium<br>sodium silicate<br>(nanoclay) | | Biocompatible | Yes | Yes | Yes | Yes | Yes | {6}------------------------------------------------ There are no differences between the subject and predicate devices with respect to intended use. The subject devices are similar in technology to the predicate devices in that the products are topical hydrogels which encourage autolytic debridement and protect the wound against contamination. The products contain viscosity enhancing agents which impart hydrogel characteristics, as well as hypochlorous acid and/or sodium hypochlorite as preservatives. Expanded testing consistent with USP <51> supports that the Free Available Chlorine (FAC) inhibits microbial contamination within the product. As supported by biocompatibility testing according to ISO10093-1, hydrogels are biologically safe. ## Non-Clinical Testing The following studies have been performed to support the safety and effectiveness of the product: - . Stability Studies - Biocompatibility studies in accordance with ISO10993-1 ● - Antimicrobial/Preservative Effectiveness Testing per modified USP <51> ● #### Conclusion Based on the analysis presented above and the results of performance testing, the subject devices are as safe and effective as the predicate devices. Therefore, it is concluded that the subject devices are substantially equivalent to the identified predicate devices. ------------------This space intentionally left blank-------------------------------------------------------------------------------------------------------------------------
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...