← Product Code [GCJ](/productcode/GCJ) · K150781

# Tissue Removal Pouch (K150781)

_Boehringer Laboratories, LLC · GCJ · Jun 19, 2015 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K150781

## Device Facts

- **Applicant:** Boehringer Laboratories, LLC
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** Jun 19, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The Tissue Removal Pouch is indicated for the removal of tissue during surgical procedures and for the containment of tissue during extracorporeal manual morcellation. The Tissue Removal Pouch is contraindicated for laparoscopic power morcellation during gynecologic procedures. The Tissue Removal Pouch is contraindicated for use with powered cutting devices (e.g., power morcellators, electrosurgical and laser instruments), and when, in the judgment of the physician, use of such a device would be contrary to the best interest of the patient.

## Device Story

Tissue Removal Pouch is a laparoscopic retrieval device consisting of a polymer film bag, a two-piece flip ring, and a guard. Used during laparoscopic surgery; device inserted into peritoneal space via existing port site. Resected tissue placed inside bag for isolation; bag opening exteriorized. Flip ring halves attached to bag opening; bag rolled around ring to draw tissue toward port site. Tissue removed en bloc or via extracorporeal manual morcellation; guard protects bag during manual morcellation. Device is single-use and sterile. Benefits include tissue containment and protection during removal, preventing potential spread of tissue during manual morcellation. Operated by surgeons in clinical/OR settings.

## Clinical Evidence

Bench testing only. Testing included film impermeability, bubble leak, weld strength, puncture resistance, burst testing, opening ring pull testing, guard cut resistance, simulated use testing, biocompatibility testing, dye penetration, and viral penetration testing to demonstrate barrier adequacy.

## Technological Characteristics

Device consists of a polymer film bag, a two-piece flip ring, and a guard. Single-use, sterile. Mechanical retrieval system. Biocompatible materials. No electronic components or software.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Applied Medical Tissue Containment System ([K142427](/device/K142427.md))
- Applied Medical Specimen Retrieval System ([K060051](/device/K060051.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol of medicine, with three human profiles incorporated into the design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 19, 2015

Boehringer Laboratories, LLC Mr. Christopher Radl Engineering and Product Development Manager 300 Thoms Drive Phoenixville, Pennsylvania 19460

Re: K150781

Trade/Device Name: Tissue Removal Pouch Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: GCJ Dated: June 10, 2015 Received: June 10, 2015

Dear Mr. Radl:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

# Jennifer R. Stevenson -S

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known)

K150781

Device Name

Tissue Removal Pouch

Indications for Use (Describe)

The Tissue Removal Pouch is indicated for the removal of tissue during surgical procedures and for the containment of tissue during extracorporeal manual morcellation.

Contraindications:

The Tissue Removal Pouch is contraindicated for laparoscopic power morcellation during gynecologic procedures.

The Tissue Removal Pouch is contraindicated for use with powered cutting devices (e.g., power morcellators, electrosurgical and laser instruments), and when, in the judgment of the physician, use of such a device would be contrary to the best interest of the patient.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary

Prepared: June 9, 2015

#### APPLICANT INFORMATION:

Name: Boehringer Laboratories. LLC Address: 300 Thoms Drive Phoenixville, PA 19460 Phone: 610-278-0900 Fax: 610-278-0907 Contact: Christopher Radl; Engineering, Product Development Manager

Signature: Date:

d June 9, 2015

## TRADE NAME:

Tissue Removal Pouch

## COMMON NAME:

Laparoscopic Retrieval Device

#### Device Classification:

Accessory to Endoscope, Class II Product Code: GCJ Regulation: 876.1500 Endoscope and Accessories Classification Panel: General & Plastic Surgery

#### PREDICATE DEVICES:

| Primary Predicate:    | Applied Medical Tissue Containment System | K142427 |
|-----------------------|-------------------------------------------|---------|
| Additional Predicate: | Applied Medical Specimen Retrieval System | K060051 |

#### Device Description:

The Tissue Removal Pouch includes a polymer film bag, a flip ring packaged in two halves, and a guard. It is provided sterile and is for single patient use.

The bag has a single large opening held open by a thin, opening ring. During laparoscopic surgery, the bag is placed within the peritoneal space via an existing port site. Tissue that has been resected is placed within the bag for isolation and removal from the patient. The bag opening is then exteriorized. The two halves of the flip ring are attached to each other and to the bag opening.

The bag is rolled around the flip ring by inverting the flip ring in order to draw the resected tissue toward the port site. The resected tissue is then removed en bloc (for smaller specimens) or by extracorporeal manual morcellation (for larger specimens). The guard may be used to protect the bag during manual morcellation. The bag is removed following the removal of sufficient tissue to be pulled out through the port site.

#### INTENDED USE:

The Tissue Removal Pouch is indicated for the removal of tissue during surgical procedures and for the containment of tissue during extracorporeal manual morcellation.

The Tissue Removal Pouch is contraindicated for laparoscopic power morcellation during gynecologic procedures.

{4}------------------------------------------------

The Tissue Removal Pouch is contraindicated for use with powered cutting devices (e.g., power morcellators, electrosurgical and laser instruments), and when, in the physician, use of such a device would be contrary to the best interest of the patient.

## Comparison Summary:

The Tissue Removal Pouch and predicate devices are intended for tissue containment and removal during laparoscopic surgery. All devices are single use, and sterile, with a biocompatible polymer film bag portion, that has a single opening incorporating an opening mechanism. Furthermore, the bag portion of all devices is inserted into the peritoneal cavity of the patient, loaded with a specimen, and exteriorization. All devices provide some mechanism to assist in the removal of the bag and the specimen.

The following aspects involve slight differences between the Tissue Removal Pouch and one or both of the predicate devices:

- Mechanism of bag insertion
- . Mechanism of bag exteriorization
- Morcellation protection
- Bag volume

## Non-Clinical Testing:

The following non-clinical tests have been included with this submission:

- Film Impermeability Testing: ● Verifies film impermeability.
- . Bubble Leak Testing: Demonstrates that the bag of the Tissue Removal Pouch is leak free.
- . Weld Strenath Testing: Demonstrates adequate weld strength.
- . Puncture Resistance Testing: Demonstrates adequate resistance to puncture.
- . Burst Testina: Demonstrates adequate burst strength.
- . Opening Ring Pull Testing: Demonstrates adequate pull strength.
- . Guard Cut Resistance Testing: Demonstrates adequate resistance to cutting.
- . Simulated Use Testing Demonstrates that the Tissue Removal Pouch performs as intended.
- . Biocompatibility Testing Demonstrates that the device is biocompatible.
- Dye Penetration Testing . Demonstrates that the Tissue Removal Pouch is leak free.
- . Viral Penetration Testing Demonstrates that the Tissue Removal Pouch is and adeguate barrier to cellular migration.

## Conclusions:

As evidenced by the provided information, the Tissue Removal Pouch is as safe and effective as, and performs as well as, or better than, the legally marketed predicate devices.

---

**Source:** [https://fda.innolitics.com/device/K150781](https://fda.innolitics.com/device/K150781)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).