The MicroAire LipoFilter System is used in the aspiration, harvesting, filtering and transferring of autologous adipose tissue for aesthetic body contouring. If the harvested fat is to be re-implanted, the harvested fat is only to be used without any additional manipulation.
Device Story
Single-use, closed-loop tissue collection system; comprises medical-grade canister, vacuum port, collection port, tissue port, and lid. Used with standard liposuction aspiration pump. Tissue harvested from patient enters canister via collection port; physician removes waste materials via vacuum port by closing valve; remaining fatty tissue transferred to syringes via tissue port for autologous re-injection. Used in clinical settings by physicians for aesthetic body contouring. Benefits patient by providing sterile, filtered autologous fat for re-implantation.
Clinical Evidence
Bench testing only. Includes gamma radiation sterility validation (ISO 11137-1/2), shelf-life validation (ISO 11607-1/2), biocompatibility (ISO 10993-1), functionality testing, ISO 10079-1 testing, tubing connection strength, vacuum leak testing, adipocyte viability, and aging studies.
Technological Characteristics
Single-use, closed-loop canister system. Materials: medical-grade biocompatible plastics. Sterilization: Gamma radiation (SAL 10^-6). Volume: up to 2500mL. Connectivity: interfaces with standard liposuction aspiration pumps via tubing. Standards: ISO 11137-1/2, ISO 11607-1/2, ISO 10993-1, ISO 10079-1.
Indications for Use
Indicated for aspiration, harvesting, filtering, and transferring autologous adipose tissue for aesthetic body contouring in patients requiring fat re-implantation without additional manipulation.
Regulatory Classification
Identification
A suction lipoplasty system is a device intended for aesthetic body contouring. The device consists of a powered suction pump (containing a microbial filter on the exhaust and a microbial in-line filter in the connecting tubing between the collection bottle and the safety trap), collection bottle, cannula, and connecting tube. The microbial filters, tubing, collection bottle, and cannula must be capable of being changed between patients. The powered suction pump has a motor with a minimum of 1/3 horsepower, a variable vacuum range from 0 to 29.9 inches of mercury, vacuum control valves to regulate the vacuum with accompanying vacuum gauges, a single or double rotary vane (with or without oil), a single or double diaphragm, a single or double piston, and a safety trap.
Special Controls
*Classification.* Class II (special controls). Consensus standards and labeling restrictions.
Lipisystems Aquavage Model AV2000 & 1200 (K092284)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an emblem featuring a stylized depiction of a human figure, with three profiles merging to form a single, abstract shape. The emblem is meant to symbolize health and human services.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 22, 2015
MicroAire Surgical Instruments LLC Mr. Chris Spofford Regulatory Affairs and Compliance Officer 3590 Grand Forks Boulevard Charlottesville, Virginia 22911
Re: K150779
Trade/Device Name: Microaire LipoFilter System Regulation Number: 21 CFR 878.5040 Regulation Name: Suction lipoplasty system Regulatory Class: Class II Product Code: MUU Dated: March 25, 2015 Received: March 26, 2015
Dear Mr. Spofford:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{2}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
for
Sincerely yours,
# David Krause -S
- Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## Indications for Use
510(k) Number (if known) K150779
Device Name Microaire LipoFilter System
Indications for Use (Describe)
The MicroAire LipoFiter System is used in the aspiration, filtering and transferring of autologous adipose tissue for aesthetic body contouring. If the harvested fat is to be re-implanted fat is only to be used without any additional manipulation.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
- 1) Preparation Date: March 24, 2015
2) Submitted by: MicroAire Surgical Instruments LLC 3590 Grand Forks Boulevard Charlottesville, Virginia 22911 Owner/Operator #: 9004658
3) Contact Person/Prepared by:
Chris Spofford Regulatory Affairs and Compliance Officer MicroAire Surgical Instruments (434) 975-8344 Chriss@microaire.com
3590 Grand Forks Boulevard Charlottesville, Virginia 22911
- 4) Device Identification:
Trade Name: MicroAire LipoFilter System Common Name: Lipoplasty Suction System Classification: Suction lipoplasty system (21 CFR 878.5040, Product Code MUU)
- 5) Predicate Device:
Sound Surgical Technologies - Origins LipoHarvesting System (K120328) Medical Device Resource Corporation - Lipisystems Aquavage Model AV2000 & 1200 (K092284)
- 6) Device Description: The MicroAire LipoFilter System consists of a single use, closed loop tissue collection system comprised of a medical grade canister, vacuum port, collection port, tissue port and lid intended to be used with a standard liposuction aspiration system pump to collect fatty tissue for aesthetic body contouring. As the tissue is harvested from the patient, it enters the canister via the collection port in the canister lid. The physician removes unwanted waste materials from the collection system via the vacuum port by closing a valve. This process leaves fatty tissue that can be transferred to syringes via the tissue port for autologous fat re-injection.
- 7) Intended Use: The MicroAire LipoFilter System is used in the aspiration, harvesting, filtering and transferring of autologous adipose tissue for aesthetic body contouring. If the harvested fat is to be re-implanted, the harvested fat is only to be used without any additional manipulation.
{5}------------------------------------------------
## 8) Comparison to Predicate:
| FEATURES | DEVICES | | |
|------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | MicroAire LipoFilter System | Origins LipoHarvesting System | Medical Device Resource Corp<br>Lipisystems Aquavage |
| Device<br>Description: | The MicroAire LipoFilter<br>System consists of a single<br>use, closed loop tissue<br>collection system<br>comprised of a medical<br>grade canister, vacuum<br>port, collection port, tissue<br>port and lid intended to be<br>used with a standard<br>liposuction aspiration<br>system pump to collect<br>fatty tissue for aesthetic<br>body contouring. As the<br>tissue is harvested from the<br>patient, it enters the<br>canister via the collection<br>port in the canister lid. The<br>physician removes<br>unwanted waste materials<br>from the collection system<br>via the vacuum port by<br>closing a valve. This process<br>leaves fatty tissue that can<br>be transferred to syringes<br>via the tissue port for<br>autologous fat re-injection. | The Origins LipoHarvesting<br>System consists of a<br>reusable, closed loop<br>tissue collection system<br>comprised of a medical<br>grade canister, a drain<br>port and a lid intended to<br>be used with a standard<br>liposuction aspiration<br>pump to collect fatty<br>tissue for aesthetic body<br>contouring. As the tissue is<br>harvested from the<br>patient, it enters the<br>collection canister via a<br>port in the canister lid. The<br>physician removes<br>unwanted waste materials<br>from the collection system<br>via the drain port at the<br>base. This process leaves<br>fatty tissue that can be<br>transferred to syringes for<br>autologous fat re-<br>injection. | Device function:<br>1. Contains port<br>interfaces<br>between:<br>canister to<br>tubing and<br>aspirator to<br>canister.<br>2. Sterile tubing<br>to connect the<br>interfaces.<br>3. Funnel to<br>interface port<br>to tubing.<br>4. Syringe to<br>collect fat.<br>Device Design:<br>1. Contents<br>subjected to<br>sterility.<br>Material used:<br>Plastic canister,<br>syringe, silicone<br>tubing. |
| Picture of<br>Product | Image: MicroAire LipoFilter System | Image: Origins LipoHarvesting System | Image: Medical Device Resource Corp Lipisystems Aquavage |
| Intended Use: | The MicroAire LipoFilter<br>System is used in the<br>aspiration, harvesting,<br>filtering and transferring of<br>autologous adipose tissue<br>for aesthetic body<br>contouring. If the harvested<br>fat is to be re-implanted,<br>the harvested fat is only to<br>be used without any<br>additional manipulation. | The Origins LipoHarvesting<br>System is used in the<br>aspiration, harvesting,<br>filtering and transferring<br>of autologous tissue. If<br>harvested fat is to be re-<br>implanted, the harvested<br>fat is only to be used<br>without any additional<br>manipulation. | For use in aspirating<br>subcutaneous fatty<br>tissue including<br>autologous fat<br>collection. |
| Technology<br>Comparison: | The MicroAire LipoFilter<br>System employs the same<br>technological characteristics<br>as the predicate device. | The Origins LipoHarvesting<br>System employs the same<br>technological<br>characteristics as the<br>predicate devices. | The Lipisytems<br>Aquavage employs the<br>same technological<br>characteristics as the<br>predicate device. |
| Suction<br>Source(s) | Aspiration device | Aspiration device | Aspiration device |
| Volume Range | Up to 2500mL | 100 to 2,000 mL | 2000cc or 1200cc |
| Shipped sterile | Yes | No, sterilized by user prior to<br>use | Yes |
| Initial method | Gamma radiation | Not applicable | Gamma radiation |
| Sterility<br>assurance level | $10^{-6}$ | $10^{-6}$ | $10^{-6}$ |
| Disposable or<br>reusable | Single use, disposable | Reusable | Single use, disposable |
| Resterilization<br>method | Not applicable | Autoclave | Not applicable |
## Table 1: Substantial Equivalence Comparison Chart
{6}------------------------------------------------
## 9) Pre-Clinical Testing
The MicroAire LipoFilter System has been tested in various ways to demonstrate equivalence to the predicate devices including Gamma Radiation validation, shelf life validation, biocompatibility, and various performance testing. Specifics include the following:
{7}------------------------------------------------
Validated for sterility using gamma radiation. Test results and analyses indicate that the MicroAire LipoFilter System sterility complies in accordance with ISO 11137-1:2006 and ISO 11137-2:2006.
Validated for a 2 year shelf life. Test results and analyses indicate that the MicroAire LipoFilter System packaging complies in accordance with ISO 11607-1: 2006 and ISO 11607-2: 2006.
The MicroAire LipoFilter System has patient contact materials and is made from medical grade biocompatible materials. Test results and analyses indicate that the MicroAire LipoFilter System materials' complies in accordance with ISO 10993-1: 2009.
The MicroAire LipoFilter System was tested for performance in accordance with its predetermined specifications on file at MicroAire which includes functionality testing, ISO 10079-1 testing, tubing connection strength testing, vacuum leak testing, adipocyte viability testing, and Aging Studies.
#### 10) Conclusion:
The MicroAire LipoFilter System uses decantation to separate the tissue which is the same method as the two identified predicate devices. The MicroAire LipoFilter System does not have any new intended uses as compared to the two predicate devices. The MicroAire LipoFilter System has a slightly larger volume (2500 mL) whereas the predicate devices have a volume of up to 2000 mL which is insignificant given the intended use of the devices. The MicroAire LipoFilter System is provided sterile whereas the predicate devices are either provided sterile and/or sterilized by the user. The results of these activities demonstrate that the MicroAire LipoFilter System is considered substantially equivalent to the two predicate devices.
The results of the evaluation above demonstrate that the MicroAire LipoFilter System is considered substantially equivalent to the two predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.