Snap-Off Self-Compressive Screw

K150772 · Neosteo · HWC · May 8, 2015 · Orthopedic

Device Facts

Record IDK150772
Device NameSnap-Off Self-Compressive Screw
ApplicantNeosteo
Product CodeHWC · Orthopedic
Decision DateMay 8, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3040
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Snap-Off Self-Compressive Screws are intended for the fixation of bone fractures and for bone reconstruction in forefoot surgery.

Device Story

Snap-Off Self-Compressive Screws are metallic bone fixation fasteners used in forefoot surgery. Implants consist of titanium alloy screws available in various lengths. Device functions via threaded design to provide bone compression during fixation. Used by orthopedic surgeons in clinical settings. Output is physical stabilization of bone fragments. Benefits include mechanical fixation of fractures and reconstruction sites.

Clinical Evidence

No clinical data. Bench testing only. Mechanical testing included resistance to torsion and pull-out strength per ASTM F543.

Technological Characteristics

Titanium alloy (ASTM F136). Threaded screw design for bone compression. Mechanical performance validated via ASTM F543 Annex 1 (torsion) and Annex 3 (pull-out strength).

Indications for Use

Indicated for patients requiring bone fracture fixation or bone reconstruction during forefoot surgery.

Regulatory Classification

Identification

A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three heads in profile, facing to the right. The bird is positioned to the right of a circular text element that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a sans-serif font. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 Neosteo % Mr. J.D. Webb The Orthomedix Group, Inc. 1001 Oakwood Blvd Round Rock, Texas 78681 Re: K150772 Trade/Device Name: Snap-off Self-compressive Screw Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC Dated: March 13, 2015 Received: April 1, 2015 Dear Mr. J. D. Webb: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in May 8, 2015 {1}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Mark N. Melkerson -S Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use K150772 Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page. 510(k) Number (if known) Device Name Snap-Off Self-Compressive Screws Indications for Use (Describe) The Snap-Off Self-Compressive Screws are intended for the fixation of bone reconstruction in forefoot surgery. Type of Use (Select one or both, as applicable) X | Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) ### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. ### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for Neosteo. The logo consists of an orange spiral design on the left and the word "neosteo" in orange on the right. Below the word "neosteo" is the phrase "EFFICIENT MOBILITY" in a smaller font size. #### l. SUBMITTER'S INFORMATION ### A. 510(k) Owner NEOSTEO Mallève 2A 1 Boulevard Jean Moulin 44 100 Nantes, France Tel: +33 (0)2 36 56 96 70 Fax: +33 (0)2 51 70 61 34 ## B. Contact Person JD Webb The Orthomedix Group, Inc. 1001 Oakwood Blvd Round Rock, TX 78681 Tel: 512 388 0199 Fax: 512 692 3699 Email: jdwebb@orthomedix.net ### C. Date of Preparation of the 510(k) Summary 19th March 2015 {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Neosteo. The logo consists of an orange spiral design on the left, with the word "neosteo" in orange to the right of the spiral. Below the word "neosteo" is the phrase "EFFICIENT MOBILITY" in a smaller font, also in orange. The spiral design appears to be made up of a series of dots connected by lines. #### DEVICE IDENTIFICATION II. | | "Screw" type devices | |----------------------------|------------------------------------------------| | Trade or proprietary name | Snap-Off Self-Compressive Screw | | Common or usual name | Self-compressive screw | | Classification regulation | 21 CFR 888.3040 | | Proposed Regulatory Class | Class II | | Panel | 87 "Orthopedic" | | Product code | HWC | | Primary Predicate Device | Spin® Snap-Off Screw (K991477) from NewDeal SA | | Secondary Predicate Device | Self-Compressive Screws (K131471) from Neosteo | #### III. DEVICE DESCRIPTION The Snap-Off Self-Compressive Screw consists of screws available in several lengths. All the implants are made of titanium alloy. The fixation is provided thanks to the threading of the screw, which allows compression. ## A. Materials Snap-Off Self-Compressive Screw: Titanium alloy per ASTM F136 {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for Neosteo. The logo consists of an orange spiral design on the left and the word "neosteo" in orange on the right. Below the word "neosteo" is the phrase "EFFICIENT MOBILITY" in a smaller font. # 510(k) Summary #### IV. INTENDED USE The Snap-Off Self-Compressive Screws are intended for the fixation of bone fractures and for bone reconstruction in forefoot surgery. ### SUMMARY OF OF TECHNOLOGICAL CHARACTERISTICS / SUBSTANTIAL V. EQUIVALENCE The Snap-Off Self-Compressive Screw is substantially equivalent to the predicate devices in terms of intended use, design, materials used, mechanical safety and performances. #### VI. NON-CLINICAL TEST SUMMARY Snap-Off Self-Compressive Screw: The following mechanical tests were performed: - Resistance to torsion according to ASTM F543 Annex 1 o - Pull out strength according to ASTM F543 Annex 3 o The results of these testing indicate that the current Snap-Off Self-Compressive Screw is equivalent to predicate devices. #### VII. CLINICAL TEST SUMMARY No clinical studies were performed. #### VIII. CONCLUSIONS NON-CLINICAL AND CLINICAL NEOSTEO considers the current Snap-Off Self-Compressive Screw to be equivalent to the predicate devices listed above. This conclusion is based on the devices' similarities in principles of operation, technology, materials and indications for use.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...