VitreaView is a medical image viewing and information distribution application that provides access, through the internet and within the enterprise to multi-modality softcopy medical images, reports, and other patient-related information. This data is hosted within disparate archives and repositories for diagnosis, review, communication, and reporting of DICOM and non-DICOM data. Display monitors used for reading medical images for diagnostic purposes must comply with the applicable regulatory approvals and quality control requirements for their use and maintenance. Lossy compressed mammography images and digitized film screen images must not be reviewed for primary image interpretations. When accessing VitreaView from a mobile device, images viewed are for informational purposes only and not intended for diagnostic use.
Device Story
VitreaView is a cross-browser, zero-footprint universal medical image viewer; enables clinicians to access multi-modality DICOM/non-DICOM images, reports, and patient data from disparate archives. Operates via web browser without client-side installation; integrates with EHR, EMR, HIE, and PHR systems. Supports diagnostic-quality review, multi-dimensional visualization, and treatment planning. Features include markup tools, comparative review, volume rendering, and cine playback. Used by physicians and clinicians in clinical environments; mobile access restricted to informational use. Healthcare providers use the viewer to access images in context with patient records, facilitating collaboration and clinical decision-making. Benefits include improved access to imaging data and streamlined diagnostic workflows.
Clinical Evidence
No clinical studies were required. Evidence consists of bench testing, software verification (performance and safety), and validation (simulated usability testing). External validation by radiologists confirmed diagnostic quality of images based on brightness, sharpness, and artifact assessment.
Technological Characteristics
Zero-footprint, web-based software; HTML/JavaScript. Supports DICOM/non-DICOM data. Features: MPR, 3D volume rendering, annotation, measurement, and linking. Connectivity: Networked/Cloud-integrated. Security: HTTPS/SSL encryption, LDAP/Active Directory authentication. Standards: DICOM (PS 3.1-3.20), ISO 14971:2007 (Risk Management), IEC 62304:2006 (Software Life Cycle).
Indications for Use
Indicated for medical image viewing and information distribution for diagnosis, review, communication, and reporting of multi-modality DICOM and non-DICOM data. Not for primary interpretation of lossy compressed mammography or digitized film screen images. Mobile access is for informational purposes only, not diagnostic use.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name encircling a symbol. The symbol is a stylized representation of three human profiles facing right, stacked on top of each other. The profiles are connected by flowing lines, creating a sense of unity and movement.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 6, 2015
Vital Images, Inc. % Mr Parthiv Shah Sr. Regulatory Affairs Specialist 5850 Opus Parkway, Suite 300 MINNETONKA MN 55343-4414
Re: K150738 Trade/Device Name: VitreaView Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: March 18, 2015 Received: March 20, 2015
Dear Mr. Shah:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert A Ochs
Robert Ochs. Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K150738
Device Name VitreaView
# Indications for Use (Describe)
VitreaView is a medical image viewing and information distribution application that provides access, through the internet and within the enterprise to multi-modality softcopy medical images, reports, and other pation. This data is hosted within disparate archives and repositories for diagnosis, review, communication, and reporting of DICOM and non-DICOM data.
Display monitors used for reading medical images for diagnostic purposes must comply with the applicable regulatory approvals and quality control requirements for their use and maintenance.
Lossy compressed mammography images and digitized film screen images must not be reviewed for primary image interpretations.
When accessing VitreaView from a mobile device, images viewed are for informational purposes only and not intended for diagnostic use.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
# FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "VITAL" in a bold, sans-serif font. The letters are black, except for the dot above the "i", which is a slightly lighter shade. Below the word "VITAL" is the text "A Toshiba Medical Systems Group Company" in a smaller, red font.
# 510(k) Summary
This 510(k) summary is submitted in accordance with the requirements of 21 C.F.R. Part 807.92(c)
| Basis for the Submission: | Modifications to a legally marketed device which are not qualified for a<br>Special 510(k) notification |
|----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| Submitter: | Vital Images, Inc.<br>5850 Opus Parkway<br>Suite 300<br>Minnetonka, MN, 55343-4414 |
| Establishment<br>Registration: | 2134213 |
| Contact Person: | Parthiv Shah<br>Sr. Regulatory Affairs Specialist<br>Phone : 952-487-9574<br>Fax: 952-487-9510<br>E-mail: pshah@vitalimages.com |
| 510(k) Type: | Traditional |
| Summary Date: | March 30, 2015 |
| Device Trade Name: | VitreaView Software |
| Device Common<br>Name / Regulatory<br>Description: | Picture Archiving and Communications System |
| Device<br>Classification<br>Name: | System, Image Processing, Radiological |
| Regulation Number: | 21 CFR 892.2050 |
| Product Code: | LLZ |
| Regulatory<br>Classification: | Class II |
| Device Panel: | Radiology |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "VITAL" in black font. Underneath the word "VITAL" is the phrase "A Toshiba Medical Systems Group Company" in red font. The font of the word "VITAL" is sans-serif. The dot above the "i" in "VITAL" is a circle.
# Predicate Devices:
| Predicate Device(s) | Manufacturer | FDA 510(k) number |
|---------------------|--------------------|-------------------|
| VitreaView | Vital Images, Inc. | K122136 |
| ViTALConnect | Vital Images, Inc. | K071362 |
# Device Description:
VitreaView is a cross-browser, cross-platform, zero-footprint universal medical image viewer solution capable of displaying both DICOM and non-DICOM medical images for diagnosis, review, communication, and reporting.
VitreaView enables clinicians and other medical professionals to access patients' medical images with integrations into a variety of medical record systems, such as Electronic Health Record (EHR), Electronic Medical Record (EMR), Health Information Exchange (HIE), Personal Health Record (PHR), and image exchange systems. It supports the physician in medical image viewing and treatment planning.
VitreaView offers medical professionals a universal viewer for accessing multi-dimensional imaging data in context with reports from enterprise patient health information databases, fosters collaboration, and provides workflows and interfaces appropriate for referring physicians and clinicians. IT departments will not have to install client systems, due to the zero footprint/zero download nature of VitreaView. VitreaView offers scalability to add new users as demand grows, may be deployed in a virtualized environment, and is designed to be integrated with enterprise patient health information databases.
When accessing VitreaView from a mobile device, images viewed are for informational purposes only and not intended for diagnostic use.
# Intended Use / Indications for Use:
VitreaView is a medical image viewing and information distribution application that provides access, through the internet and within the enterprise to multi-modality softcopy medical images, reports, and other patient-related information. This data is hosted within disparate archives and repositories for diagnosis, review, communication, and reporting of DICOM and non-DICOM data.
Display monitors used for reading medical images for diagnostic purposes must comply with the applicable regulatory approvals and quality control requirements for their use and maintenance.
Lossy compressed mammography images and digitized film screen images must not be reviewed for primary image interpretations.
When accessing VitreaView from a mobile device, imaqes viewed are for informational purposes only and not intended for diagnostic use.
# Intended for Disease / Condition / Patient Population:
VitreaView is medical image viewer software. Therefore, particular information of intended for disease, condition, and patient population is not applicable.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "VITAL" in black, with a red tagline underneath that says "A Toshiba Medical Systems Group Company". The word "VITAL" is written in a bold, sans-serif font, with a dot above the "i". The tagline is written in a smaller, serif font.
# Software Changes (from the last 510(k) clearance K122136):
# Key changes:
- Query on demand searches archives in real-time directly from the source without the need for an index
- Color and Greyscale Softcopy Presentation State (CSPS and GSPS) ●
- Easy study navigation and study sharing via email, including bookmarking ●
- View DICOM key image objects that have been saved with a study
- . Comparative review allows users to view multiple studies and to track and quantify changes
- . Markup tools allow you to create measurements and add annotations, with indication if off-screen
- . Window/Level sets specific window-level, grayscale-inverted, a user preset or stored-DICOM
- Volume rendering presets allow you to change the appearance of the 3D view
- Enhanced notifications for study changes and study processing .
- Linking is available across modalities and can be applied to duplicated series. A default ● set of linking options are remembered by modality and applied to linked series, including: navigation, window/level and zoom per modality.
- . Auto-Cine enables users of cine to quickly see the dynamic view of studies most often viewed in cine mode
- Provides diagnostic quality image review to medical professionals .
# Other minor enhancements:
- Annotation Indicator ●
- Markup Scrollbar Indicator for easy access to key slices .
- Bookmarks automatically save when users log out ●
- Query retrieve progress indicator
- Text annotation is available with the Arrow tool ●
- Arrow tool is available from the Ruler toolbar ●
- Ability to review the report and move through the VitreaView studies ●
- Calibration handling for projection modalities ●
- . Windows 8.1 support
- Internet Explorer 11 support ●
- Windows 8 support ●
- Internet Explorer 10 support ●
- Administrators can remove changed or deleted studies from the VitreaView Study List by ● authorized permissions
- . The updated URL encryption method provides additional security for patient health information (PHI)
- Support for automatic LDAP group mapping ●
- . Structured reports can be created from metadata during the initial study load
- The patient demographics display within the viewport has been refined to show more ● information when the browser is resized
- . Custom actions are available to define specific search behavior
- Improved selection of the viewport layout
- MINT storage can be optimized to minimize overhead
- Improved access to related images and documents ●
- . The status of Pending has been added to Administration Change Sets
- Ability to redirect the user to another URL .
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "VITAL" in bold, black letters. Below the word "VITAL" is the phrase "A Toshiba Medical Systems Group Company" in a smaller, red font. The word "VITAL" is the main focus of the image.
- Ability to use a custom logout hook ●
- . Carousel swipe navigation
- . Ability to disable pixel measurements
- . Ability to disable logout
# Substantial Equivalence Comparison:
#### . Regulatory Comparison
| Characteristic | Modified VitreaView<br>Software<br>(Submission Subject) | 510(k) Cleared<br>VitreaView Software<br>(K122136)<br>(Predicate Device-1) | ViTALConnect<br>(K071362)<br>(Predicate Device-2) | Noted<br>Differences |
|------------------------|---------------------------------------------------------|----------------------------------------------------------------------------|------------------------------------------------------|--------------------------------|
| Classification<br>Name | Picture Archiving and<br>Communications<br>System | Picture Archiving and<br>Communications<br>System | Picture Archiving<br>and<br>Communications<br>System | None |
| Regulatory<br>Number | 892.2050 | 892.2050 | 892.2050 | None |
| Product Code | LLZ | LLZ | LLZ | None |
| Classification | Class II | Class II | Class II | None |
| Review Panel | Radiology | Radiology | Radiology | None |
| Decision Date | Under Review | September 7, 2012 | May 30, 2007 | Both predicates<br>are cleared |
#### Intended Use / Indications for Use Comparison with the Predicate Device-1 .
| Characteristic | Modified VitreaView<br>Software<br>(Submission Subject) | 510(k) Cleared<br>VitreaView Software<br>(K122136)<br>(Predicate Device-1) | Noted Differences |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use /<br>Indications for<br>Use | VitreaView is a<br>medical image viewing<br>and information<br>distribution application<br>that provides access,<br>through the internet<br>and within the<br>enterprise to multi-<br>modality softcopy | VitreaView is a medical<br>image viewing and<br>information distribution<br>application that<br>provides access,<br>through the Internet<br>and within the<br>enterprise, to multi-<br>modality softcopy | The <b>only</b> changes in the<br>Intended Use statements<br>are as below:<br>• The following line from<br>the Intended Use<br>statement of already<br>cleared VitreaView |
| Characteristic | Modified VitreaView<br>Software<br>(Submission Subject) | 510(k) Cleared<br>VitreaView Software<br>(K122136)<br>(Predicate Device-1) | Noted Differences |
| medical images,<br>reports, and other<br>patient-related<br>information. This data<br>is hosted within<br>disparate archives and<br>repositories for<br>diagnosis, review,<br>communication, and<br>reporting of DICOM<br>and non-DICOM data.<br>Display monitors used<br>for reading medical<br>images for diagnostic<br>purposes must comply<br>with the applicable<br>regulatory approvals<br>and quality control<br>requirements for their<br>use and maintenance.<br>Lossy compressed<br>mammography<br>images and digitized<br>film screen images<br>must not be reviewed<br>for primary image<br>interpretations.<br>When accessing<br>VitreaView from a<br>mobile device, images<br>viewed are for<br>informational purposes<br>only and not intended<br>for diagnostic use. | medical images, reports<br>and other patient-<br>related information, that<br>may be hosted within<br>disparate archives and<br>repositories for review,<br>communication and<br>reporting of DICOM and<br>non-DICOM data.<br>VitreaView is not<br>intended for primary<br>diagnosis. When<br>accessed from a mobile<br>tablet, VitreaView is for<br>informational purposes<br>only and not intended<br>for diagnostic use.<br>Display monitors used<br>for reading medical<br>images for diagnostic<br>purposes must comply<br>with applicable<br>regulatory approvals<br>and with quality control<br>requirements for their<br>use and maintenance.<br>Lossy compressed<br>mammographic images<br>and digitized film<br>screen images must not<br>be reviewed for primary<br>image interpretations.<br>Mammographic images<br>may only be interpreted<br>using an FDA approved<br>monitor that offers at<br>least 5 MP resolution<br>and meets other<br>technical specifications<br>reviewed and accepted<br>by FDA. | 510(k) — K122136 has<br>been removed:<br>"VitreaView is not<br>intended for primary<br>diagnosis"<br>The following word has<br>been added the in the<br>updated Intended Use<br>statement:<br>"for diagnosis" | |
| Modality<br>Support | Multi-modality | Multi-modality | None |
| Characteristic | Modified VitreaView<br>Software<br>(Submission Subject) | VITALConnect<br>(K071362)<br>(Predicate Device-2) | Noted Differences |
| Intended Use /<br>Indications for<br>Use | VitreaView is a<br>medical image viewing<br>and information<br>distribution application<br>that provides access,<br>through the internet<br>and within the<br>enterprise to multi-<br>modality softcopy<br>medical images,<br>reports, and other<br>patient-related<br>information. This data<br>is hosted within<br>disparate archives and<br>repositories for<br>diagnosis, review,<br>communication, and<br>reporting of DICOM<br>and non-DICOM data.<br>Display monitors used<br>for reading medical<br>images for diagnostic<br>purposes must comply<br>with the applicable<br>regulatory approvals<br>and quality control<br>requirements for their<br>use and maintenance.<br>Lossy compressed<br>mammography<br>images and digitized<br>film screen images<br>must not be reviewed<br>for primary image<br>interpretations.<br>When accessing<br>VitreaView from a<br>mobile device, images<br>viewed are for<br>informational purposes<br>only and not intended<br>for diagnostic use. | VITALConnect, Version<br>4.1 is a medical<br>diagnostic software<br>system intended to<br>process, analyze,<br>review, and distribute<br>multi-dimensional<br>digital images acquired<br>from a variety of<br>imaging devices<br>including: CT, MR,<br>CR/DR/DX, SC, US,<br>NM, PET, XA, and RF,<br>etc.<br>VITALConnect is not<br>meant for primary<br>Image Interpretation In<br>mammography. | None<br>Both software products<br>provide diagnostic quality<br>medical image review for<br>multi-dimensional digital<br>images acquired from a<br>variety of imaging devices. |
| Characteristic | Modified VitreaView<br>Software<br>(Submission Subject) | ViTALConnect<br>(K071362)<br>(Predicate Device-2) | Noted Differences |
| Modality<br>Support | Multi-modality | Multi-modality | None |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "VITAL" in bold, black letters. Below the word is the text "A Toshiba Medical Systems Group Company" in a smaller, red font. The dot on the "i" in "VITAL" is a circle.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "VITAL" in bold, black letters. Below the word "VITAL" is the phrase "A Toshiba Medical Systems Group Company" in a smaller, red font. The dot above the "i" in "VITAL" is a perfect circle.
#### Intended Use / Indications for Use Comparison with the Predicate Device-2 ●
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows the word "VITAL" in large, bold, black letters. Below the word "VITAL" is the text "A Toshiba Medical Systems Group Company" in a smaller, red font. The dot on the "i" in "VITAL" is a perfect circle.
#### Device Description Comparison with the Predicate Device-1 •
| Characteristic | Modified VitreaView<br>Software<br>(Submission Subject) | 510(k) Cleared<br>VitreaView Software<br>(K122136)<br>(Predicate Device-1) | Noted Differences |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device<br>Description | VitreaView is a cross-<br>browser, cross-<br>platform, zero-footprint<br>universal medical<br>image viewer solution<br>capable of displaying<br>both DICOM and non-<br>DICOM medical<br>images for diagnosis,<br>review,<br>communication, and<br>reporting.<br>VitreaView enables<br>clinicians and other<br>medical professionals<br>to access patients'<br>medical images with<br>integrations into a<br>variety of medical<br>record systems, such<br>as Electronic Health<br>Record (EHR),<br>Electronic Medical<br>Record (EMR), Health<br>Information Exchange<br>(HIE), Personal Health<br>Record (PHR), and<br>image exchange<br>systems. It supports<br>the physician in<br>medical image viewing | VitreaView is a cross-<br>browser, cross-<br>platform, zero-footprint<br>universal medical<br>image viewer solution<br>capable of displaying<br>both DICOM and non-<br>DICOM medical<br>images.<br>VitreaView enables<br>clinicians and other<br>medical professionals<br>to access patients'<br>medical images with<br>integrations into a<br>variety of medical<br>record systems, such<br>as Electronic Health<br>Record (EHR),<br>Electronic Medical<br>Record (EMR), Health<br>Information Exchange<br>(HIE), Personal Health<br>Record (PHR), and<br>image exchange<br>systems. It supports the<br>physician in medical<br>image viewing and<br>treatment planning. | The only change in the<br>Device Description<br>statements is as below:<br>●<br>The support for<br>diagnostic quality<br>medical image<br>review has been<br>added into the<br>subject device. |
| Characteristic | Modified VitreaView<br>Software<br>(Submission Subject) | 510(k) Cleared<br>VitreaView Software<br>(K122136)<br>(Predicate Device-1) | Noted Differences |
| | and treatment<br>planning.<br><br>VitreaView offers<br>medical professionals<br>a universal viewer for<br>accessing multi-<br>dimensional imaging<br>data in context with<br>reports from enterprise<br>patient health<br>information databases,<br>fosters collaboration,<br>and provides<br>workflows and<br>interfaces appropriate<br>for referring physicians<br>and clinicians. IT<br>departments will not<br>have to incur time to<br>install client systems,<br>due to the zero<br>footprint/zero-<br>download nature of<br>VitreaView. VitreaView<br>offers scalability to<br>add new users as<br>demand grows, may<br>be deployed in a<br>virtualized<br>environment, and is<br>designed to be<br>integrated with<br>enterprise patient<br>health information<br>databases.<br><br>When accessing<br>VitreaView from a<br>mobile device, images<br>viewed are for<br>informational purposes<br>only and not intended<br>for diagnostic use. | VitreaView offers<br>medical professionals a<br>universal viewer for<br>accessing imaging data<br>in context with reports<br>from enterprise patient<br>health information<br>databases, fosters<br>collaboration, and<br>provides workflows and<br>interfaces appropriate<br>for referring physicians<br>and clinicians. IT<br>departments will not<br>have to incur time to<br>install client systems,<br>due to the zero<br>footprint, zero-<br>download nature of<br>VitreaView. VitreaView<br>offers scalability to add<br>new users as demand<br>grows, may be<br>deployed in a<br>virtualized environment,<br>and is designed to be<br>integrated with<br>enterprise patient<br>health information<br>databases.<br><br>When accessed the<br>VitreaView software<br>from a mobile device,<br>the image viewed on<br>the VitreaView software<br>is for informational<br>purposes only and not<br>intended for diagnostic<br>use. | |
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows the word "VITAL" in black font. Below the word "VITAL" is the phrase "A Toshiba Medical Systems Group Company" in red font. The letters in the word "VITAL" are all capitalized except for the "i", which is lowercase and has a dot above it.
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the word "VITAL" in a bold, sans-serif font. The "i" in "VITAL" has a circular dot above it. Below the word "VITAL" is the text "A Toshiba Medical Systems Group Company" in a smaller, red font.
| Characteristic | Modified VitreaView<br>Software<br>(Submission Subject) | VITALConnect<br>(K071362)<br>(Predicate Device-2) | Noted Differences |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device<br>Description | VitreaView is a cross-browser, cross-platform, zero-footprint universal medical image viewer solution capable of displaying both DICOM and non-DICOM medical images for diagnosis, review, communication, and reporting.<br><br>VitreaView enables clinicians and other medical professionals to access patients' medical images with integrations into a variety of medical record systems, such as Electronic Health Record (EHR), Electronic Medical Record (EMR), Health Information Exchange (HIE), Personal Health Record (PHR), and image exchange systems. It supports the physician in medical image viewing and treatment planning.<br><br>VitreaView offers medical professionals a universal viewer for accessing multi-dimensional imaging data in context with reports from enterprise patient health information databases, fosters collaboration | The VITALConnect system is a medical diagnostic device that allows the processing, review, analysis, communication and media interchange of multi-dimensional digital images acquired from a variety of imaging devices.<br><br>The VITALConnect system provides multi-dimensional visualization of digital images to aid clinicians in their analysis of anatomy and pathology. The VITALConnect system user interface follows typical clinical workflow patterns to process, review, and analyze digital images. | None<br><br>Both software products provide diagnostic quality medical image review for multi-dimensional digital images acquired from a variety of imaging devices. |
| Characteristic | Modified VitreaView<br>Software<br>(Submission Subject) | VITALConnect<br>(K071362)<br>(Predicate Device-2) | Noted Differences |
| | and provides<br>workflows and<br>interfaces appropriate<br>for referring physicians<br>and clinicians. IT<br>departments will not<br>have to incur time to<br>install client systems,<br>due to the zero<br>footprint/ zero-<br>download nature of<br>VitreaView. VitreaView<br>offers scalability to<br>add new users as<br>demand grows, may<br>be deployed in a<br>virtualized<br>environment, and is<br>designed to be<br>integrated with<br>enterprise patient<br>health information<br>databases. | | |
| | When accessing<br>VitreaView from a<br>mobile device, images<br>viewed are for<br>informational purposes<br>only and not intended<br>for diagnostic use. | | |
#### . Device Description Comparison with the Predicate Device-2
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the word "VITAL" in a bold, sans-serif font. The "i" in "VITAL" has a circular dot above it. Below the word "VITAL" is the text "A Toshiba Medical Systems Group Company" in a smaller, red font.
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image shows the word "VITAL" in large, bold, black letters. Below the word "VITAL" is the phrase "A Toshiba Medical Systems Group Company" in a smaller, red font. The word "VITAL" is the main focus of the image, with the company affiliation provided as additional context.
#### Technological Characteristics Similarity with the Predicate Device-1 ●
| Modified VitreaView<br>Software<br>(Submission<br>Subject)<br>Technological<br>Characteristic | Description | 510(k) Cleared<br>VitreaView Software<br>(K122136)<br>(Predicate Device-1) | Explanation of<br>Differences |
|---------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|-------------------------------|
| Annotation and<br>Measurement Tools | Line Angle Ruler Arrow | Same | Same |
| User Installation<br>Requirements | Runs within browser using HTML and JavaScript only No installation is required on user's machine | Same | Same |
| Data Type Supported | DICOM Non-DICOM | Same | Same |
| Image View/<br>Manipulation | Image Zoom Pan Window Level Auto Window Level Reset Scout Lines Image Rotate Image Flip Magnify Image Invert Image Cine | Same | Same |
| Data Encryption | HTTPS SSL | Same | Same |
| Patient Demographic<br>Display | Capable of displaying patient demographic information | Same | Same |
| Linking | Co-planar linking:<br>Autolink Manual | Same | Same |
| Modified VitreaView<br>Software<br>(Submission<br>Subject)<br>Technological<br>Characteristic | Description | 510(k) Cleared<br>VitreaView Software<br>(K122136)<br>(Predicate Device-1) | Explanation of<br>Differences |
| User and Password<br>Control | Users can be<br>managed via an<br>internal database,<br>active directory, or<br>parent application | Same | Same |
| Data Security | Stored on server | Same | Same |
| Audit Trail | Audit trail logged | Same | Same |
| User Management | Database structure<br>allows mapping users<br>to groups internally or<br>mapping external<br>groups (AD, parent<br>application) to internal<br>groups and role | Same | Same |
| Transmission Modes | Via the web with<br>Internet browsers | Same | Same |
| File Type Used | • JPEG for<br>Lossy data<br>• PNG for<br>Lossless data | Same | Same |
| MPR Viewing | This viewing feature<br>enables the display of<br>reformatted CT and<br>MR images into axial,<br>coronal and sagittal<br>orientations. | Same | Same |
| 3D Volume Rendered<br>Viewing | This viewing feature<br>enables the display of<br>3D perspective views<br>of CT and MR image<br>sets that have been<br>transformed into<br>volumes. It also<br>provides presets to<br>enable users to alter<br>the visualization | Same | Same |
| Modified VitreaView<br>Software<br>(Submission<br>Subject)<br>Technological<br>Characteristic | Description | 510(k) Cleared<br>VitreaView Software<br>(K122136)<br>(Predicate Device-1) | Explanation of<br>Differences |
| | parameters of the 3D<br>views to highlight<br>features. | | |
| Active Target Tool | This viewing feature<br>provides a facility to<br>view a single target<br>location within<br>multiple images. | Same | Same |
| Crosshair Navigation<br>and Synchronization: | This viewing feature<br>provides a facility to<br>synchronize and<br>scroll through<br>multiple views at the<br>same time. | Same | Same |
| Ability to clone<br>images side by side | Ability to clone<br>images side by side. | Same | Same |
| Ability to close an<br>image by clicking an<br>"X" in the upper-left<br>portion of the<br>viewport | Ability to close an<br>image by clicking an<br>"X" in the upper-left<br>portion of the<br>viewport. | Same | Same |
| Ability to select locale<br>and language<br>settings on the login<br>screen | Ability to select locale<br>and language settings<br>on the login screen. | Same | Same |
| Ability to customize<br>the columns in the<br>study directory by<br>selecting the<br>dropdown arrow on<br>the right side of each<br>column. | Ability to customize<br>the columns in the<br>study directory by<br>selecting the<br>dropdown arrow on<br>the right side of each<br>column. | Same | Same |
| Help Tips | Proactive help tips<br>appear for 10-15<br>seconds to educate<br>users on certain<br>functionality that may | Same | Same |
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image shows the word "VITAL" in bold, black letters. Below the word is the phrase "A Toshiba Medical Systems Group Company" in a smaller, red font. The text appears to be a company logo.
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image shows the word "VITAL" in a bold, sans-serif font. The "i" in "VITAL" has a circular dot above it. Below the word "VITAL" is the text "A Toshiba Medical Systems Group Company" in a smaller, red font.
{16}------------------------------------------------
Image /page/16/Picture/0 description: The image shows the word "VITAL" in large, bold, black letters. Below the word "VITAL" is the phrase "A Toshiba Medical Systems Group Company" in a smaller, red font. The word "VITAL" is the main focus of the image, with the company affiliation provided as additional context.
| Modified VitreaView<br>Software<br>(Submission<br>Subject)<br>Technological<br>Characteristic | Description | 510(k) Cleared<br>VitreaView Software<br>(K122136)<br>(Predicate Device-1) | Explanation of<br>Differences |
|-----------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------|-------------------------------|
| | not be obvious to a<br>new user. | | |
| Support for TIF Files | VitreaView can<br>display TIF files. | Same | Same |
| Tablet support for<br>information purpose<br>only<br>(Not for diagnostic<br>use) | This viewing feature<br>provides access of<br>VitreaView software<br>on various iOS and<br>Android tablet<br>devices through the<br>default internet<br>browser. Key<br>features are:<br>Two-finger<br>pinch to zoom<br>and pan Touch and<br>drag to scroll Double-tap to<br>access<br>Gesture<br>menu Tap Carousel<br>thumbnail,<br>then tap<br>Image Pane<br>to…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.