← Product Code [MCW](/productcode/MCW) · K150732

# Diamondback 360 Peripheral Orbital Atherectomy System (K150732)

_Cardiovascular Systems, Inc. · MCW · Apr 13, 2015 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K150732

## Device Facts

- **Applicant:** Cardiovascular Systems, Inc.
- **Product Code:** [MCW](/productcode/MCW.md)
- **Decision Date:** Apr 13, 2015
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 870.4875
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

The Diamondback 360 Peripheral Orbital Atherectomy System is a percutaneous orbital atherectomy system indicated for use as therapy in patients with occlusive atherosclerotic disease in peripheral arteries and who are acceptable candidates for percutaneous transluminal atherectomy. The OAS supports removal of stenotic material from artificial arteriovenous dialysis fistulae (AV shunt). The OAS is a percutaneous orbital atherectomy system indicated as a therapy in patients with occluded hemodialysis grafts who are acceptable candidates for percutaneous transluminal angioplasty.

## Device Story

Percutaneous orbital atherectomy system for removal of stenotic material from peripheral arteries and AV dialysis fistulae. Device utilizes eccentrically rotating diamond-coated crown to sand/remove stenotic material; particles are small enough for bodily absorption. System components include Orbital Atherectomy Device (OAD), ViperWire Advance guide wire, Saline Infusion Pump (SIP), and ViperSlide lubricant. Used by physicians in clinical settings. Modifications include reduced shaft diameter and updated saline sheath/glue plug design. Output is physical removal of vessel wall obstructions, facilitating improved blood flow. Benefits include minimally invasive treatment of occlusive disease.

## Clinical Evidence

Bench testing only. Testing included system life, stall, temperature, tensile, flexibility, torque, orbit, guide wire brake, saline flow, track, and packaging/distribution testing. Biocompatibility testing performed per ISO 10993-5 (cytotoxicity), ISO 10993-10 (sensitization/irritation), ISO 10993-11 (acute systemic toxicity/pyrogenicity), and ISO 10993-4 (hemocompatibility).

## Technological Characteristics

Orbital atherectomy system featuring diamond-coated eccentrically rotating crown. Components: OAD, guide wire, saline infusion pump, lubricant. Materials validated per ISO 10993 standards. Mechanical operation via rotational energy. Modifications include reduced diameter drive shaft, drive shaft tip bushing, and saline sheath.

## Regulatory Identification

An intraluminal artery stripper is a device used to perform an endarterectomy (removal of plaque deposits from arterisclerotic arteries.)

## Predicate Devices

- Stealth 360°® Orbital PAD System, Model PRD-SC30-125 ([K110389](/device/K110389.md), [K122987](/device/K122987.md), & [K131092](/device/K131092.md))
- Diamondback 360® Peripheral Orbital Atherectomy System, Model DBP-125MICRO145 ([K133399](/device/K133399.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized image of three faces in profile, stacked on top of each other. The faces are oriented to the right and are connected to form a single, flowing shape. The logo is surrounded by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 13, 2015

Cardiovascular Systems, Inc. c/o Mr. Blair Buth Regulatory Affairs Manager 1225 Old Highway 8 NW Saint Paul, MN 55112

Re: K150732

Trade/Device Name: Diamondback 360® Peripheral Orbital Atherectomy System Regulation Number: 21 CFR 870.4875 Regulation Name: Intraluminal Artery Stripper Regulatory Class: Class II Product Code: MCW Dated: March 19, 2015 Received: March 20, 2015

Dear Mr. Buth:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office

of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Kenneth J. Cavanaugh -S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K150732

Device Name

Diamondback 360® Peripheral Orbital Atherectomy System

Indications for Use (Describe)

The Diamondback 360 Peripheral Orbital Atherectomy System is a percutaneous orbital atherectomy system indicated for use as therapy in patients with occlusive atherosclerotic disease in peripheral arteries and who are acceptable candidates for percutaneous transluminal atherectomy.

The OAS supports removal of stenotic material from artificial arteriovenous dialysis fistulae (AV shunt). The OAS is a percutaneous orbital atherectomy system indicated as a therapy in patients with occluded hemodialysis grafts who are acceptable candidates for percutaneous transluminal angioplasty.

Type of Use (Select one or both, as applicable)

| <div> <span> <span style="font-size: 20px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)                 </span> </div> | <div> <span> <span style="font-size: 20px;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C)                 </span> </div> |
|----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# Section. 6 510(k) Summary

| Company Name:              | Cardiovascular Systems, Inc.<br>1225 Old Highway 8<br>Saint Paul, MN 55112 |  |                                                                                                                                 |
|----------------------------|----------------------------------------------------------------------------|--|---------------------------------------------------------------------------------------------------------------------------------|
| Contact:                   | Blair Buth                                                                 |  |                                                                                                                                 |
| Phone:                     | (651) 259-2032                                                             |  |                                                                                                                                 |
| Fax:                       | (651) 305-7734                                                             |  |                                                                                                                                 |
| Summary Date:              | March 19, 2015                                                             |  |                                                                                                                                 |
| Trade Name:                | Diamondback 360® Peripheral Orbital Atherectomy System                     |  |                                                                                                                                 |
| Product Code:              | MCW—Catheter, Peripheral, Atherectomy                                      |  |                                                                                                                                 |
| Classification Regulation: | 21 CFR 870.4875—Intraluminal Artery Stripper                               |  |                                                                                                                                 |
| Classification:            | II                                                                         |  |                                                                                                                                 |
| Predicate Devices:         |                                                                            |  |                                                                                                                                 |
|                            | 510(k) Numbers:<br>Manufacturer:<br>Trade Name:                            |  | K110389, K122987, & K131092<br>Cardiovascular Systems, Inc.<br>Stealth 360°® Orbital PAD System, Model<br>PRD-SC30-125          |
|                            | 510(k) Number:<br>Manufacturer:<br>Trade Name:                             |  | K133399<br>Cardiovascular Systems, Inc.<br>Diamondback 360® Peripheral Orbital<br>Atherectomy System, Model DBP-<br>125MICRO145 |

#### 6.1 Description of Device

The Diamondback 360 Peripheral Orbital Atherectomy System (OAS) is intended for use in the treatment of peripheral arteries and A-V graft (shunt) stenosis.

The OAS provides a method of removing stenotic material from peripheral arteries and A-V grafts. The Diamondback 360 uses an eccentrically rotating sanding surface (crown) to remove stenotic material on the vessel wall. The stenotic particles that are removed are small enough to be absorbed by the body.

{4}------------------------------------------------

The Diamondback 360 Peripheral Orbital Atherectomy System consists of the following components:

- 1) Orbital Atherectomy Device (OAD)
- 2) Atherectomy Guide Wire (e.g., ViperWire Advance)
- 3) Saline Infusion Pump (SIP)
- 4) Atherectomy Lubricant (e.g., ViperSlide)

#### 6.2 Description of Change

The 1.25 Solid Peripheral Orbital Atherectomy Device was modified to reduce the diameter of the shaft (drive shaft, drive shaft tip bushing, and saline sheath) including modification of the crown for compatibility with the smaller diameter drive shaft. In addition, the saline sheath material and design for the 1.25 Solid and 1.25 Micro OAD configurations was changed and an updated glue plug component was added to accomadate the new saline sheath design.

#### 6.3 Intended Use

The Diamondback 360 Peripheral Orbital Atherectomy System is a percutaneous orbital atherectomy system indicated for use as therapy in patients with occlusive atherosclerotic disease in peripheral arteries and who are acceptable candidates for percutaneous transluminal atherectomy.

The OAS supports removal of stenotic material from artificial arteriovenous dialysis fistulae (AV shunt). The OAS is a percutaneous orbital atherectomy system indicated as a therapy in patients with occluded hemodialysis grafts who are acceptable candidates for percutaneous transluminal angioplasty.

#### 6.4 Technology

The Diamondback 360 Peripheral Orbital Atherectomy System provides a method of removing or reducing occlusive atherosclerotic or stenotic material. The OAS applies a diamond coated, eccentrically rotating surface to remove stenotic material on the vessel wall. The stenotic particles that are removed are small enough to be absorbed by the body. This is the same technology that was cleared to market for use in the peripheral arteries for Stealth 360° in 510(k) K110389. K122987. K131092 and for Diamondback 360° Peripheral per K133399, and K133999.

{5}------------------------------------------------

#### 6.5 Performance Data

The Diamondback 360 Peripheral Orbital Atherectomy System was evaluated using the following performance bench testing to confirm the performance characteristics as compared to the predicate device.

- System Life Testing including guide wire ●
- Stall Testing
- o Introducer Compatibility Testing
- Temperature Testing ●
- Tensile Testing
- o Flexibility Testing
- o Delivered Torque Testing
- Orbit Testing ●
- Guide Wire Brake Testing
- Saline Flow Testing ●
- Track Testing
- Packaging/Simulated Distribution Testing ●
- . Biocompatibility
	- Cytotoxicity per ISO 10993-5 o
	- Sensitization per ISO 10993-10 o
	- Irritation per ISO 10993-10 o
	- Acute Systemic Toxicity per ISO 10993-11 o
	- Pyrogenicity per ISO 10993-11 o
	- Hemocompatibility per ISO 10993-4 o

All test results demonstrate that the materials chosen, the manufacturing process, and the design utilized for the Diamondback 360 Peripheral Orbital Atherectomy System met the established specifications necessary for consistent performance during its intended use.

#### 6.6 Conclusions

Diamondback 360 Peripheral Orbital Atherectomy System met all The predetermined acceptance criteria of design verification and validation testing as specified by applicable standards, test protocols, and/or customer inputs. Testing results demonstrate that the Diamondback 360 Peripheral Orbital Atherectomy System is substantially equivalent to the legally marketed predicate devices and does not raise any new safety or effectiveness questions.

---

**Source:** [https://fda.innolitics.com/device/K150732](https://fda.innolitics.com/device/K150732)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
