K150727 · Tokuyama Dental Corporation · EMA · Jun 17, 2015 · Dental
Device Facts
Record ID
K150727
Device Name
ESTECEM
Applicant
Tokuyama Dental Corporation
Product Code
EMA · Dental
Decision Date
Jun 17, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3275
Device Class
Class 2
Indications for Use
The ESTECEM device is an adhesive resin cement for the cementation of crowns, bridges, inlays, and onlays made of glass/oxide ceramics (porcelain, zirconia and alumina), metals/alloys (precious and non-precious) and resin materials including inorganic filler (composite materials). The ESTECEM device is indicated for: - Repair of fractured porcelain fused to metal crowns and all ceramic restorations - Cementation of veneers - Cementation of adhesion bridges - Cementation of metal or resin cores, metal or glass-fiber posts
Device Story
ESTECEM is a dual-cure (light and/or self-cure) radiopaque adhesive resin cement system used by dental professionals in clinical settings. The system consists of a paste (A/B) dispensed via automix syringe and a primer kit (ESTELINK bond and Tokuyama Universal Primer). The primer enhances adhesion to prosthetic materials and tooth structure. The clinician applies the primer and cement to the restoration and tooth surface; the material then undergoes chemical and/or photo-polymerization to bond the restoration. This process secures indirect restorations, improving structural integrity and esthetics for the patient.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench testing of physical properties (film thickness, working/setting time, flexural/tensile strength, water sorption, color stability, radio-opacity) and biocompatibility (ISO 10993-1, -3, -5, -10, -11; ISO 4049; ISO 7405).
Technological Characteristics
Dual-cure (light/self-cure) adhesive resin cement. Components: paste (A/B) and primer kit (ESTELINK bond, Tokuyama Universal Primer). Radiopaque. No energy source. Non-sterile. Testing standards: ISO 4049:2009, ISO 7405:2008, ISO 10993-1:2009, ISO 10993-3, ISO 10993-5, ISO 10993-10, ISO 10993-11.
Indications for Use
Indicated for dental cementation of crowns, bridges, inlays, onlays, veneers, and cores/posts made of glass/oxide ceramics, metals/alloys, or resin composites; also indicated for repair of fractured porcelain-fused-to-metal or all-ceramic restorations.
Regulatory Classification
Identification
Zinc oxide-eugenol is a device composed of zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp. Dental cement other than zinc oxide-eugenol is a device composed of various materials other than zinc oxide-eugenol intended to serve as a temporary tooth filling or as a base cement to affix a temporary tooth filling, to affix dental devices such as crowns or bridges, or to be applied to a tooth to protect the tooth pulp.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 17, 2015
Tokuyama Dental Corporation c/o Mr. Keith A. Barritt Fish & Richardson P.C. 1425 K Street NW Suite 1100 Washington, DC 20005
Re: K150727
Trade/Device Name: ESTECEM Regulation Number: 21 CFR 872.3275 Regulation Name: Dental Cement Regulatory Class: II Product Code: EMA Dated: March 13, 2015 Received: March 20, 2015
#### Dear Mr. Barritt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely vours.
Image /page/1/Picture/8 description: The image shows a signature that reads "Tina Kiang-S". The signature is written in a cursive style, with the first name "Tina" appearing above the last name "Kiang-S". There is a large looping design to the left of the name. The signature is in black ink on a white background.
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
K150727
Device Name ESTECEM
#### Indications for Use (Describe)
The ESTECEM device is an adhesive resin cementation of crowns, bridges, inlays, and onlays made of glass/oxide ceramics (porcelain, zirconia and alumina), metals/alloys (precious) and resin materials including inorganic filler (composite materials). The ESTECEM device is indicated for:
- Repair of fractured porcelain fused to metal crowns and all ceramic restorations
- Cementation of veneers
- Cementation of adhesion bridges
- Cementation of metal or resin cores, metal or glass-fiber posts
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY Tokuyama Dental Corporation
# ESTECEM Adhesive Resin Cement
The following information is provided pursuant to 21 CFR 807.92.
# 807.92(a)(1): Submitter's Name and Address
# (i) 510(k) Submitter
Tokuyama Dental Corporation 38-9 Taitou 1-chome, Taitou-ku Tokyo 110-0016 Japan Phone: 011-81-3-3835-2261
# (ii) 510(k) Submitter Contact
Keith A. Barritt Fish & Richardson P.C. 1425 K Street, N.W., Suite 1100 Washington, DC 20005 Phone: (202) 783-5070 Facsimile: (202) 783-2331 barritt@fr.com Email:
# (iii) Preparation Date
March 9, 2015
# 807.92(a)(2): Name of Device
Trade or Proprietary Name: Common Name: Classification Name: Product Code: Regulation:
ESTECEM dental adhesive resin cement cement, dental EMA 21 CFR 872.3275
{4}------------------------------------------------
#### 807.92(a)(3): Predicate Devices
The ESTECEM adhesive resin cement device is substantially equivalent for purposes of FDA medical device regulations to multiple predicate devices with respect to performance and biocompatibility. namely to the Primary Predicate BISTITE II DC (K#011685, Health Canada license no. 60061) and Reference Devices CLEARFIL ESTHETIC CEMENT & DC BOND (K#062410, Health Canada license no. 73128); MULTILINK AUTOMIX (K#123397, Health Canada license no. 79221); and DUOLINK II (K#101787, Health Canada license no. 30968). Additional information regarding the predicate devices appears at the end of this 510(k) Summary.
#### 807.92(a)(4): Device Description
The ESTECEM device is a dual-cure (light and/or self-cure), radiopaque, adhesive resin cement system with excellent handling, esthetic and adhesive properties to teeth and all prosthetic materials. The main components of the ESTECEM device are (1) the paste (A/B) and (2) primer kit. The paste and primer are packaged in plastic syringes and plastic bottles, respectively. The device also comes with a mixing dish (dispensing well), disposable applicator, and cement mixing tip.
The paste is dispensed in an automix syringe and is available in various shades (universal, clear, brown and white-opaque). The primer kit components consist of ESTELINK (bond A/bond B) and Tokuyama Universal Primer (A/B). The ESTELINK bond promotes the adhesion of the paste to the tooth structure. The Tokuyama Universal Primer enhances the adhesion of the paste to prosthetic materials.
The ESTECEM device does not come sterilized and is not intended to be sterilized prior to use.
#### 807.92(a)(5): Intended Use
The ESTECEM device is an adhesive resin cement for the cementation of crowns, bridges, inlays, and onlays made of glass/oxide ceramics (porcelain, zirconia and alumina), metals/alloys (precious and non-precious) and resin materials including inorganic filler (composite materials). The ESTECEM device is indicated for:
- Repair of fractured porcelain fused to metal crowns and all ceramic restorations
- Cementation of veneers
- Cementation of adhesion bridges
- Cementation of metal or resin cores, metal or glass-fiber posts
{5}------------------------------------------------
# 807.92(a)(6): Technological Characteristics
The ESTECEM device has the same basic technological characteristics in terms of design, material, and chemical composition as the predicate devices identified above. The ESTECEM device does not have an energy source. Although the chemical compositions are not identical, the material properties, principles of operation, and performance characteristics of the ESTECEM device are comparable to the selected predicates, as demonstrated in part by the non-clinical performance bench testing described below.
# 807.92(b)(1): Non-clinical Testing
Non-clinical testing of the physical properties of the ESTECEM device were conducted, including film thickness, working time, setting time, flexural strength, water sorption. color stability, radio-opacity, tensile strength, and biocompatibility testing in accordance with ISO 4049:2009, ISO 7405:2008, and ISO 10993-1:2009, ISO 10993-3 (genotoxicity), ISO 10993-5 (cytotoxicity), ISO 10993-10 (sensitization and irritation), and ISO 10993-11 (systemic toxicity).
# 807.92(b)(2): Clinical Testing
There were no clinical tests performed for the ESTECEM device.
# 807.92(b)(3): Conclusions from Testing
Based on the non-clinical testing conducted of the physical properties of the ESTECEM device and the biocompatibility testing as described above, it is concluded that the ESTECEM device is as safe, as effective, and performs as well as or better than the predicate devices.
{6}------------------------------------------------
# Predicate devices
| Device name | ESTECEM | BISTITE II DC | CLEARFIL<br>ESTHETIC CEMENT<br>& DC BOND 1) | MULTILINK<br>AUTOMIX | DUOLINK II 2) |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Tokuyama Dental<br>Corporation | Tokuyama Dental<br>Corporation | Kuraray Noritake<br>Dental Inc. | Ivoclar Vivadent AG | Bisco, Inc. |
| 510(K) no. | pending | K011685 | K062410 | K123397 | K101787 |
| Licence no.<br>(Health<br>Canada) | pending | 60061 | 73128 | 79221 | 30968 |
| Device<br>classification<br>name | Cement, Dental | Cement, Dental | Cement, Dental | Material, Tooth Shade,<br>Resin | Cement, Dental |
| Product code | EMA | EMA | EMA | EBF | EMA |
| Indication for<br>use | 1. Cementation of crowns,<br>bridges, inlays, and onlays<br>made of glass/oxide ceramics<br>(porcelain, zirconia and<br>alumina), metals/alloys<br>(precious and non-precious)<br>and resin materials including<br>inorganic filler (composite<br>materials)<br>2. Repair of fractured<br>porcelain fused to metal<br>crowns and all ceramic<br>restorations<br>3. Cementation of veneers<br>4. Cementation of adhesion<br>bridges<br>5. Cementation of metal or<br>resin cores, metal or glass-<br>fiber posts | For use as dental cement in<br>various applications,<br>including luting, repair of<br>fractured porcelains or crown<br>and bridge resins, cementation<br>of adhesive bridge prostheses<br>and CR core posts | 1. Cementation of crowns,<br>bridges, inlays and onlays<br>made of porcelain, ceramic,<br>hybrid ceramics, composite<br>resin or metal<br>2. Cementation of veneers<br>3. Cementation of adhesion<br>bridges<br>4. Cementation of metal cores,<br>resin cores, metal posts or<br>glass-fiber posts | Multilink Automix is used for<br>the permanent cementation of<br>indirect restorations where a<br>strong bond is desired:<br>Inlays, onlays, crowns,<br>bridges and root posts made<br>of:<br>- Metal and metal ceramics<br>- All-ceramics, in particular<br>opaque zirconium oxide<br>ceramics<br>- Composites and fibre-<br>reinforced composites | 1. All indirect restorations (i.e.<br>crowns.<br>bridges, inlay, and onlays<br>fabricated from<br>metal composite, porcelain,<br>ceramic, zirconia, alumina,<br>etc.)<br>2. All endodontic posts (i.e.<br>fiber, composite and metal)<br>3. All abutments (i.e. screws) |
| Components | - Paste (A and B)<br>Primer Kit<br>- Bond A/Bond B) and<br>Tokuyama Universal Primer<br>(A/B) | - Bistite II Paste (A and B)<br>- Primer 1 A<br>- Primer 1 B<br>- Primer 2<br>- Metaltite<br>- Tokuso Ceramic Primer A<br>- Tokuso Ceramic Primer B<br>- Air Barrier | - Paste (A and B)<br>- Try In Paste | - Multilink Automix (Base and<br>Catalyst)<br>- Multilink Primer A<br>- Multilink Primer B | - Duo-Link Universal (Base<br>and Catalyst) |
| Principle of<br>operation | Dental cement that is cured by<br>chemical and/or photo<br>polymerization reaction.<br>(Dual-cure) | Dental cement that is cured by<br>chemical and/or photo<br>polymerization reaction.<br>(Dual-cure) | Dental cement that is cured by<br>chemical and/or photo<br>polymerization reaction.<br>(Dual-cure) | Dental cement that is cured by<br>chemical and/or photo<br>polymerization reaction.<br>(Dual-cure) | Dental cement that is cured by<br>chemical and/or photo<br>polymerization reaction.<br>(Dual-cure) |
1) This device is presumably marketed with the proprietary name as "CLEARFIL ESTHETIC CEMENT EX."
2) This device is presumably marketed with the proprietary name as "DUO-LINK UNIVERSAL."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.