Salem Sump Dual Lumen Stomach Tube with Multi-functional Port
Applicant
Covidien
Product Code
PIF · Gastroenterology, Urology
Decision Date
Oct 8, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.5980
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
Intended for gastric decompression and delivery of fluids, including irrigation, nutritional supplements, and medication during the time period that gastric decompression is required. The device is intended for patients with age of two-year and older.
Device Story
Double lumen gastric sump tube; utilizes multi-functional port for alternating gastric decompression or fluid delivery. Input: suction source for decompression; syringe or feeding set for fluid delivery. Output: gastric drainage or delivered fluids/meds. Used in clinical settings by healthcare providers. Features male ENFit connector compliant with AAMI/CN3:2014 to prevent misconnections. Benefits patient by enabling safe, standardized enteral access and decompression while reducing risk of tubing misconnections.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including biocompatibility (ISO 10993-1), stability testing, usability/human factors testing, and physical/functional performance testing of the ENFit connector (leakage, stress cracking, axial load, unscrewing, and misconnection assessment).
Technological Characteristics
Double lumen sump tube; PVC construction with barium sulfate and print ink. Multi-functional port uses polycarbonate, polyethylene, polypropylene, cyanoacrylate adhesive, silicone rubber/lubricant, and copolyester. Features male ENFit connector (AAMI/CN3:2014). Sizes 10Fr–18Fr; lengths 36–48 inches. Sterilization via gamma irradiation and ethylene oxide.
Indications for Use
Indicated for gastric decompression and delivery of fluids (irrigation, nutritional supplements, medication) in patients aged two years and older requiring gastric decompression.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol consists of a stylized human figure with three heads, representing the department's focus on health and human well-being.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 8, 2015
Covidien Wei Zhao, M.D. Senior Director, Regulatory Affairs 15 Hampshire Street Mansfiled. MA 02048
Re: K150711
> Trade/Device Name: Salem Sump™ Dual Lumen Stomach Tube with Multi-functional Port Regulation Number: 21 CFR\$ 876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: II Product Code: PIF, FEG Dated: September 3, 2015 Received: September 4, 2015
Dear Wei Zhao,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
Page 2 - Wei Zhao, M.D.
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
# Benjamin R. Fisher -S
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K150711
#### Device Name
Salem Sump™ Dual Lumen Stomach Tube with Multi-functional Port
Indications for Use (Describe)
Intended for gastric decompression and delivery of fluids, including irrigation, nutritional supplements, and medication during the time period that gastric decompression is required. The device is intended for patients with age of two-year and older.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 2.0 510(k) Summary
Salem Sump™ Dual Lumen Stomach Tube with Multi-functional Port
In accordance with section 513(i) of the SMDA and as defined in 21CFR Part 807.92 this summary is submitted by:
Covidien 15 Hampshire Street Mansfield, MA 02048 Date Prepared: January 18, 2015
#### a. Contact Person
Wei Zhao Senior Director, Regulatory Affairs Covidien Telephone: (508) 261-8404 Fax: (508) 261-8461
#### b. Name of Medical Device
Common Name: Tube, double lumen for intestinal decompression and/or intubation
U.S. FDA Classification Product Code: FEG, PIF
U.S. Regulation Description: Gastrointestinal tube and accessories, 21 CFR 876.5980
Proprietary / Trade Name: Salem Sump™ Dual Lumen Stomach Tube with Multi-functional Port
## c. Identification of Legally Marketed Device(s)
Next Generation Salem Sump, K040388
{4}------------------------------------------------
# d. Device comparison summary
| Device Comparison Summary | | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Predicate Device | Proposed Device | |
| Device Name | Next Generation Salem<br>Sump with Multi-port<br>Connector | Salem Sump™ Dual Lumen<br>Stomach Tube with Multi-<br>Functional Port (with ENFit<br>connector) |
| Intended use | Intended for gastric<br>decompression and delivery of<br>fluids, including irrigation,<br>nutritional supplements, and<br>medication during the time<br>period that gastric<br>decompression is required. | Intended for gastric<br>decompression and delivery of<br>fluids, including irrigation,<br>nutritional supplements, and<br>medication during the time<br>period that gastric<br>decompression is required. The<br>device is intended for patients<br>with age of two-year and older. |
| Sterility | Sterile by gamma irradiation | Sterile by gamma irradiation<br>and ethylene oxide |
| Technological Characteristics | The device is a gastric sump<br>tube which utilizes a<br>multifunctional port to allow<br>alternating use for gastric<br>decompression or delivery of<br>fluids, including irrigation,<br>nutritional supplements, and<br>medication. The connection<br>for delivery of fluids<br>supports multiple types of<br>devices and connectors. | The device is a gastric sump<br>tube which utilizes a<br>multifunctional port to allow<br>alternating use for gastric<br>decompression or delivery of<br>fluids, including irrigation,<br>nutritional supplements, and<br>medication. The connection<br>for delivery of fluids is via a<br>male ENFit connector. |
| Design | The product is a double<br>lumen sump tube available in<br>5 sizes, 10Fr – 18Fr, and<br>lengths of 36 – 48 inches.<br>The device is equipped with<br>a Multi-functional port to<br>allow alternating use for<br>gastric decompression (via a<br>suction connector) or<br>delivery of fluids, including<br>irrigation, nutritional<br>supplements, and medication<br>via a catheter tip syringe or a<br>feeding set with a step<br>connector. | The product is a double<br>lumen sump tube available in<br>5 sizes, 10Fr – 18Fr, and<br>lengths of 36 – 48 inches.<br>The device is equipped with<br>a Multi-functional port to<br>allow alternating use for<br>gastric decompression (via a<br>suction connector) or<br>delivery of fluids, including<br>irrigation, nutritional<br>supplements, and medication<br>via a syringe or feeding set<br>equipped with a female<br>ENFit connector. |
| Materials / Chemical<br>Composition | Sump Tube<br>● Polyvinyl chloride | Sump Tube<br>● Polyvinyl chloride |
| | | |
| (PVC)<br>• Barium Sulfate<br>• print ink | (PVC)<br>• Barium Sulfate<br>• print ink | |
| Multi-functional port<br>• Polycarbonate<br>• Polyethylene<br>• Polypropylene<br>• Cyanoacrylate adhesive<br>• Silicone rubber<br>• Silicone lubricant<br>• Print ink | Multi-functional port<br>• Polycarbonate<br>• Polyethylene<br>• Polypropylene<br>• Cyanoacrylate adhesive<br>• Silicone rubber<br>• Silicone lubricant<br>• Print ink<br>• Copolyester | |
{5}------------------------------------------------
#### Discussion of technological differences e.
The Salem Sump™ Dual Lumen Stomach Tube with Multi-functional Port are intended for gastric decompression and delivery of fluids, including irrigation, nutritional supplements, and medication during the time period that gastric decompression is required. The primary technological difference is the incorporation of the new ENFit connector which is compliant to AAMI/CN3:2014 (PS) Part 3. This connector is part of an industry wide effort to address misconnections by adopting a uniform connector that has been engineered to meet the objective of AAMI/ANSI/ISO 80369-1, small-bore connectors for liquids and gases in healthcare applications - part 1: general requirements. The additional changes implemented in manufacturing technologies are incidental to the primary change, and do not alter device performance.
- f. Discussion of Nonclinical testing
- The following risk management activities were executed to identify the potential hazards, evaluate their risks and develop the risk mitigation methods:
- ENFit connector: Risk management report o
- Proposed device: Design Failure Modes and Effects Analysis (FMEA) and risk o analysis.
- Biocompatibility testing in accordance with ISO 10993-1:2009. Biological Evaluation of ● medical Devices- Part 1: Evaluation and Testing has demonstrated the biological safety of parts of the medical device which may indirectly contact the patient, and is consistent with FDA "Draft Guidance for Industry and FDA staff, Use of international Standard ISO 10993 'Biological Evaluation of medical Devices Part 1: Evaluation and Testing,"' issued on April 23, 2013. Similar testing had been conducted for the predicate device.
{6}------------------------------------------------
- Stability testing of the proposed device evaluated the key performance properties of the feeding set after accelerated aging in support of the expiration date which will be applied to the device.
- Usability and human factors testing was conducted as part of the design of the ENFit . connector.
- Device tensile properties were evaluated for its durability.
- . The following tests were performed to demonstrate the physical and functional performance of the ENFit connector:
- fluid leakage o
- stress cracking о
- resistance to separation from axial load O
- resistance to separation from unscrewing O
- resistance to overriding о
- disconnection from unscrewing O
- dimensional analysis o
- Misconnection assessment of ENFit connector and suctioning port was conducted to . prove the incompatibility of the ENFIt small bore connector to other medical devices to reduce the risk of misconnection.
- g. Clinical testing
Clinical evaluations were not relied upon for evidence of safety and effectiveness, or for the determination of substantial equivalence.
#### h. Conclusions
This information provided within this pre-market notification demonstrates that the modified Salem Sump™ Dual Lumen Stomach Tube with Multi-Functional Port with ENFit Connectors is as safe, as effective, and performs as well as or better than the legally marketed device, therefore I find the subject device to be substantially equivalent to the predicate device. The addition of the ENFit connector, which is compliant with AAMI/CN3:2014 (PS) Part 3 is intended to improve device performance by addressing the risk of misconnections.
End of Summary
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.