Frozen clinical specimens from 1,930 consecutive outpatients presenting to emergency or ambulatory departments
The study validated the clinical performance (sensitivity, specificity, NPV, PPV) of the INNOVANCE D-Dimer assay on the Sysmex CS-5100 for the exclusion of pulmonary embolism (PE) using real-world patient samples.
INNOVANCE D-Dimer PE Exclusion Validation Study; Multi-center, prospective collection of clinical specimens with retrospective analysis; Follow-up/Duration: 3-month follow-up for PE development
1,467 outpatients presenting to emergency or ambulatory departments with suspected PE; Sample Size: 1,467 (final analysis); Number of Sites: Multi-center (US and OUS)
Imaging methods (spiral CT and/or VQ scan)
Sensitivity, specificity, negative predictive value (NPV), and positive predictive value (PPV) for PE exclusion
Indications for Use
The Sysmex® CS-5100 is a fully automated blood coagulation analyzer intended for in vitro diagnostic use using plasma collected from venous blood samples in 3.2% sodium citrate tubes to analyze clotting, chromogenic and immunoassay methods in the clinical laboratory. For determination of: - Prothrombin Time (PT) seconds and PT INR with Dade® Innovin® - Activated Partial Thromboplastin Time (APTT) with Dade® Actin® FSL - Fibrinogen (Fbg) with Dade® Thrombin Reagent - Antithrombin (AT) with INNOVANCE® Antithrombin - D-dimer with INNOVANCE® D-Dimer The performance of this device has not been established in neonate and pediatric patient populations.
Device Story
Sysmex CS-5100 is a fully automated blood coagulation analyzer for clinical laboratories. It processes venous blood plasma samples collected in 3.2% sodium citrate tubes. The device utilizes clotting, chromogenic, and immunochemical methods to determine PT, PT INR, APTT, Fibrinogen, Antithrombin, and D-dimer levels. Samples are loaded into racks and automatically transported to aspiration positions; the system features random access, liquid level sensing, and barcode reading for samples and reagents. Analysis is performed using transmitted light detection (absorbance) via a halogen lamp. Results are displayed on an Information Processing Unit (IPU) touch panel, printed, or transmitted to a host computer. The device supports normal and micro-sample modes. By providing rapid, automated coagulation testing, the device assists clinicians in diagnosing and monitoring hemostatic disorders and excluding pulmonary embolism, thereby facilitating timely clinical decision-making.
Clinical Evidence
Clinical evidence includes a multi-center method comparison (n=469 for PT, n=361 for D-dimer) against the predicate, showing high correlation (r=0.996-0.999). Reproducibility studies (20-day, 3-site) demonstrated total CVs generally <7.5%. A prospective multi-center study (n=1467) evaluated D-dimer for Pulmonary Embolism (PE) exclusion, achieving 98.0% sensitivity and 99.7% NPV at a 0.50 mg/L FEU cutoff, confirming clinical performance comparable to the predicate.
Technological Characteristics
Automated coagulation analyzer; uses transmitted light detection (absorbance) at 340-800 nm via halogen lamp. Materials include pneumatic units, sample/reagent probes, and cuvette handling systems. Features include liquid level sensing, barcode readers, and bidirectional interface (CA, ASTM, CS protocols). Temperature control: detector 37°C ± 0.5°C; reagent probe 37.5°C ± 0.5°C. Reagent cooling: 4°C-15°C. Software manages automated pipetting, dilution, and analysis modes (normal/micro).
Indications for Use
Indicated for in vitro diagnostic use in clinical laboratories to analyze clotting, chromogenic, and immunoassay coagulation parameters (PT, PT INR, APTT, Fibrinogen, Antithrombin, D-dimer) using citrated human plasma. Performance not established for neonate or pediatric populations.
Regulatory Classification
Identification
A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.
Special Controls
*Classification.* Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized human figure or a series of overlapping profiles. The symbol is composed of three curved lines that create the impression of faces or heads in profile.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 11, 2016
Siemens Healthcare Diagnostics Products GmbH Ms. Donna Noeh Regulatory Manager, US Affairs Emil-von-Behring-Str. 76 35041 Marburg, Germany
Re: K150678
Trade/Device Name: Sysmex CS-5100 Regulation Number: 21 CFR 864.5425 Regulation Name: Multipurpose system for in vitro coagulation studies Regulatory Class: Class II Product Code: JPA Dated: December 10, 2015 Received: December 11, 2015
Dear Ms. Noah:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{1}------------------------------------------------
CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Leonthena R. Carrington -S
Leonthena R. Carrington, MS, MBA, MT(ASCP) Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K150678
Device Name Sysmex CS-5100
#### Indications for Use (Describe)
The Sysmex® CS-5100 is a fully automated blood coagulation analyzer intended for in vitro diagnostic use using plasma collected from venous blood samples in 3.2% sodium citrate tubes to analyze clotting, chromogenic and immunoassay methods in the clinical laboratory.
For determination of:
- Prothrombin Time (PT) seconds and PT INR with Dade® Innovin®
- · Activated Partial Thromboplastin Time (APTT) with Dade® Actin® FSL
- · Fibrinogen (Fbg) with Dade® Thrombin Reagent
- Antithrombin (AT) with INNOVANCE® Antithrombin
- · D-dimer with INNOVANCE® D-Dimer
The performance of this device has not been established in neonate and pediatric patient populations.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "SIEMENS" in a bold, sans-serif font. The color of the text is a teal or turquoise shade. The letters are evenly spaced and appear to be a logo or brand name.
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92 and follows the FDA guidance "The 510(k) Program: Evaluating Substantial Equivalence in Premarket Notifications [510(k)]", issued July 28, 2014.
## 1. Submitter
Siemens Healthcare Diagnostics Products GmbH Emil-von-Behring-Str. 76 35041 Marburg, Germany Contact Person: Donna Noeh Email: donna.noeh@siemens.com Phone: + 49 6421 39 5107 Facsimile: + 49 6421 39 4977 Date Prepared: January 7, 2016 2. Device Sysmex CS-5100 Name of Device: Common or Usual Name: Automated Coagulation Instrument Multipurpose system for in vitro coagulation studies (21 CFR Classification Name: 864.5425) Regulatory Class: 2 Product Code: JPA 510(k) Review Panel Hematology 3. Predicate Device Name of Device: Sysmex CA-1500 (K011235) Common or Usual Name: Automated Coagulation Instrument Classification Name: Multipurpose system for in vitro coagulation studies (21 CFR 864.5425) Regulatory Class: 2 Product Code: JPA Hematology 510(k) Review Panel
The predicate has not been subject to a design-related recall for any of the applications associated with this Premarket Notification.
No reference devices were used in this submission.
{5}------------------------------------------------
# 4. Device Description / Test Principle
The Sysmex CS-5100 is an automated blood coagulation instrument which can analyze samples using clotting, chromogenic and immunoassay methods. Analysis results are displayed on the Information Processing Unit (IPU) screen. They can be printed on external printers or transmitted to a host computer. Sold separately from the instrument are the associated
- Reagents ●
- Controls
- Calibrators ●
- Consumable materials ●
The subject of this 510(k) notification are reagent applications which perform the coagulation tests Prothrombin Time (PT) seconds and PT INR with Dade® Innovin®; Activated Partial Thromboplastin Time (APTT) with Dade® Actin® FSL; Fibrinogen (Fbg) with Dade® Thrombin Reagent; Antithrombin (AT) with INNOVANCE® Antithrombin; and D-dimer with INNOVANCE® D-Dimer.
The analysis principles used on the instrument are reflected by the reagent application testing provided in this 510(k) notification and is described in the below table.
| Table of Sysmex CS-5100 Analysis Principles | | |
|---------------------------------------------|------------------------------------------------|---------------------------------|
| Reagent | Application | Methodology |
| Dade® Innovin® | PT, Prothrombin Time<br>(seconds) | Clotting<br>(extrinsic pathway) |
| Dade® Innovin® | PT, Prothrombin Time (INR) | Clotting, calculated |
| Dade® Actin® FSL | APTT, Activated Partial<br>Thromboplastin Time | Clotting<br>(intrinsic pathway) |
| Dade® Thrombin<br>Reagent | Fibrinogen quantitation | Clotting<br>(common pathway) |
| INNOVANCE®<br>Antithrombin | Antithrombin quantitation | Chromogenic |
| INNOVANCE® D-<br>Dimer | D-dimer quantitation | Immunochemical |
The intended Environment of Use is a clinical central/hospital laboratory.
{6}------------------------------------------------
# Front View of the Instrument
Image /page/6/Figure/2 description: This image shows a diagram of a machine with several parts labeled with numbers. The top part of the machine is labeled (1), and there are buttons labeled (4), (5), and (6). The middle section has a part labeled (7), and the bottom section has parts labeled (3) and (8), with a button labeled (2) on the side.
- (1) Light shield lid: Open this cover to set reagents, perform maintenance, etc.
- (2) Power switch: Turns the power ON/OFF.
- (3) Left door: Holds the Pneumatic Unit inside. Open this door to adjust the positive pressure (0.22 MPa).
- (4) Alarm indicator LED: Indicates the instrument status.
- (5) Mechanical stop switch: Press this switch to immediately stop the instrument's mechanical movement.
- (6) Start button: Press this button to immediately start an analysis. This button is the same as the [Start] button on the IPU toolbar.
- (7) Sampler: Automatically transports samples that are set in the sample rack to the aspiration position.
- (8) Right door: Open the door for maintenance or to discard cuvettes.
{7}------------------------------------------------
# Informational Processing Unit
Image /page/7/Figure/2 description: The image shows a desktop computer with a monitor, CPU, keyboard, and mouse. The monitor is labeled as (1), the CPU is labeled as (2), the keyboard is labeled as (3), and the mouse is labeled as (4). The computer is a desktop model, with the CPU placed under the monitor and the keyboard and mouse connected to the CPU.
- (1) Touch panel display: Displays the IPU screen. It can also be used as a touch panel.
- (2) IPU Main Unit: This is the Main Unit of IPU.
- (3) Keyboard: Used to operate the IPU together with the touch panel.
- (4) Mouse: Used to operate the IPU together with the touch panel.
### Options and Accessories
Options and accessories that can be used for this instrument are as follows:
- (1) Waste tank (with float switch for waste tank): Waste fluids discharged from the Main Unit enter this tank.
- (2) Wand barcode reader: Reads barcodes to input sample numbers, rack numbers and reagent IDs.
- (3) 2D barcode reader: Reads barcodes to input calibrator's or reagent's assay sheet values, normal values and ISI values, and control's targets/limits.
- (4) IPU cart: The IPU (which includes the keyboard, PC and touch panel display), and the tanks for waste, rinse and CA Clean II can be placed on this cart.
- (5) External indicator light: The status of the instrument is indicated with a red, yellow or green light that can be seen when the operator is not directly in front of the instrument.
- (6) IPU holder: This is an optional holder for the IPU which includes the keyboard, PC and touch panel display which can be installed on the right side of the instrument to minimize the instrument footprint.
{8}------------------------------------------------
The instrument is capable of measuring in the following analysis modes:
- (1) Normal mode: Samples for all the analyses including re-analyses are taken into the instrument at the same time and analyzed. In a normal mode, a capped sample tube analysis can be performed. Automatic re-analysis can also be performed.
- (2) Micro-sample mode: Samples set in the sampler or STAT holder are taken into the instrument for each analysis through a secondary dispensing sample probe. When measurements are to be performed in Micro mode, sample tubes must be uncapped. The instrument detects capped tubes automatically and displays an error message. This analysis mode can be performed with less sample volume than normal mode (consult instruction manual for further information). However, automatic re-analysis cannot be performed.
# 5. Intended Use / Indications for Use
The Sysmex CS-5100 is a fully automated blood coagulation analyzer intended for in vitro diagnostic use using plasma collected from venous blood samples in 3.2% sodium citrate tubes to analyze clotting, chromogenic and immunoassay methods in the clinical laboratory.
For determination of:
- Prothrombin Time (PT) seconds and PT INR with Dade® Innovin® ●
- Activated Partial Thromboplastin Time (APTT) with Dade® Actin® FSL
- Fibrinogen (Fbg) with Dade® Thrombin Reagent ●
- Antithrombin (AT) with INNOVANCE® Antithrombin ●
- D-dimer with INNOVANCE® D-Dimer. ●
The performance of this device has not been established in neonate and pediatric patient populations.
### 6. Comparison of Technological Characteristics with the Predicate Device
Both the subject and predicate instruments employ the same technological characteristics in that they automatically analyze various clotting tests using reagents, calibrators and controls previously cleared for automated coagulation analyzers. The reagents perform at least equally well on both the subject and predicate instruments. At a high level, the devices have the following same technological elements:
### Similarities between the CS-5100 and CA-1500
| Similarities between Sysmex CS-5100 and Sysmex CA-1500 | | |
|--------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------|
| Analyzer Component | Proposed Device<br>Sysmex CS-5100 | Predicate Device<br>Sysmex CA-1500 |
| Regulatory Classification | JPA, Class 2<br>System, Multipurpose for in<br>vitro coagulation studies | Same |
| Sample Type | Human plasma<br>3.2% sodium citrate | Same |
| Similarities between Sysmex CS-5100 and Sysmex CA-1500 | | |
| Analyzer Component | Proposed Device | Predicate Device |
| | Sysmex CS-5100 | Sysmex CA-1500 |
| Application type | Clotting Applications:<br>Prothrombin Time (PT) with<br>Dade® Innovin®;<br>Activated Partial Thromboplastin<br>Time (APTT) with Dade® Actin®<br>FSL;<br>Fibrinogen (Clauss) with Dade®<br>Thrombin Reagent<br><br>Chromogenic Application:<br>Antithrombin with<br>INNOVANCE® Antithrombin<br><br>Immuno-chemical Application:<br>D-dimer<br>with INNOVANCE® D-Dimer<br> | Same |
| Specimen Processing | Automatic Pipetting and Dilution | Same |
| Random Access | Yes | Same |
| Liquid Level Sensing | Yes - reagent and sample | Same |
| Bar code Reader | Sample + reagent | Same |
| STAT Testing | Yes | Same |
| Sampling Capabilities | Normal and Micro Mode | Same |
| Sample Volumes in Normal<br>Mode (Plasma) | PT with Dade® Innovin® 50 µL<br>APTT with Dade® Actin® FSL<br>50 µL<br>Fibrinogen with Dade®<br>Thrombin Reagent 10 µL<br>Antithrombin with<br>INNOVANCE® Antithrombin<br>10 µL<br>D-dimer with INNOVANCE® D-Dimer | Same |
| Similarities between Sysmex CS-5100 and Sysmex CA-1500 | | |
| Analyzer Component | Proposed Device | Predicate Device |
| | Sysmex CS-5100 | Sysmex CA-1500 |
| Sample Volumes in Micro<br>Mode (Plasma) | PT with Dade® Innovin® 50 µL<br>APTT with Dade® Actin® FSL<br>50 µL<br>Fibrinogen with Dade®<br>Thrombin Reagent 10 µL | Same |
| Rinse & Buffer Solutions | | |
| On-board | CA-CLEAN I | Same |
| External | CA-CLEAN II | Same |
| | Dade Owren's Buffer | Same |
| | Water | Same |
| Light Source | | |
| Chromogenic | Halogen Lamp | Same |
| Immuno-chemical | Halogen Lamp | Same |
| Wavelengths used in Analysis | Antithrombin with<br>INNOVANCE® Antithrombin<br>(405 nm) | Same |
| Temperature Control | Sample incubation well: 37℃ ±<br>1.0°C | Same |
{9}------------------------------------------------
{10}------------------------------------------------
{11}------------------------------------------------
There are no technological differences between the subject and predicate devices. However the following minor changes exist between the subject and predicate devices:
| Differences between Sysmex CS-5100 and Sysmex CA-1500 | | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Analyzer Component | Proposed Device | Predicate Device |
| | Sysmex CS-5100 | Sysmex CA-1500 |
| Intended Use Statement | The Sysmex CS-5100 is a fully<br>automated blood coagulation analyzer<br>intended for in vitro diagnostic use<br>using plasma collected from venous<br>blood samples in 3.2% sodium citrate<br>tubes to analyze clotting, chromogenic<br>and immunoassay methods in the<br>clinical laboratory.<br>For determination of:<br>• Prothrombin Time (PT) seconds and<br>PT INR with Dade® Innovin®<br>• Activated Partial Thromboplastin<br>Time (APTT) with Dade®<br>Actin® FSL<br>• Fibrinogen (Fbg) with Dade®<br>Thrombin Reagent<br>• Antithrombin (AT) with<br>INNOVANCE® Antithrombin<br>• D-dimer with INNOVANCE® D-<br>Dimer<br>The performance of this device has not<br>been established in neonate and<br>pediatric patient populations. | The intended use of the Sysmex CA-<br>1500 is as a fully automated,<br>computerized blood plasma<br>coagulation analyzer for in vitro<br>diagnostic use in clinical<br>laboratories.<br>The instrument uses citrated<br>human plasma to perform the<br>following parameters and<br>calculated parameters:<br>Clotting Analysis Prameters:<br>Prothrombin Time (PT); Activated<br>Partial Thromboplastin Time<br>(APTT); Fibrinogen (Clauss);<br>Batroxobin Time; Extrinsic Factors<br>(II, V, VII, X); Intrinsic Factors<br>(VIII, IX, XI, XII); Protein C.<br>Chromogenic Analysis<br>Parameters: Antithrombin III;<br>Factor VIII; Plasminogen;<br>Heparin; Protein C; α2-<br>Antiplasmin.<br>Immunologic Analysis Parameters:<br>D-dimer.<br>Calculated Parameters: PT Ratio;<br>PT INR; PT %; Derived Fibrinogen;<br>Factor Assays % Activity |
| Differences between Sysmex CS-5100 and Sysmex CA-1500 | | |
| | Proposed Device | Predicate Device |
| Analyzer Component | Sysmex CS-5100 | Sysmex CA-1500 |
| Operating Principle | | |
| Clotting | Transmitted Light Detection<br>(Absorbance) at 340, 405, 575, 660 or<br>800 nm. Wavelengths 340 and 575 are<br>technically available but not validated in<br>combination with the intended<br>applications. | Scattered Light Detection at 660 nm |
| Chromogenic | Transmitted Light Detection<br>(Absorbance) at 340, 405, 575, 660,<br>800 nm. Wavelengths 340, 575, 660,<br>and 800 are technically available but not<br>validated in combination with the<br>intended applications. | Transmitted Light Detection<br>(Absorbance) at 405, 575, 800 nm |
| Immuno-chemical | Transmitted Light Detection<br>(Absorbance) at 340, 405, 575, 660 or<br>800 nm. Wavelengths 340, 405, 575,<br>and 800 are technically available but not<br>validated in combination with the<br>intended applications. | Transmitted Light Detection<br>(Absorbance) at 405, 575, or 800 nm |
| Wavelengths* used in Analysis<br>*The default wavelength is normally<br>used to generate the reported value of<br>the measurement. The sub-wavelength | Clotting Applications:<br>PT (seconds) Dade® Innovin®<br>(Default = 660 nm; Sub-<br>Wavelength= 800 nm)<br><br>PT (INR) with Dade® Innovin®<br>(Default = 660 nm; Sub-Wavelength=<br>800 nm)<br><br>APTT with Dade® Actin® FSL<br>Activated PTT Reagent (Default =<br>660 nm; Sub-Wavelength= 800 nm) | Clotting Applications:<br>PT (seconds) with Dade®<br>Innovin® Default = 660 nm; Sub-<br>Wavelength= none)<br><br>PT (INR) with Dade® Innovin®<br>(Default = 660 nm; Sub-<br>Wavelength= none)<br><br>APTT with Dade® Actin® FSL<br>Activated PTT Reagent (Default<br>= 660 nm; Sub-Wavelength=<br>none) |
| is run in parallel. If a light intensity<br>error occurs by using the default<br>wavelength the value from the sub-<br>wavelength is used automatically. | Fibrinogen with Dade® Thrombin<br>Reagent (Default = 405 nm; Sub-<br>Wavelength= none)<br><br>Immuno-chemical Application:<br>D-dimer with INNOVANCE®<br>D-Dimer (Default = 660 nm; Sub-<br>Wavelength= none) | Fibrinogen with Dade® Thrombin<br>(Default = 660 nm;<br>Sub-Wavelength= none)<br><br>Immuno-chemical Application:<br>D-dimer with INNOVANCE® D-<br>Dimer (Default = 800 nm; Sub-<br>Wavelength= none) |
| Differences between Sysmex CS-5100 and Sysmex CA-1500 | | |
| Analyzer Component | Proposed Device<br>Sysmex CS-5100 | Predicate Device<br>Sysmex CA-1500 |
| Light Source<br>Clotting | Halogen Lamp | Light Emitting Diode |
| Probes | 2 Sample probes;<br>3 Reagent probes | 1 Sample probe;<br>1 Reagent probe |
| Cap Piercing | Cap Piercer only | Both options available:<br>Cap Piercer and Non-Cap Piercer |
| Temperature Control | -Detector : 37 °C ± 0.5 °C<br>-Reagent probe : 37.5 °C ± 0.5 °C | -Detector: 37°C ± 1.0°C<br>- Reagent probe: 37°C ± 1.0°C |
| Reagent Cooling | 10°C ± 2°C, when ambient temperature is 20°C - 28°C.<br>During operation 4°C -15°C, when ambient temperature is 15°C – 30°C | 15°C ± 2°C, when ambient temperature is 15°C - 30°C |
| Pipetting Capabilities | Reagent probe:<br>20 – 200 μL<br>Sample probe:<br>4 – 270 μL | Reagent probe:<br>3 – 200 μL<br>Sample probe:<br>5 – 450 μL |
| Sample Volumes in Micro<br>Mode (Plasma) | Antithrombin with INNOVANCE® Antithrombin 14 μL<br>D-dimer with INNOVANCE® D-Dimer 15 µL | Antithrombin with INNOVANCE® Antithrombin 10 μL<br>D-dimer with INNOVANCE® D-Dimer 13 µL |
| Bidirectional Interface<br>communication protocols | CA-, ASTM-, CS- Protocol | CA-, ASTM-Protocol |
# Differences between CS-5100 and CA-1500
{12}------------------------------------------------
{13}------------------------------------------------
The above described differences do not raise new questions as to safety and effectiveness of the new device.
# 7. Performance Data
The following performance data were provided in support of the substantial equivalence determination.
#### 7.1 Method comparison
Method comparison studies designed according to EP09-A3 CLSI Guideline "Measurement Procedure Comparison and Bias Estimation Using Patient Samples" were conducted at three external sites in the United States, all sites using the same protocol.
{14}------------------------------------------------
Samples were measured on both the predicate device (Sysmex CA-1500) as well as the new device (Sysmex CS-5100), in random order to eliminate any inherent bias. Results were compared by Passing-Bablok regression analysis as well as Bland-Altman plots. Results from each application met the pre-established acceptance criteria. The following summary of Passing-Bablok regression shows that the proposed and predicate devices provide equivalent results when used in a clinical setting.
| Sysmex CS-5100: Method Comparison Summary Table, Passing-Bablok regression | | | | |
|-------------------------------------------------------------------------------------------------------|--------------------------------------------|--------------------------------------------|--------------------------------------------|--------------------------------------------|
| Application<br>(measuring interval) | Site 01 | Site 02 | Site 03 | Sites Combined |
| Prothrombin<br>Time using<br>Dade® Innovin®<br>(8.7 - 90.0<br>seconds) | n=125<br>$y = 1.000x + 0.400$<br>r = 0.997 | n=209<br>$y = 1.000x + 0.000$<br>r = 0.999 | n=135<br>$y = 0.987x +0.374$<br>r = 0.999 | n=469<br>$y = 1.000x + 0.100$<br>r = 0.998 |
| Prothrombin<br>Time (INR) using<br>Dade® Innovin®<br>(0.93 - 8.00 INR) | n=122<br>$y = 1.044x - 0.037$<br>r = 0.997 | n=208<br>$y = 1.030x-0.038$<br>r = 0.999 | n=135<br>$y = 1.039x - 0.019$<br>r = 0.999 | n=465<br>$y = 1.047x-0.047$<br>r = 0.999 |
| Activated Partial<br>Thromboplastin<br>Time using<br>Dade® Actin®<br>FSL<br>(20.0 - 139.0<br>seconds) | n=126<br>$y = 1.029x - 1.112$<br>r = 0.994 | n=210<br>$y = 1.017x-1.203$<br>r = 0.998 | n=130<br>$y = 1.027x - 1.225$<br>r = 0.994 | n=466<br>$y = 1.026x - 1.315$<br>r = 0.996 |
| Fibrinogen<br>quantitation<br>using Dade®<br>Thrombin Reagent<br>(50 - 860 mg/dL) | n=145<br>$y = 1.052x -4.466$<br>r = 0.994 | n=91<br>$y = 1.028x + 5.491$<br>r = 0.996 | n=132<br>$y = 0.982x + 9.889$<br>r = 0.998 | n=368<br>$y = 1.018x + 4.633$<br>r = 0.995 |
| Antithrombin<br>quantitation<br>using<br>INNOVANCE®<br>Antithrombin<br>(9.0 - 128.0% of<br>norm) | n=135<br>$y = 0.990x-0.413$<br>r = 0.996 | n=120<br>$y = 0.957x + 1.825$<br>r = 0.994 | n=126<br>$y = 0.985x - 0.359$<br>r = 0.997 | n=381<br>$y = 0.980x + 0.222$<br>r = 0.996 |
{15}------------------------------------------------
| Sysmex CS-5100: Method Comparison Summary Table, Passing-Bablok regression | | | | |
|---------------------------------------------------------------------------------------|---------------------------------|-----------------------------------|-----------------------------------|-----------------------------------|
| D-dimer<br>quantitation<br>using<br>INNOVANCE®<br>D-Dimer (0.19 –<br>35.20 mg/L FEU¹) | n=137 | n= 108 | n=116 | n=361 |
| | $y = 1.058x-0.013$<br>r = 0.997 | $y = 1.000x - 0.035$<br>r = 0.997 | $y = 1.000x + 0.000$<br>r = 0.998 | $y = 1.021x - 0.007$<br>r = 0.996 |
#### 7.2 Reproducibility Studies
Twenty-day precision studies were performed at two external sites in Germany and one external site in the United States. Testing followed the scheme of two runs per day, with two replicates per run, at each of the three sites according to CLSI EP05-A2 "Evaluation of Precision Performance of Quantitative Measurement Methods". The order of the analysis of parameter, samples and quality control samples for each run and day varied to avoid an inherent bias to the study. One calibration curve of each calibrated application was used in the study. Within Run, Between Run, Between Day, and Total With-in Site was calculated. The data is summarized in the following tables.
| Sysmex CS-5100: Reproducibility Summary Table, Within Run | | | | | |
|-------------------------------------------------------------------------------------------------|----------------------------------------|--------------------------------|--------------------------------|--------------------------------|----------------------------|
| Application<br>(CRR/<br>Clotting time range) | Sample<br>Range (mean<br>of all sites) | Site 01<br>Within Run<br>(%CV) | Site 02<br>Within Run<br>(%CV) | Site 03<br>Within Run<br>(%CV) | Sites<br>Combined<br>(%CV) |
| Prothrombin Time<br>using Dade® Innovin®<br>(8.7 - 90.0 seconds) | 9.27 - 79.67<br>seconds | 0.64 – 1.85 | 0.47 – 0.98 | 0.51 - 3.61 | 0.59 — 2.35 |
| Prothrombin Time<br>(INR) using Dade®<br>Innovin® (0.93 - 8.00<br>INR) | 1.02 - 7.76<br>INR | 0.62 - 2.21 | 0.45 - 0.92 | 0.44 - 1.31 | 0.51 - 1.50 |
| Activated Partial<br>Thromboplastin Time<br>using Dade® Actin®<br>FSL<br>(20.0 - 139.0 seconds) | 22.02 -<br>123.25<br>seconds | 0.71 - 5.51 | 0.51 - 3.05 | 0.56 - 2.67 | 0.76 - 3.93 |
| Fibrinogen<br>quantitation using<br>Dade® Thrombin<br>Reagent<br>(50 - 860 mg/dL) | 56.2 - 752.6<br>mg/dL | 1.14 - 6.18 | 0.99 - 3.27 | 1.41 - 2.72 | 1.20 - 4.20 |
<sup>1</sup> D-dimer results are reported in fibrinogen equivalent units (FEU).
{16}------------------------------------------------
| Sysmex CS-5100: Reproducibility Summary Table, Within Run | | | | | |
|--------------------------------------------------------------------------------------------|-------------------------------|-------------|-------------|-------------|-------------|
| Antithrombin<br>quantitation using<br>INNOVANCE®<br>Antithrombin<br>(9.0 - 128.0% of norm) | 10.17 -<br>119.41% of<br>norm | 0.71 - 2.66 | 0.81 - 3.45 | 0.97 - 3.13 | 0.92 - 3.11 |
| D-dimer quantitation<br>using INNOVANCE®<br>D-Dimer (0.19 - 35.20<br>mg/L FEU) | 0.25 - 33.81<br>mg/L FEU | 0.91 - 2.15 | 0.80 - 1.56 | 1.27 – 2.51 | 1.08 – 2.07 |
| Sysmex CS-5100: Reproducibility Summary Table, Between Run | | | | | |
|-----------------------------------------------------------------------------------------------|----------------------------------------|---------------------------------|---------------------------------|---------------------------------|----------------------------|
| Application<br>(CRR/<br>Clotting time range) | Sample Range<br>(mean of all<br>sites) | Site 01<br>Between Run<br>(%CV) | Site 02<br>Between Run<br>(%CV) | Site 03<br>Between Run<br>(%CV) | Sites<br>Combined<br>(%CV) |
| Prothrombin Time<br>using Dade®<br>Innovin®<br>(8.7 - 90.0 sec) | 9.27 - 79.67<br>seconds | 0.12 - 1.73 | 0.00 - 1.15 | 0.00 - 1.45 | 0.35 - 1.42 |
| Prothrombin Time<br>(INR) using Dade®<br>Innovin® (0.93 –<br>8.00 INR) | 1.02 – 7.76<br>INR | 0.00 - 2.05 | 0.00 - 1.12 | 0.53 - 1.30 | 0.25 - 1.37 |
| Activated Partial<br>Thromboplastin<br>Time using Dade®<br>Actin® FSL<br>(20.0 - 139.0 sec) | 22.02 - 123.25<br>seconds | 0.00 - 0.97 | 0.00 - 2.50 | 0.29 - 2.66 | 0.34 - 2.40 |
| Fibrinogen<br>quantitation using<br>Dade® Thrombin<br>Reagent<…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.