← Product Code [NUC](/productcode/NUC) · K150615

# Vaginal Moisturizing Gel (K150615)

_Combe Incorporated · NUC · May 1, 2015 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K150615

## Device Facts

- **Applicant:** Combe Incorporated
- **Product Code:** [NUC](/productcode/NUC.md)
- **Decision Date:** May 1, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.5300
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

The device is a personal lubricant, for penial application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity, and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms.

## Device Story

Vaginal Moisturizing Gel is a personal lubricant provided in pre-filled applicators for ease of use. It is intended for penile and/or vaginal application to moisturize and lubricate during intimate sexual activity. The device supplements natural body lubrication and is compatible with natural rubber latex and polyisoprene condoms. It is not compatible with polyurethane condoms. The product is intended for over-the-counter use. The gel formulation is identical to the predicate device, with the primary modification being the delivery system (pre-filled applicators). Safety is maintained through adherence to product specifications including pH, osmolality, and antimicrobial effectiveness. Biocompatibility of the applicator was confirmed via extractable and leachable testing.

## Clinical Evidence

Bench testing only. No clinical data provided. Testing included biocompatibility of the applicator (extractable/leachable analysis per USP <661>, GC-MS, ICP-MS) and condom compatibility testing per ASTM D7661-10.

## Technological Characteristics

Personal lubricant gel; pH 4.5-5.0. Delivered via pre-filled applicators. Applicator biocompatibility assessed per USP <661>. Condom compatibility tested per ASTM D7661-10. Product specifications include appearance, odor, viscosity, osmolality, and microbial limits (TAMC, TYMC, absence of pathogens).

## Regulatory Identification

A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.

## Special Controls

*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.

## Predicate Devices

- Internal Hydrating Gel ([K141718](/device/K141718.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 1, 2015

Combe Incorporated Pushpa Rao, Ph.D., D.A.B.T., R.A.C. Sr. Director Global Regulatory Affairs/Product Safety 1101 Westchester Avenue White Plains, NY 10604

Re: K150615

> Trade/Device Name: Vaginal Moisturizing Gel Regulation Number: 21 CFR§ 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: NUC Dated: March 9, 2015 Received: March 10, 2015

Dear Pushpa Rao,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

Page 2 - Pushpa Rao, Ph.D., D.A.B.T., R.A.C.

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

# Joyce M. Whang -S

for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known)

## K150615

Device Name Vaginal Moisturizing Gel

#### Indications for Use (Describe)

The device is a personal lubricant, for penial application, intended to moisturize and lubricate, to enhance the ease and comfort of intimate sexual activity, and supplement the body's natural lubrication. This product is compatible with natural rubber latex and polyisoprene condoms.

Type of Use (Select one or both, as applicable)

] Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

{3}------------------------------------------------

| 510(k) Owner:                         | Combe Incorporated<br>1101 Westchester Avenue<br>White Plains, NY 10604                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|---------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person                        | Pushpa Rao, Ph.D., D.A.B.T., R.A.C.<br>Sr. Director Global Regulatory Affairs/Product Safety<br>Email: prao@combe.com<br>Phone: (914) 461-4458<br>Fax: (914) 697-7764                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Name of Device                        | Trade name – Vaginal Moisturizing Gel Classification<br>name - Personal lubricant Citation: 21 CFR 884.5300<br>Product Code: NUC Class: II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Predicate Device                      | Internal Hydrating Gel (K141718)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Device Description                    | Vaginal Moisturizing Gel is identical to the predicate device except<br>that it is presented in 8 pre-filled applicators for easy use.<br><br>Vaginal Moisturizing Gel is intended for use as a personal lubricant.<br>This product is compatible with natural rubber latex and polyisoprene<br>condoms. This device is not compatible with polyurethane condoms.<br>The following parameters are included as part of the product<br>specification:<br>• Appearance<br>• Color<br>• Odor<br>• pH<br>• Viscosity<br>• Osmolality<br>• Antimicrobial effectiveness<br>• Total Aerobic Microbial Count (TAMC)<br>• Total Yeast and Mold Count (TYMC)<br>• Absence of Pathogenic Organisms (at minimum Pseudomonas<br>aeruginosa, Staphylococcus aureus, and Candida albicans)<br><br>Vaginal Moisturizing Gel has a pH of 4.5-5.0 and a shelf life of 18 |
| Indications for<br>Use                | The device is a personal lubricant, for penile and/or vaginal application,<br>intended to moisturize and lubricate, to enhance the ease and comfort of<br>intimate sexual activity, and supplement the body's natural lubrication.<br>This product is compatible with natural rubber latex and polyisoprene<br>condoms.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Technological<br>Characteristics      | The proposed device, Vaginal Moisturizing Gel has the identical<br>technological characteristics as the predicate device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Biocompatibility<br>Data              | Biocompatibility of the new applicator was supported with a study of<br>extractable and leachable materials from the applicator. The new<br>applicator was subjected to extractable testing according to USP <661>:<br>Physicochemical Tests for Plastics. Extracts were also analyzed using<br>infrared analysis, gas chromatography coupled with mass spectrometry<br>(GC-MS), and inductively coupled plasma mass spectroscopy. Extracted<br>molecules were only detected after aggressive extraction with ethanol as<br>determined by GC-MS. Risk analysis showed that ethanol extractable<br>materials were in amounts much lower than the tolerable intake of any<br>component.                                                                                                                                                                 |
| Performance Testing-<br>Non- Clinical | The proposed device was tested for condom compatibility according to<br>ASTM D7661-10 and found to be compatible with natural rubber latex<br>and polyisoprene condoms.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Conclusions                           | Based on the results of non-clinical testing, the proposed device,<br>Vaginal Moisturizing Gel, is substantially equivalent to the predicate<br>device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |

{4}------------------------------------------------

# 510(k) Summary

---

**Source:** [https://fda.innolitics.com/device/K150615](https://fda.innolitics.com/device/K150615)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
