Annuy Latex Patient Examination Glove (powdered and powder-free); Annuy Nitrile Patient Examination Glove (powdered and powder-free)
K150612 · Anhui Anyu Latex Products Co., Ltd. · LYY · Nov 16, 2015 · General Hospital
Device Facts
Record ID
K150612
Device Name
Annuy Latex Patient Examination Glove (powdered and powder-free); Annuy Nitrile Patient Examination Glove (powdered and powder-free)
Applicant
Anhui Anyu Latex Products Co., Ltd.
Product Code
LYY · General Hospital
Decision Date
Nov 16, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner.
Device Story
Disposable patient examination gloves; made of natural rubber latex or nitrile butadiene rubber latex; ambidextrous; cuffed; non-sterile; single-use. Act as physical barrier between examiner and patient to prevent contamination. Used in clinical settings by healthcare providers. Protects both examiner and patient from cross-contamination during examinations.
Clinical Evidence
Bench testing only. No clinical data included. Performance verified against ASTM D3578-10 (latex), ASTM D6319-10 (nitrile), ASTM D5151-11 (holes), ASTM D6124-11 (powder), ASTM D5712-10 (protein), and ISO 10993-10 (biocompatibility).
Technological Characteristics
Materials: Natural rubber latex or nitrile butadiene rubber latex. Polymer coating (polyurethane/polyacrylic acid) on powder-free versions. Non-sterile. Ambidextrous. Sizes: S, M, L. Standards: ASTM D3578-10, D6319-10, D5151-11, D6124-11, D5712-10, ISO 10993-10.
Indications for Use
Indicated for use as a disposable medical device worn on the examiner's hand or fingers to prevent cross-contamination between patient and examiner.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
Latex Examination Gloves (Powdered and Powder-Free) (K072145)
Nitrile Examination Gloves (Powdered and Powder-Free) (K072145)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of a human figure, represented by three overlapping profiles, suggesting a sense of community and support. The profiles are arranged in a way that creates a sense of depth and movement. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 16, 2015
Anhui Anyu Latex Products Co., Ltd. c/o Doris Dong Shanghai CV Technology Co., Ltd. Room 1706. No. 128 Songle Rd., Songjiang Area Shanghai 201600 CHINA
Re: K150612 Trade/Device Name: "Annuy Latex Patient Examination Glove (Powdered and Powder-free)" & "Annuy Nitrile Patient Examination Glove (Powdered and Powder-free)" Regulation Number: 21 CFR 880.6250 Regulation Name: Patient examination glove Regulatory Class: Class I Product Code: LYY, LYZ Dated: October 4, 2015 Received: October 16, 2015
Dear Ms. Dong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
**Tejashri Purohit-Sheth, M.D.**
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K150612
#### Device Name
Annuy Latex Patient Examination Glove (powdered and powder-free); Annuy Nitrile Patient Examination Glove (powdered and powder-free)
Indications for Use (Describe)
A patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or fingers to prevent contamination between patient and examiner.
Type of Use (Select one or both, as applicable)
] Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows a logo with a circular design. The top half of the circle is blue, while the bottom half is green. There are several curved lines in the green section that resemble blades of grass or leaves. The logo is set against a light blue background.
# Section 5 510(k) Summary
[As required by 21 CFR 807.92]
#### 1. Submission Information:
| 510(k) Number: | K150612 |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date: | Sept 30th, 2015 |
| Type of 510(k) Submission: | Traditional |
| Basis for 510(k) Submission: | New device |
| Submitter/Manufacturer: | Anhui Anyu Latex Products Co., Ltd<br>East Qinji Road & North Xinghua Road, Bengbu, Anhui, 233010 China |
| Contactor: | Doris Dong (Consultant)<br>Shanghai CV Technology Co., Ltd.<br>Room 1706, No. 128 Songle Rd., Songjiang Area, Shanghai, 201600 China<br>E-mail: doris_d@126.com<br>Tel: 86 21-31261348 / Fax: 86 21-37824346 |
#### 2. Device Description:
| Proprietary Name: | Annuy Latex Patient Examination Glove (powdered and powder-free);<br>Annuy Nitrile Patient Examination Glove (powdered and powder-free) |
|----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common Name: | Latex Examination Gloves (powdered and powder free);<br>Nitrile Examination Gloves (powdered and powder free) |
| Classification Name: | Glove, patient examination, latex;<br>Glove, patient examination, poly |
| Product Code: | LYY & LZA |
| Device Class: | I |
| Regulation Number: | 21 CFR 880.6250 |
| Review Panel: | General Hospital |
| Indications for use: | A patient examination glove is a disposable device intended for medical purposes that<br>is worn on the examiner's hand or fingers to prevent contamination between patient and<br>examiner |
Device Description:
Annuy Latex Patient Examination Glove (powdered and powder-free) is a glove made from natural rubber that covers the hand up to the wrist. It is cuffed and equally wearable on either hand, free from differentiation between the left hand and the right. It has three sizes, i.e., large, medium, and small. All variations share the same natural color, creamy white. The glove is non-sterile and is for single use only, to be discarded after each examination is performed. It acts as a barrier between the examiner and the subject being examined in order to prevent contamination between them.
Annuy Nitrile Patient Examination Glove (powdered and powder-free) is a glove made from nitrile butadiene rubber Latex that covers the hand up to the wrist. It is cuffed and equally wearable on either hand, free from differentiation between the left hand and the right. It has three sizes, i.e. large, medium, and small. All variations share the same blue color. The glove is non-sterile and is for single use only, to be discarded after each examination is performed. It acts as a barrier between the examiner and the subject being examined in order to prevent contamination between them.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows a logo with a circular shape. The top half of the circle is blue, and the bottom half is green. There are several curved lines in the center of the circle, resembling a plant or a bird's wing. The logo is simple and clean, with a focus on natural elements.
### 3. Predicate Device Identification
#### 510(k) Number: K072145
#### Product Name:
Latex Examination Gloves (Powdered and Powder-Free)
Nitrile Examination Gloves (Powdered and Powder-Free)
Manufacturer: Zhanjiang Jiali Glove Products Co., LTD
#### 4. Performance Test Data Summary
#### • The Powdered Latex Patient Examination Gloves
| Characteristics | Standard | | Test Data | |
|-----------------------------|----------------------------------------------------------|-------------------------------------|---------------------------------------------------------------------------------------------------|-------------------------------------|
| Dimension | ASTM D3578-10 | | | |
| Palm width | Size | (mm) | (mm) | |
| | Small | 80±10 | 80~84 | |
| | Medium | 95±10 | 93~97 | |
| | Large | 111±10 | 107~109 | |
| Overall Length | Size | (mm) | (mm) | |
| | Small | 220 min | 240~246 | |
| | Medium | 230 min | 241~248 | |
| | Large | 230 min | 240~249 | |
| Finger Thickness | Size | (mm) | (mm) | |
| | Small | 0.08 min | 0.09~0.11 | |
| | Medium | 0.08 min | 0.09~0.11 | |
| | Large | 0.08 min | 0.09~0.12 | |
| Palm Thickness | Size | (mm) | (mm) | |
| | Small | 0.08 min | 0.10~0.12 | |
| | Medium | 0.08 min | 0.10~0.12 | |
| | Large | 0.08 min | 0.09~0.12 | |
| Physical Properties | ASTM D3578-10 | | | |
| | Before Aging | After aging at 70±2 °C<br>166±2 hrs | Before Aging | After aging at 70±2 °C<br>166±2 hrs |
| - Tensile Strength | 18Mpa min | 14Mpa min | 20.3~35.1Mpa | 16.7~24.0Mpa |
| - Ultimate Elongation | 650% min | 500% min | 961~1233% | 748~1021% |
| - Stress at 500% Elongation | 5.5Mpa max | -- | 4.6~5.4Mpa | -- |
| Freedom from Holes | ASTM D5151-11 | | Passed Standard Acceptance Criteria<br>Holes at Inspection Level I, AQL 2.5 | |
| Powder Amount | ASTM D6124-11, ≤ 10mg/dm² | | 2.34 mg/dm², Meets ASTM D6124-11<br>Results generated values below 10mg/dm² of<br>residual powder | |
| Protein Content | ASTM D5712-10, ≤200µg/dm² Aqueous<br>extractable protein | | Aqueous extractable protein: 72.1 mcg /dm²,<br>Meets ASTM D5712-10 Results generated | |
values below 200µg/dm² of aqueous extractable
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows a logo with a circular design. The top half of the circle is light blue, while the bottom half is green. Within the green portion, there are several curved lines that resemble leaves or blades of grass. The overall design is simple and clean, with a focus on natural elements.
| | | protein content |
|------------------|-------------------------------------------------------------------|----------------------------------------------------|
| Biocompatibility | Primary Skin Irritation in rabbits ISO<br>10993-10:2010-08-01 | Under the condition of the test, not an irritant. |
| | | |
| | Dermal sensitization in the guinea pig ISO<br>10993-10:2010-08-01 | Under the condition of the test, not a sensitizer. |
| | | |
#### • The Powder-free Latex Patient Examination Gloves
| Characteristics | Standard | | Test Data | |
|--------------------------------|----------------------------------------------------------|-------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------|
| Dimension | ASTM D3578-10 | | | |
| Palm width | Size | (mm) | | (mm) |
| | Small | 80±10 | | 80~84 |
| | Medium | 95±10 | | 93~97 |
| | Large | 111±10 | | 107~109 |
| Overall Length | Size | (mm) | | (mm) |
| | Small | 220 min | | 240~246 |
| | Medium | 230 min | | 241~248 |
| | Large | 230 min | | 241~248 |
| Finger Thickness | Size | (mm) | | (mm) |
| | Small | 0.08 min | | 0.09~0.12 |
| | Medium | 0.08 min | | 0.09~0.11 |
| | Large | 0.08 min | | 0.09~0.12 |
| Palm Thickness | Size | (mm) | | (mm) |
| | Small | 0.08 min | | 0.10~0.12 |
| | Medium | 0.08 min | | 0.10~0.12 |
| | Large | 0.08 min | | 0.08~0.12 |
| | | | | |
| Physical Properties | ASTM D3578-10 | | | |
| | Before Aging | After aging at 70±2 °C<br>166±2 hrs | Before Aging | After aging at 70±2 °C<br>166±2 hrs |
| - Tensile Strength | 18Mpa min | 14Mpa min | 18.4~32.3Mpa | 18.7~24.5Mpa |
| - Ultimate Elongation | 650% min | 500% min | 955~1148% | 797~976% |
| - Stress at 500%<br>Elongation | 5.5Mpa max | -- | 4.8~5.4Mpa | -- |
| | | | | |
| | | | | |
| Freedom from Holes | ASTM D5151-11 | | Passed Standard Acceptance Criteria<br>Holes at Inspection Level I, AQL 2.5 | |
| Powder residue | ASTM D6124-11, ≤2mg/glove | | 0.68 mg/glove, Meets ASTM D6124-11<br>Results generated values below 2mg/glove of<br>residual powder | |
| Protein Content | ASTM D5712-10, ≤200µg/dm2 Aqueous<br>extractable protein | | Aqueous extractable protein: 70.8 mcg/dm2,<br>Meets ASTM D5712-10, Results generated<br>values below 200µg/dm² of Aqueous extractable<br>protein content | |
| Biocompatibility | Primary Skin Irritation in rabbits ISO | | Under the condition of the test, not an irritant. | |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows a logo with a circular design. The top half of the circle is blue, while the bottom half is green. Within the green portion, there are several curved lines that resemble blades of grass or leaves. The logo has a clean and modern look.
| 10993-10:2010-08-01 | |
|-------------------------------------------------------------------|----------------------------------------------------|
| Dermal sensitization in the guinea pig ISO<br>10993-10:2010-08-01 | Under the condition of the test, not a sensitizer. |
### • The Powdered Nitrile Patient Examination Gloves
| • The Powdered Nitrile Patient Examination Gloves | | | | |
|---------------------------------------------------|-------------------------------------------------------------------|-------------------------------------|---------------------------------------------------------------------------------------------------|-------------------------------------|
| Characteristics | Standard | | | Test Data |
| Dimension | ASTM D6319-10 | | | |
| Palm width | Size | (mm) | | (mm) |
| | Small | 80±10 | | 81~85 |
| | Medium | 95±10 | | 95~99 |
| | Large | 110±10 | | 110~114 |
| Overall Length | Size | (mm) | | (mm) |
| | Small | 220 min | | 251~255 |
| | Medium | 230 min | | 250~262 |
| | Large | 230 min | | 259~264 |
| Finger Thickness | Size | (mm) | | (mm) |
| | Small | 0.05 min | | 0.08 |
| | Medium | 0.05 min | | 0.08 |
| | Large | 0.05 min | | 0.08 |
| Palm Thickness | Size | (mm) | | (mm) |
| | Small | 0.05 min | | 0.08 |
| | Medium | 0.05 min | | 0.08 |
| | Large | 0.05 min | | 0.08 |
| | | | | |
| Physical Properties | ASTM D6319-10 | | | |
| | Before Aging | After aging at 70±2 °C<br>166±2 hrs | Before Aging | After aging at 70±2 °C<br>166±2 hrs |
| - Tensile Strength | 14Mpa min | 14Mpa min | 17.5~31.1Mpa | 15.7~18.5Mpa |
| - Ultimate Elongation | 500% min | 400% min | 761~948% | 421~839% |
| | | | | |
| Freedom from Holes | ASTM D5151-11 | | Passed Standard Acceptance Criteria | |
| | | | Holes at Inspection Level G-1, AQL 2.5 | |
| Powder Amount | ASTM D6124-11, ≤ 10mg/dm² | | 1.86 mg/dm², Meets ASTM D6124-11<br>Results generated values below 10mg/dm² of<br>residual powder | |
| | | | | |
| | | | | |
| Biocompatibility | Primary Skin Irritation in rabbits ISO<br>10993-10:2010-08-01 | | Under the condition of the test, not an irritant. | |
| | Dermal sensitization in the guinea pig ISO<br>10993-10:2010-08-01 | | Under the condition of the test, not a sensitizer. | |
| | | | | |
### • The Powder-free Nitrile Patient Examination Gloves
| Characteristics | Standard | Test Data |
|-----------------|---------------|-----------|
| Dimension | ASTM D6319-10 | |
| Palm width | Size (mm) | (mm) |
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows a logo with a circular shape. The top half of the circle is blue, and the bottom half is green. Inside the circle, there are several curved lines that resemble leaves or blades of grass. The lines are also green and are arranged in a fan-like pattern.
| | Small | 80±10 | | 81~85 |
|-----------------------|------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------|-------------------------------------------------------------------------------------------------------------|-------------------------------------|
| | Medium | 95±10 | | 94~99 |
| | Large | 110±10 | | 110~114 |
| Overall Length | Size | (mm) | | (mm) |
| | Small | 220 min | | 252~257 |
| | Medium | 230 min | | 255~260 |
| | Large | 230 min | | 257~263 |
| Finger Thickness | Size | (mm) | | (mm) |
| | Small | 0.05 min | | 0.08 |
| | Medium | 0.05 min | | 0.08 |
| | Large | 0.05 min | | 0.08 |
| Palm Thickness | Size | (mm) | | (mm) |
| | Small | 0.05 min | | 0.08 |
| | Medium | 0.05 min | | 0.08 |
| | Large | 0.05 min | | 0.08 |
| | | | | |
| Physical Properties | ASTM D6319-10 | | | |
| | Before Aging | After aging at 70±2 °C<br>166±2 hrs | Before Aging | After aging at 70±2 °C<br>166±2 hrs |
| - Tensile Strength | 14Mpa min | 14Mpa min | 18.8~29.5Mpa | 16.2~21.0Mpa |
| - Ultimate Elongation | 500% min | 400% min | 745~945% | 445~773% |
| | | | | |
| Freedom from Holes | ASTM D5151-11 | | Passed Standard Acceptance Criteria<br>Holes at Inspection Level G-1, AQL 2.5 | |
| Powder residue | ASTM D6124-11, ≤2mg/glove | | 0.72 mg/glove, Meets ASTM D6124-11<br>Results generated values below 2mg/glove of<br>residual powder | |
| Biocompatibility | Primary Skin Irritation in rabbits ISO<br>10993-10:2010-08-01<br>Dermal sensitization in the guinea pig ISO<br>10993-10:2010-08-01 | | Under the condition of the test, not an irritant.<br><br>Under the condition of the test, not a sensitizer. | |
## 5. Substantial Equivalent Based on Assessment of Non-Clinical Performance Data:
Bench tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
1) ASTM D3578-10, Standard Specification for Rubber Examination Gloves.
2) ASTM D5151-11, Standard Test Method for Detection of Holes in Medical Gloves.
3) ASTM D6124-11, Standard Test Method for Residual Powder on Medical Gloves.
4) ASTM D5712-10, Standard Test Method for the Analysis of Aqueous Extractable Protein in Natural Rs
Products Using the Modified Lowry Method.
5) ASTM D6319-10, Standard Specification for Nitrile Examination Gloves for Medical Application.
6) ISO 10993-10, Biological Evaluation of Medical Devices - Part 10: Tests for Irritation and Skin Sensitization.
## 6. Substantial Equivalent Based on Assessment of Clinical Performance Data:
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows a logo with a circular design. The logo features a blue circle with a green, stylized image of a plant or bird inside. The green image has several curved lines that resemble leaves or feathers. The overall design is simple and modern.
Clinical data was not including in this submission.
## 7. Substantially Equivalent Comparison Conclusion
## • For Powdered Latex Patient Examination Gloves
| Parameters | | New Device | Predicate Device |
|------------|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | 510(k) Number | K150612 | K072145 |
| 2 | Marketing<br>clearance date | -- | Feb 26th, 2008 |
| 3 | Device Name | Annuy Powdered Latex Patient Examination<br>Glove | Powdered Latex Patient Examination Glove |
| 4 | Manufacturer | Anhui Anyu Latex Products Co., Ltd | Zhanjiang Jiali Glove Products Co., LTD |
| 5 | Product Code | LYY | LYY |
| 6 | Regulation No. | 21 CFR 880.6250 | 21 CFR 880.6250 |
| 7 | Class | I | I |
| 8 | Intended use | A patient examination glove is a disposable<br>device intended for medical purposes that is<br>worn on the examiner's hand or finger to<br>prevent contamination between patient and<br>examiner. | A patient examination glove is a disposable<br>device intended for medical purposes that is<br>worn on the examiner's hand or finger to<br>prevent contamination between patient and<br>examiner. |
| 9 | Material | Natural Rubber Latex | Natural Rubber Latex |
| 10 | Design | Single use<br>Non-sterile<br>Powdered<br>Ambidextrous<br>Cuffed | Single use<br>Non-sterile<br>Powdered<br>Ambidextrous<br>Cuffed |
| 11 | Description | A device covering the hand and wrist area.<br>Gloves have separate sheaths or openings for<br>each finger and the thumb. | A device covering the hand and wrist area.<br>Gloves have separate sheaths or openings<br>for each finger and the thumb. |
| 12 | Size | Provided in sizes small, medium and large | Provided in sizes small, medium and large |
| 13 | Specifications | Powdered Latex Examination Gloves Meet<br>ASTM D3578-10 | Powdered Latex Examination Gloves Meet<br>ASTM D3578-05 |
| 14 | Dimensions<br>- Overall Length | Meets ASTM D3578-10<br>Small: 220mm min<br>Medium: 230mm min<br>Large: 230mm min | Meets ASTM D3578-05<br>Small: 220mm min<br>Medium: 230mm min<br>Large: 230mm min |
| 15 | Dimensions<br>- Palm Width | Meets ASTM D3578-10<br>Small: 80±10mm<br>Medium: 95±10mm<br>Large: 111±10mm | Meets ASTM D3578-05<br>Small: 80±10mm<br>Medium: 95±10mm<br>Large: 111±10mm |
| 16 | Dimensions<br>- Finger Thickness | Meets ASTM D3578-10<br>0.08mm min | Meets ASTM D3578-05<br>0.08mm min |
| | Dimensions<br>- Palm Thickness | Meets ASTM D3578-10<br>0.08mm min | Meets ASTM D3578-05<br>0.08mm min |
| 17 | Physical Properties | Meets ASTM D3578-10 | Meets ASTM D3578-05 |
| | | Before Aging: | Before Aging: |
| | | - Tensile Strength: 18Mpa min | - Tensile Strength: 18Mpa min |
| | | - Ultimate Elongation: 650% min | - Ultimate Elongation: 650% min |
| | | - Stress at 500% Elongation: 5.5Mpa max | - Stress at 500% Elongation: 5.5Mpa max |
| | | Meets ASTM D3578-10 | Meets ASTM D3578-05 |
| | | After aging at 70±2 °C 166±2 hrs: | After aging at 70±2 °C 166±2 hrs: |
| | | - Tensile Strength: 14Mpa min | - Tensile Strength: 14Mpa min |
| | | - Ultimate Elongation: 500% min | - Ultimate Elongation: 500% min |
| 18 | Freedom from | Meets ASTM D3578-10 and ASTM D5151-11 | Meets ASTM D3578-05 and ASTM |
| | Holes | Holes at Inspection Level I, AQL 2.5 | D5151-06 |
| 19 | Powder Amount | Meets ASTM D6124-11 | Meets ASTM D6124-06 |
| | | Below 10mg/dm² of residual powder | Below 10mg/dm² of residual powder |
| 20 | Protein Content | Meets ASTM D5712-10, | Meets ASTM D5712-05 |
| | | Below 200µg/dm² of aqueous extractable | Below 200µg/dm² of aqueous extractable |
| | | protein content | protein content |
| 21 | Absorbable donning | Cornstarch, | Cornstarch, |
| | or dusting powder | Powdered with absorbable dusting powder, | Powdered with absorbable dusting powder, |
| | | U.S.P. | U.S.P. |
| 22 | Biocompatibility | Meets ISO 10993-10:2010; | Meets ISO 10993-10:2006; |
| | | Skin Irritation: Under the condition of the | Skin Irritation: Under the condition of the |
| | | test, not an irritant. | test, not an irritant. |
| | | Skin Sensitization:Under the condition of the | Skin Sensitization:Under the condition of |
| | | test, not a sensitizer. | the test, not a sensitizer. |
| 23 | Color | Creamy white, no colorant | Same |
| 24 | Labeling Features | Include the required labeling: | Include the required labeling: |
| | | • Examination Gloves | • Examination Gloves |
| | | • Non-sterile | • Single Use Only |
| | | • Single Use Only | • Powdered |
| | | • Powdered | • Size, Serial No., Manufacturing Date and |
| | | • Ambidextrous | Quantity |
| | | • Natural rubber latex | • Natural rubber latex |
| | | • Device color: Creamy white | • Protein Label Claims |
| | | • Allergy warning | • Allergy warning |
| | | • Contents, Size, Lot No. and Manufacturing | • Manufacturer Name and Address |
| | | Date | |
| | | • Storage conditions | |
| | | • Manufacturer Name and Address | |
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows a logo with a circular design. The top half of the circle is blue, while the bottom half is green. Within the green portion, there are several curved lines that resemble blades of grass or stylized leaves. The overall design is simple and modern, suggesting a connection to nature or the environment.
# • For Powder-free Latex Patient Examination Gloves
| | Parameters | New Device | Predicate Device |
|-------------------------------------------------------------------------------------------------------|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | 510(k) Number | K150612…
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.