OVA1 Next Generation

K150588 · Vermillion, Inc. · ONX · Mar 18, 2016 · Obstetrics/Gynecology

Device Facts

Record IDK150588
Device NameOVA1 Next Generation
ApplicantVermillion, Inc.
Product CodeONX · Obstetrics/Gynecology
Decision DateMar 18, 2016
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.6050
Device ClassClass 2
AttributesReal-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K150588 · Mar 18, 2016OVA1 Next GenerationVermillion, Inc.Banked serum samples from the prospective, multi-site OVA500 pivotal studyBanked samples from a prior prospective study were used to conduct a side-by-side clinical validation of the OVA1 Next Generation test against the predicate device, comparing sensitivity and specificity in a cohort of 493 evaluable subjects.Banked samples; Clinical validation; Retrospective analysis; Ovarian adnexal mass

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
OVA500 Study (banked sample subset); Retrospective analysis of banked samples from a prospective, multi-site pivotal study; Follow-up/Duration: Not applicable; Study Period: Samples from a study published in Feb 2013Women over age 18 with an ovarian adnexal mass for which surgery was planned; Sample Size: 493 evaluable subjectsPredicate OVA1 device (K081754)Sensitivity and specificity for malignancy detection

Indications for Use

The OVA1 Next Generation test is a qualitative serum test that combines the results of five immunoassays into a single numeric result. It is indicated for women who meet the following criteria: over age 18, ovarian adnexal mass present for which surgery is planned, and not yet referred to an oncologist. The OVA1 Next Generation test is intended to be part of the preoperative evaluation to aid in assessing whether a woman who presents with an ovarian adnexal mass is at high or low likelihood of finding malignancy at surgery. The OVA1 Next Generation test must be interpreted in conjunction with an independent clinical and imaging evaluation. The test is not intended as a screening or stand-alone diagnostic assay.

Device Story

Device uses serum samples to measure five biomarkers (APO, CA 125 II, FSH, HE4, TRF) via Roche cobas 6000 system. Biomarker values are processed by OvaCalc v4.0.0 software to generate a single numeric risk score (0.0–10.0). A cutoff of 5.0 classifies patients as low or high risk. Used in clinical settings by physicians to aid preoperative evaluation of adnexal masses. Output assists clinicians in determining if a patient requires referral to an oncologist. Benefits include improved specificity over predicate, potentially reducing unnecessary surgeries or delayed diagnoses.

Clinical Evidence

Clinical validation used 493 evaluable samples from the prospective, multi-site OVA500 study. Bench testing included precision, reproducibility, sample stability, and interference studies. Results showed ~14% improvement in overall specificity with physician assessment (PA) compared to predicate. Sensitivity was clinically equivalent (~93.5% vs 95.7%). Standalone specificity improved by ~16%. Precision (%CV) was 1.54% (vs 4.09% for predicate).

Technological Characteristics

Uses five immunoassays (APO, CA 125 II, FSH, HE4, TRF) on Roche cobas 6000 system. APO/TRF measured turbidimetrically; CA 125 II/FSH/HE4 measured via chemiluminescence. Software (OvaCalc v4.0.0) calculates score. No electronic connection between analyzer and software. Standardized against WHO/CRM reference materials. Software is standalone PC-based application.

Indications for Use

Indicated for women >18 years old with an ovarian adnexal mass scheduled for surgery and not yet referred to an oncologist. Used as an aid to assess malignancy likelihood when independent clinical/radiological evaluation is non-indicative. Not for screening or stand-alone diagnosis.

Regulatory Classification

Identification

An ovarian/adnexal mass assessment test system is a device that measures one or more proteins in serum or plasma. It yields a single result for the likelihood that an adnexal pelvic mass in a woman, for whom surgery is planned, is malignant. The test is for adjunctive use, in the context of a negative primary clinical and radiological evaluation, to augment the identification of patients whose gynecologic surgery requires oncology expertise and resources.

Special Controls

The device is classified as Class II under regulation 21 CFR 866.6050 with special controls. The special control guidance document, "Class II Special Controls Guidance Document: Ovarian Adnexal Mass Assessment Score Test System" accompanies this device.

*Classification.* Class II (special controls). The special control for this device is FDA's guidance document entitled “Class II Special Controls Guidance Document: Ovarian Adnexal Mass Assessment Score Test System.” For the availability of this guidance document,*see* § 866.1(e).(c) *Black box warning.* Under section 520(e) of the Federal Food, Drug, and Cosmetic Act these devices are subject to the following restriction: A warning statement must be placed in a black box and must appear in all advertising, labeling, and promotional material for these devices. That warning statement must read:

Predicate Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY A. 510(k) Number: K150588 B. Purpose for Submission: New device C. Measurand: Score based on 5 serum analytes D. Type of Test: Software algorithm that combines five immunoassays into a single score E. Applicant: Vermillion, Inc. F. Proprietary and Established Names: OVA1 Next Generation G. Regulatory Information: 1. Regulation section: 21 CFR §866.6050, Ovarian adnexal mass assessment score test system 2. Classification: Class II 3. Product code: ONX, Serum, algorithm, ovarian cancer assessment test 4. Panel: Immunology (82) {1} # H. Intended Use: # 1. Intended use(s): The OVA1 Next Generation test is a qualitative serum test that combines the results of five immunoassays into a single numeric result. It is indicated for women who meet the following criteria: over age 18, ovarian adnexal mass present for which surgery is planned, and not yet referred to an oncologist. The OVA1 Next Generation test is an aid to further assess the likelihood that malignancy is present when the physician's independent clinical and radiological evaluation does not indicate malignancy. The test is not intended as a screening or stand-alone diagnostic assay. PRECAUTION: The OVA1 Next Generation test should not be used without an independent clinical and imaging evaluation and is not intended to be a screening test or to determine whether a patient should proceed to surgery. Incorrect use of the OVA1 Next Generation test carries the risk of unnecessary testing, surgery, and/or delayed diagnosis. # 2. Indication(s) for use: Same as Intended use # 3. Special conditions for use statement(s): For prescription use only # 4. Special instrument requirements: For use on Roche cobas® 6000 system # I. Device Description: The OVA1 Next Generation test consists of software, instruments, assays and reagents. The software incorporates the results of five serum biomarker concentrations from immunoassays run separately to calculate a single, unitless numeric result indicating a low or high risk of ovarian malignancy. The biomarkers and corresponding immunoassays and calibrators used to generate the numeric result (OVA1 Next Generation score) are: | Analyte | Reagent and Calibrator | Instrument | | --- | --- | --- | | Apolipoprotein A-1 (APO) | cobas APO A1 | Roche cobas® 6000: | | | C.f.a.s. Lipids | Roche cobas® c501 | | CA 125 II | cobas CA 125 II | Roche cobas® 6000: | | | CA 125 II Cal Set | Roche cobas® e601 | | Follicle Stimulating Hormone (FSH) | cobas FSH | Roche cobas® 6000: | | | FSH Cal Set II | Roche cobas® e601 | | Human epididymis protein 4 (HE4) | cobas HE4 | Roche cobas® 6000: | | | HE4 Cal Set | Roche cobas® e601 | {2} 3 | Analyte | Reagent and Calibrator | Instrument | | --- | --- | --- | | Transferrin (TRF) | cobas Transferrin C.f.a.s. Proteins | Roche cobas® 6000: Roche cobas® c501 | The biomarker immunoassays and reagents are sold separately from the OVA1 Next Generation software (OvaCalc). All immunoassays are run on the Roche cobas® 6000 system according to the manufacturer’s instructions as detailed in the product insert for each reagent. Users are instructed to use only qualified lot numbers for the immunoassays as posted on www.vermillion.com. Roche cobas® 6000 system is a fully automated, software-controlled system for clinical chemistry and immunoassay analysis. The OvaCalc software (v4.0.0) contains a proprietary algorithm that utilizes the results (values) from the five biomarker immunoassays. The assay values from the cobas® 6000 system are either imported into OvaCalc through a .csv file or manually entered into the OvaCalc user interface to generate an OVA1 Next Generation score between 0.0 and 10.0. **OVA1 Next Generation score:** Low probability of malignancy  Risk score &lt; 5.0 High probability of malignancy  Risk score ≥ 5.0 **J. Substantial Equivalence Information:** 1. Predicate device name(s) and 510(k) number(s): Vermillion OVA1, K081754 2. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | New Device OVA1 Next Generation | Predicate OVA1 Test | | Intended Use/ Indication for Use | The OVA1 Next Generation test is a qualitative serum test that combines the results of five immunoassays into a single numeric result. It is indicated for women who meet the following criteria: over age 18, ovarian adnexal mass present for which surgery is planned, and not yet referred to an oncologist. | The OVA1 Test is a qualitative serum test that combines the results of five immunoassays into a single numerical score. It is indicated for women who meet the following criteria: over age 18; ovarian adnexal mass present for which surgery is planned, and not yet referred to an oncologist. | | | The OVA1 Next Generation test is an aid to further assess the likelihood that malignancy is present when the physician’s independent clinical and radiological evaluation does not indicate malignancy. | The OVA1 Test is an aid to further assess the likelihood that malignancy is present when the physician’s independent clinical and radiological evaluation does not indicate malignancy. The test is not intended as a screening or stand-alone diagnostic assay. | {3} | Differences | | | | --- | --- | --- | | Item | New Device OVA1 Next Generation | Predicate OVA1 Test | | Analytes | Roche Elecsys APO, CA 125, TRF, FSH, HE4 | Roche Elecsys CA 125 and Siemens BN II APO, TRF, Prealbumin, β2 Microglobulin | | Equation used for test | One equation with one cut-off | One equation with two cut-offs depending on menopausal status | | Clinical cutoff | Pre-menopausal and Post-menopausal: • OVA1 risk score < 5.0 Low probability for malignancy • OVA1 risk score ≥ 5.0 High probability for malignancy | Pre-menopausal: • OVA1 risk score < 5.0 Low probability for malignancy • OVA1 risk score ≥ 5.0 High probability for malignancy Post-menopausal: • OVA1 risk score < 4.4 Low probability for malignancy • OVA1 risk score ≥ 4.4 High probability for malignancy | | Platform | Roche cobas e601 (CA125, FSH and HE4) Roche cobas c501 (APO and TRF) | Roche Elecsys 2010 (CA 125) Siemens BNII (APO, TRF, Prealbumin, β2 Microglobulin) | K. Standard/Guidance Document Referenced (if applicable): {4} - FDA Guidance “Class II Special Controls Guidance Document: Ovarian Adnexal Mass Assessment Score Test System.” - FDA Guidance “Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.” - ISO 14971:2012 Medical Devices-Application of Risk Management to Medical Devices, International Organization for Standardization. - CLSI guideline EP05-A2, “Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline.” - CLSI guideline EP07-A2, “Interference Testing in Clinical Chemistry; Approved Guideline – Second Edition.” L. Test Principle: The individual assays for APO and TRF each contain a biomarker specific polyclonal antibody which forms an immune complex with the target when reacted with a serum specimen. The levels of immune complexes can be measured turbidimetrically and are proportional to the concentration of biomarker in the serum specimen for each specific assay. The individual assays for CA 125 II, FSH and HE4 each use two mouse monoclonal antibodies to their respective biomarkers. The quantity of each biomarker present is then measured by chemiluminescence emission. The Cobas 6000 is an automated analyzer with electrochemiluminescence detection. The amount of analyte in each assay is determined against the calibration curve. Each assay uses its own specific calibrator and controls. The user enters results of the five analytes manually into an Excel spreadsheet together with the headers needed by OvaCalc Software. There is no physical or electronic connection between the immunoassay devices and the OvaCalc Software. Using an algorithm and the values of these five analytes, the OvaCalc Software generates a single unit-less numerical score from 0.0 to 10.0. M. Performance Characteristics (if/when applicable): 1. Analytical performance: All results met the manufacturer’s pre-determined acceptance criteria. a. Precision: Precision performance of the OVA1 Next Generation test was evaluated in accordance with CLSI guideline EP05-A2 “Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline.” Five pooled serum samples spanning the OVA1 Next Generation score range (three close to the cut-offs, one low and one high value) were tested over 20 days, two runs per day, and two replicates per run by multiple operators (equal to 80 replicate determinations per pool). There were no un-evaluable results. Total percent coefficient of variation (%CV) for all pools is 1.54%. {5} | Specimen | n | OVA1 Next Generation (Mean score) | Within-run | | Between-run | | Total | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | | | SD | CV% | SD | CV% | SD | %CV | | 1 | 80 | 3.30 | 0.05 | 1.62 | 0.01 | 0.34 | 0.07 | 1.95 | | 2 | 80 | 4.11 | 0.07 | 1.68 | 0.01 | 0.27 | 0.09 | 2.06 | | 3 | 80 | 5.08 | 0.13 | 2.57 | 0.06 | 1.25 | 0.16 | 3.16 | | 4 | 80 | 8.16 | 0.05 | 0.61 | 0.00 | 0.00 | 0.06 | 0.67 | | 5 | 80 | 8.50 | 0.00 | 0.00 | 0.00 | 0.00 | 0.00 | 0.00 | Reproducibility: Five pooled serum samples spanning the OVA1 Next Generation score range (three close to the cut-offs, one low and one high value) were tested in duplicate, two runs each day, over six non-consecutive days, by two operators at each of three sites. Each operator performed the test on three nonconsecutive days, i.e., operator 1 ran the test on days one, three, and five; operator 2 ran the test on days two, four, and six. At each site, the test was run with the same lots of calibrators, kit reagents, and controls for the duration of the study. Each operator performed the complete analysis of the OVA1Next Generation test for the day on the cobas® 6000 and imported the five biomarkers into Vermillion's OvaCalc Software for generation of the OVA1Next Generation score. For an assessment of the components of imprecision in the reproducibility study, the mean score, SD and %CV of each of the five pools were estimated and shown in the table below. The overall %CV including all sites was 1.63%. | Parameters | Specimen | | | | | | | --- | --- | --- | --- | --- | --- | --- | | | | 1 | 2 | 3 | 4 | 5 | | OVA1 Next Generation test results | | | | | | | | Mean score | | 3.28 | 4.08 | 5.09 | 8.16 | 8.50 | | Repeatability (within-run) | SD | 0.05 | 0.06 | 0.11 | 0.05 | 0.00 | | | %CV | 1.44 | 1.38 | 2.25 | 0.56 | 0.00 | | Between-run | SD | 0.02 | 0.07 | 0.10 | 0.03 | 0.00 | | | %CV | 0.68 | 1.59 | 2.03 | 0.39 | 0.00 | | Between-day | SD | 0.00 | 0.00 | 0.07 | 0.00 | 0.00 | | | %CV | 0.00 | 0.00 | 1.43 | 0.00 | 0.00 | | Between-operator | SD | 0.00 | 0.00 | 0.04 | 0.00 | 0.00 | | | %CV | 0.00 | 0.00 | 0.69 | 0.00 | 0.00 | | Between-site | SD | 0.00 | 0.02 | 0.03 | 0.04 | 0.00 | | | %CV | 0.00 | 0.55 | 0.67 | 0.48 | 0.00 | | Reproducibility (total) | SD | 0.05 | 0.09 | 0.18 | 0.06 | 0.00 | | | %CV | 1.59 | 2.15 | 3.43 | 0.67 | 0.00 | | Individual biomarker immunoassay results | | | | | | | | APO, mg/dL | | | | | | | | Mean concentration | | 157.96 | 153.09 | 166.68 | 149.69 | 133.07 | | Repeatability (within-run) | SD | 1.45 | 1.57 | 2.16 | 1.79 | 1.91 | | | %CV | 0.92 | 1.03 | 1.30 | 1.19 | 1.43 | | Between-run | SD | 2.30 | 2.19 | 2.67 | 1.87 | 2.30 | | | %CV | 1.45 | 1.43 | 1.60 | 1.25 | 1.73 | {6} | Between-day | SD | 0.50 | 1.60 | 0.00 | 0.00 | 0.00 | | --- | --- | --- | --- | --- | --- | --- | | | %CV | 0.32 | 1.04 | 0.00 | 0.00 | 0.00 | | Between-operator | SD | 0.55 | 0.72 | 0.91 | 0.00 | 0.00 | | | %CV | 0.35 | 0.47 | 0.55 | 0.00 | 0.00 | | Between-site | SD | 0.87 | 0.37 | 1.64 | 1.08 | 0.00 | | | %CV | 0.55 | 0.24 | 0.98 | 0.72 | 0.00 | | Reproducibility (total) | SD | 2.88 | 3.19 | 3.77 | 2.73 | 2.98 | | | %CV | 1.82 | 2.08 | 2.26 | 1.82 | 2.24 | | CA 125 II, IU/mL | | | | | | | | Mean concentration | | 13.58 | 22.39 | 38.34 | 246.80 | 868.49 | | Repeatability (within-run) | SD | 0.30 | 0.43 | 0.55 | 3.16 | 11.16 | | | %CV | 2.19 | 1.90 | 1.44 | 1.28 | 1.29 | | Between-run | SD | 0.00 | 0.00 | 0.15 | 1.51 | 0.00 | | | %CV | 0.00 | 0.00 | 0.38 | 0.61 | 0.00 | | Between-day | SD | 0.13 | 0.18 | 0.14 | 0.00 | 5.13 | | | %CV | 0.96 | 0.78 | 0.37 | 0.00 | 0.59 | | Between-operator | SD | 0.00 | 0.00 | 0.08 | 0.00 | 3.72 | | | %CV | 0.00 | 0.00 | 0.21 | 0.00 | 0.43 | | Between-site | SD | 0.02 | 0.19 | 0.46 | 3.77 | 14.32 | | | %CV | 0.15 | 0.85 | 1.19 | 1.53 | 1.65 | | Reproducibility (total) | SD | 0.32 | 0.48 | 0.70 | 4.67 | 17.27 | | | %CV | 2.38 | 2.16 | 1.83 | 1.89 | 1.99 | | FSH, mIU/mL | | | | | | | | Mean concentration | | 25.74 | 29.08 | 32.48 | 31.84 | 34.66 | | Repeatability (within-run) | SD | 0.23 | 0.26 | 0.35 | 0.31 | 0.43 | | | %CV | 0.90 | 0.89 | 1.08 | 0.97 | 1.24 | | Between-run | SD | 0.28 | 0.35 | 0.42 | 0.30 | 0.37 | | | %CV | 1.10 | 1.19 | 1.28 | 0.95 | 1.06 | | Between-day | SD | 0.37 | 0.42 | 0.31 | 0.50 | 0.43 | | | %CV | 1.42 | 1.45 | 0.96 | 1.57 | 1.23 | | Between-operator | SD | 0.00 | 0.00 | 0.00 | 0.00 | 0.00 | | | %CV | 0.00 | 0.00 | 0.00 | 0.00 | 0.00 | | Between-site | SD | 0.33 | 0.28 | 0.37 | 0.36 | 0.46 | | | %CV | 1.29 | 0.98 | 1.13 | 1.12 | 1.31 | | Reproducibility (total) | SD | 0.58 | 0.64 | 0.69 | 0.71 | 0.80 | | | %CV | 2.25 | 2.21 | 2.13 | 2.24 | 2.30 | | HE4, pmol/L | | | | | | | | Mean concentration | | 60.33 | 55.96 | 81.10 | 213.18 | 532.00 | | Repeatability (within-run) | SD | 0.66 | 0.65 | 0.81 | 1.76 | 5.57 | | | %CV | 1.09 | 1.15 | 1.00 | 0.82 | 1.05 | | Between-run | SD | 0 | 0.13 | 0.52 | 0.51 | 1.96 | | | %CV | 0 | 0.23 | 0.65 | 0.24 | 0.37 | | Between-day | SD | 0.49 | 0.40 | 0.61 | 2.03 | 5.98 | | | %CV | 0.81 | 0.72 | 0.75 | 0.95 | 1.12 | {7} | Between-operator | SD | 0.00 | 0.00 | 0.00 | 0.00 | 0.00 | | --- | --- | --- | --- | --- | --- | --- | | | %CV | 0.00 | 0.00 | 0.00 | 0.00 | 0.00 | | Between-site | SD | 1.07 | 1.05 | 1.45 | 3.27 | 8.95 | | | %CV | 1.77 | 1.87 | 1.79 | 1.53 | 1.68 | | Reproducibility (total) | SD | 1.19 | 1.153 | 1.65 | 3.81 | 11.12 | | | %CV | 1.98 | 2.06 | 2.03 | 1.79 | 2.09 | | TRF, mg/dL | | | | | | | | Mean concentration | | 288.69 | 276.82 | 276.75 | 255.80 | 225.31 | | Repeatability (within-run) | SD | 5.28 | 4.88 | 4.45 | 4.36 | 3.93 | | | %CV | 1.83 | 1.76 | 1.61 | 1.70 | 1.75 | | Between-run | SD | 5.49 | 3.47 | 3.78 | 4.10 | 2.34 | | | %CV | 1.90 | 1.25 | 1.37 | 1.60 | 1.04 | | Between-day | SD | 0.00 | 0.00 | 1.77 | 1.30 | 0.00 | | | %CV | 0.00 | 0.00 | 0.64 | 0.51 | 0.00 | | Between-operator | SD | 0.00 | 0.00 | 2.06 | 0.00 | 0.52 | | | %CV | 0.00 | 0.00 | 0.74 | 0.00 | 0.23 | | Between-site | SD | 1.56 | 3.25 | 1.08 | 3.87 | 1.49 | | | %CV | 0.54 | 1.17 | 0.39 | 1.51 | 0.66 | | Reproducibility (total) | SD | 7.69 | 6.54 | 6.43 | 6.88 | 4.76 | | | %CV | 2.67 | 2.36 | 2.32 | 2.69 | 2.11 | Lot-to-lot variability was evaluated and changes in the biomarkers $\leq \pm 3\%$ due to different reagent lots resulted in difference of sensitivity of OVA1 Next Generation score ranging from $-2.2\%$ to $3.3\%$ and specificity of OVA1 Next Generation score ranging from $-5.5\%$ to $4.5\%$ . # b. Linearity/assay reportable range: For each analyte, measurement linearity (as claimed in the package inserts for the individual analytes and shown in the table below) was demonstrated for measurement intervals corresponding to those used in the OVA1 Next Generation test. | Analyte | Measuring Range | | --- | --- | | Apolipoprotein A-1 (APO) | 20–400 mg/dL | | CA 125 II | 0.6–5000 IU/mL | | Follicle Stimulating Hormone (FSH) | 0.10–200 mIU/mL | | Human epididymis protein 4 (HE4) | 15.0–1500 pmol/L | | Transferrin (TRF) | 10–520 mg/dL | c. Traceability, Stability, Expected values (controls, calibrators, or methods): i. Traceability {8} Each biomarker immunoassay uses its own calibrator and controls. For APO, the method has been standardized against IFCC SP1-01 (WHO IRP October 1992). For TRF, the method has been standardized against BCR470/CRM 470 (Reference Preparation for Proteins un Human Serum). For CA 125 II, the method is standardized against Enzymun-Test CA 125 II method which in turn has been standardized against the CA125 II RIA from Fujirebio Diagnostics. For FSH, the method is standardized against Enzymun-Test FSH method which in turn has been standardized against the $2^{\text{nd}}$ IRP WHO reference standard 79/549. For HE4, the method is standardized against the HE4 EIA method from Fujirebio Diagnostics, Inc. # ii. Reagent Stability Closed/open vial and on-board: For each biomarker immunoassay used in the OVA1 Next Generation test, users are instructed to refer to the storage and stability information in the package insert. This information is summarized in the table below. | Analyte | Stability Data | | --- | --- | | Apolipoprotein A-1 (APO) | Closed vial at 2–8°C: up to the expiration date On-board in use and refrigerated: 12 weeks | | CA 125 II | Closed vial at 2–8°C: up to the expiration date Open vial at 2–8°C: 12 weeks On-board: 6 weeks | | Follicle Stimulating Hormone (FSH) | Closed vial at 2–8°C: up to the expiration date Open vial at 2–8°C: 12 weeks On-board: 8 weeks | | Human epididymis protein 4 (HE4) | Closed vial at 2–8°C: up to the expiration date Open vial at 2–8°C: 12 weeks On-board: 28 days | | Transferrin (TRF) | Closed vial at 2–8°C: up to the expiration date On-board in use and refrigerated: 12 weeks | # iii. Specimen Stability Seven serum pools spanning the OVA1 Next Generation risk score range (two close to the cut-off, one low and four high values) were tested to determine sample storage and freeze/thaw stability. The mean, SD, mean change, and %mean change from Day 0 (the day of specimen collection) were used to describe the OVA1 Next Generation risk score for each pool stored refrigerated up to nine days or four freeze/thaw cycles. The results support the following stability claims: fresh serum stored can be stored up to eight days between $2 - 8^{\circ}\mathrm{C}$ and may be subjected to four freeze/thaw cycles prior to testing. # d. Detection limit: The limits of detection and limits of quantitation reported in each assay's package insert are summarized in the table below. They were confirmed and incorporated into the algorithm such {9} that results outside of the measuring interval are not imported and do not yield an OVA1 Next Generation score. | Analyte | Detection limit | | --- | --- | | Apolipoprotein A-1 (APO) | 0.03 g/L (1.07 μmol/L) | | CA 125 II | 0.600 IU/mL | | Follicle Stimulating Hormone (FSH) | 0.100 mIU/mL | | Human epididymis protein 4 (HE4) | LoB: 5.0 pmol/L LoD: 15.0 pmol/L LoQ: 20.0 pmol/L | | Transferrin (TRF) | 0.1 g/L (1.26 μmol/L) | # e. Analytical specificity: # i. Endogenous Interference: Three pooled serum samples with low ( $\sim 3.28$ ), medium ( $\sim 5.15$ ) and high ( $\sim 8.50$ ) OVA1 Next Generation scores were evaluated for interference by hemoglobin, bilirubin (conjugated and unconjugated), triglycerides and rheumatoid factor. The effect of each interfering substance on the OVA1 Next Generation score was assessed using a mean of four repeated measurements on each sample spiked with the potentially interfering substances and compared to control measurements of samples without the interfering substances but with the same amount of solvent. Based on the manufacturer's acceptance limits for non-interference, none of the interfering substances demonstrated significant interference on the OVA1 Next Generation score up to the concentrations evaluated. | | OVA1 Next Generation scores | | | | | --- | --- | --- | --- | --- | | | | Low | Medium | High | | Interference substance | Test concentration | Mean change (%) in score compared to control (95% CI of difference) | | | | Bilirubin Unconjugated | 0.3 g/L | 0.92 (-0.10 to 0.15) | -0.96 (-0.32 to 0.22) | 0.0 (0.0 to 0.0) | | | 0.9 g/L | 2.94 (-0.02 to 0.22) | -1.23 (-0.34 to 0.19) | 0.0 (0.0 to 0.0) | | Bilirubin Conjugated | 0.3 g/L | 1.52 (-0.11 to 0.21) | 1.87 (-0.25 to 0.45) | 0.0 (0.0 to 0.0) | | | 0.9 g/L | 4.02 (-0.01 to 0.31) | 2.52 (-0.20 to 0.50) | 0.0 (0.0 to 0.0) | | Hemoglobin | 5.0 g/L | 1.52 (-0.06 to 0.16) | 1.36 (-0.17 to 0.32) | 0.0 (0.0 to 0.0) | | | 9.0 g/L | 0.61 (-0.09 to 0.14) | 1.26 (-0.17 to 0.32) | 0.0 (0.0 to 0.0) | | Triglycerides | 2.0 g/L | -0.61 (-0.12 to 0.07) | 4.37 (0.01 to 0.44) | 0.0 (0.0 to 0.0) | | | 4.6 g/L | -0.93 | -0.95 | 0.0 | {10} | | | (-0.12 to 0.07) | (-0.27 to 0.17) | (0.0 to 0.0) | | --- | --- | --- | --- | --- | | | 10.0 g/L | 0.0 (-0.10 to 0.10) | -0.37 (-0.24 to 0.19) | 0.0 (0.0 to 0.0) | | Rheumatoid Factor | 250 IU/mL | -1.54 (-0.12 to 0.02) | -1.91 (-0.34 to 0.14) | 0.0 (0.0 to 0.0) | | | 1000 IU/mL | -0.61 -0.10 to 0.05 | -3.6 -0.44 to 0.04 | 0.0 (0.0 to 0.0) | f. Assay cut-off: See clinical cut-off 2. Comparison studies: a. Method comparison with predicate device: For analysis of agreement with the predicate device, a total of 493 preoperatively collected serum specimens from pre-menopausal and post-menopausal women presenting with an adnexal mass requiring surgical intervention (refer to the section 'Clinical Studies' for a description of the clinical study and patient demographics) were assayed on both OVA1 Next Generation and OVA1 Test to generate OVA1 scores. All women had a physician's pre-surgical assessment (PA) by a non-gynecological oncologist and clinical prediction of malignant ovarian tumor (prior to surgery) using the combination of PA and OVA1 score (dual assessment) if either the PA was malignant or the OVA1 score indicated high risk for finding malignancy on surgery. The comparison of performance for risk stratification between dual assessment of PA with OVA1 Next Generation (PA + OVA1 Next Generation) and dual assessment of PA with OVA1 Test (PA + OVA1 Test) for all evaluable subjects, and malignant and benign cases as determined by pathology is summarized in the table below. Results showed that PA+OVA1 Next Generation and PA+OVA1 Test agreed on 187 high risk cases and 168 low risk cases for a total percentage agreement of 355 of 493 cases (72%). For risk stratification agreement of malignant cases, PA+OVA1 Next Generation and PA+OVA1 Test agreed on 88 high risk cases and 2 low risk cases (misclassified) for a total percentage agreement of 86 of 92 cases (93.5%). For benign cases, PA+OVA1 Next Generation and PA+OVA1 Test agreed on the classification of 166 of 401 benign cases (41% of all benign cases) but incorrectly classified 103 benign cases as high risk (26%). PA+OVA1 Next Generation correctly classified 94 benign cases as low risk which PA+OVA1 Test classified as high risk (23% of benign cases correctly classified by PA+OVA1 Next Generation but not PA+OVA1 Test). PA+OVA1 Next Generation incorrectly classified 38 benign cases as high risk which PA+OVA1 Test classified as low risk (9% of benign cases correctly classified by PA+OVA1 Test but not PA+OVA1 Next Generation). Overall, PA+OVA1 Next Generation showed a net improvement of 14% in the classification of benign subjects. {11} 12 | All Evaluable Subjects | | | | | | --- | --- | --- | --- | --- | | | Dual Assessment of PA with OVA1 Test | | Total | | | | | High risk | | Low risk | | Dual Assessment of PA with OVA1 Next Generation | High risk | 187 | 40 | 227 | | | Low risk | 98 | 168 | 266 | | Total | | 285 | 208 | 493 | | Positive Percent Agreement: 65.6% (87/285) 95% CI: 59.9% to 70.9% Negative Percent Agreement: 80.8% (168/208) 95% CI: 74.9% to 85.5% Total Percent Agreement: 72.0% (355/493) 95% CI: 67.9% to 75.8% | | | | | | Malignant Cases | | | | | | | Dual Assessment of PA with OVA1 Test | | Total | | | | | High risk | | Low risk | | Dual Assessment of PA with OVA1 Next Generation | High risk | 84 | 2 | 86 | | | Low risk | 4 | 2 | 6 | | Total | | 88 | 4 | 92 | | Total Percent Agreement: 93.5% (86/92) 95% CI: 86.5% to 97.0% | | | | | | Benign Cases | | | | | | | Dual Assessment of PA with OVA1 Test | | Total | | | | | High risk | | Low risk | | Dual Assessment of PA with OVA1 Next Generation | High risk | 103 | 38 | 141 | | | Low risk | 94 | 166 | 260 | | Total | | 197 | 204 | 401 | | Total Percent Agreement: 67.1% (269/401) 95% CI: 62.3% to 71.5% | | | | | The proportions of high and low risk classifications between PA+OVA1 Next Generation and PA+OVA1 Test stratified by pre-menopausal subjects is summarized in the table below: | Pre-menopausal Subjects | | | | | | --- | --- | --- | --- | --- | | | Dual Assessment of PA with OVA1 Test | | Total | | | | | High risk | | Low risk | | Dual Assessment of PA with OVA1 Next Generation | High risk | 84 | 24 | 108 | | | Low risk | 45 | 123 | 168 | | Total | | 129 | 147 | 276 | {12} 13 | Positive Percent Agreement: 65.1% (84/129) 95% CI: 56.6% to 72.8% Negative Percent Agreement : 83.7% (123/147) 95% CI: 76.9% to 88.8% Total Percent Agreement : 75.0% (207/276) 95% CI: 69.6% to 79.7% | | | | | | --- | --- | --- | --- | --- | | Malignant Cases | | | | | | | Dual Assessment of PA with OVA1 Test | | Total | | | | | High risk | | Low risk | | Dual Assessment of PA with OVA1 Next Generation | High risk | 27 | 1 | 28 | | | Low risk | 2 | 1 | 3 | | Total | | 29 | 2 | 31 | | Total Percent Agreement: 90.3% (28/31) 95% CI: 75.1% to 96.7% | | | | | | Benign Cases | | | | | | | Dual Assessment of PA with OVA1 Test | | Total | | | | | High risk | | Low risk | | Dual Assessment of PA with OVA1 Next Generation | High risk | 57 | 23 | 80 | | | Low risk | 43 | 122 | 165 | | Total | | 100 | 145 | 245 | | Total Percent Agreement: 73.1% (179/245) 95% CI: 67.2% to 78.2% | | | | | The proportions of high and low risk classifications between PA+OVA1 Next Generation and PA+OVA1 Test stratified by post-menopausal subjects is summarized in the table below: | Post-menopausal Subjects | | | | | | --- | --- | --- | --- | --- | | | Dual Assessment of PA with OVA1 Test | | Total | | | | | High risk | | Low risk | | Dual Assessment of PA with OVA1 Next Generation | High risk | 103 | 16 | 119 | | | Low risk | 53 | 45 | 98 | | Total | | 156 | 61 | 217 | | Positive Percent Agreement: 66.0% (103/156) 95% CI: 58.3% to 73.0% Negative Percent Agreement: 73.8% (45/61) 95% CI: 61.6% to 83.2% Total Percent Agreement: 68.2% (148/217) 95% CI: 61.7% to 74.0% | | | | | | Malignant Cases | | | | | | | Dual Assessment of PA with OVA1 Test | | Total | | | | | High risk | | Low risk | | Dual Assessment of PA with OVA1 Next Generation | High risk | 57 | 1 | 58 | | | Low risk | 2 | 1 | 3 | | Total | | 59 | 2 | 61 | | Total Percent Agreement: 95.1% (58/61) 95% CI:86.5% to 98.3% | | | | | | Benign Cases | | | | | | | | Dual Assessment of PA with | | | | | | OVA1 Test | | | {13} | | OVA1 Test | | Total | | | --- | --- | --- | --- | --- | | | | High risk | | Low risk | | Dual Assessment of PA with OVA1 Next Generation | High risk | 46 | 15 | 61 | | | Low risk | 51 | 44 | 95 | | Total | | 97 | 59 | 156 | | Total Percent Agreement: 57.7% (90/156) 95% CI: 49.8% to 65.2% | | | | | b. Matrix comparison: Serum is the only claimed matrix for each of the five analytes evaluated. 3. Clinical studies: a. Clinical sensitivity and specificity: The ability of OVA1 Next Generation to contribute to the physician's pre-surgical assessment (PA) was evaluated in two studies. The first study used archived serum specimens collected from a prospective, multi-site clinical study for pre-menopausal and post-menopausal women presenting with an adnexal mass requiring surgical intervention (Bristow et al. 2013; Ovarian malignancy risk stratification of the adnexal mass using a multivariate index assay. Gynecologic Oncology 128; 252-259). The clinical study subject enrollment centers are representative of institutions where ovarian tumor subjects potentially undergo a gynecologic examination. The specimens were collected at 27 non-gynecologic oncology practices. A total of 493 women (age range 18-87) were evaluable in the study. For each patient, an initial PA was completed by a non-gynecological oncologist, providing the assessment of the patient's mass as benign (negative) or malignant (positive) based upon the information available to the non-gynecological oncologist during their work-up of the patient. The corresponding histopathology reports were collected after surgery. Using a preoperatively collected serum sample, the OVA1 Next Generation test score was determined and the patient was stratified into a low or high risk group for finding malignancy on surgery. For each patient, the OVA1 Next Generation test result was compared to the pathology report from biopsied tissue. Malignancy included epithelial ovarian cancer (EOC), other primary ovarian malignancy, ovarian tumor with low malignant potential, malignancy extending to, or metastatic to ovaries, and malignancy that neither arose in nor involved the ovaries. A positive result, indicating high probability of malignancy for pre-menopausal and post-menopausal subjects was defined as an OVA1 Next Generation high risk result (risk score value of 5.0 or greater). Subject enrollment demographics and pathology diagnoses are presented in the table below. {14} | | All Evaluable | Pre-menopausal | Post-menopausal | | --- | --- | --- | --- | | N | 493 | 276 | 217 | | Age, years | | | | | Mean (SD) | 48.6 (14.16) | 39.5 (8.96) | 60.2 (10.74) | | Median | 48 | 41 | 60 | | Range (min, max) | (18,87) | (18,60) | (33,87) | | Ethnicity/race, n (%) | | | | | White | 347 (70.4) | 173 (62.7) | 174 (80.2) | | Black or African American | 81 (16.4) | 54 (19.6) | 27 (12.4) | | Hispanic or Latino | 46 (9.3) | 36 (13.0) | 10 (4.6) | | Asian | 13 (2.6) | 8 (2.9) | 5 (2.3) | | Native Hawaiian/Pacific Islander | 1 (0.2) | 1 (0.4) | 0 (0.0) | | Other | 5 (1.0) | 4 (1.4) | 1 (0.5) | | No. of pregnancies, n (%) | | | | | None | 80 (16.2) | 56 (20.3) | 24 (11.1) | | 1 | 86 (17.4) | 52 (18.8) | 34 (15.7) | | 2 | 131 (26.6) | 70 (25.4) | 61 (28.1) | | 3 | 94 (19.1) | 50 (18.1) | 44 (20.3) | | 4 or more | 102 (20.7) | 48 (17.4) | 54 (24.9) | | Histopathological classification, n (%) | | | | | Benign ovarian conditions | 401 (81.3) | 245 (88.8) | 156 (71.9) | | Epithelial ovarian cancer (EOC) | 60 (12.2) | 18 (6.5) | 42 (19.4) | | Other primary ovarian malignancies (non-EOC) | 5 (1.0) | 5 (1.8) | 0 (0.0) | | Ovarian tumors with low malignant potential (LMP) | 17 (3.4) | 5 (1.8) | 12 (5.5) | | Non-primary ovarian malignancies with involvement of the ovaries | 6 (1.2) | 2 (0.7) | 4 (1.8) | | Non-primary ovarian malignancies with no involvement of ovaries | 4 (0.8) | 1 (0.4) | 3 (1.4) | | Tumor stage, n (% of all primary malignant ovarian tumor) | | | | | Stage 1 | 28 (43.1) | 9 (39.1) | 19 (45.2) | | Stage 2 | 7 (10.8) | 2 (8.7) | 5 (11.9) | | Stage 3 | 25 (38.5) | 10 (43.5) | 15 (35.7) | | Stage 4 | 5 (7.7) | 2 (8.7) | 3 (7.1) | # Performance for Pre-menopausal and Post-menopausal Subjects Combined Among 493 subjects, there were 92 subjects with malignancy by pathology and 401 subjects with no malignancy by pathology. OVA 1 Next Generation performance was compared with clinical pathology: {15} | Malignancy by Pathology | | | | | | --- | --- | --- | --- | --- | | | | Physician Assessment (PA) | | Total | | OVA1 Next Generation | Positive | 66 | 18 | 84 | | | Negative | 2 | 6 | 8 | | Total | | 68 | 24 | 92 | | No Malignancy by Pathology | | | | | | | | Physician Assessment (PA) | | Total | | OVA1 Next Generation | Positive | 12 | 112 | 124 | | | Negative | 17 | 260 | 277 | | Total | | 29 | 372 | 401 | The analysis examined whether patient referral to gynecologic oncologist is supported when dual assessment is determined to be positive (either OVA1 Next Generation or PA is positive or both are positive). A dual assessment is negative when both OVA1 Next Generation and PA are negative. | Performance | Physician Assessment (PA) | Dual Assessment (PA and OVA1 Next Generation result used) | | --- | --- | --- | | Sensitivity | 73.9% (68/92) | 93.5% (86/92) | | 95% CI | 64.1% to 81.8% | 86.5% to 97.0% | | Specificity | 92.8 (372/401) | 64.8% (260/401) | | 95% CI | 89.8% to 94.9% | 60.0% to 69.4% | | PPV | 70.1% (68/97) | 37.9% (86/227) | | 95% CI | 61.9% to 77.3% | 34.5% to 41.4% | | NPV | 93.9% (372/396) | 97.7 (260/266) | | 95% CI | 91.8% to 95.7% | 95.4 to 98.9 | | Prevalence | 18.7% (92/493) | | With dual assessment, sensitivity for malignancy increased from $73.9\%$ to $93.5\%$ . Specificity for malignant diagnoses decreased from $92.8\%$ to $64.8\%$ with dual assessment. PPV of the dual assessment decreased from $70\%$ to $38\%$ . However, NPV of the dual assessment significantly increased from $93.9\%$ to $97.7\%$ , supporting improved performance by dual assessment. The confidence interval for the observed $3.8\%$ increase was $1.8\%$ to $6.2\%$ , Performance characteristics of the OVA1 Next Generation test in combination with PA for the 493 patients classified according to their menopausal status are presented separately in the tables below: ## Pre-menopausal subjects Among 276 pre-menopausal subjects, there were 31 subjects with malignancy by pathology and 245 subjects with no malignancy by pathology. OVA 1 Next Generation was compared to the PA alone: {16} 17 | Malignancy by Pathology | | | | | | --- | --- | --- | --- | --- | | | | Physician Assessment (PA) | | Total | | OVA1 Next Generation | Positive | 23 | 5 | 28 | | | Negative | 0 | 3 | 3 | | Total | | 23 | 8 | 31 | | | | | | | | No Malignancy by Pathology | | | | | | | | Physician Assessment (PA) | | Total | | OVA1 Next Generation | Positive | 5 | 65 | 70 | | | Negative | 10 | 165 | 175 | | Total | | 15 | 230 | 245 | With dual assessment, sensitivity for malignancy increased from 74.2% to 90.3%. PPV of the dual assessment decreased from 60.5% to 25.9% and NPV of the dual assessment increased from 96.6% to 98.2%. The confidence interval for the observed 1.6% increase was 0.1% to 3.7%. | Performance | Physician Assessment (PA) | Dual Assessment (PA and OVA1 Next Generation result used) | | --- | --- | --- | | Sensitivity | 74.2% (23/31) | 90.3% (28/31) | | 95% CI | 56.8% to 86.3% | 75.1% to 96.7% | | Specificity | 93.9% (230/245) | 67.3% (165/245) | | 95% CI | 90.1% to 96.3% | 61.2% to 72.9% | | PPV | 60.5% (23/38) | 25.9% (28/108) | | 95% CI | 47.0% to 72.3% | 21.5% to 30.2% | | NPV | 96.6% (230/238) | 98.2% (165/168) | | 95% CI | 94.4% to 98.3% | 95.9% to 99.5% | | Prevalence | 11.2% (31/276) | | ## Post-menopausal subjects Among 217 post-menopausal subjects, there were 61 subjects with malignancy by pathology and 156 subjects with no malignancy by pathology. OVA 1 Next Generation was compared to the PA alone: | Malignancy by Pathology | | | | | | --- | --- | --- | --- | --- | | | | Physician Assessment (PA) | | Total | | OVA1 Next Generation | Positive | 43 | 13 | 56 | | | Negative | 2 | 3 | 5 | | Total | | 45 | 16 | 61 | | | | | | | {17} | No Malignancy by Pathology | | | | | | --- | --- | --- | --- | --- | | | | Physician Assessment (PA) | | Total | | OVA1 Next Generation | Positive | 7 | 47 | 54 | | | Negative | 7 | 95 | 102 | | Total | | 14 | 142 | 156 | With dual assessment, sensitivity for malignancy increased from $73.8\%$ to $95.1\%$ . PPV of the dual assessment decreased from $76.3\%$ to $48.7\%$ and NPV of the dual assessment significantly increased from $89.9\%$ to $96.9\%$ . The confidence interval for the observed $7.0\%$ increase was $2.8\%$ to $11.5\%$ . | Performance | Physician Assessment (PA) | Dual Assessment (PA and OVA1 Next Generation result used) | | --- | --- | --- | | Sensitivity | 73.8% (45/61) | 95.1% (58/61) | | 95% CI | 61.6% to 83.2% | 86.5% to 98.3% | | Specificity | 91.0% (142/156) | 60.9% (95/156) | | 95% CI | 85.5% to 94.6% | 53.1% to 68.2% | | PPV | 76.3% (45/59) | 48.7% (58/119) | | 95% CI | 65.9% to 84.5% | 43.8%% to 54.2% | | NPV | 89.9% (142/158) | 96.9% (95/98) | | 95% CI | 85.8% to 93.3% | 92.0% to 99.0% | | Prevalence | 28.1% (61/217) | | # Performance Characteristics for OVA1 Next Generation Alone The OVA1 Next Generation test is not for use as a stand-alone test. Clinical/imaging evaluation is needed in order to identify patients in the intended use population (women over age 18, ovarian adnexal mass present for which surgery is planned, and not yet referred to an oncologist), for whom the performance characteristics of the test are established. These tables below compare the performance of the OVA1 Next Generation test to that of the predicate OVA1 Test as stand-alone tests, despite the fact that both are intended to be used in conjunction with PA. {18} | Clinical Performance of OVA1 Next Generation and OVA1 Test Stand-Alone for All Evaluable Subjects and Stratified by Menopausal status | | | | | | | | --- | --- | --- | --- | --- | --- | --- | | | OVA1 Next Generation | | | OVA1 Test | | | | | All Evaluable subjects | Pre-menopausal women | Post-menopausal women | All Evaluable subjects | Pre-menopausal women | Post-menopausal women | | All malignancies | | | | | | | | Sensitivity | 91.3% (84/92) | 90.3% (28/31) | 91.8% (56/61) | 92.4% (85/92) | 93.5% (29/31) | 91.8% (56/61) | | 95% CI | 83.8% to 95.5% | 75.1% to 96.7% | 82.2% to 96.4% | 85.1% to 96.3% | 79.3% to 98.2% | 82.2% to 96.4% | | All benign | | | | | | | | Specificity | 69.1% (277/401) | 71.4% (175/245) | 65.4% (102/156) | 53.6% (215/401) | 61.6% (151/245) | 41.0% (64/156) | | 95% CI | 64.4% to 73.4% | 65.5% to 76.7% | 57.6% to 72.4% | 47.8% to 58.4% | 55.4% to 67.5% | 33.6% to 48.9% | | | | | | | | | | PPV | 40.4% (84/208) | 28.6% (28/98) | 50.9% (56/110) | 31.4% (85/271) | 23.6% (29/123) | 37.8% (56/148) | | 95% CI | 36.5% to 44.3% | 23.7% to 33.5% | 45.3% to 56.8% | 28.7% to 34.0% | 19.9% to 27.0% | 34.2% to 41.7% | | NPV | 97.2% (277/285) | 98.3% (175/178) | 95.3% (102/107) | 96.8% (215/222) | 98.7% (151/153) | 92.8% (64/69) | | 95% CI | 94.9% to 98.6% | 95.7% to 99.5% | 90.3% to 97.9% | 94.0% to 98.5% | 95.8% to 99.7% | 85.2% to 96.7% | | a- Characterization evaluated stand-alone risk stratification versus cutoff, without regard to results of physician assessment. OVA1 Next Generation is not intended as a stand-alone diagnostic test. | | | | | | | Sensitivity of OVA1 Next Generation was 91.3%, sensitivity of OVA1 Test was 92.4% and difference was -1.1% with 95% CI: -7.5%-5.3%. Specificity of OVA1 Next Generation was 69.1%, specificity of OVA1™ Test was 53.6% and difference was 15.5% with 95% CI: 9.8%-21.1%. The tables below provide further characterization of the OVA1 Next Generation stand-alone performance stratified by histological subtype of malignancy and stage of primary ovarian malignancy for all evaluable subjects, pre-menopausal and post-menopausal women. {19} | Sensitivity of OVA1 Next Generation and OVA1 Test for Histological Subtype of Malignancy | | | | | | | | --- | --- | --- | --- | --- | --- | --- | | | OVA1 Next Generation | | | OVA1 Test | | | | | All Evaluable subjects | Pre-menopausal women | Post-menopausal women | All Evaluable subjects | Pre-menopausal women | Post-menopausal women | | EOC | | | | | | | | Sensitivity | 95.0% (57/60) | 100% (18/18) | 92.9% (39/42) | 95.0% (57/60) | 100% (18/18) | 92.9% (39/42) | | 95% CI | 86.3% to 98.3% | 82.4% to 100% | 81.0% to 97.5% | 86.3% to 98.3% | 82.4% to 100% | 81.0% to 97.5% | | Non-EOC Malignancies | | | | | | | | Sensitivity | 80.0% (4/5) | 80.0% (4/5) | n/a | 80.0% (4/5) | 80.0% (4/5) | n/a | | 95% CI | 37.6% to 96.4% | 37.6% to 96.4% | | 37.6% to 96.4% | 37.6% to 96.4% | | | Low Malignant Potential | | | | | | | | Sensitivity | 82.4% (14/17) | 80.0% (4/5) | 83.3% (10/12) | 82.4% (14/17) | 80.0% (4/5) | 83.3% (10/12) | | 95% CI | 59.0% to 93.8% | 37.6% to 96.4% | 55.2% to 95.3% | 59.0% to 93.8% | 37.6% to 96.4% | 55.2% to 95.3% | | Malignancies Metastatic to the Ovaries | | | | | | | | Sensitivity | 100% (6/6) | 100% (2/2) | 100% (4/4) | 100% (6/6) | 100% (2/2) | 100% (4/4) | | 95% CI | 61.0% to 100% | 34.2% to 100% | 51.0% to 100% | 61.0% to 100% | 34.2% to 100% | 51.0% to 100% | | Other non-Ovarian Malignancies | | | | | | | | Sensitivity | 75.0% (3/4) | 0.0% (0/1) | 100% (3/3) | 100% (4/4) | 100% (1/1) | 100% (3/3) | | 95% CI | 30.1% to 95.4% | 0.0% to 79.3% | 43.9% to 100% | 51.0% to 100% | 20.7% to 100% | 43.9% to 100% | | a- Characterization evaluated stand-alone risk stratification versus cutoff, without regard to results of physician assessment. OVA1 Next Generation is not intended as a stand-alone diagnostic test. | | | | | | | There were no differences in the number of primary ovarian malignancies detected between OVA1 Next Generation and OVA1 Test (61/65 vs. 61/65). One fewer non-ovarian malignancy in a pre-menopausal woman was classified as high risk by OVA1 Next Generation (sensitivity of $75\%$ or $3/4$ for OVA1 Next Generation vs. $100\%$ or $4/4$ for OVA1 Test). One fewer Stage I primary ovarian malignancy (sensitivity of $85.7\%$ or $24/28$ for OVA1 Next Generation vs. $89.3\%$ or $25/28$ for OVA1 Test) and one more Stage III primary ovarian was detected (sensitivity of $100\%$ or $25/25$ for OVA1 Next Generation vs. $96\%$ or $24/25$ for OVA1 Test). The single stage I missed and the single stage III case picked up by OVA1 Next Generation were both in the postmenopausal subgroup. {20} | Sensitivity of OVA1 Next Generation and OVA1 Test for Stage of Primary Ovarian Malignancy | | | | | | | | --- | --- | --- | --- | --- | --- | --- | | | OVA1 Next Generation | | | OVA1 Test | | | | | All Evaluable subjects | Pre-menopausal women | Post-menopausal women | All Evaluable subjects | Pre-menopausal women | Post-menopausal women | | Stage I | | | | | | | | Sensitivity | 85.7% (24/28) | 88.9% (8/9) | 84.2% (16/19) | 89.3% (25/28) | 88.9% (8/9) | 89.5% (17/19) | | 95% CI | 68.5% to 94.3% | 56.5% to 98.0% | 62.4% to 94.5% | 72.8% to 96.3% | 56.5% to 98.0% | 68.6% to 97.1% | | Stage II | | | | | | | | Sensitivity | 100% (7/7) | 100% (2/2) | 100% (5/5) | 100% (7/7) | 100% (2/2) | 100% (5/5) | | 95% CI | 64.6% to 100% | 34.2% to 100% | 56.6% to 100% | 64.6% to 100% | 34.2% to 100% | 56.6% to 100% | | Stage III | | | | | | | | Sensitivity | 100% (25/25) | 100% (10/10) | 100% (15/15) | 96.0% (24/25) | 100% (10/10) | 93.3% (14/15) | | 95% CI | 86.7% to 100% | 72.2% to 100% | 79.6% to 100% | 80.5% to 99.3% | 72.2% to 100% | 70.2% to 98.8% | | Stage IV | | | | | | | | Sensitivity | 100% (5/5) | 100% (2/2) | 100% (3/3) | 100% (5/5) | 100% (2/2) | 100% (3/3) | | 95% CI | 56.6% to 100% | 34.2% to 100% | 43.9% to 100% | 56.6% to 100% | 34.2% to 100% | 43.9% to 100% | | a. Characterization evaluated stand-alone risk stratification versus cutoff, without regard to results of physician assessment. OVA1 Next Generation is not intended as a stand-alone diagnostic test. | | | | | | | One fewer Stage I primary ovarian malignancy (sensitivity of 85.7% or 24/28 for OVA1 Next Generation vs. 89.3% or 25/28 for OVA1 Test) and one more Stage III primary ovarian was detected (sensitivity of 100% or 25/25 for OVA1 Next Generation vs. 96% or 24/25 for OVA1 Test). The single stage I missed and the single stage III case picked up by OVA1 Next Generation were both in the post-menopausal subgroup. In the second study, in addition to the long-term archived serum specimens collected from the multi-site clinical study described above, sensitivities and specificities of OVA1 Next Generation and OVA1 Test were compared using serum specimens that been stored at -65 °C to -85 °C and tested for OVA1 Next Generation and OVA1 Test no more than one year after collection. These samples were acquired from prospective studies that recruited premenopausal and postmenopausal women presenting with an ovarian adnexal mass requiring surgical intervention. The purpose of the comparison was to demonstrate that for samples archived less than one year prior to testing, the OVA1 Next Generation test showed similar sensitivity {21} and specificity when compared with the OVA1 Test. This blinded study included 28 patients confirmed by pathology to have primary ovarian malignancy, along with 105 patients with benign conditions. A comparison of OVA1 Next Generation and OVA1 Test standalone sensitivity in this set of samples is shown in the tables below. | | OVA1 Next Generation | OVA1 Test | Difference OVA1 Next Generation - OVA1 Test | | --- | --- | --- | --- | | All subjects | | | | | Sensitivity | 78.6% (22/28) | 82.1% (23/28) | -3.6% (1/28) | | 95% CI | 60.5% to 89.9% | 64.4% to 92.1% | -19.2% to 12.0% | | | | | | | Specificity | 74.3% (78/105) | 57.1% (60/105) | 17.2% (18/105) | | 95% CI | 65.2% to 81.7% | 47.6% to 66.2% | 7.1% to 27.2%* | | a- Characterization evaluated stand-alone risk stratification versus cutoff, without regard to results of physician assessment. OVA1 Next Generation is not intended as a stand-alone diagnostic test. * - Performance was considered statistically different if the 95% CI of the difference did not bound or contain zero. | | | | | Stage | N | OVA1 Next Generation Sensitivity% | OVA1 Test Sensitivity% | | --- | --- | --- | --- | | I | 10 | 90.0% (9/10) | 90.0% (9/10) | | II | 1 | 100% (1/1) | 100% (1/1) | | III | 9 | 88.9% (8/9) | 88.9% (8/9) | | IV | 3 | 66.7% (2/3) | 100% (3/3) | | Not Staged | 5 | 40.0% (2/5) | 40.0% (2/5) | | a- Characterization evaluated stand-alone risk stratification versus cutoff, without regard to results of physician assessment. OVA1 Next Generation is not intended as a stand-alone diagnostic test. | | | | $b$ . Other clinical supportive data (when a. is not applicable): # 4. Clinical cut-off: The results are interpreted as follows: Low probability of malignancy OVA1 Next Generation risk score $&lt; 5.0$ High probability of malignancy OVA1 Next Generation risk score $\geq 5.0$ {22} 5. Expected values/Reference range: The reference interval of OVA1 Next Generation test was determined in 68 pre-menopausal and 84 post-menopausal healthy women (total = 152 evaluable subjects). Ages ranged from 18 to 91 years and represented whites (84.9%), Hispanic/Latino (7.2%) and African American (5.3%) subjects. The mean, SD, median, range and 5th to 95th percentile of OVA1 Next Generation scores and the OVA1 Next Generation test results are shown for each group in the table below. It is recommended that each laboratory establish its own reference range for the population of interest. | OVA1 Next Generation Scores and Results in Healthy Subjects | | | | | --- | --- | --- | --- | | | All Evaluable Subjects | Pre-menopausal Women | Post-menopausal Women | | n (%) | 152 (100) | 68 (44.7) | 84 (55.3) | | Mean age (SD) | 51.0 (13.75) | 39.2 (8.23) | 60.5 (9.22) | | Median age | 51 | 41 | 59 | | OVA1 Next Generation score | | | | | Mean (SD) | 3.94 (0.984) | 3.72 (0.938) | 4.12 (0.989) | | Median | 3.90 | 3.60 | 4.05 | | Range (min, max) | (2.2, 7.1) | (2.2, 6.1) | (2.5, 7.1) | | Reference interval (5th, 95th percentiles) | (2.5, 5.9) | (2.4, 5.3) | (2.9, 5.9) | | OVA1 Next Generation result, n (%) | | | | | Positive | 23 (15.1%) | 9 (13.2%) | 14 (16.7%) | | Negative | 129 (84.9%) | 59 (86.8%) | 70 (83.3%) | Expected Values in Non-Ovarian Malignancy Condition: To evaluate the performance of the OVA1 Next Generation test in subjects with other disease conditions, patients with cancer conditions other than ovarian cancer (bladder cancer, breast cancer, cervical cancer, colon cancer, endometrial cancer, lung cancer, leukemia and lymphoma) and patients with non-cancer conditions (autoimmune disease, cardiac disease, diabetes, endometriosis, hepatitis and kidney diseases) were evaluated. | Evaluable Specimens from Subjects with non-Ovarian Cancers and Other Conditions | | | --- | --- | | Bladder cancer | 20 | | Breast cancer | 40 | | Cervical cancer | 20 | | Colon cancer | 40 | | Endometrial cancer | 40 | | Leukemia | 11 | | Lung cancer | 40 | | Lymphoma | 10 | | Autoimmune disease | 20 | | Cardiac disease | 20 | {23} | Evaluable Specimens from Subjects with non-Ovarian Cancers and Other Conditions | | | --- | --- | | Diabetes | 40 | | Endometriosis | 40 | | Hepatitis | 20 | | Kidney disease | 20 | | Pregnant women | 20 | | All specimens | 401 | The mean, median, standard deviation, $5^{\text{th}}$ to $95^{\text{th}}$ percentiles as observed in the data are shown for each condition group. Using the defined cut-off of 5.0 for OVA1 Next Generation scores, the number of positive $(\geq 5.0)$ and negative $(&lt; 5.0)$ cases is presented below: | OVA1 Next Generation Scores and Results in Subjects with Non-Ovarian Cancers | | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | --- | | | Bladder Cancer | Breast Cancer | Cervical Cancer | Colon Cancer | Endometrial Cancer | Leukemia | Lung Cancer | Lymphoma | | n | 20 | 40 | 20 | 40 | 40 | 11 | 40 | 10 | | OVA1 Next Generation score, statistics | | | | | | | | | | Mean (SD) | 5.32 (1.83) | 4.03 (1.22) | 6.56 (1.91) | 5.11 (1.73) | 5.45 (1.72) | 6.66 (1.30) | 5.02 (1.38) | 6.10 (1.91) | | Median | 4.8 | 3.9 | 7.7 | 4.6 | 4.8 | 7.2 | 4.9 | 6.0 | | 5th, 95th percentiles | 2.9, 8.2 | 2.8, 6.6 | 3.5, 8.5 | 3.0, 7.7 | 3.1, 8.1 | 4.4, 8.1 | 3.1, 7.5 | 2.4, 8.5 | | Range min, max | 2.8, 8.5 | 2.6, 8.4 | 2.6, 8.5 | 2.4, 8.1 | 3.0,8.1 | 4.4, 8.1 | 2.8, 7.8 | 2.4, 8.5 | | OVA1 Next Generation test result, n | | | | | | | | | | Positive | 10 | 6 | 13 | 18 | 20 | 10 | 18 | 8 | | Negative | 10 | 34 | 7 | 22 | 20 | 1 | 22 | 2 | | % negative results | 50 | 85 | 35 | 55 | 50 | 9.1 | 55 | 20 | | OVA1 Next Generation Scores and Results in Subjects with Conditions Other than Cancers | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | | | Autoimmune Disease | Cardiac Disease | Diabetes | Endometriosis | Hepatitis | Kidney Disease | Pregnant Women | | n | 20 | 20 | 40 | 40 | 20 | 20 | 20 | | OVA1 Next Generation score, statistics | | | | | | | | | Mean (SD) | 5.52 (1.86) | 6.12 (1.58) | 4.72 (1.67) | 4.35 (1.38) | 5.19 (1.78) | 6.65 (1.37) | 5.31 (0.35) | | Median | 5.9 | 6.4 | 4.2 | 4.2 | 5.1 | 7.2 | 5.3 | | 5th, 95th percentiles | 2.8, 8.2 | 3.6, 8.1 | 2.5, 8.1 | 2.5, 7.2 | 3.0, 7.9 | 3.8, 8.3 | 4.8, 6.0 | | Range min, max | 2.4, 8.3 | 3.3, 8.3 | 2.0, 8.1 | 2.2, 7.9 | 2.7, 7.9 | 3.3, 8.3 | 4.5, 6.2 | | | | | | | | | | {24} | OVA1 Next Generation test result, n | | | | | | | | | --- | --- | --- | --- | --- | --- | --- | --- | | Positive | 11 | 15 | 14 | 11 | 11 | 18 | 19 | | Negative | 9 | 5 | 26 | 29 | 9 | 2 | 1 | | % negative results | 45 | 25 | 65 | 72.5 | 45 | 10 | 5 | The number of cases is small within each of the examined diseases and conditions, but the results suggest that caution is warranted when interpreting OVA1 Next Generation results for pregnant women and patients with cervical cancer, leukemia, lymphoma, cardiac disease, kidney disease. **N. Proposed Labeling:** The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. **O. Conclusion:** The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...