← Product Code [OAP](/productcode/OAP) · K150578

# Capillus 272 Office Pro (K150578)

_Capillus, LLC · OAP · Jun 9, 2015 · Physical Medicine · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K150578

## Device Facts

- **Applicant:** Capillus, LLC
- **Product Code:** [OAP](/productcode/OAP.md)
- **Decision Date:** Jun 9, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 890.5500
- **Device Class:** Class 2
- **Review Panel:** Physical Medicine
- **Attributes:** Therapeutic

## Indications for Use

The Capillus272 OfficePro is indicated to promote hair growth in females who have androgenic alopecia and Ludwig-Savin Classifications of I- II; and with Fitzpatrick Classification of Skin Phototypes I to IV. The Capillus272 OfficePro is intended for use in a dry, indoor environment.

## Device Story

Device is a low-level laser therapy (LLLT) system for promoting hair growth; consists of 272 red diode lasers (650 nm) mounted in a helmet on a stand. Intended for in-office use; operated by clinic staff or patients under supervision. Device connects to power via adapter; lacks battery. Safety interlock automatically pauses therapy if head moves outside radiation zone and resumes when repositioned. Audible tones signal start (2 beeps) and end (1 long beep) of therapy. Output is red laser light delivered to scalp. Benefits include non-invasive hair growth promotion for specified alopecia patients.

## Clinical Evidence

Bench testing only. Performance testing confirmed compliance with design specifications and verification that all functions operate as intended. No new clinical data provided; substantial equivalence is supported by the established safety and efficacy profile of the predicate and reference devices, which utilize the same LLLT technology.

## Technological Characteristics

272 red diode lasers (650 nm) housed in a helmet. Powered by ~12V DC adapter. Safety interlock system for head positioning. Audible tone feedback. Class 3R laser system per IEC standards. Non-portable, stand-mounted configuration.

## Regulatory Identification

An infrared lamp is a device intended for medical purposes that emits energy at infrared frequencies (approximately 700 nanometers to 50,000 nanometers) to provide topical heating.

## Special Controls

*Classification.* Class II (special controls). The device, when it is an infrared therapeutic heating lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 890.9.

## Predicate Devices

- iGrow II Hair Growth System
- Sunetics Clinical Laser "G" or "W2326"
- Capillus 272 Pro ([K143199](/device/K143199.md))

## Reference Devices

- Hairmax Lasercomb ([K140931](/device/K140931.md), [K141567](/device/K141567.md), [K132646](/device/K132646.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the top half of the circle. Inside the circle is a stylized image of three human profiles facing right, with the profiles overlapping and connected to a flowing, ribbon-like shape below.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 9, 2015

Capillus. LLC Ms. Patricia Schnoor Quality Manager 1430 South Dixie Highway, Suite 304 Coral Gables, Florida 33146

Re: K150578

Trade/Device Name: Capillus272 OfficePro Regulation Number: 21 CFR 890.5500 Regulation Name: Infrared lamp Regulatory Class: Class II Product Code: OAP Dated: March 24, 2015 Received: March 25, 2015

Dear Ms. Schnoor:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{1}------------------------------------------------

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

# Jennifer R. Stevenson -S

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K150578

Device Name Capillus272 OfficePro

Indications for Use (Describe)

Indications of Describer
Promote hair growth in females diagnosed with androgenetic alopedia presenting with Ludwig-Savin Classifications I-II and Fitzpatrick Skin Phototypes I-IV.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

# Capillus, LLC

#### Date Prepared:

### June 2, 2015

## Submitter's Contact Information:

| Patricia Schnoor                                             |
|--------------------------------------------------------------|
| Capillus LLC<br>1430 S. Dixie Hwy.<br>Coral Gables, FL 33146 |
| (786) 888-1874<br>(305) 418-7581                             |
|                                                              |

#### Name of Device and Name / Address of Sponsor: Trade Name: Capillus272 OfficePro

| Common or Usual Name:        | Low level laser for promotion of hair growth                                     |
|------------------------------|----------------------------------------------------------------------------------|
| Classification Name:         | LLLT per 21 CFR 878.4810                                                         |
| Classification Code:         | OAP (Laser, Comb, Hair)                                                          |
| Sponsor Contact Information: | Patricia Schnoor<br>Capillus LLC<br>1430 S. Dixie Hwy.<br>Coral Gables, FL 33146 |
| Telephone:                   | (786) 888-1874                                                                   |

### Predicate Devices:

| Device Trade Name                      | Manufacturer                               |
|----------------------------------------|--------------------------------------------|
| iGrow II Hair Growth System            | Apira Science, Inc.                        |
| Sunetics Clinical Laser "G" or "W2326" | Sunetics international Marketing Group LLC |
| Capillus 272 Pro                       | Capillus, LLC                              |
|                                        |                                            |
|                                        |                                            |

# Reference Devices:

| Device Trade Name | Manufacturer            |
|-------------------|-------------------------|
| Hairmax Lasercomb | Lexington International |

# Intended Use / Indications for Use:

The Capillus272 OfficePro is indicated to promote hair growth in females who have androgenic alopecia and Ludwig-Savin Classifications of I- II; and with Fitzpatrick Classification of Skin Phototypes I to IV.

The Capillus272 OfficePro is intended for use in a dry, indoor environment.

{4}------------------------------------------------

#### Technological Characteristics

The Capillus272 OfficePro is exactly the same as the Capillus272 Pro in every way except the following:

- 1. The Capillus272 OfficePro is the same hat, but mounted on a stand. The same electrical design has been applied (i.e. adapter steps down to ~12V DC; but no battery as it is not intended to be portable).
The Capillus272 Pro consists of 272 red, visible light, diode lasers operating at 650 nanometers, configured within an outer helmet and protective inner liner, and configured for portable battery and adapter. The Capillus272 OfficePro is intended for in-office use and is not intended to be portable. The same hat is mounted on a stand and the adapter is connected directly to the cap through wiring threaded through the stand.

Just as for the previous version, the helmet system will automatically pause therapy if the subject's head is moved outside of the zone of radiation and will resume therapy when the correct head position is re-established. This is achieved by a safety interlock. At the beginning and end of a therapy session, audible tones are delivered to the user, indicating that therapy has begun (2 beeps) or ended (one long beep). The system is powered by rechargeable Nickel Metal Hydride battery cells (Capillus272 Pro only) assembled into a proprietary pack. Both the battery pack and charger are fully compliant to recognized, international standards.

#### Performance Data:

Performance testing was conducted to confirm compliance to design specifications; all functions have been verified to operate as designed. All acceptance criterion were met by the device. See Section 12 for standards met and test reports.

The performance data included (Section 12) demonstrates that the Capillus272 Pro has the same or similar laser wavelength, output power, output beam, energy type, laser field, treatment area, and energy delivered (J/cm') as the declared predicates. Just as for all predicates (K143199) and reference devices (K140931, K141567, K132646), there are no reported adverse events for this technology.

#### Substantial Equivalence

The Capillus272 OfficePro is the same technology used by Capillus272 Pro (K143199) and all reference devices. The Capillus272 OfficePro is as safe and effective as the Capillus272 Pro, the iGrow, and Sunetics as well as other reference devices in its class, such as the Hairmax Lasercomb.

One key question remains - does the Capillus272 OfficePro demonstrate substantial equivalence to the other devices in its class? The sponsor believes that with the exception of the configuration of the optical elements, the predicate devices are the same devices in form, function safety and efficacy as the Capillus272 OfficePro. The Hairmax Lasercomb, offered as a reference, is proof of the functionality and acceptability of the first device to be cleared by the FDA in the category of OAP, both technically and clinically. The sponsor believes that the difference in the physical appearance, number of diodes, or in the method of delivering the radiant energy of the systems is of no

{5}------------------------------------------------

consequence and does not affect the therapeutic value or the safety profile.

All compliant LLLT systems which use red light diode lasers are classified as class 3R laser systems by the IEC standard for allowable emission levels, which is a recognized standard by the FDA as well, and the adverse event profile is the same.

For these reasons, the Capillus272 OfficePro satisfies the FDA's substantial equivalence requirements with respect to intended use, and technological and design characteristics.

With the classification of OAP, the sponsor respectfully proposes that the FDA has acknowledged that Low-Level Laser/Light Therapy is a viable modality for treating androgenic alopecia in both genders and that the red light lasers in class 3R, used in the Capillus272 Pro and Capillus272 OfficePro, are substantially equivalent to the predicates. Additionally, no new safety or efficacy concerns are raised due to the minor differences present between devices.

---

**Source:** [https://fda.innolitics.com/device/K150578](https://fda.innolitics.com/device/K150578)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
