← Product Code [LRK](/productcode/LRK) · K150566

# Oniris (K150566)

_Sas Oniris · LRK · Mar 21, 2016 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K150566

## Device Facts

- **Applicant:** Sas Oniris
- **Product Code:** [LRK](/productcode/LRK.md)
- **Decision Date:** Mar 21, 2016
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.5570
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

The Oniris device is indicated in the treatment of snoring and/or mild to moderate obstructive sleep apnea in adults.

## Device Story

Oniris is a mandibular advancement device (MAD) used during sleep to treat snoring and mild to moderate obstructive sleep apnea. It consists of two injection-molded thermoplastic trays customized via a 'boil-and-bite' process to fit the patient's teeth without dental impressions. The trays are linked by interconnecting rods that allow for mandibular advancement (protrusion) up to 11 mm relative to the maxilla. This advancement increases upper airway opening. The device is intended for prescription use; it is cleaned daily with water and a toothbrush. The design includes a rod-locking mechanism to prevent accidental detachment during use. By maintaining the lower jaw in a protruded position, the device reduces airway obstruction, potentially improving sleep quality, reducing morning fatigue, and decreasing apnea/hypopnea indices.

## Clinical Evidence

A one-arm pilot study in patients with moderate sleep apnea demonstrated a reduction in apnea index and oxygen desaturation index. Subjective patient evaluation reported positive impacts on snoring, morning fatigue, and sleep quality.

## Technological Characteristics

Intraoral mandibular advancement device; two-part injection-molded thermoplastic trays; thermoforming resin; interchangeable rods for titration (up to 11 mm protrusion); non-sterile; single-patient use. Physical properties tested include density, tensile strain, tensile modulus, impact strength, softening temperature, melt flow rate, and Vicat Softening Point.

## Regulatory Identification

Intraoral devices for snoring and intraoral devices for snoring and obstructive sleep apnea are devices that are worn during sleep to reduce the incidence of snoring and to treat obstructive sleep apnea. The devices are designed to increase the patency of the airway and to decrease air turbulence and airway obstruction. The classification includes palatal lifting devices, tongue retaining devices, and mandibular repositioning devices.

## Special Controls

*Classification.* Class II (special controls). The special control for these devices is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intraoral Devices for Snoring and/or Obstructive Sleep Apnea; Guidance for Industry and FDA.”

## Predicate Devices

- SomnoGuard® SP Soft ([K121761](/device/K121761.md))

## Reference Devices

- Zquiet® Mouthpiece ([K093407](/device/K093407.md))
- SomnoGuard Series (SomnoGuard, 2.0, AP, and AP Pro) ([K061688](/device/K061688.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 21, 2016

SAS Oniris Mr. Thibault Vincent CEO 704 Avenue Roger Salengro Chaville, 92370 France

Re: K150566 Trade/Device Name: Oniris Regulation Number: 21 CFR 872.5570 Regulation Name: Intraoral Devices for Snoring and Intraoral Devices for Snoring and Obstructive Sleep Apnea Regulatory Class: II Product Code: LRK Dated: February 8, 2016 Received: February 10, 2016

Dear Mr. Vincent:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Tina Kiang -S

for Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

| DEPARTMENT OF HEALTH AND HUMAN SERVICES |  |
|-----------------------------------------|--|
| Food and Drug Administration            |  |
| Indications for Use                     |  |

| Form Approved:           | OMB No. 0910-0120 |
|--------------------------|-------------------|
| Expiration Date:         | January 31, 2017  |
| See PRA Statement below. |                   |

| 510(k) Number (if known) | K150566 |
|--------------------------|---------|
| Device Name              | ONIRIS  |

Indications for Use (Describe)

The Oniris device is indicated in the treatment of snoring and/or mild to moderate obstructive sleep apnea in adults.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)

Page 1 of 1

PSC Publishing Services (30) 443-6740 EF

{3}------------------------------------------------

# 510(k) Summary K150566

Device Trade Name : Oniris

#### Submitter:

SAS Oniris 147 avenue Paul Doumer 92500 RUEIL-MALMAISON FRANCE

Phone: +33 1 47 16 17 17 Fax: +1 47 16 20 11

Contact person: Thibault Vincent

Device Common Name : Mandibular advancement device

Date summary prepared : 18/03/2016

Device Classification Name : Intraoral Devices for snoring and obstructive sleep apnea

Device classification : Class II Product Classification : 872.5570 Product Code : LRK

Predicate devices:

Primary predicate : SomnoGuard® SP Soft. K121761 Reference predicates : Zquiet® Mouthpiece. K093407 SomnoGuard Series (SomnoGuard, 2.0, AP, and AP Pro). K061688

### Description of the device :

The Oniris Mandibular advancement device is an intraoral device for use during up to 8 hours of sleep, and composed of two injection molded thermoplastic trays overmoulded with thermoforming resin and customized to fit over the patient's teeth. The trays are attached and articulated through interconnecting rods to allow for mandibular advancement in relation to the maxilla up to 11 mm of protrusion. Mandibular advancement is the mechanism of action to increase upper airway opening to reduce snoring and mild to moderate sleep apnea.

Image /page/3/Picture/14 description: The image shows a clear plastic oral appliance. The appliance consists of two parts: a horseshoe-shaped piece that fits over the upper teeth and a rectangular piece that fits over the lower teeth. The two pieces are connected by two small plastic arms. The appliance is designed to be worn at night to help prevent teeth grinding and clenching.

The Oniris device is formed to upper and lower teeth without the need of taking a patient's dental impressions. Oniris device is a "boil-&-bite" device, the trays are customizable.

This mandibular advancement device maintains performance for 12 to 24 months depending on its conditions of use and maintenance as well as the patient's salivary composition.

{4}------------------------------------------------

Oniris mandibular device is proposed in two sizes to be adapted to different sizes of jaw (size 2 in the same box). The practitioner/dentist has to choose the most comfortable size for the patient.

## Intended use:

The Oniris device is indicated in the treatment of snoring and/or mild to moderate obstructive sleep apnea in adults.

### Technological characteristics compared to predicates

The principle of mandibular advancement for treatment of snoring/or obstructive sleep apnea is well known and there are a number of predicate devices. The advancement of the lower jaw, enables opening the upper airway and reduce snoring.

The Oniris Mandibular advancement device has the same technical characteristics as the primary predicate devices as summarized in table. Reference predicates supports substantial equivalence for the protrusion up to 11 mm and the possibility to mouth breath during device wearing.

|                                            | Oniris mandibular<br>advancement<br>K150566                                                                                          | SomnoGuard® SP<br>Soft<br>K121761                                                                                                                                  | Zquiet ®<br>K093407<br>Reference predicate                                                                                                      | SomoGuard AP®<br>K061688<br>Reference predicate                                                                                                                                                    |
|--------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use                        | The Oniris device is<br>indicated in the<br>treatment of snoring<br>and/or mild to<br>moderate obstructive<br>sleep apnea in adults. | The SomnoGuard®<br>SP Soft is intended<br>for the treatment of<br>night-time snoring<br>and mild to<br>moderate<br>obstructive sleep<br>apnea (OSAS) in<br>adults. | The ZQuiet<br>mandibular<br>advancement<br>device is intended<br>for the treatment of<br>nighttime<br>snoring in adults 1<br>8 years are older. | The SomnoGuard series<br>of mandibular<br>advancement devices is<br>intended for the<br>treatment of night-time<br>snoring and mild to<br>moderate obstructive<br>sleep apnea (OSAS)<br>in adults. |
| Dental impressions                         | « Boil & bite »<br>appliance                                                                                                         | « Boil & bite »<br>appliance                                                                                                                                       |                                                                                                                                                 | « Boil & bite »<br>appliance :                                                                                                                                                                     |
| Mandibular<br>advancement                  | Oniris Device comes<br>with 9 connectors of<br>different length<br>enables up to 11mm<br>protusion                                   | SomnoGuard® SP<br>Soft with 6<br>connectors of<br>different length<br>enables up to 10<br>mm protusion                                                             | ZQuiet® employs a<br>buccal cam<br>mechanism                                                                                                    | SomnoGuard AP<br>comprises two-part<br>titratable enables up to<br>12 mm protusion.                                                                                                                |
| Sterility                                  | Non sterile                                                                                                                          | Non sterile                                                                                                                                                        | Non sterile                                                                                                                                     | Non sterile                                                                                                                                                                                        |
| Design                                     | Two independent<br>thermoplastic trays<br>linked to each other<br>by a coupling<br>protrusion<br>mechanism.                          | Two independent<br>thermoplastic trays<br>linked to each other<br>by a coupling<br>protrusion<br>mechanism.                                                        | Upper and lower<br>dental orthotics<br>with a buccal cam<br>mechanism.                                                                          | SomnoGuard AP<br>consists of two<br>independent<br>thermoplastic trays,<br>both parts are linked to<br>each other by a coupling<br>protusion mechanism                                             |
| Single patient use                         | X                                                                                                                                    | X                                                                                                                                                                  | X                                                                                                                                               | X                                                                                                                                                                                                  |
| Cleaned with water<br>and toothbrush daily | X                                                                                                                                    | X                                                                                                                                                                  | X                                                                                                                                               | X                                                                                                                                                                                                  |
| Interchangeable rods                       | Fixed by pressure on<br>lower tray; angular<br>rotation on pin tray                                                                  | Fixed by pressure<br>on lower and upper<br>tray.                                                                                                                   |                                                                                                                                                 |                                                                                                                                                                                                    |

{5}------------------------------------------------

The thermoplastic material used for tray is different, and the connecting mechanism of interchangeable rods Oniris device presents some differences compared to the primary predicate. The difference consists on larger shape stopper on lower tray to decrease the possibility of rod detachment. A slot has been added on the shape stopper to facilitate the fitting by pressure. The change of this rod is possible after angular rotation of the rod along its axis. This angular rotation is impossible to reproduce when wearing in mouth, and is a security to avoid detachment of rod during wearing.

On upper tray the rod is fitting like a key system, the rod rotate around the axis in order to reach the functional location, while the rod is fixed by pressure on the lower tray primary predicate.

The submission includes physical property specifications for materials used in the fabrication of the device, including density, tensile strain, tensile modulus, impact strength, softening temperature, melt flow rate, and Vicar Softening Point.

#### Clinical Data

In a one arm pilot study investigating the use of Oniris device in patients with moderate sleep apnea demonstrated a reduction of apnea index and oxygen desaturation index and was subjectively evaluated by patients to provide positive impact on snoring, morning fatigue and sleep quality.

#### Conclusions :

The Oniris device function is a similar manner to other comparative predicate devices and the intended use is the same. All are prescription devices indicated for patients suffering and/or mild to moderate obstructive apnea.

The information in this 510(k) submission demonstrate the Oniris device are substantially equivalent to the predicate devices. Fundamental scientific technology of the device is identical to the referenced predicate device. Rationale discussion on design differences of interchangeable rod has been conducted in the 510(k) Submission.

---

**Source:** [https://fda.innolitics.com/device/K150566](https://fda.innolitics.com/device/K150566)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
