K150535 · Dentsply International, Inc. · ELC · Jun 16, 2015 · Dental
Device Facts
Record ID
K150535
Device Name
Cavitron Touch Ultrasonic Scaling System
Applicant
Dentsply International, Inc.
Product Code
ELC · Dental
Decision Date
Jun 16, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.4850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Used for ultrasonic procedures: - All general supra and subgingival scaling applications . - Periodontal debridement for all types of periodontal diseases .
Device Story
Cavitron Touch™ is an ultrasonic scaling system for dental procedures; used for supra/subgingival scaling and periodontal debridement. System inputs include clinician settings via digital touch screen and foot pedal activation. Device transforms electrical energy into ultrasonic vibrations at the handpiece tip; includes Sustained Performance System (SPS) for power adjustment. Produces mechanical scaling action and lavage. Operated by dental professionals in clinical settings. Features Bluetooth-enabled Tap-On™ wireless foot pedal for hands-free activation and temporary power boost. Touch screen provides visual feedback on power, battery, and maintenance status. Benefits include ergonomic control, efficient scaling, and reduced cross-contamination risk via flat, easy-to-clean surfaces.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including AAMI ANSI ES 60601-1 safety, IEC 60601-1-2 EMC, software validation, and internal functional testing of handpiece, touch screen, and wireless connectivity.
Technological Characteristics
Ultrasonic scaler; 28.5-31.5 kHz frequency range; 100-240 VAC input. Features digital touch screen interface, Bluetooth connectivity, and autoclavable handpiece. Includes Sustained Performance System (SPS) for automated power adjustment. Complies with AAMI ANSI ES 60601-1 and IEC 60601-1-2 standards.
Indications for Use
Indicated for patients requiring general supra and subgingival scaling and periodontal debridement for all types of periodontal diseases.
Regulatory Classification
Identification
An ultrasonic scaler is a device intended for use during dental cleaning and periodontal (gum) therapy to remove calculus deposits from teeth by application of an ultrasonic vibrating scaler tip to the teeth.
Predicate Devices
Cavitron® RF Ultrasonic Scaler System with Sterimate Handpiece (K052334)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized eagle-like figure with three heads, representing health, human services, and the USA. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the figure.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 16, 2015
Dentsply International, Inc. Ms. Helen Lewis Director of Regulatory Affairs 221 West Philadelphia St., Suite 60W York, Pennsylvania 17401
Re: K150535
Trade/Device Name: Cavitron Touch™ Ultrasonic Scaling System Regulation Number: 21 CFR 872.4850 Regulation Name: Ultrasonic Scaler Regulatory Class: II Product Code: ELC Dated: March 17, 2015 Received: March 19, 2015
Dear Ms. Lewis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina
Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital. Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Appendix A Indications for Use Statement
网址
通
{3}------------------------------------------------
## SECTION 4. INDICATIONS FOR USE STATEMENT
K150535 510(k) Number (if known):
Device Name: Cavitron TouchTM Ultrasonic Scaling System
Indications for Use:
Used for ultrasonic procedures:
- All general supra and subgingival scaling applications .
- Periodontal debridement for all types of periodontal diseases .
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{4}------------------------------------------------
### DENTSPLY International
World Headauarters Susquehanna Commerce Center 221 West Philadelphia Street PO Box 872 York, PA 17405-2558 (717) 849-4593 Fax (717) 849-4343
### SECTION 5. 510(k) SUMMARY
for
Cavitron Touch™ Ultrasonic Scaling System
#### 1. Submitter Information:
DENTSPLY International Inc. Susquehanna Commerce Center 221 West Philadelphia Street York, PA 17401
| Contact Person: | Helen Lewis |
|-------------------|--------------|
| Telephone Number: | 717-487-1332 |
| Fax Number: | 717-849-4343 |
27 February 2015 Date Prepared:
### 2. Device Name:
.
●
- Cavitron Touch™ Ultrasonic Scaling System Proprietary Name:
- Classification Name: Ultrasonic Scaler
- CFR Number: 872.4850 .
- Device Class: .
- Product Code: ELC .
### 3. Predicate Device:
| Predicate Device Name | 510(k) | Company Name |
|----------------------------------------------------------------------|---------|--------------------------------|
| Cavitron® RF Ultrasonic<br>Scaler System with Sterimate<br>Handpiece | K052334 | DENTSPLY International<br>Inc. |
II
### 4. Description of Device:
The Cavitron Touch™ Ultrasonic Scaling System is an ultrasonic scaling unit that is used in all general supra and subgingival scaling applications and periodontal debridement for all types of periodontal diseases. The Cavitron Touch™ Ultrasonic Scaling System includes a digital touch screen interface and Bluetooth technology. This device is equipped with the Tap-On™ Wireless Foot Pedal with Tap-On™ technology. Tap-On™ Wireless Foot Pedal was cleared under premarket notification K130862 in July 2013.
{5}------------------------------------------------
### 1. Indications for Use:
Used for ultrasonic procedures:
- All general supra and subgingival scaling applications ●
- . Periodontal debridement for all types of periodontal diseases
#### Substantial Equivalence: 6.
### Technological Characteristics
| Proposed Device<br>Cavitron Touch™ Ultrasonic Scaling<br>System | Predicate Device<br>Cavitron® RF Ultrasonic Scaler System<br>with Sterimate Handpiece (K052334) | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use: | Indications for Use: | |
| Used for ultrasonic procedures | Used for ultrasonic procedures | |
| All general supra and subgingival<br>scaling applications Periodontal debridement for all<br>types of periodontal diseases | All general supra and subgingival<br>scaling applications Periodontal debridement for all<br>types of periodontal diseases Endodontic procedures | |
| Features:<br>All functions are touch screen<br>activated, including the power<br>setting, power presets, rinse, purge,<br>scaling, and lock/unlock. All<br>indicators are also visible on the<br>touch screen, such as battery,<br>service, water filter replacement,<br>boost, manual, and Tap-On. | Features:<br>All functions are controlled<br>through the analog power switch or<br>a button located on the front cover.<br>These functions include power<br>setting, rinse, purge, and scaling.<br>All indicators show on the<br>information panel, including<br>battery, service, boost, manual, and<br>Tap-On. | |
| An adjustment collar on the<br>handpiece offers the option of finer<br>water control for precise and<br>convenient adjustment of lavage on<br>the handpiece to the preferred<br>setting. | An adjustment collar on the<br>handpiece offers the option of finer<br>water control for precise and<br>convenient adjustment of lavage on<br>the handpiece to the preferred<br>setting. | |
| The power range setting can be<br>activated in two different ways.<br>The clinician can touch the screen<br>to move the circle up or down to<br>the desired setting, or select one of<br>the three power preset button<br>options that are preset to the<br>clinician's desired power setting. | The power range setting is<br>controlled by a knob. | |
| Proposed Device | Predicate Device | |
| Cavitron Touch™ Ultrasonic Scaling<br>System | Cavitron® RF Ultrasonic Scaler System<br>with Sterimate Handpiece (K052334) | |
| by lines as well as color on the<br>touch screen. | | |
| Depressing the Tap-On™ Wireless<br>Foot Pedal to the second positions<br>initiates Temporary Boost to<br>increase the stroke of the insert by<br>at least 25% when at min power. | Depressing the Tap-On™ Wireless<br>Foot Pedal to the second position<br>initiates Temporary Boost for a<br>10% increase in power. | |
| When the power boost feature is<br>activated, the word "boost" will<br>show in the "informational<br>bubble". | When the power boost feature is<br>activated, the boost indicator will<br>light up with a pink color<br>signifying that an additional 10%<br>of power is in use. | |
| The handpiece is autoclavable/<br>sterilizable as well as detachable<br>from the unit. | The handpiece is autoclavable/<br>sterilizable as well as detachable<br>from the unit. | |
| The unit is equipped with a round,<br>battery operated, ergonomically<br>designed, Tap-On™ Wireless Foot<br>Pedal. If preferred, the Tap-On™<br>Wireless Foot Pedal can be<br>attached by wire. The Tap-On™<br>Wireless Foot Pedal contains two<br>positions: one to activate the scaler<br>and the other to activate the boost<br>feature. | The unit is equipped with a round,<br>battery operated, ergonomically<br>designed, Tap-On™ Wireless Foot<br>Pedal. If preferred, the Tap-On™<br>Wireless Foot Pedal can be<br>attached by wire. The Tap-On™<br>Wireless Foot Pedal contains two<br>positions: one to activate the scaler<br>and the other to activate the boost<br>feature. | |
| The battery indicator is always<br>present on the screen. This allows<br>the clinician to see how much<br>battery life remains in the wireless<br>Tap-On™ Wireless Foot Pedal<br>before it must be recharged. | A battery symbol will light up on<br>the interface indicating that the<br>battery in the Tap-On™ Wireless<br>Foot Pedal needs to be replaced. | |
| The Tap-On™ Wireless Foot Pedal<br>allows the clinician to use a single<br>Tap-On™ the foot pedal to activate<br>or deactivate continuous scaling.<br>This permits the clinician to rest<br>their foot during the procedure. | The Tap-On™ Wireless Foot Pedal<br>allows the clinician to use a single<br>Tap-On™ the Tap-On™ Wireless<br>Foot Pedal to activate or deactivate<br>continuous scaling. This permits<br>the clinician to rest their foot<br>during the procedure. | |
| When the Tap-On™ Wireless Foot<br>Pedal technology is engaged, the<br>word "Tap-On" will appear in the<br>"Informational Bubble". | | |
| | Proposed Device | Predicate Device |
| | Cavitron Touch™ Ultrasonic Scaling<br>System | Cavitron® RF Ultrasonic Scaler System<br>with Sterimate Handpiece (K052334) |
| - | To deactivate the Tap-On™<br>technology, the clinician must push<br>the Tap-On™ button located on the<br>interface. | To deactivate the Tap-On™<br>technology, the clinician must<br>press the Purge and Turbo<br>simultaneously for a few seconds. |
| - | The Purge cycle allows the<br>clinician to purge the water lines<br>by pressing the Purge button<br>located on the touch screen. The<br>Purge screen appears and displays<br>the cycle countdown of 120<br>seconds or 2 minutes. After the 2<br>minute cycle is complete, the<br>screen will return to the Scale<br>screen. | The Purge cycle allows the<br>clinician to purge the water lines.<br>A single push button is used to<br>activate the 2 minute purge cycle.<br>The button will stay illuminated<br>until the cycle is completed. |
| - | The Rinse mode allows the<br>clinician to rinse the area for the<br>patient where needed without the<br>tip moving. Pressing the Rinse<br>button on the touch screen will<br>bring up the Rinse screen. The tip<br>will not move and the system will<br>only rinse. Once finished with the<br>Rinse mode, the clinician can press<br>the Scale button on the touch<br>screen to return to the Scale screen. | The Rinse mode allows the<br>clinician to rinse the area for the<br>patient where needed without the<br>tip moving. To initiate the Rinse<br>mode, the clinician will turn the<br>knob to the Rinse icon until they<br>feel it click. Once the knob is<br>properly in place, the tip will not<br>move and the system will only<br>rinse. To return to scaling, turn the<br>knob off of the Rinse icon. |
| - | Sustained Performance System<br>(SPS), similar to cruise control,<br>will adjust the power when needed. | Sustained Performance System<br>(SPS), similar to cruise control,<br>will adjust the power when needed. |
| - | The flat surfaces of the device<br>make it easy to clean to minimize<br>the risk of cross-contamination. A<br>disposable barrier will be provided<br>to cover the screen. | This device is easy to clean. To<br>clean underneath the knob, simply<br>remove the knob from the device. |
| - | When the water filter needs to be<br>replaced, the Water Filter Indicator<br>Light will light up on the touch<br>screen. | |
| - | The device is internally equipped<br>with a water solenoid, which is<br>where the water comes into the<br>unit and continues to flow to the | The device is internally equipped<br>with a water solenoid, which is<br>where the water comes into the<br>unit and continues to flow to the |
| Proposed Device | Predicate Device | |
| Cavitron Touch™ Ultrasonic Scaling<br>System | Cavitron® RF Ultrasonic Scaler System<br>with Sterimate Handpiece (K052334) | |
| handpiece. | handpiece. | |
| - When the device needs to be<br>serviced, the Service Indicator<br>Light will appear on the touch<br>screen. The 4 pre-programmed<br>service indicators are: missing<br>handpiece, unit over temperature,<br>filter replacement, and battery<br>status. | - When the device needs to be<br>serviced, the Service Indicator<br>Light will illuminate on the<br>interface. | |
| - This device has Bluetooth wireless<br>technology capabilities. | - This device has radio frequency<br>(RF) wireless technology<br>capabilities. | |
| - The half-life of a white LED<br>backlight, which is used in the<br>LCD Screen of the device, is<br>approximately 20,000 hours. | | |
| - The water range setting indicator<br>goes from 1 – 6 with half<br>markings. | - The water range setting indicator<br>goes from 1 – 6 with half<br>markings. | |
| - It is recommended the plastic<br>water filter be replaced once a<br>month. | - It is recommended the plastic<br>water filter be replaced once a<br>month | |
| - The system operates over a<br>frequency range of 28.5 - 31.5<br>kHz. | - The system operates over a<br>frequency range of 28.5 - 31.5<br>kHz. | |
| - The input voltage of the scaler is<br>100 - 240 VAC, 50 to 60 Hz. | - The input voltage of the scaler is<br>100 - 240 VAC, 50 to 60 Hz. | |
| - The system provides a means for<br>control of power (tip stroke) and<br>lavage flow rates. | - The system provides a means for<br>control of power (tip stroke) and<br>lavage flow rates. | |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
The Cavitron Touch™ Ultrasonic Scaling System is an ultrasonic scaler which is intended to be used for all general supra and subgingival scaling applications as well as periodontal debridement for all types of periodontal diseases. The Cavitron Touch™ Ultrasonic Scaling System has the same intended use, incorporates the same fundamental technology, and has similar indications for use as the predicate Cavitron® RF Ultrasonic Scaler System with Sterimate Handpiece cleared under premarket notification K052334. The differences in the table above do not raise any new or different types of questions of safety and effectiveness.
{9}------------------------------------------------
- 7. Non-Clinical Performance Data.
Performance testing focused on verification of design, ultrasonic scaling performance, and function of the Cavitron Touch® Ultrasonic Scaler. Below is a summary of the testing performed:
- AAMI ANSI ES 60601-1 Medical Electrical Equipment Part 1: General requirements . for basic safety and essential performance (2005; reapproved 2012)
- . IEC 60601-1-2 Medical Electrical Equipment PART 1-2: General Requirements for Basic Safety and Essential Performance Collateral Standard: Electromagnetic Compatibility (2007)
- . FDA - Guidelines for the Content of Premarket Submissions for Software Contained in Medical Devices- Software Validation
- . Internal specification and testing for handpiece operation, including lavage control
- . Internal specification and testing for autocycle operation
- Internal specification and testing for touch screen and user interface functionality
- . Internal specification and testing for radio frequency distance for Tap-On™ Wireless Foot Pedal (K130862)
The results of these performance tests support the substantial equivalence of the proposed Cavitron Touch™ Ultrasonic Scaling System with the predicate Cavitron® RF Ultrasonic Scaler System with Sterimate Handpiece (K052334).
New biocompatibility testing was not required to support the substantial equivalence between the proposed Cavitron Touch™ Ultrasonic Scaling System and the predicate device, Cavitron® RF Ultrasonic Scaler System with Sterimate Handpiece (K052334). Biocompatibility testing was submitted with the predicate device, Cavitron® RF Ultrasonic Scaler System with Sterimate Handpiece and cleared under premarket notification K052334. The Cavitron® Steri-Mate™ Sterilizable Handpiece was previously cleared as Jet-Mate™ Sterilizable Handpiece under premarket notification K023697.
#### 1. Clinical Performance Data
No data from human clinical studies has been included to support the substantial equivalence of the Cavitron Touch™ Ultrasonic Scaling System.
#### 2. Conclusion Regarding Substantial Equivalence
The Cavitron Touch™ Ultrasonic Scaling System is an ultrasonic scaler which is intended to be used for all general supra and subgingival scaling applications as well as periodontal debridement for all types of periodontal diseases. The Cavitron Touch™ Ultrasonic Scaling System has the same intended use, incorporates the same fundamental technology, and has similar indications for use as the predicate Cavitron® RF Ultrasonic Scaler System with Sterimate Handpiece cleared under premarket notification K052334. Testing to verify the performance of the Cavitron Touch™ Ultrasonic Scaling System shows that the key features and performance of the Cavitron Touch™ Ultrasonic Scaling System are substantially equivalent to the predicate device, Cavitron® RF Ultrasonic Scaler System with Sterimate Handpiece (K052334).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.