In-use testing was conducted to demonstrate the efficacy of the ASTRA reprocessor in a clinical environment, confirming the device achieves complete kill of microorganisms during routine clinical reprocessing of ultrasound probes.
The GUS ASTRA TEE™ automated reprocessor facilitates high-level disinfection and rinsing of 1 or 2 transesophageal ultrasound probes using FDA cleared and PCI Medical approved high-level liquid disinfectants. The GUS ASTRA VR™ automated reprocessor facilitates high-level disinfection and rinsing of 1 or 2 endovaginal and/or endorectal ultrasound probes using FDA cleared and PCI Medical approved high-level liquid disinfectants.
Device Story
ASTRA Product Family (ASTRA TEE and ASTRA VR) are floor-standing electromechanical reprocessors for enzymatically pre-cleaned ultrasound probes. Devices control high-level liquid disinfectant (HLD) temperature, soak time, and rinse cycles. Operator scans HLD barcode and probe ID; system loads preset parameters. Sensors monitor HLD/rinse levels, temperature, and probe presence. Door sensor initiates cycle. Output includes automated disinfection and rinsing; data retrievable via USB. Used in clinical settings by trained personnel. Benefits include standardized, automated disinfection process with tracking of critical parameters, improving disinfection outcomes compared to manual methods.
Clinical Evidence
Bench testing only. Simulated use testing demonstrated ≥6 Log reduction of M. terrae. In-use testing confirmed complete kill. Residual testing per ISO 10993-5 showed cytotoxic effect ≤ 2. Electrical safety and EMC testing passed per UL 61010-1/CSA C22.2 and EN 61326-1.
Technological Characteristics
Floor-standing electromechanical unit. Materials: PVC, polyethylene, polypropylene, EPDM, stainless steel, polycarbonate, polyurethane, polysulfone, ABS, silicone, silver, polyester. Sensing: temperature sensors, microswitches for probe detection, optical sensors for HLD/rinse levels. Energy: electrical. Connectivity: USB port for data retrieval. Software: microprocessor-controlled, fixed parameters based on HLD. Sterilization: high-level liquid disinfection. Standards: UL 61010-1, IEC 62304, ISO 10993-5.
Indications for Use
Indicated for high-level disinfection and rinsing of 1 or 2 transesophageal, endovaginal, or endorectal ultrasound probes in non-sterile healthcare settings by trained personnel.
Regulatory Classification
Identification
A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.
Predicate Devices
CS MEDICAL TD-100 Transesophageal Probe Disinfector (K051305)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three wing-like shapes extending from its body. The bird is positioned above the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA", which is arranged in a circular fashion around the bird.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 30, 2017
PCI Medical, Inc. Mr. Kevin Mader Director of Quality and Regulatory 6 Winter Avenue Deep River, CT 06417
Re: K150504
Trade/Device Name: GUS ASTRA TEE™ Transesophageal Probe Reprocessor GUS ASTRA VRTM Endovaginal/Endorectal Probe Reprocessor Regulation Number: 21 CFR 892.1570 Regulation Name: Diagnostic Ultrasonic Transducer Regulatory Class: Class II Product Code: PSW Dated: October 16, 2015 Received: October 16, 2015
Dear Mr. Kevin Mader:
This letter corrects our substantially equivalent letter of November 16, 2015.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
# Page 2 - Mr. Kevin Mader
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
Tina
Kiang -S
Tina Kiang, Ph.D. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K150504
Device Name
GUS ASTRA TEE™ Transesophageal Probe Reprocessor
Indications for Use (Describe)
The GUS ASTRA TEE™ automated reprocessor facilitates high-level disinfection and rinsing of 1 or 2 transsophageal ultrasound probes using FDA cleared and PCI Medical approved high-level liquid disinfectants.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Indications for Use
510{k} Number (if known) K | 50504
Device Name
GUS ASTRA VR™ Endovaginal/Endorectal Probe Reprocessor
Indications for Use (Describe)
The GUS ASTRA VRTM automated reprocessor facilitates high-level disinfortion and rinsing of 1 or 2 endovaginal and/or endorectal ultrasound probes using FDA cleared and PCI Medical approved high-level liquid disinfectants.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
区 Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows a logo for PCI Medical. The logo consists of the letters "PCI" stacked on top of the word "MEDICAL". A plus sign is located above the letter "I" in "PCI". The text and symbol are in black and white.
# 510(k) SUMMARY
| Submitter | PCI Medical, Inc.<br>6 Winter Avenue<br>Deep River, CT 06417 |
|-------------------------|------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Kevin Mader<br>Director of Quality & Regulatory<br>k.mader@pcimedical.com<br>(800) 862-3394 ext. 247<br>(860) 526-3081 (fax) |
| Date Prepared | November 13th 2015 |
| Trade Name(Common Name) | |
| Regulation Name | Diagnostic Ultrasonic Transducer, 21 CFR 892.1570 |
| Classification | Class II |
| Product Code | ITX |
| Predicate Device(s) | CS MEDICAL TD-100 Transesophageal Probe Disinfector<br>K051305 |
Intended Use
cilitates high-level disinfection and rinsing of 1 or 2 transesophageal ultrasound probes using FDA cleared and PCI Medical approved high-level liquid disinfectants.
ates high-level disinfection and rinsing of 1 or 2 endovaginal/endorectal ultrasound probes using FDA cleared and PCI Medical approved high-level liquid disinfectants.
The intended use statements use similar language to the predicate TD-100 device (K051305). The differences in wording used does not raise new questions of safety or efficacy.
{5}------------------------------------------------
# Device Description
The ASTRA Product Family of automated reprocessors are floor standing electromechanical lab equipment that facilitate high-level liquid disinfection of ultrasound probes that have been enzymatically pre-cleaned.
# Model types
The GUS ASTRA family of products contai
# Physical Characteristics
The two models are floor standing units with the following physical characteristics:
- . ● . . Weight: 73 lbs. (33.1 kg) . . ●
- . Weight: 73 lbs. (33.1 kg)
### Materials
materials. Only the shape and length of the disinfection chamber and the probe cable restraining designs differ in order to properly accommodate the differences between TEE and VR ultrasound probes. Materials for various components include:
- ◆ ● (PVC), Polyethylene, Polypropylene, EPDM, . Stainless Steel, Polycarbonate, Polyurethane, Polysulfone . ● Polycarbonate, ABS, Polypropylene. ● Sulfide resin, Acryl resin, Polycarbonate resin, stainless steel, silicone, silver, polyester. 6 Winter Avenue Deep River Connecticut, 06443 - Tel: (800) 862-3394 Fax: (860) 526-3081
www.pcimedical.com
{6}------------------------------------------------
-
- ●
- .
-
- . .
.
-
compliant electronics, PVC, Copper
- Polysulfone, Polypropylene, Silicone
Component selection was carefully performed to select materials that have appropriate electrical safety rating, fire safety rating, and to be compatible with select FDA cleared HLDs.
# Differences
Both reprocessors are derived from the same base unit (identical software, electronics, plumbing, valves, sensors, warming pad, cabinet, etc.) which is configurable to become either a er demand. To configure a base unit, TEE and VR specific components (e.g. disinfection chamber, chamber door, cable holder, etc.) are
used to accommodate the differences between TEE and VR ultrasound probe geometries.
# Device Operational Overview
The ASTRA Product Family of automated reprocessors are designed to control the temperature of the high-level liquid disinfectant (HLD), the disinfection time, and the rinse cycles used to
automatically loads the appropriate preset parameters based on the barcode of the scanned HLD bottle. After the probes are enzymatically pre-cleaned per hospital protocol, reprocessing in the ASTRA is initiated through the use of a simple user interface. The operator presses Enter to start the cycle and uses the onboard barcode reader to scan and load 1 or 2 probes and to s MRC status at the next prompt. The user will
on completion of the disinfection cycle. The
ASTRA systems use select FDA cleared high-level liquid disinfectant to safely and effectively disinfect medical devices. By automating the disinfection process through validated software and hardware, the ASTRA Product Family of automated reprocessors help users improve their disinfection outcomes by controlling and tracking the critical process parameters and recording the key operational data.
The GUS ASTRA TEE and GUS ASTRA VR are automated reprocessors which are identical with the exception of the disinfection chamber configuration necessary to compensate for two different types of ultrasound probes. The GUS ASTRA TEE is an automated reprocessor which enables high-level disinfection and rinsing for transesophageal (TEE) ultrasound probes. The GUS ASTRA VR is an automated reprocessor which enables high-level disinfection and rinsing for endovaginal and/or endorectal ultrasound probes.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for PCI Medical. The logo consists of the letters "PCI" in a bold, sans-serif font, with the word "MEDICAL" written in a smaller font below. A plus sign is located above the letter "i" in "PCI". The logo is black and white.
Both ASTRA systems utilize predefined programs based upon the high-level liquid disinfectant (HLD) used during the disinfection and potable water rinsing cycles. The system controls the HLD temperature and cycle times which are fixed, no user configuration is available. The device will perform the defined automated disinfection and rinsing cycles based upon the HLD specifications. The microprocessor controlled interface monitors temperature and liquid levels of the HLD in the disinfection chamber and the reservoir. Optical sensors are used for HLD reservoir presence and probe identification. Additional user inputs include unique probe identification (for disinfection records) and the results entry of the Minimum Recommended Concentration (MRC) test using test strips and minimum concentration levels as defined by the HLD manufacturer. A sensor located on the disinfection chamber door is used to initiate the cycle after the user has input the required information. User retrievable data/information regarding reprocessing is accessible through a USB port.
# Performance Characteristics
The ASTRA Product Family of automated reprocessors utilizes the same technology and highlevel disinfection approach employed by the predicate device, the TD-100 cleared under K051305. It should be noted that TD-100 system is comprised of the instrument and proprietary TD-5 HLD. The TD-100 controls time, temperature, and rinse parameters using the TD-5 proprietary HLD to achieve a ≥6 Log reduction. The ASTRA controls time, temperature, and rinse parameters to facilitate a ≥6 Log reduction by the FDA approved HLD. As a result, this submission primarily focuses on the performance specification of the instrument and controls (time, temperature, and rinse) common to both devices.
The ASTRA system and the TD-100 both control the HLD temperature, soak time, and rinse cycle to facilitate ≥6 Log reduction of M. terrae on enzymatically cleaned probes.
| Device Name | Manufacturer | 510(k) | Performance Attribute |
|-------------------------------------------------------|------------------|---------|-----------------------------------------------------------------------------------------------------------------------------------------|
| TD-100<br>Transesophageal Probe<br>Disinfector System | CS Medical, Inc. | K051305 | Control the HLD temperature,<br>soak time, and rinse cycle to<br>achieve ≥6 Log reduction of M.<br>terrae using proprietary TD-5<br>HLD |
# Performance Specifications
The ASTRA Product Family of automated reprocessors share common hardware and software with minor differences necessary to accommodate the different ultrasound probes (TEE and VR). The intended use and function of the automated reprocessors are the same and they are substantially equivalent to the TD-100 probe disinfector in design and function. The ASTRA
{8}------------------------------------------------
series of automated reprocessors control the HLD temperature, soak time, and rinse cycle to facilitate ≥6 Log reduction of M. terroe on enzymatically cleaned probes at each site tested, thus achieving equivalent performance to the TD-100 instrument.
| Device Name | Manufacturer | 510(k) | Performance Attribute |
|--------------------|-------------------|---------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Probe Reprocessors | PCI Medical, Inc. | K150504 | Control the HLD temperature,<br>soak time, and rinse cycle to<br>facilitate ≥6 Log reduction of<br>M. terrae at each site tested<br>using FDA cleared Metricide<br>OPA. |
# Clinical/Non-Clinical Performance Testing Summary
- . The ASTRA system maintains time and temperature required for disinfection as per HLD
- . The ASTRA system passes Safety and EMC testing
- The ASTRA system passes simulated use and residual testing ●
- . The ASTRA system passes In-use testing
### Technology Comparison to Predicate Devices
The ASTRA Product Family of automated reprocessors utilizes the same technology employed by the predicate device, the TD-100 cleared under K051305. The ASTRA system and the TD-100 both control the HLD temperature, soak time, and rinse cycle for the safe and effective disinfection of ultrasound probes.
| Summary of Predicate Comparisons | | | | |
|-------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Characteristic | GUS ASTRA TEE<br>PCI medical | GUS ASTRA VR<br>PCI medical | TD-100<br>CS Medical | |
| <b>Regulatory</b> | | | | |
| 510(k) Number | K150504 | K150504 | K051305 | |
| Product Code | ITX | ITX | ITX | |
| Regulation<br>Number | 21 CFR 892.1570 | 21 CFR 892.1570 | 21 CFR 892.1570 | |
| | | | N/A<br>ASTRA TEE and ASTRA VR same as<br>predicate. | |
| | | | ASTRA TEE and ASTRA VR same as<br>predicate. | |
| | | Summary of Predicate Comparisons | | |
| Characteristic | GUS ASTRA TEE<br>PCI medical | GUS ASTRA VR<br>PCI medical | TD-100<br>CS Medical | |
| Regulation<br>Name | Diagnostic<br>ultrasonic<br>transducer | Diagnostic<br>ultrasonic<br>transducer | Diagnostic<br>ultrasonic<br>transducer | ASTRA TEE and ASTRA VR same as<br>predicate. |
| Intended Use/<br>Indications for<br>Use | The GUS ASTRA<br>reprocessor<br>facilitates high-level<br>disinfection and<br>rinsing of 1 or 2<br>transesophageal<br>ultrasound probes<br>using FDA cleared<br>and PCI Medical<br>approved high-level<br>liquid disinfectants. | automated<br>reprocessor<br>facilitates high-level<br>disinfection and<br>rinsing of 1 or 2<br>endovaginal and/or<br>endorectal<br>ultrasound probes<br>using FDA cleared<br>and PCI Medical<br>approved high-level<br>liquid disinfectants. | The TD-100<br>disinfector is<br>designed to provide<br>high-level<br>disinfection of<br>Transesophageal<br>(TEE) ultrasound<br>probes. The system<br>uses the TD-5<br>disinfectant, which<br>is designed to be<br>used only with the<br>TD-100. The<br>disinfectant bottles<br>cannot be reused in<br>the system.<br>TD-5 is intended for<br>use as a single use<br>high-level<br>disinfectant to be<br>used exclusively in<br>the TD-100 for high<br>level disinfection of<br>TEE ultrasound<br>probes.<br>The TD-5 High Level<br>Disinfectant and TD-<br>100 disinfector<br>system is intended<br>for use by qualified<br>individuals trained<br>in its use. | ASTRA TEE and ASTRA VR similar<br>to predicate:<br>Both are intended for<br>use with FDA cleared<br>High-level Liquid<br>Disinfectants (HLDs) to<br>disinfect ultrasound<br>probes and are to be<br>used by trained<br>personnel in non-sterile<br>healthcare settings<br>ASTRA TEE:<br>The ASTRA TEE can<br>reprocess up to 2 TEE<br>probes; TD-100 does 1<br>TEE probe at a time<br>The ASTRA TEE uses<br>select FDA cleared<br>HLDS; TD-100 uses<br>proprietary HLD<br>The ASTRA TEE re-uses<br>HLD; TD-100 uses a<br>single-use bottle<br>ASTRA VR:<br>The ASTRA VR con<br>reprocess up to 2<br>endovaginal/endorectal<br>ultrasound probes; TD-<br>100 does 1 TEE probe at<br>a time<br>The ASTRA VR uses<br>select FDA cleared<br>HLDs; TD-100 uses<br>proprietary HLD<br>The ASTRA VR re-uses<br>HLD; TD-100 uses a<br>single-use bottle |
| Prescription<br>Use | No | No | No | ASTRA TEE and ASTRA VR same as<br>predicate. |
| Types of probes | Transesophageal<br>(TEE) probes | Endorectal,<br>endovaginal (VR)<br>probes | Transesophageal<br>probes | (TD-100)<br>ASTRA VR<br>equivalent to TD-100 as both<br>reprocess non-lumened<br>ultrasound probes of various<br>geometries |
| | Summary of Predicate Comparisons | | | |
| Characteristic | GUS ASTRA TEE<br>PCI medical | GUS ASTRA VR<br>PCI medical | TD-100<br>CS Medical | |
| | | | | |
| Theory of<br>Operation | • Control time,<br>temperature<br>and fluid levels<br>to maintain<br>optimal<br>conditions for<br>disinfection<br>and rinsing as<br>established by<br>HLD<br>manufacturers | • Control time,<br>temperature<br>and fluid levels<br>to maintain<br>optimal<br>conditions for<br>disinfection<br>and rinsing as<br>established by<br>HLD<br>manufacturers | • Control time,<br>temperature<br>and fluid levels<br>to maintain<br>optimal<br>conditions for<br>disinfection<br>and rinsing as<br>established by<br>HLD<br>manufacturer | ASTRA TEE and ASTRA VR same as<br>predicate. |
| Disinfection<br>time control | • Timing Chip &<br>Software | • Timing Chip &<br>Software | • Electronics &<br>Software | ASTRA TEE and ASTRA VR<br>Substantially equivalent to TD-<br>100. Time control parameters are<br>contingent to HLD Type and<br>manufacturer specifications. |
| Disinfection<br>temperature<br>control | • Temperature<br>sensor &<br>Software | • Temperature<br>sensor &<br>Software | • Electronics &<br>Software | ASTRA TEE and ASTRA VR<br>Substantially equivalent to TD-<br>100. Temperature control<br>parameters are contingent to HLD<br>Type and manufacturer<br>specifications. |
| HLD warming<br>capability | • Yes (Warmer<br>pad) | • Yes (Warmer<br>pad) | • Yes (Unknown) | ASTRA TEE and ASTRA VR<br>Substantially equivalent to TD-<br>100. |
| Liquid Level<br>Control<br>(HLD and rinse) | • Sensors<br>(upper/lower)<br>& Software | • Sensors<br>(upper/lower)<br>& Software | • Sensor (upper)<br>& Software<br>Fixed volume<br>single use<br>bottle (HLD) | ASTRA TEE and ASTRA VR<br>Substantially equivalent to TD-<br>100. HLD Level control<br>TEE and ASTRA VR use sensors to<br>detect when HLD level is met.<br>Predicate uses a pre-measured<br>HLD single use bottle<br>Rinse Water Level: ASTRA TEE and<br>ASTRA VR same as predicate. |
| Overflow<br>Protection | • Sensor<br>(overflow) &<br>Software | • Sensor<br>(overflow) &<br>Software | • Sensor<br>(overflow) &<br>Software | ASTRA TEE and ASTRA VR same as<br>predicate. |
| Probe<br>Detection | • Sensor<br>(microswitch)<br>& Software | • Sensor<br>(microswitch)<br>& Software | • Sensor<br>(microswitch)<br>& Software | ASTRA TEE and ASTRA VR same as<br>predicate. |
| HLD Vapor<br>Control | • Door, fan &<br>filter | • Door, fan &<br>filter | • Door, fan &<br>filter | ASTRA TEE and ASTRA VR same as<br>predicate. |
| User Interface | • Text based<br>interface<br>4 button<br>keypad<br>USB port | • Text based<br>interface<br>4 button<br>keypad<br>USB port | • Text based<br>interface<br>15 button<br>keypad<br>Printer | ASTRA TEE and ASTRA VR<br>Substantially equivalent to TD-<br>100. ASTRA TEE and ASTRA VR<br>utilize USB port to download data<br>while predicate uses printable<br>data. ASTRA TEE and ASTRA VR |
| Characteristic | GUS ASTRA TEE<br>PCI medical | GUS ASTRA VR<br>PCI medical | TD-100<br>CS Medical | |
| | Barcode<br>Scanner | Barcode<br>Scanner | | utilize Barcode Scanner to capture<br>input data while the predicate<br>device allows manual input<br>through keypad. |
| HLD MRC<br>Verification | Operator must<br>verify MRC | Operator must<br>verify MRC | Operator must<br>verify HLD<br>expiration date | ASTRA TEE and ASTRA VR<br>Substantially equivalent to TD-<br>100. |
| Material<br>Components | Diaphragm<br>Pump Solenoid valves Polyethylene<br>tubing Push Fit fittings Powder coated<br>sheet metal<br>cabinet with<br>ABS plastic<br>components | Diaphragm<br>Pump Solenoid valves Polyethylene<br>tubing Push Fit fittings Powder coated<br>sheet metal<br>cabinet with<br>ABS plastic<br>components | Diaphragm<br>Pump Solenoid valves Polyethylene<br>tubing Fittings ABS Plastic<br>cabinet | ASTRA TEE and ASTRA VR<br>Substantially equivalent to TD-<br>100. |
| Compatible<br>HLDS | Uses FDA<br>cleared HLDS<br>(Metricide<br>OPA) per<br>specifications…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.