Integra® (Jarit®, Ruggles™-Redmond™, Miltex®, MeisterHand®) Kerrison Rongeurs are manually operated instruments indicated for cutting or biting bone during surgery involving the skull or spinal column.
Device Story
Integra Kerrison Rongeurs are reusable, manually operated stainless steel surgical instruments. Used by surgeons in clinical settings for cutting or biting bone during skull or spinal procedures. Devices feature various bite sizes, jaw openings, cutting angles, and shaft lengths; some models include Detach shafts. Instruments are provided non-sterile and are designed to be sterilized by the user. The device functions through mechanical force applied by the surgeon to the handle, which actuates the cutting tip. This submission specifically updates the Instructions for Use to allow sterilization in either assembled or disassembled configurations, supported by existing validation data.
Clinical Evidence
Bench testing only. Validation of pre-vacuum steam sterilization per ANSI/AAMI ST79:2010 and ANSI/AAMI/ISO 14937:2009 (270°F/132°C, 4-minute exposure, 20-minute dry time) was performed and passed.
Technological Characteristics
Manual, non-electrical, reusable surgical instrument. Materials: 420 stainless steel; surface coatings include Titanium Nitride (TiN), TiAlN, Dicronite DL-5, ZrN, PTFE, and Silicone (Elastosil® LR3003/80). Dimensions: 1-6 mm bite sizes, 9-15.5 mm jaw openings, 4.75-15" shaft lengths. Sterilization: Steam (autoclave).
Indications for Use
Indicated for cutting or biting bone during surgery involving the skull or spinal column.
Regulatory Classification
Identification
A manual rongeur is a manually operated instrument used for cutting or biting bone during surgery involving the skull or spinal column.
Predicate Devices
Integra® Jarit®, Ruggles™ -Redmond™, Miltex®, and MeisterHand® Kerrison Rongeurs (K092227)
Submission Summary (Full Text)
{1}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 16, 2015
Integra York PA, Inc. Ms. Stephanie Sheeslev Senior Manager, Regulatory Affairs 589 Davies Drive York, Pennsylvania 17402
Re: K150428
> Trade/Device Name: Integra® (Jarit®, Ruggles™-Redmond™, Miltex®, MeisterHand®) Kerrison Rongeurs Regulation Number: 21 CFR 882.4840 Regulation Name: Manual Rongeur Regulatory Class: Class II Product Code: HAE Dated: February 18, 2015 Received: February 19, 2015
Dear Ms. Sheesley,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{2}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Carlos L. Pena-S
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K150428
### Device Name
Integra® (Jarit®, Ruggles™-Redmond™, Miltex®, MeisterHand®) Kerrison Rongeurs
Indications for Use (Describe)
Integra® (Jarit®, Ruggles™-Redmond™, Miltex®, MeisterHand®) Kerrison Rongeurs are manually operated instruments indicated for cutting or biting bone during surgery involving the skull or spinal column.
Type of Use (*Select one or both, as applicable*)|X Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
#### I. SUBMITTER
Integra York PA, Inc. 589 Davies Drive York, PA 17402 USA
Phone: (717) 717-840-3522 Fax: (717) 840-9347
Contact Person: Stephanie Sheesley Date Prepared: April 15, 2015
#### II. DEVICE
Name of Device: Integra® (Jarit®, Ruggles™-Redmond™, Miltex®, MeisterHand®) Kerrison Rongeurs Common Name: Kerrison Rongeur Classification Name: Manual, Rongeur (21 CFR 882.4840) Regulatory Class: Class II Product Code: HAE
#### III. PREDICATE DEVICE
Integra® Jarit®, Ruggles™ -Redmond™, Miltex®, and MeisterHand® Kerrison Rongeurs, K092227
This predicate device has not been subject to a design-related recall.
#### DEVICE DESCRIPTION IV.
Integra® Kerrison Rongeurs are reusable stainless steel instruments that are sterilizable and packaged non-sterile. Devices are available with the following features: with or without proprietary surface treatments; 1-6 mm bite sizes; 9 - 15.5 mm jaw openings; 40° and 90° up/down cutting angles: regular and thin/low profile footplates: standard and eiector tips: 4.75 -15" shaft lengths; and various handle and shaft styles, including Detach®. Integra® Kerrison Rongeurs are distributed under the following brand names: Jarit®, Ruggles™ -Redmond™, Miltex®, and MeisterHand®.
#### V. INDICATIONS FOR USE
Integra® (Jarit®, Ruggles™-Redmond™, Miltex®, MeisterHand®) Kerrison Rongeurs are manually operated instruments indicated for cutting or biting bone during surgery involving the skull or spinal column.
{5}------------------------------------------------
### VI. COMPARISON OF TECHNOLOGICAL CHRACTERISTICS WITH THE PREDICATE DEVICE
| | Integra York PA, Inc.<br>Integra® Kerrison Rongeurs | Integra Medical Instrument Group<br>(now Integra York PA, Inc.)<br>Integra® Kerrison Rongeurs | |
|------------------------------------|-----------------------------------------------------|-----------------------------------------------------------------------------------------------|--|
| 510(k) # | Subject of Submission | K092227 | |
| Class | II | II | |
| Pro Code | HAE | HAE | |
| Regulation # | 882.4840 | 882.4840 | |
| Classification | Manual, Rongeur | Manual, Rongeur | |
| Indications for Use | Kerrison Rongeurs are manually | Kerrison Rongeurs are manually | |
| | operated instruments indicated for | operated instruments indicated for | |
| | cutting or biting bone during surgery | cutting or biting bone during surgery | |
| | involving the skull or spinal column. | involving the skull or spinal column. | |
| Contact Materials | 420 Stainless steel | 420 Stainless steel | |
| | Surface coatings: Titanium | Surface coatings: Titanium | |
| | Nitride(TiN); Diamond Coat and Hard- | Nitride(TiN); Diamond Coat and Hard- | |
| | Coat™ (TiAIN): Stealth Coat and | Coat™ (TiAIN); Stealth Coat and | |
| | Smooth-Coat™ (Dicronite DL-5); | Smooth-Coat™ (Dicronite DL-5); | |
| | Ultra-Coat™ (ZrN); PTFE | Ultra-Coat™ (ZrN); PTFE | |
| | (polytetrafluoroethylene [Teflon®]); | (polytetrafluoroethylene [Teflon®]); | |
| | Silicone (Elastosil® LR3003/80 A,B) | Silicone (Elastosil® LR3003/80 A,B) | |
| Instructions for Use | Sterilize in either the Assembled or | Sterilize in Disassembled. Open | |
| | Disassembled, Open Position. | Position. | |
| | Successful Sterilization Validation | | |
| | supports Assembled and Open | | |
| | Configuration | | |
| Technological | Identical to Predicate | Manual, non-electrical, non-sterile, | |
| Characteristics | | reusable | |
| Malleability | Non-malleable | Non-malleable | |
| Sterility | Non-sterile | Non-sterile | |
| Sterilization | Sterilizable | Sterilizable | |
| Utility | Reusable | Reusable | |
| Energy Source | N/A (manual) | N/A (manual) | |
| Image of Example | | | |
| Kerrison Rongeur -<br>Detach Shaft | | | |
| | | | |
| | | | |
The below table provides a comparison of the subject devices and the predicate devices.
There are no differences in technology, materials, intended use, or design between the subject devices and the above predicates. The only change is with respect to the labeling. The Instructions For Use for the predicate device stated that the device was to be sterilized in the disassembled but open configuration. The Instructions for Use for the subject devices allows the devices to be sterilized in the either the disassembled, but open configuration. This change is supported by a successful sterilization validation.
{6}------------------------------------------------
#### VII. PERFORMANCE DATA
No performance standards have been promulgated under Section 514 of the Food. Drug and Cosmetic Act for these devices. No new testing was required to be performed in support of this labeling change. This Special 510(k) pre-market notification includes the sterilization validation submitted with the predicate device submission #K092227 as a reference only.
The below testing is provided as a reference to support the labeling change subject to this Special 510(k). This testing is not new and was previously included in the 510(k) submission #K092227 and was deemed acceptable to support the proposed labeling change per Pre-Submission #Q141112.
| Testing Performed | Results |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Pre-Vacuum (wrapped) Steam Sterilization Validation per ANSI/AAMI ST79:2010<br>& A1:2010 and ANSI/AAMI/ISO 14937:2009 at 270°F (132°C) with an Exposure<br>Time of 4 minutes and Drying Time of 20 minutes | Pass |
No biocompatibility testing was performed as a result of this Special 510(k) as the subject devices are not under-going a change in materials - this submission is for a labeling change only.
## VIII. CONCLUSIONS
The labeling was the only change submitted under this Special 510(k) and supports the substantial equivalence to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.