← Product Code [GAT](/productcode/GAT) · K150327

# Dunamis Force DFX Suture, Size 5, Dunamis Force DFX Suture Size 2 (K150327)

_Dunamis, LLC · GAT · Jun 29, 2015 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K150327

## Device Facts

- **Applicant:** Dunamis, LLC
- **Product Code:** [GAT](/productcode/GAT.md)
- **Decision Date:** Jun 29, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.5000
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Dunamis Force DFX Sutures, Polyethylene Non-Absorbable Surgical Sutures are indicated for use in approximation and or ligation of soft tissues including use in cardiovascular surgeries and the use of allograft tissue for orthopaedic surgeries

## Device Story

Dunamis Force DFX Suture is a sterile, single-use, non-absorbable surgical suture. It consists of a co-braid of Ultra High Molecular Weight Polyethylene (UHMWPE) and blue monofilament polypropylene. The device is used by surgeons in clinical settings for soft tissue approximation and ligation, including cardiovascular and orthopaedic procedures. It is provided as a pre-attached, single-armed needle suture. The device functions as a mechanical fastener to hold tissues together during healing. It is contract-manufactured to the same specifications as the predicate device.

## Clinical Evidence

No clinical data. Substantial equivalence is supported by the fact that the device is contract-manufactured to the identical specifications as the predicate device (K063778). Performance is supported by USP testing for non-absorbable sutures, ISO 10993-1 biocompatibility testing, and stability testing previously cleared for the predicate.

## Technological Characteristics

Co-braid of UHMWPE and blue monofilament polypropylene. Colorant: [phthalocyaninato (2-)] copper (<0.5% by weight). USP Size 2 and Size 5. Sterile, single-use. Meets USP requirements for non-absorbable sutures (except diameter).

## Regulatory Identification

Nonabsorbable poly(ethylene terephthalate) surgical suture is a multifilament, nonabsorbable, sterile, flexible thread prepared from fibers of high molecular weight, long-chain, linear polyesters having recurrent aromatic rings as an integral component and is indicated for use in soft tissue approximation. The poly(ethylene terephthalate) surgical suture meets U.S.P. requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be provided uncoated or coated; and it may be undyed or dyed with an appropriate FDA listed color additive. Also, the suture may be provided with or without a standard needle attached.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.

## Predicate Devices

- Teleflex Force Fiber Polyethylene Non-Absorbable Surgical Sutures ([K063778](/device/K063778.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 29, 2015

Dunamis, LLC % Mr. Robert O. Dean Lexamed 705 Front Street Toledo, Ohio 43605

Re: K150327

Trade/Device Name: Dunamis Force DFX Suture Regulation Number: 21 CFR 878.5000 Regulation Name: Nonabsorbable poly(ethylene terephthalate) surgical suture. Regulatory Class: Class II Product Code: GAT Dated: March 25, 2015 Received: March 31, 2015

Dear Mr. Dean:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{1}------------------------------------------------

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# David Krause -S

- Binita S. Ashar, M.D., M.B.A., F.A.C.S. for Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure

{2}------------------------------------------------

| DEPARTMENT OF HEALTH AND HUMAN SERVICES |  |
|-----------------------------------------|--|
| Food and Drug Administration            |  |

#### Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.

510(k) Number (if known)

Device Name

Dunamis Force DFX Suture

Indications for Use (Describe)

The Dunamis Force DFX Sutures, Polyethylene Non-Absorbable Surgical Sutures are indicated for use in approximation and or ligation of soft tissues including use in cardiovascular surgeries and the use of allograft tissue for orthopaedic surgeries

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW .*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect. of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

FORM FDA 3881 (8/14)

Page 1 of 1

PSC Pollohing Samlowi (JWD #45-6749 EIF

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Dunamis Medical. The word "Dunamis" is in gray, with the word "Medical" in a smaller font below it. To the right of "Dunamis" is an orange swoosh above the words "inspiring movement" in orange. 

#### 510K Summary - Rev. 2

- 1. Submitter: Dr. Prithvi Raj Chavan
Dunamis, LLC. 693 Sherling Lake Rd. Greenville, AL. 36037 (p) 731-217-2533 Email: dr.raj76@gmail.com

- 2. Date Prepared: 04/21/15
# 3. Name of Device:

- Proprietary Trade Name: Dunamis Force DFX Suture —
- Common Name: Suture, Polyethylene Synthetic Non-Absorbable Surgical Suture
- Classification Name: Non-absorbable poly (ethylene terephthalate) surgical suture
- Code: GAT ー
- Regulation:878.5000

# 4. Identification of the legally marketed device (predicate):

Teleflex Force Fiber Polyethylene Non-Absorbable Surgical Sutures, K063778.

- 5. Description of the device: The Dunamis Force DFX Sutures are sterile single use sutures are used in approximation and/or ligation of soft tissues, including use in cardiovascular surgeries and the use of allograft tissue for orthopaedic surgeries. The technical design configuration for both USP Size 2 and Size 5 are as follows:
#### USP Size 2 (metric 5)

White/Blue Co-braid - Ultra High Molecular Weight Polyethylene (UHMWPE) Suture USP Size 2 (oversized) x 36" length, single armed HC-5 needle.

Packaged 1(one) strand per pouch, 12 pouches per box, sterile.

#### USP Size 5 (metric 7)

White/Blue Co-braid - Ultra High Molecular Weight Polyethylene (UHMWPE) Suture USP Size 5 (oversized) x 36" length, single armed HS48 needle.

Packaged 1(one) strand per pouch, 12 pouches per box, sterile

#### Materials:

Co-Braid of UHMWPE and Blue Monofilament Polypropylene Non-absorbable Surgical Sutures meets USP requirements except for diameter.

Polymer/Colorant: Ultra High Molecular Weight Polyethylene (white): None

Polypropylene (blue): [bhthalocyaninato (2-)] copper with a concentration not to exceed 0.5% by weight.

- 6. Indications for Use: The Dunamis Force DFX Sutures, Polyethylene Non-Absorbable Surgical Sutures are indicated for use in approximation and/or ligation of soft tissues including use in cardiovascular surgeries and the use of allograft tissue for orthopaedic surgeries.

{4}------------------------------------------------

- 7 . Comparison of Technological Characteristics: The Dunamis Force DFX Sutures are contract manufactured for Dunamis by Teleflex Medical to the same specifications for the Force Fiber sutures cleared in K063778; therefore, the device characteristics are identical to the predicate. The only differences between the Dunamis device and the predicate device are the brand name, part number, and "manufactured for" indication on the product labeling.
- 8. Description of Device Design and Test Methods: Dunamis LLC. has established a contract manufactured product specifications that has been accepted by Teleflex which documents and secures the product configuration and required test methods to assure the device is equivalent in performance and to all the elements of Teleflex 510k K063778.

# USP Size 2 (metric 5)

White/Blue Co-braid - Ultra High Molecular Weight Polyethylene (UHMWPE) Suture USP Size 2 (oversized) x 36" length, single armed HC-5 needle.

Packaged 1(one) strand per pouch, 12 pouches per box, sterile.

# USP Size 5 (metric 7)

White/Blue Co-braid - Ultra High Molecular Weight Polyethylene (UHMWPE) Suture USP Size 5 (oversized) x 36" length, single armed HS48 needle.

Packaged 1(one) strand per pouch, 12 pouches per box, sterile

# Materials:

Co-Braid of UHMWPE and Blue Monofilament Polypropylene Non-absorbable Surgical Sutures meets USP requirements except for diameter.

Polymer/Colorant: Ultra High Molecular Weight Polyethylene (white): None

Polypropylene (blue): [bhthalocyaninato (2-)] copper with a concentration not to exceed 0.5% by weight.

#### Performance Testing:

Dunamis has established contract manufactured product specifications with Telefical that documents the product configuration and required test methods to assure the device is identical in performance and to all the elements of Teleflex 510(k) K063778. Because the sutures are manufactured to the same specifications as those cleared in K063778, data provided in K063778 may be leveraged to support equivalence, including testing in accordance to the USP for non-absorbable sutures, biocompatibility testing performed in accordance with ISO 10993-1, and stability testing. No additional testing was performed on the contractmanufactured product.

- 9. Conclusions: The Dunamis Force DFX Sutures are identical to the predicate Force Fiber sutures cleared in K063778 with respect to indications for use, materials, design, and manufacturing processes. The sutures are contract manufactured for Dunamis by Teleflex to the same product specifications cleared in K063778. Therefore, the Dunamis Force DFX Sutures are substantially equivalent to the Force Fiber sutures cleared in K063778.

---

**Source:** [https://fda.innolitics.com/device/K150327](https://fda.innolitics.com/device/K150327)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
