← Product Code [LZO](/productcode/LZO) · K150302

# PROFEMUR Preserve Classic Stem (K150302)

_Microport Orthopedics, Inc. · LZO · Apr 10, 2015 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K150302

## Device Facts

- **Applicant:** Microport Orthopedics, Inc.
- **Product Code:** [LZO](/productcode/LZO.md)
- **Decision Date:** Apr 10, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3353
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The PROFEMUR® Preserve Classic Stems are indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions: 1. non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia; 2. inflammatory degenerative joint disease such as rheumatoid arthritis; 3. correction of functional deformity; and, 4. revision procedures where other treatments or devices have failed. The PROFEMUR® Preserve Classic Stems are single use components, intended for use in conjunction with associated ceramic or metal femoral heads as part of uncemented total hip arthroplasty.

## Device Story

Monolithic femoral hip stem; forged titanium alloy; used in uncemented total hip arthroplasty. Implanted by orthopedic surgeons in clinical settings. Provides structural support for hip joint; replaces diseased or damaged femoral head/neck. Available in twelve sizes with straight or varus neck offsets. Features unalloyed titanium plasma coating for bone integration. Used with compatible modular femoral heads and acetabular shells. Benefits patient through pain reduction and improved hip function.

## Clinical Evidence

No clinical data provided. Substantial equivalence supported by bench testing, including range of motion (ISO 21535) and proximal fatigue testing (ISO 7206-6). Distal fatigue testing (ISO 7206-4) was verified via comparison to a modular stem with similar geometry. FEA models were previously validated in K141235.

## Technological Characteristics

Monolithic femoral component; forged Ti6Al4V titanium alloy (ASTM F620); unalloyed titanium plasma spray coating (ASTM F1580 powder). Twelve sizes; straight and varus neck offsets. Uncemented fixation. Mechanical device; no energy source or software.

## Regulatory Identification

A hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis is a device intended to be implanted to replace a hip joint. This device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. The two-part femoral component consists of a femoral stem made of alloys to be fixed in the intramedullary canal of the femur by impaction with or without use of bone cement. The proximal end of the femoral stem is tapered with a surface that ensures positive locking with the spherical ceramic (aluminium oxide, A12 03 ) head of the femoral component. The acetabular component is made of ultra-high molecular weight polyethylene or ultra-high molecular weight polyethylene reinforced with nonporous metal alloys, and used with or without bone cement.

## Predicate Devices

- PROFEMUR® Preserve Stem ([K112080](/device/K112080.md))

## Reference Devices

- DYNASTY® Biofoam Shell ([K122382](/device/K122382.md))
- PROFEMUR® TL Classic Hip Stem ([K123688](/device/K123688.md))
- PROFEMUR® TL Classic Long Neck Hip Stems ([K140676](/device/K140676.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three faces in profile, overlapping each other to create a sense of depth and unity. The faces are simple and abstract, with no distinct features other than the outline of the head and neck.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

MicroPort Orthopedics. Incorporated Mr. Matt Paul Regulatory Affairs Project Manager 5677 Airline Road Arlington, Tennessee 38002

April 10, 2015

Re: K150302 Trade/Device Name: PROFEMUR® Preserve Classic Stem Regulation Number: 21 CFR 888.3353 Regulation Name: Hip joint metal/ceramic/polymer semi-constrained cemented or nonporous uncemented prosthesis Regulatory Class: Class II Product Code: LZO Dated: February 5, 2015 Received: February 6, 2015

Dear Mr. Paul:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing

{1}------------------------------------------------

## Page 2 - Mr. Matt Paul

(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Mark N. Melkerson -S

Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

510(k) Number (if known)

K150302 (page 1/1)

Device Name PROFEMUR® Preserve Classic Stem

Indications for Use (Describe)

The PROFEMUR® Preserve Classic Stems are indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:

1. non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia;

2. inflammatory degenerative joint disease such as rheumatoid arthritis;

3. correction of functional deformity; and,

4. revision procedures where other treatments or devices have failed.

The PROFEMUR® Preserve Classic Stems are single use components, intended for use in conjunction with associated ceramic or metal femoral heads as part of uncemented total hip arthroplasty.

Type of Use (Select one or both, as applicable)

2 Prescription Use {Part 21 CFR 801 Subpart D}

Over-The-Counter Use (21 CFR 801 Subpart C)

PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.

### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo and contact information for MicroPort Orthopedics. The logo features the company name in blue text with a red star in a blue square. Below the logo is the company's address, which is 5677 Airline Road, Arlington, TN 38002, and their website, ortho.microport.com.

# 510(k) Summary of Safety and Effectiveness

In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807, this information serves as a Summary of Safety and Effectiveness for the use of the PROFEMUR® Preserve Classic Stems.

| Submitted by:                        | MicroPort Orthopedics Inc.<br>5677 Airline Rd, Arlington TN, 38002<br>Phone: 901-867-4350<br>Fax: 901-451-6018                    |
|--------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| Date:                                | February 5, 2015                                                                                                                  |
| Contact Person:                      | Matt Paul<br>Requlatory Affairs Project Manager                                                                                   |
| Proprietary Name:                    | PROFEMUR® Preserve Classic Stem                                                                                                   |
| Common Name:                         | Hip Stem                                                                                                                          |
| Classification Name and Reference:   | 888.3353 LZO Hip joint metal/ceramic/polymer semi<br>constrained cemented or nonporous, uncemented prosthesis -<br>Class II       |
| Subject Product Code and Panel Code: | Orthopedics/87/ LZO                                                                                                               |
| Predicate Device:                    | PROFEMUR® Preserve Stem (K112080)                                                                                                 |
| Reference Devices:                   | DYNASTY® Biofoam Shell (K122382)<br>PROFEMUR® TL Classic Hip Stem (K123688)<br>PROFEMUR® TL Classic Long Neck Hip Stems (K140676) |

{4}------------------------------------------------

#### l. Device Description

The PROFEMUR® Preserve Classic Stems are monolithic femoral components manufactured from a forged titanium alloy (ASTM F620). The hip stems are designed for use in cementless total hip arthroplasty and possess identical indications for use to the predicates (K112080). The design features of the PROFEMUR® Preserve Classic Stems are summarized below:

- . Stems are forged from Ti6Al4V alloy conforming to ASTM F620
- Stems possess unalloyed titanium plasma coating manufactured from powder ● conforming to ASTM F1580
- . Neck offsets are offered in Straight and Varus 8 options
- Stems are offered in twelve sizes for both neck offsets ●

#### II. Indications for Use

Identically to the predicate, the PROFEMUR® Preserve Classic Stems are indicated for use in total hip arthroplasty for reduction or relief of pain and/or improved hip function in skeletally mature patients with the following conditions:

- 1. non-inflammatory degenerative joint disease such as osteoarthritis, avascular necrosis, ankylosis, protrusio acetabuli, and painful hip dysplasia;
- 2. inflammatory degenerative joint disease such as rheumatoid arthritis;
- 3. correction of functional deformity; and,
- 4. revision procedures where other treatments or devices have failed.

The PROFEMUR® Preserve Classic Stems are single use components, intended for use in uncemented total hip arthroplasty.

PROFEMUR® Preserve Classic is intended for use with the sterile compatible components listed in Table 1 below, identical to the predicate.

{5}------------------------------------------------

|                              | Table 1. Compatible Components                                                |  |
|------------------------------|-------------------------------------------------------------------------------|--|
| 510(k)                       | Intended Combinations                                                         |  |
| Modular Femoral Neck Sleeves |                                                                               |  |
| K072656                      | Short to X-long neck sleeves to use with ceramic-polyethylene with ID 38-46mm |  |
| Modular Femoral Heads        |                                                                               |  |
| K953025                      | Metal-Polyethylene, OD 28mm XXL                                               |  |
| K932222                      | Metal-Polyethylene, OD 28mm                                                   |  |
| K920593                      | Ceramic-Polyethylene, OD 28mm                                                 |  |
| K925512                      | Ceramic-Polyethylene, OD 28mm                                                 |  |
| K893685                      | Ceramic-Polyethylene with ID 28-36mm                                          |  |
| K002149                      | Metal-Polyethylene with ID 22-36mm, OD 42-68mm                                |  |
| K051348                      | Metal-Polyethylene with ID 36-56mm                                            |  |
| K072656                      | Ceramic-Polyethylene with ID 38-46mm                                          |  |
| Acetabular Shell and Inserts |                                                                               |  |
| K002149                      | Metal-Poly with ID 22-36mm, OD 42-68mm                                        |  |
| K043073                      | Metal-Polyethylene and Ceramic-Polyethylene with ID 22-36mm, OD 46-68mm       |  |
| K043099                      | Metal-Polyethylene and Ceramic-Polyethylene with ID 22-36mm, OD 46-68mm       |  |
| K052026                      | Metal-Polyethylene and Ceramic-Polyethylene with ID 28-36mm                   |  |
| K061547                      | Metal-Polyethylene with ID 32-42mm, OD 50-58mm                                |  |
| K070785                      | Metal-Polyethylene and Ceramic-Polyethylene with ID 28-46mm, OD 46-68mm       |  |
| K082924                      | Metal-Polyethylene with ID 28-54mm, OD 46-76mm                                |  |
| K122382                      | Metal-Polyethylene with OD 46-76 mm                                           |  |

#### III. Nonclinical Testing

The subject PROFEMUR® Preserve Classic Stems were evaluated for range of motion in accordance with ISO 21535. The stems were also successfully evaluated by proximal fatigue testing according to ISO 7206-6. The subject stems were verified to be applicable to ISO 7206-4 distal fatigue testing that was performed on a modular stem possessing a similar geometry. The FEA models used in comparison were previously validated in testing submitted in K141235. Successful results of testing support the expectation that the device will perform as intended under normal physiological loading conditions, and support the substantial equivalence of the PROFEMUR® Preserve Classic Stem.

#### IV. Clinical Testing

Clinical data was not provided for the subject devices.

{6}------------------------------------------------

#### V. Technological Characteristics

The PROFEMUR® Preserve Classic Stems are designed to provide geometric sizing options that resemble a subset of those available with the predicate device. Furthermore, the subject devices are made from an identical titanium alloy (ASTM F620) and possess an identical titanium plasma spray coating design specification and material (ASTM F1580 powder) to the predicate device. Coating process and vendor information are identical to those provided for reference device K123688.

For all associated instruments that do contain colorants, detailed color additive information was provided for the same instruments in K140676.

#### VI. Substantial Equivalence Information

The intended use and indications for use of the PROFEMUR® Preserve Classic Stems are identical to those of the predicate devices (K112080). The materials of the subject devices are identical to those of the predicate devices. The fundamental scientific technology of the modified devices has not changed relative to that of the predicate devices. The safety and effectiveness of the subject devices are adequately supported by the substantial equivalence information, materials information, and analysis data provided within this Premarket Notification. For more details on subject and predicate comparison, see the comparison table located in the Device Description (Tab 014).

#### VII. Conclusion

The design features, subject and predicate testing, materials information, and analysis data provided in this premarket notification adequately support the substantial equivalence of the PROFEMUR® Preserve Classic Stem.

---

**Source:** [https://fda.innolitics.com/device/K150302](https://fda.innolitics.com/device/K150302)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
