The EZvue UV (ocufilcon D) Spherical Soft Contact Lenses for daily wear are indicated for the correction of refractive error in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia. The lens may be worn by persons who exhibit refractive astigmatism of 2.00 diopters or less where the astigmatism does not interfere with visual acuity. The EZvue Colors (ocufilcon D) Spherical Soft Contact Lenses for daily wear are indicated for the correction of refractive error in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia. The lens may be worn by persons who exhibit refractive astigmatism of 2.00 diopters or less where the astigmatism does not interfere with visual acuity. The lens is available clear or tinted and may be used to enhance or alter the apparent color of the eye. Daily wear replacement schedules may vary from patient and should be decided by eyecare practitioners in consultation with their patients. Frequent/Planned Replacement Wear: Eyecare practitioners may prescribe any of the above lenses for frequent/planned replacement wear, with cleaning disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfecting system. Disposable Wear: Eyecare practitioners may prescribe any of the above lenses for single use Daily Disposable Wear. When Prescribed for Daily Disposable Wear the lens is to be discarded after each removal.
Device Story
Soft contact lenses (ocufilcon D) function as refractive media to focus light on retina; hemispherical shells with molded spherical base curves and front surfaces. EZvue UV incorporates UV-absorbing monomer; EZvue Colors features iris-patterned tinting using FDA-listed color additives imprinted on non-optical zone. Lenses worn by patients to correct refractive errors; prescribed by eyecare practitioners. Hydrated state (55% water) required for proper optical performance; conform to corneal curvature. Benefits include vision correction and potential cosmetic enhancement. No electronic or software components.
Clinical Evidence
No clinical data provided. Safety and effectiveness established via non-clinical testing, including in vitro and in vivo toxicology, biocompatibility, and shelf-life stability studies conducted under GLP.
Technological Characteristics
Material: ocufilcon D (hydrophilic co-polymer of 2-HEMA and methacrylic acid, cross-linked with EGDMA). Water content: 55±2%. Oxygen permeability: 16 x 10^-11 (cm²/sec)(mlO₂)/(ml x mmHg @ 35°C). Refractive index: 1.410. UV-blocking: >95% UVB, >50% UVA (EZvue UV). Manufacturing: fully-molded. Sterilization: blister package. No software or electronic components.
Indications for Use
Indicated for aphakic and non-aphakic persons with non-diseased eyes requiring correction of myopia, hyperopia, or refractive astigmatism ≤ 2.00 diopters. Available for daily wear, frequent/planned replacement, or single-use daily disposable wear.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol of three human profiles facing right, arranged in a cascading manner.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
April 15, 2015
EyeReg Consulting, Inc. I-SEE VISION TECHNOLOGY Inc. Mr. Bret Andre Official Correspondent 474 NE 61st PL Hillsboro, OR 97124
Re: K150293
Trade/Device Name: Ezvue Uv, Soft Daily Wear Contact Lenses (ocufilcon D), Ezvue Colors, Soft Daily Wear Contact Lenses (oculfilcon D) Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (Hydrophilic) Contact Lens Regulatory Class: Class II Product Code: LPL. MVN Dated: February 3, 2015 Received: February 6, 2015
Dear Mr. Andre:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification"
(21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## Kesia Y. Alexander -A
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose, and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known) K150293
#### Device Name EZvue UV, Soft Daily Wear Contact Lenses (ocufilcon D)
EZvue Colors, Soft Daily Wear Contact Lenses (ocufilcon D)
Indications for Use (Describe)
The EZvue UV (ocufilcon D) Spherical Soft Contact Lenses for daily wear are indicated for the correction of refractive error in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia. The lens may be worn by persons who exhibit refractive astigmatism of 2.00 diopters or less where the astigmatism does not interfere with visual acuity.
The EZvue Colors (ocufilcon D) Spherical Soft Contact Lenses for daily wear are indicated for the correction of refractive error in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia. The lens may be worn by persons who exhibit refractive astigmatism of 2.00 diopters or less where the astigmatism does not interfere with visual acuity. The lens is available clear or tinted and may be used to enhance or alter the apparent color of the eye.
Daily wear replacement schedules may vary from patient and should be decided by eyecare practitioners in consultation with their patients.
Frequent/Planned Replacement Wear:
Eyecare practitioners may prescribe any of the above lenses for frequent/planned replacement wear, with cleaning disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfecting system.
Disposable Wear:
Eyecare practitioners may prescribe any of the above lenses for single use Daily Disposable Wear. When Prescribed for Daily Disposable Wear the lens is to be discarded after each removal.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| The assigned 510(k) number is: | <b>K150293</b> |
|--------------------------------|----------------------------------------------------------------------------------------------------------------------------|
| Applicant information: | |
| Date Prepared: | 03/30/2015 |
| Name: | I-SEE VISION TECHNOLOGY INC.<br>1F, No.45, Lane 2, Sec. 2,<br>Kuang-Fu Road, Hsinchu 30071,<br>Taiwan, R.O.C. |
| Contact Person: | Isaac Huang<br>Special Assistant to President |
| Phone number: | +886-3-5752822 |
| Consultant: | Bret J Andre<br>EyeReg Consulting, Inc.<br>474 NE 61st Pl<br>Hillsboro, OR 97124<br>United States |
| Phone number: | (503) 372-5226 |
| Device Information: | |
| Device Classification: | Class II |
| Product Code: | LPL; MVN |
| Classification Name: | Soft (hydrophilic) Contact Lens (21 CFR 886.5925) |
| Trade Name: | EZvue UV, Soft Daily Wear Contact Lenses<br>(ocufilcon D)<br>EZvue Colors, Soft Daily Wear Contact Lenses<br>(ocufilcon D) |
{5}------------------------------------------------
## Equivalent Devices:
The EZvue UV (ocufilcon D) and EZvue Colors (ocufilcon D), Soft Daily Wear Contact Lenses are substantially equivalent to the following predicate devices:
Predicate devices:
- "EZyue (ocufilcon D)" . By I-See Vision Technology Inc. 510(k) number: K082879 Primary Predicate
- "42 UV (hefilcon A)" . By Optical Connection, Inc. 510(k) number: K040900 Reference Predicate
- "Frequency Colors (methafilcon A)" . By Coopervision, Inc. 510(k) number: K001090 Reference Predicate
## EZvue UV(ocufilcon D) Device Description:
The EZvue UV soft contact lenses are hemispherical shells with molded spherical base curves and molded front surfaces.
The nonionic lens material (ocufilcon D) is a hydrophilic co-polymer of 2-Hydroxyethyl methacrylate (2-HEMA) and methacrylic acid, cross-linked with ethylene glycol dimethacrylate (EGDMA), plus an initiator. The co-polymer consists of 45% ocufilcon D and 55% water by weight when immersed in saline solution. The (ocufilcon D) name has been adopted by the United States Adopted Names Council (USAN).
A UV absorbing monomer-2-(Benzoyl-3-hydroxyphenoxy)ethyl acrylate-is incorporated in the contact lens material to block UV radiation. The UV blocking for EZvue UV averages >95% in the UVB range of 280nm - 315nm and >50% in the UVA range of 316nm - 380mm.
EZvue UV lenses are available clear and visibility tinted (for handling) using C.I. Reactive Blue Dye #19. In the hydrated state, the lens conforms to the curvature of the eye covering the cornea and extending slightly beyond the limbus forming a transparent or colored optical surface. The (ocufilcon D) soft hydrophilic contact lens has a spherical back surface. The hydrophilic properties of the lens require that it be maintained in a fully hydrated state in a solution compatible with the eye. If the lens dries out, it will become hard and appear somewhat warped; however, it will return to its proper configuration when completely rehydrated in the proper storage solution.
{6}------------------------------------------------
## EZvue Colors (ocufilcon D) Device Description:
The EZvue Colors soft contact lenses are hemispherical shells with molded spherical base curves and molded front surfaces.
The nonionic lens material (ocufilcon D) is a hydrophilic co-polymer of 2-Hydroxyethyl methacrylate (2-HEMA) and methacrylic acid, cross-linked with ethylene glycol dimethacrylate (EGDMA), plus an initiator. The co-polymer consists of 45% ocufilcon D and 55% water by weight when immersed in saline solution. The (ocufilcon D) name has been adopted by the United States Adopted Names Council (USAN).
EZvue Colors lenses are tinted to enhance or alter the apparent color of the eye. Lenses are tinted with a combination of one or more of the following 'listed' color additives:
| Name of Colorant | Listing |
|-------------------------|------------------|
| C.I. Reactive Blue 19 | 21 CFR § 73.3127 |
| C.I. Pigment Green 7 | 21 CFR § 73.3124 |
| Iron oxides | 21 CFR § 73.3125 |
| C.I. Reactive Yellow 86 | 21 CFR § 73.3121 |
| C.I. Reactive Black 5 | 21 CFR § 73.3127 |
| C.I. Pigment Violet 23 | 21 CFR § 73.3107 |
| Titanium dioxide | 21 CFR § 73.3126 |
Lenses that contain a unique tinting pattern are subsequently processed to incorporate the 'listed' color additives, and contain only the amount of color additive needed to accomplish the intended coloring effect. When producing the tinted lenses, the manufacturing process alters and/or changes the specifications to the clear version of a contact lens by imprinting a listed color additive(s) on the non-optical zone area of the lens, which corresponds to the location of the iris. The imprinted color pattern is surrounded on both sides by layers of contact lens material (ocufilcon D). As part of the manufacturing process, the lenses containing the color additives are thoroughly washed to remove any excess color additives. The color additives used are not removed by lens handling and cleaning/disinfecting procedures. Except for affecting the amount of light transmittance through the lens in the location that corresponds to the iris, the coloring process does not alter the original characteristics of the pre-tinted lens. The tinting pattern has a Clear Pupil diameter of 5.5 to 9.0 mm.
In the hydrated state, the lens conforms to the curvature of the eye covering the cornea and extending slightly beyond the limbus forming a transparent or colored optical surface. The (ocufilcon D) soft hydrophilic contact lens has a spherical back surface. The hydrophilic properties of the lens require that it be maintained in a fully hydrated state in a solution compatible with the eye. If the lens dries out, it will become hard and appear somewhat warped; however, it will return to its proper configuration when completely rehydrated in the proper storage solution.
{7}------------------------------------------------
The physical properties of the EZvue UV (ocufilcon D) and EZvue Colors (ocufilcon D). Soft Daily Wear Contact Lenses are:
| Refractive Index | 1.410 |
|-------------------------------|----------------------------------------------------------------------|
| Light Transmission (clear/UV) | greater than 93% |
| Light Transmission (color) | greater than 93% (clear region corresponding to pupil) |
| Surface Character | hydrophilic |
| Water Content | 55±2% |
| Specific Gravity | 1.05 (hydrated) |
| Oxygen Permeability | 16 x 10-11 (cm²/sec)(mlO₂)/(ml x mmHg @ 35°C)) (revised Fatt method) |
The hydrophilic characteristics allow aqueous solution to enter the lens, and in its fully hydrated state the lens is approximately 55% water by weight. The lenses will be manufactured in the spherical design configuration with the following features and properties:
| Chord Diameter: | 13.60 mm to 15.00 mm |
|-------------------|----------------------------------|
| Center Thickness: | 0.060 mm to 0.200 mm |
| Base Curve: | 8.40 mm to 9.10 mm |
| Power Range | |
| - Sphere Power: | -12.00D to +8.00D in 0.25D steps |
## Intended Use:
The EZvue UV (ocufilcon D) Spherical Soft Contact Lenses for daily wear are indicated for the correction of refractive error in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia. The lens may be worn by persons who exhibit refractive astigmatism of 2.00 diopters or less where the astigmatism does not interfere with visual acuity.
The EZvue Colors (ocufilcon D) Spherical Soft Contact Lenses for daily wear are indicated for the correction of refractive error in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia. The lens may be worn by persons who exhibit refractive astigmatism of 2.00 diopters or less where the astigmatism does not interfere with visual acuity. The lens is available clear or tinted and may be used to enhance or alter the apparent color of the eye.
Daily wear replacement schedules may vary from patient and should be decided by eyecare practitioners in consultation with their patients.
Frequent/Planned Replacement Wear:
Eyecare practitioners may prescribe any of the above lenses for frequent/planned replacement wear, with cleaning disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be disinfected using a chemical disinfecting system.
{8}------------------------------------------------
#### K150293
## Disposable Wear:
Eyecare practitioners may prescribe any of the above lenses for single use Daily Disposable Wear. When Prescribed for Daily Disposable Wear the lens is to be discarded after each removal.
## Testing:
- Non-clinical Testing A series of in vitro and in vivo preclinical toxicology and biocompatibility tests were performed to assess the safety and effectiveness of the EZvue UV (ocufilcon D) and EZvue Colors (ocufilcon D), Soft Daily Wear Contact Lenses packaged in blister packages. All non-clinical toxicology tests were conducted in accordance with the GLP regulation. All other testing was conducted according to valid scientific protocols.
Test results of the non-clinical testing on the EZvue UV (ocufilcon D) and EZvue Colors (ocufilcon D), Soft Daily Wear Contact Lenses demonstrate that:
- . Lenses supplied in blister packages are sterile for the indicated shelf-life,
- . The packaging material and extracts are not toxic and not irritating,
- Clinical Data The clinical performance of the (polymacon) lens material has been previously established, and therefore was not required for this 510(k).
## Conclusions Drawn from Studies
#### Validity of Scientific Data
Several laboratories under Good Laboratory Practice regulations conducted toxicology studies, microbiology, chemistry, shelf-life stability studies and followed scientific protocols. The data were determined to be scientifically valid under 21 CFR 860.7
## Substantial Equivalence
Information presented in this Premarket Notification establishes that the EZvue UV (ocufilcon D) and EZvue Colors (ocufilcon D), Soft Daily Wear Contact Lenses is as safe and effective as the predicate device when used in accordance with the labeled directions for use and for the requested indication.
## Risks and Benefits
The risks of the subject device are the same as those normally attributed to the wearing of daily wear soft contact lenses. The benefits to the patient are the same as those for other daily wear soft contact lenses.
{9}------------------------------------------------
## Substantial Equivalence:
The EZvue UV/ EZvue Colors Soft Contact Lens will be manufactured according to specified process controls and a cGMP quality assurance program currently in place.
The final packaging and sterilization of the lenses will be carried out in accordance with the currently established procedures for EZvue (K082879), the predicate device.
The established safety profile of the EZvue UV/ EZvue Colors contact lens material is equivalent to the predicate devices identified previously. Being similar with respect to indications for use, materials, physical construction and safety & effectiveness to the predicate devices, this meets the requirements per section 510(k) of the act regarding substantial equivalence and does not raise different questions of safety and effectiveness than the predicate device identified above.
The following matrix illustrates the equivalencies of the EZvue UV(ocufilcon D) / EZvue Colors (ocufilcon D) Soft Contact Lenses for Daily Wear, as well as the substantial equivalent predicate devices.
{10}------------------------------------------------
## Substantial Equivalence Matrix
| | I-See Vision<br>Technology Inc.<br>EZvue UV<br>(Subject Device) | I-See Vision<br>Technology Inc.<br>EZvue Colors<br>(Subject Device) | I-See Vision<br>Technology Inc.<br>EZvue<br>(Primary Predicate<br>Device) | Optical<br>Connection, Inc.<br>42 UV<br>(Reference<br>Predicate Device) | Coopervision,<br>Inc.<br>Frequency<br>Colors<br>(Reference Predicate<br>Device) |
|------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Soft Contact Lenses for<br>daily wear are indicated<br>for the correction of<br>refractive error in<br>aphakic and not aphakic<br>persons with non-<br>diseased eyes with<br>myopia or hyperopia<br>and/or possesses<br>refractive astigmatism<br>not exceeding 2.00<br>Diopters, and/or are<br>presbyopic. | Soft Contact Lenses for<br>daily wear are indicated<br>for the correction of<br>refractive error in<br>aphakic and not aphakic<br>persons with non-<br>diseased eyes with<br>myopia or hyperopia<br>and/or possesses<br>refractive astigmatism<br>not exceeding 2.00<br>Diopters, and/or are<br>presbyopic. The lens<br>may be used to enhance<br>or alter the apparent<br>color of the eye. | Soft Contact Lenses<br>for daily wear are<br>indicated for the<br>correction of<br>refractive error in<br>aphakic and not<br>aphakic persons with<br>non-diseased eyes<br>with myopia or<br>hyperopia and/or<br>possesses refractive<br>astigmatism not<br>exceeding 2.00<br>Diopters, and/or are<br>presbyopic. | Soft Contact Lenses<br>for daily wear are<br>indicated for the<br>correction of<br>refractive error in<br>aphakic and not<br>aphakic persons with<br>non-diseased eyes<br>with myopia or<br>hyperopia and/or<br>possesses refractive<br>astigmatism not<br>exceeding 2.00<br>Diopters, and/or are<br>presbyopic. | Soft Contact Lenses<br>for daily wear to<br>enhance or alter the<br>apparent color of<br>the eye, including<br>ocular masking, in<br>either sighted or<br>non-sighted eyes<br>that require<br>prosthetic contact<br>lens for the<br>management of<br>conditions such as<br>corneal, iris, or lens<br>abnormalities. |
| Functionality | Same as predicate<br>device | Same as predicate<br>device | The contact<br>lenses act as a<br>refractive<br>medium that<br>focus light rays<br>from near and<br>distant objects<br>on the retina | The contact<br>lenses act as a<br>refractive<br>medium that<br>focus light rays<br>from near and<br>distant objects on<br>the retina | The contact<br>lenses act as a<br>refractive<br>medium that<br>focus light rays<br>from near and<br>distant objects<br>on the retina |
| Indications | Daily Wear, Soft<br>(hydrophilic)<br>Contact Lens | Daily Wear, Soft<br>(hydrophilic)<br>Contact Lens | Daily Wear, Soft<br>(hydrophilic)<br>Contact Lens | Daily Wear, Soft<br>(hydrophilic)<br>Contact Lens | Daily Wear, Soft<br>(hydrophilic)<br>Contact Lens |
| Production<br>Method | Fully-molded | Fully-molded | Fully-molded | Fully-molded | Fully-molded |
| USAN name | ocufilcon D | ocufilcon D | ocufilcon D | hefilcon A | methafilcon A |
| Water Content<br>(%) | 55±2% | 55±2% | 55±2% | 42±2% | 55±2% |
| Oxygen<br>Permeability | $16 x 10^{-11}$<br>(cm²/sec)(mlO₂)/(ml x<br>mmHg @ 35°C))<br>(revised Fatt method) | $16 x 10^{-11}$<br>(cm²/sec)(mlO₂)/(ml x<br>mmHg @ 35°C))<br>(revised Fatt method) | $16 x 10^{-11}$<br>(cm²/sec)(mlO₂)/(ml<br>x mmHg @ 35°C))<br>(revised Fatt method) | | $15.04 x 10^{-11}$<br>(cm²/sec)(mlO₂)/(ml<br>x mmHg @ 35°C)) |
| UV<br>Transmittance | <5% UVB<br>(280nm - 315nm)<br><50% UVA<br>(316nm - 380nm) | | | < 10% UVB | |
| Specific Gravity | 1.05 (hydrated) | 1.05 (hydrated) | 1.05 (hydrated) | 1.039 (hydrated) | |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.