← Product Code [PIF](/productcode/PIF) · K150286

# Medline Enteral Feeding Sets (K150286)

_Medline Industries, Inc. · PIF · Apr 27, 2015 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K150286

## Device Facts

- **Applicant:** Medline Industries, Inc.
- **Product Code:** [PIF](/productcode/PIF.md)
- **Decision Date:** Apr 27, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5980
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

Medline Enteral Feeding Sets are intended to dispense liquid nutrition (feeding solutions) at a user controlled rate. These enteral feeding sets interface with a patient's feeding tube and may use gravity or an enteral feeding pump to dispense feeding solution. The enteral feeding sets may include a bag to contain the feeding solution and/or spike to connect to a pre-filled container.

## Device Story

Medline Enteral Feeding Sets consist of single-use, disposable tubing, drip chambers, piercing spikes, roller clamps, and ENFit distal connectors. Sets interface with enteral feeding pumps (e.g., Alcor Sentinel, Nestle Compat) or operate via gravity to deliver liquid nutrition from a bag or pre-filled container to a patient's feeding tube. Operated by clinicians or caregivers in clinical or home settings. The device facilitates controlled-rate delivery of nutritional solutions. The primary design feature is the integration of ISO 80369-1 compliant ENFit connectors to mitigate the risk of misconnections with non-enteral medical devices. The device provides a physical conduit for nutrition; it does not perform automated analysis or clinical decision-making.

## Clinical Evidence

No clinical data. Evidence consists of bench testing, including human factors validation for enteral connectors, FMEA, and performance testing (leakage, stress cracking, axial load, unscrewing, overriding, and disconnection). Biocompatibility testing performed per ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation and sensitization).

## Technological Characteristics

Single-use, non-sterile, disposable enteral feeding sets. Components include enteral feeding bag, tubing, drip chamber, piercing spike, roller clamp, and ISO 80369-1 compliant ENFit distal connectors. Materials are identical to the predicate. No electronic components or software.

## Regulatory Identification

A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).

## Special Controls

*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.

## Predicate Devices

- Generica Medical International Enteral Feeding Sets ([K133077](/device/K133077.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles a stylized human figure.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

April 27, 2015

Medline Industries, Inc. Stephanie Blair Regulatory Affairs Specialist One Medline Place Mundelein, IL 60060

Re: K150286

> Trade/Device Name: Medline Enteral Feeding Sets Regulation Number: 21 CFR§ 876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: II Product Code: PIF, KNT Dated: April 13, 2015 Received: April 14, 2015

Dear Stephanie Blair,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

Image /page/1/Picture/8 description: The image shows the name "Herbert P. Lerner -S" in a clear, sans-serif font. The text is horizontally aligned and appears to be part of a document or title. The letters are uniformly sized and spaced, contributing to the legibility of the name.

for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

#### Indications for Use

510(k) Number (if known)

K150286

Device Name Medline Enteral Feeding Sets

Indications for Use (Describe)

Medline Enteral Feeding Sets are intended to dispense liquid nutrition (feeding solutions) at a user controlled rate. These enteral feeding sets interface with a patient's feeding tube and may use gravity or an enteral feeding pump to dispense feeding solution. The enteral feeding sets may include a bag to contain the feeding solution and/or spike to connect to a pre-filled container.

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Medline, a healthcare company. The logo features the word "MEDLINE" in a bold, sans-serif font, with a stylized white starburst symbol above the text. The background is a solid dark blue color, and there is a gray bar at the top right corner of the image.

ledline Industries, Inc. One Medline Place Mundelein, IL 60060

510(k) Notification Medline Enteral Feeding Sets All information on this page is confidential.

# TRADTIONAL 510(k) SUMMARY [AS REQUIRED BY 21 CFR 807.92(a)]

### SUBMITTER / 510(k) SPONSOR

Medline Industries, Inc. One Medline Place Mundelein, IL 60060 Registration Number: 1417592

### CONTACT PERSON

Stephanie Blair Regulatory Affairs Specialist Phone: 847-643-3690 Email: SBlair@medline.com

#### SUMMARY PREPARATION DATE

April 22, 2015

### TYPE OF 510(k) SUBMISSION

Traditional

### DEVICE NAME / CLASSIFICATION

Trade/Proprietary Name: Medline Enteral Feeding Sets Common Name: Tubes, Gastrointestinal and Accessories Classification Name: Gastrointestinal Tubes with Enteral Specific Connectors Product Code: PIF / KNT Device Class: II Classification Regulation: 21 CFR 876.5980 Classification Panel: Gastroenterology and Urology

### PREDICATE DEVICE

Generica Medical International Enteral Feeding Sets (K133077)

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image is a logo for Medline. The logo is white text on a blue background. The text "MEDLINE" is in a sans-serif font. There is a white star-like shape behind the text. The star shape has four points, two pointing up and down, and two pointing left and right.

Medline Industries, Inc. One Medline Place Mundelein, IL 60060

## INDICATIONS FOR USE

Medline Enteral Feeding Sets are intended to dispense liquid nutrition (feeding solutions) at a user controlled rate. These enteral feeding sets interface with a patient's feeding tube and may use gravity or an enteral feeding pump to dispense feeding solution. The enteral feeding sets may include a bag to contain the feeding solution and/or spike to connect to a pre-filled container.

### DEVICE DESCRIPTION

Medline Enteral Feeding Sets consist of the following three configurations:

- 1.) Medline Enteral Feeding Pump Bag Set (1000ml)
- 2.) Medline Enteral Feeding Pump Spike Set
- 3.) Medline Enteral Feeding Gravity Bag Set (1000ml)

The Medline Enteral Feeding Pump Bag Set (1000ml) is designed for use with Alcor Sentinel® and Nestle Compat® enteral feeding pump systems and features a 1000ml formula bag with 100ml graduations and a protective closure cap.

The Medline Enteral Feeding Gravity Bag Set (1000ml) uses the free flow of gravity to dispense enteral feeding solutions and features a 1000ml formula bag with 100ml graduations and a protective closure сар.

The Medline Enteral Feeding Pump Spike Set features a piercing spike used to connect the set to a prefilled container of enteral feeding solution. It is designed for use with Alcor Sentinel® and Nestle Compat® enteral feeding pump systems.

### SUMMARY OF TECHNOLOGICAL CHARACTERISTICS

Medline Enteral Feeding Sets are single-use, non-sterile and disposable enteral feeding sets consiting of an enteral feeding bag, tubing, drip chamber, piercing spike, roller clamp and Distal Tip ENFit connectors. The intended use and function of Medline Enteral Feeding Sets are identical to that of the predicate. In addition, Medline Enteral Feeding Sets are offered in the same design style configurations (pump bag set, spike set, and gravity bag set) and are constructed of the same materials as the predicate.

Variation from the predicate device exists only in the replacement of the predicate's 4-Step Distal Tip Connector with the subject's Distal Tip ENFit connector tested in accordance with ISO 80369-1. The addition of the ENFit connector to Medline Enteral Feeding Sets is intended to improve device performance and safety by addressing the risk of misconnections. In addition, this variation in connector design is part of an industry-wide effort to adopt a uniform connector for enteral devices that meets the requirements of ISO 80369-1 and is in accordance with FDA Draft Guidance for Industry: Safety

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image is a logo for Medline. The logo consists of the word "MEDLINE" in white, with a white starburst-like symbol above and below the text. The background is a solid dark blue color. The logo is simple and clean, with a modern design.

Aedline Industries. Inc One Medline Place Mundelein, IL 60060

# Considerations for 510(k) Submissions to Mitigate the Risks of Misconnections with Small-bore Connectors Intended for Enteral Applications (July 27, 2012).

Biocompatibility, chemical and functional performance testing were conducted to evaluate Medline Enteral Feeding Sets with the ENFit connector incorporated into its design. The results of these tests have demonstrated that the replacement of the predicate's 4-step Distal Tip connector with the proposed device's ENFit connector does not alter the function or performance of the devices and does not raise additional questions or compromise the safety and/or effectiveness of the device. This testing, along with the various supporting data provided in this pre-market notification, establishes that Medline Enteral Feeding Sets are substantially equivalent to the predicate device.

# SUMMARY OF PERFORMANCE TESTING

The following performance testing was conducted on Medline Enteral Feeding Sets:

- Enteral Connector Misconnection Assessment Study
- o Human Factors Validation Study: Enteral Connectors Final Report
- PG-Lock Misconnection Data with Failure Modes and Effects Analysis (FMEA) ●
- Performance Testing with and without Reference Connectors: o
	- Falling Drop Positive Pressure Liquid Leakage o
	- Stress Cracking o
	- Resistance to Separation from Axial Load O
	- Resistance to Separation from Unscrewing O
	- Resistance to overriding O
	- o Disconnection by unscrewing
- ENFit Component Verification Testing
- ENFit Bond Strength Testing ●
- Risk Analysis ●
- Pump Compatibility Testing against predicate ●
- Flow Rate Testing against predicate ●
- ENFit Connection Strength Testing
- Bioburden Testing –against predicate ●
- o Cross Cavity Metrology Testing

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the Medline logo. The logo is set against a blue background. The word "MEDLINE" is written in white, bold, sans-serif font. Above the word is a white symbol that resembles a stylized star or cross.

Medline Industries, Inc One Medline Place Mundelein, IL 60060

- Leachable and Extractable (L&E) Profile Analysis with comprehensive Toxicological ● Health Risk Assessment
- Biocompatibility Testing: ●
	- o Cytotoxicity MEM elution per ISO 10993-5;
	- Irritation Intracutaneous reactivity per ISO 10993-10; and o
	- Delayed-Type Hypersensitivity (Sensitization) Guinea Pig Maximization Test o per ISO 10993-10.

#### SUMMARY OF CLINICAL TESTING

Not applicable.

#### SUMMARY OF ANIMAL TESTING

Not applicable.

### CONCLUSION

In accordance with 21 CFR Part 807, and based on a comparison of the Indications for Use, technological characteristics, and performance data to the predicate, Medline Industries, Inc. concludes that the Medline Enteral Feeding Sets are safe, effective and substantially equivalent to the predicate.

---

**Source:** [https://fda.innolitics.com/device/K150286](https://fda.innolitics.com/device/K150286)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
