K150143 · Nurse Assist, Inc. · NGT · Aug 31, 2015 · General Hospital
Device Facts
Record ID
K150143
Device Name
Normal Saline Flush
Applicant
Nurse Assist, Inc.
Product Code
NGT · General Hospital
Decision Date
Aug 31, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5200
Device Class
Class 2
Indications for Use
This device is intended for use in flushing compatible intravenous administration sets and indwelling intravenous access devices.
Device Story
The Normal Saline Flush is a pre-filled, single-use, sterile polypropylene syringe containing 0.9% Sodium Chloride for Injection, USP. It is designed for manual operation by healthcare professionals to flush intravenous administration sets and indwelling access devices. The device is available in 12 mL syringe sizes with fill volumes of 3 mL, 5 mL, or 10 mL. It features a luer lock tip for secure connection to IV access points. By maintaining patency of IV lines, the device helps ensure the delivery of medications or fluids to the patient. It is terminally sterilized via gamma radiation and is not made with natural rubber latex.
Clinical Evidence
Bench testing only. No clinical data provided. Testing included package integrity (ISTA 2A), shelf-life stability (pH, sodium chloride concentration, heavy metals, particulate matter), and biocompatibility (ISO 10993-5, -4, -11, -10; USP <151>). Results confirmed the device is non-hemolytic, non-toxic, non-irritant, non-sensitizer, and non-pyrogenic. Endotoxin levels were <0.03 EU/ml (limit <0.25 EU/ml). Sterility was confirmed with SAL of 10^-6.
Technological Characteristics
Materials: Polypropylene (barrel, plunger, tip cap). Dimensions: 12 mL syringe with 3 mL, 5 mL, or 10 mL fill volumes. Principle: Manual fluid displacement via luer lock tip. Sterilization: Gamma radiation (SAL 10^-6). Connectivity: None. Software: None.
Indications for Use
Indicated for flushing compatible intravenous administration sets and indwelling intravenous access devices in patients requiring intravenous therapy.
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol. The logo is simple and recognizable, representing the department's role in promoting health and well-being.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 31, 2015
Nurse Assist, Inc. Mr. Brian Cox Vice President of Operations 4409 Haltom Road Haltom City, TX 76117
Re: K150143
> Trade/Device Name: Normal Saline Flush Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: II Product Code: NGT Dated: July 29, 2015 Received: July 30, 2015
Dear Mr. Cox,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Page 2 - Mr. Cox
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
*Tejashri Purohit-Sheth, M.D.*
Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K150143
Device Name Normal Saline Flush
Indications for Use (Describe)
This device is intended for use in flushing compatible intravenous administration sets and indwelling intravenous access devices.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
#### 1. Submitter's name and address:
Nurse Assist, Inc. 4409 Haltom Road Haltom City, TX 76117
### 2. Submitter's telephone number and fax number:
- Tel: (817) 231-1300 Fax: (817) 231-1500
### 3. Contact person:
Mr. Brian Cox - Vice President Operations
# 4. Date this 510(k) summary prepared:
05/14/15
### 5. Regulatory Description:
| Trade name: | Normal Saline Flush |
|-------------------------|-------------------------------|
| Common Name: | Saline IV Flush |
| Classification Name: | Saline, Vascular Access Flush |
| Regulation Description: | Intravascular Catheter |
| Regulation Number: | 21 CFR 880.5200 |
| Class: | II |
| Product Code: | NGT |
### 6. Legally marketed device to which substantial equivalence is claimed:
Primary predicate - AMUSA 0.9% Sodium Chloride Flush Syringe, K111034 Reference device - Kendall Monoject Prefill Flush Syringe, K032438
### 7. Description of the device:
The subject device is a polypropylene plastic syringe filled with 0.9% Sodium Chloride for Injection, USP, and capped with a polypropylene plastic cap. The device will be terminally sterilized by gamma radiation sterilization. The device will be marketed as a 12 mL syringe with a 3 mL, 5 mL, or 10 mL fill volume.
### 8. Intended use and indication for use:
For prescription use: This device is intended for use in flushing compatible intravenous administration sets and indwelling intravenous access devices.
{4}------------------------------------------------
### 9. Summary of technological characteristics compared to the predicate device:
| | Proposed<br>device | Primary Predicate<br>K111034 | Reference Device<br>K032438 |
|-------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| | Nurse Assist<br>Normal Saline<br>Flush | AMUSA 0.9% Sodium<br>Chloride Flush<br>Syringe | Kendall Monoject<br>Prefill Flush<br>Syringe |
| Classification Product Code | NGT | NGT | NGT |
| Intended use: | This device is<br>intended for use in<br>flushing compatible<br>intravenous<br>administration sets<br>and indwelling<br>intravenous access<br>devices. | 0.9% Sodium Chloride<br>Flush Syringes are<br>intended for use in<br>flushing compatible<br>intravenous<br>administration sets and<br>indwelling intravenous<br>access devices. Use<br>according to the<br>recommendations of the<br>manufacture for the<br>appropriate device. | The syringes are<br>intended for flushing<br>compatible<br>intravenous tubing<br>systems and<br>indwelling<br>intravascular access<br>devices. |
| Use on sterile field? | No | No | Unknown |
| Prescription only? | Yes | Yes | Yes |
| Sterile? | Yes | Yes | Yes |
| Single use only? | Yes | Yes | Yes |
| Method of Sterilization | Radiation | Radiation | Steam Autoclave |
| Shelf life | 2 years | 2 years | 2 years |
| Chemical composition | 0.9% Sodium<br>Chloride Injection,<br>USP | 0.9% Sodium Chloride<br>Injection, USP | 0.9% Sodium<br>Chloride Injection,<br>USP |
| Mechanism of dispensing | 12 mL plastic<br>syringe, luer lock tip | 12 mL plastic syringe,<br>luer lock tip | 12 mL plastic<br>syringe, luer lock tip |
| Fill Volumes | 3 mL, 5 mL, 10 mL | 3 mL, 5 mL, 10 mL | 3 mL, 5 mL, 10 mL |
| Barrel, Plunger, Tip Cap, Bag<br>Material | Polypropylene | Polypropylene | Polypropylene |
| Uses cleared syringe | Yes - K945715 | unknown | Yes - K945715 |
| Plunger Grommet Material | This product is not<br>made with natural<br>rubber latex | This product is not<br>made with natural<br>rubber latex | This product is not<br>made with natural<br>rubber latex |
## 10. Non-Clinical Performance Data
Gamma radiation sterilization has been validated for this device. This method provides an SAL of 106. The following testing was performed post-sterilization:
{5}------------------------------------------------
| NAME | TEST METHOD(S) | RESULTS OVERVIEW |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Package Integrity | Subject device to ISTA 2A 2005 conditioning and<br>testing criteria<br>Perform post-conditioning testing:<br>Visual Inspection of packaging: confirm package integrity Seal (vacuum) testing: 15 in-Hg for 15 minutes and verify no leaks Sterility: Tryptic Soy Broth & Fluid Thioglycollate Medium | Visual Inspection of packaging: no major compromise of packaging Seal testing: no leaks Sterility: confirmation of sterile barrier and product sterility—no growth present |
| Shelf Life<br>(Stability) | Post Sterility Testing at T=0, T=1 and T=2 (Note:<br>Accelerated aging for T=1 and T=2):<br>Visual: confirm no leaks, holes or cracks Seal (vacuum) testing: 15 in-Hg for 15 minutes and verify no leaks Appearance/Color: verify clear liquid Odor: confirm objectionable or unusual odors are not present pH: 4.5 – 7.0 Sodium Chloride: 0.885 – 0.945% Heavy Metals: USP <231> and <241> Sterility: Tryptic Soy Broth Particulate: USP <778> | Visual: no leaks, cracks or holes Seal: no leaks Appearance/Color: no discoloration or cloudiness Odor: no objectionable odors pH: within upper and lower limits Sodium Chloride: within upper and lower limits Heavy metals: USP <231>: less than 10 ppm (0.001%) USP <241>: less than 2 ppm Sterility: confirmation of sterile barrier and product sterility—no growth present Particulate: less than 3,000 for 10 μm and 300 for 25 μm |
| Biocompatibility | The classification of the device is: Blood Path, Indirect<br>(limited to ≤ 24 hours). The following tests were performed:<br>Cytotoxicity: ISO 10993-5:2009 Hemocompatibility: ASTM 7756:2008 and ISO 10993-4:2006 Acute Systemic Toxicity: ISO 10993-11:2006 Intracutaneous Irritation: ISO 10993-10:2010 (modified for a chemical solution) Contact Sensitization: ISO 10993-10:2010 Pyrogenicity: USP <151> and ISO 10993-11:2006 | Cytotoxicity: Grade 0 (no reactivity) Hemocompatibility: Hemolytic index of 0.0% (non-hemolytic) Acute Systemic Toxicity: Non-toxic Intracutaneous Irritation: Non-irritant Contact Sensitization: Non-sensitizer Pyrogenicity: Nonpyrogenic |
| Endotoxin | USP <85>: Limulus Amebocyte Lysate (LAL) endotoxin<br>testing using Gel Clot Method (Monograph #85). (Note: test results must be less than 0.25 EU/ml.) | All samples tested at an endpoint of 0.03 EU/ml Sensitivity of Lysate (AntiLog10 of Mean): 0.03 EU/ml |
The test results demonstrate that the device is non-hemolytic, non-toxic, a non-irritant, non-sensitizer, and non-pyrogenic. In addition to the test results noted and highlighted above, the results also demonstrate that packaging integrity and sterility were maintained, and as such, the device met specifications throughout the noted shelf life.
## 11. Conclusion
The above summarized characteristics and comparisons demonstrate that the Nurse Assist Normal Saline Flush device is as safe and effective as the predicate and reference devices. In summary, the Nurse Assist Normal Saline Flush described in this submission is substantially equivalent to the predicate and reference devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.