← Product Code [HRS](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS) · K150099

# Depuy Synthes Variable Angle Locking Hand System (1.3 mm and 2.0 mm Plates and Screws) (K150099)

_Synthes (USA) Products, LLC · HRS · Mar 23, 2015 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K150099

## Device Facts

- **Applicant:** Synthes (USA) Products, LLC
- **Product Code:** [HRS](/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS.md)
- **Decision Date:** Mar 23, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The DePuy Synthes Variable Angle Locking Hand System is intended for fracture fixation of the hand and other small bones and small bone fragments, in adults and adolescents (12-21) particularly in osteopenic bone. System indications include the following: o Open reduction and internal fixation of fractures, mal-unions, and non-unions o Following excision of benign bone tumors o Replantations and reconstructions o Arthrodeses of joints involving small bones o Osteotomies, including deformity correction such as rotation, lengthening, shortening o Pathological fractures, including impending pathologic fractures

## Device Story

System consists of stainless steel and titanium plates and screws for internal bone fixation; used in hand and small bone surgery. Plates feature locking holes, limited contact profiles, and variable angle locking technology. Surgeon implants hardware during open reduction procedures to stabilize fractures, mal-unions, or osteotomies. Variable angle locking allows screw placement at various trajectories to accommodate patient anatomy. Provides mechanical stability to bone fragments, facilitating healing and deformity correction. Used in clinical settings by orthopedic or hand surgeons.

## Clinical Evidence

Bench testing only. Fatigue strength testing performed on plate-screw constructs targeting weakest plate sections. Screw performance evaluated per ASTM F543-13. No clinical data provided; literature cited to support adolescent indication.

## Technological Characteristics

Materials: stainless steel and titanium. Features: locking holes, limited contact profiles, variable/standard locking technology, anatomic contours. Dimensions: 1.3mm and 2.0mm plate/screw systems. Testing standards: ASTM F543-13 for screws. Mechanical fixation device; no software or energy source.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Synthes Stainless Steel Modular Hand System ([K030310](/device/K030310.md))
- Synthes (USA) Modular Mini Fragment LCP System ([K063049](/device/K063049.md))
- Synthes 2.4 mm/2.7 mm Variable Angle LCP Forefoot/Midfoot System ([K100776](/device/K100776.md))
- Synthes Cortical Screws ([K112583](/device/K112583.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. The profiles are rendered in a dark color, contrasting with the white background of the seal.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 23, 2015

Synthes (USA) Products, LLC Mr. Nicholas Fountoulakis Regulatory Affairs Specialist 1301 Goshen Parkway West Chester, Pennsylvania 19380

Re: K150099

Trade/Device Name: DePuy Synthes Variable Angle Locking Hand System (1.3mm and 2.0mm Plates and Screws) Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: January 16, 2015 Received: January 20, 2015

Dear Mr. Fountoulakis:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing

{1}------------------------------------------------

### Page 2 - Mr. Nicholas Fountoulakis

(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Lori A. Wiggins -S

for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

# 510(k) Number (if known)

K150099

Device Name

DePuy Synthes Variable Angle Locking Hand System (1.3mm and 2.0mm Plates and Screws)

Indications for Use (Describe)

The DePuy Synthes Variable Angle Locking Hand System is intended for fracture fixation of the hand and other small bones and small bone fragments, in adults and adolescents (12-21) particularly in osteopenic bone.

System indications include the following:

o Open reduction and internal fixation of fractures, mal-unions, and non-unions

o Following excision of benign bone tumors

- o Replantations and reconstructions
- o Arthrodeses of joints involving small bones
- o Osteotomies, including deformity correction such as rotation, lengthening, shortening

o Pathological fractures, including impending pathologic fractures

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary

## Date Prepared: March 13, 2015

| Sponsor:                           | DePuy Synthes<br>Nicholas S. Fountoulakis<br>1301 Goshen Parkway<br>West Chester, PA 19380<br>Office: (610) 719-6553<br>Fax: (484) 356-9682                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Proprietary<br>Name:               | DePuy Synthes Variable Angle Locking Hand System (1.3 mm and 2.0 mm Plates<br>and Screws)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Classification:                    | Class II, HRS, §888.3030 – Plate, Fixation, Bone<br>Class II, HWC, §888.3040 - Screw, Fixation, Bone                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Primary<br>Predicate<br>Devices    | K030310 – Synthes Stainless Steel Modular Hand System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Additional<br>Predicate<br>Devices | K063049 – Synthes (USA) Modular Mini Fragment LCP System<br>K100776 - Synthes 2.4 mm/2.7 mm Variable Angle LCP Forefoot/Midfoot System<br>K112583 - Synthes Cortical Screws                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Device<br>Description:             | The DePuy Synthes Variable Angle Locking Hand System (1.3 mm and 2.0 mm<br>Plates and Screws) consists of stainless steel and titanium plates and screws that<br>offer screw-to-plate locking designed for various fracture modes of the hand. The<br>plates and screws contained in the DePuy Synthes Variable Angle Locking Hand<br>System (1.3 mm and 2.0 mm Plates and Screws) are offered in a range of<br>configurations to accommodate patient anatomy and surgical need. The subject<br>system contains two plate and screw sizes, 1.3 mm and 2.0 mm, general instruments,<br>and device specific instruments. The 1.3 mm plates in this submission are designed<br>to accept new 1.3 mm cortex and locking screws. The 2.0 mm plates are designed to<br>accept existing 2.0 mm cortex screws, 2.0 mm locking screws, and new 2.0 mm<br>Variable Angle (VA) locking screws. The new 2.0 mm VA locking plates and screws<br>feature existing variable angle locking technology (K100776).                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Indications for<br>Use:            | The DePuy Synthes Variable Angle Locking Hand System is intended for fracture<br>fixation of the hand and other small bones and small bone fragments, in adults and<br>adolescents (12-21) particularly in osteopenic bone.<br><br>System indications include the following:<br>Open reduction and internal fixation of fractures, mal-unions, and non-unions Following excision of benign bone tumors Replantations and reconstructions Arthrodeses of joints involving small bones Osteotomies, including deformity correction such as rotation, lengthening, shortening Pathological fractures, including impending pathologic fractures                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Substantial<br>Equivalence:        | The proposed DePuy Synthes Variable Angle (VA) Locking Hand System (1.3 mm<br>and 2.0 mm Plates and Screws) shares the same fundamental technological<br>characteristics and intended use as the predicate Synthes systems (K030310,<br>K112583, K063049, K100776). Fatigue strength testing was completed for the<br>plates included in the subject system, demonstrating equal or greater strength in<br>comparison to representative predicate plates. Plates were setup for fatigue strength<br>testing in a plate-screw construct configuration that was intended to target the<br>weakest portion of the plate by aligning the simulated fracture gap with the plate's<br>smallest cross sectional area. New 1.3 mm Cortex and Locking Screws and 2.0 mm<br>VA Locking Screws were tested using methods detailed in ASTM F543-13 in order<br>to demonstrate performance in accordance with the specifications of the standard or<br>in cases where a benchmark specification was not available, comparison was made to<br>representative predicate devices. Literature has been provided to support the revision<br>in indications, most notably the addition of the adolescent population Lastly,<br>technological characteristics of the DePuy Synthes Variable Angle System such as<br>locking holes, limited contact profiles, variable and standard locking technology,<br>anatomic contours, and size ranges are prominent throughout the representative<br>predicate systems. Based on the similarity of technical features across the predicate<br>device, results of the performance data, and literature discussions, the subject DePuy<br>Synthes Variable Angle Locking Hand System (1.3 mm and 2.0 mm Plates and<br>Screws) is substantially equivalent to the predicate devices. |

{4}------------------------------------------------

---

**Source:** [https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K150099](https://fda.innolitics.com/submissions/OR/subpart-d%E2%80%94prosthetic-devices/HRS/K150099)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
