A clinical case series was used to demonstrate the clinical performance (defect healing and new bone formation) of the Regenomer device in human subjects.
Clinical case series; Bone grafting; Periodontal defect healing
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Clinical case series; Clinical case series
Human oral, periodontal
Not applicable for this study
Defect healing and formation of new bone
Indications for Use
The Regenomer® is in general recommended for the filling of extraction sockets and periodontal defect. The indication for use depending on the device format is further classified as described in following table. Format Syringe: - Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) - Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR) Format Plug: - Filling of extraction sockets - Augmentation or reconstructive treatment of the alveolar ridge - Elevation of maxillary sinus floor Format Block: - Augmentation or reconstructive treatment of the alveolar ridge - Elevation of maxillary sinus floor
Device Story
Regenomer® is a sponge-like, absorbable, porous bone filling augmentation material derived from purified porcine type I collagen. Available in syringe (sheet), plug (bullet), and block formats. Used by dental professionals in oral and periodontal surgery to fill defects, augment alveolar ridges, or elevate the maxillary sinus floor. Acts as an osteoconductive scaffold to support new bone formation; resorbs over time. Supplied sterile via gamma irradiation for single-use. Clinical benefit includes defect healing and new bone formation without inflammation.
Clinical Evidence
Evidence includes bench testing (pH, loss on drying, amino acid content, FT-IR, SDS-PAGE), biocompatibility testing per ISO 10993 (acute systemic injection, AMES, cytotoxicity, implantation, intracutaneous reactivity, micronucleus, sensitization, oral mucosa irritation), sterility, and pyrogen testing. Preclinical performance evaluated in a canine alveolar bone defect model, demonstrating comparable bone growth and resorption rates to the predicate. Clinical case series reported successful defect healing and new bone formation without inflammation.
Technological Characteristics
Material: Purified porcine type I collagen. Form: Sponge-like, porous, absorbable. Formats: Syringe (sheet), Plug (bullet), Block. Sterilization: Gamma irradiation (SAL 10^-6). Shelf-life: 36 months. Non-pyrogenic. Biocompatible per ISO 10993.
Indications for Use
Indicated for human oral and periodontal bone grafting procedures, including filling of extraction sockets, periodontal defects (with GTR), peri-implant defects (with GBR), alveolar ridge augmentation/reconstruction, and maxillary sinus floor elevation.
Regulatory Classification
Identification
Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.
Special Controls
*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
Predicate Devices
FOUNDATION Bone Filling Augmentation Material (K040783)
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 10, 2015
Nibec Co., Ltd. c/o Takashi Yamada CEO and Regulatory Affairs Smile US 22395 South Western Avenue, Suite 304 Torrance, California 90501
Re: K150079
Trade/Device Name: Regenomer® Syringe, Regenomer® Plug, Regenomer® Block Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: II Product Code: NPM Dated: November 3, 2015 Received: November 9, 2015
Dear Takashi Yamada:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{2}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology. General Hospital, Respiratory, Infection Control, and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
## 1.0 Indications for Use Statement
## INDICATION FOR USE
| 510(k) Number: | K150079 |
|----------------|--------------------|
| Device Name: | Regenomer® Syringe |
| | Regenomer® Plug |
| | Regenomer® Block |
# INDICATIONS FOR USE:
The Regenomer® is in general recommended for the filling of extraction sockets and periodontal defect. The indication for use depending on the device format is further classified as described in following table.
| Format | Syringe | Plug | Block |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|
| Indication<br>for use | - Filling of periodontal<br>defects in conjunction with<br>products intended for<br>Guided Tissue<br>Regeneration (GTR)<br><br>- Filling of peri-implant<br>defects in conjunction with<br>products intended for<br>Guided Bone Regeneration<br>(GBR) | - Filling of extraction sockets<br><br>- Augmentation or<br>reconstructive treatment of the<br>alveolar ridge<br><br>-Elevation of maxillary sinus<br>floor | - Augmentation or<br>reconstructive treatment of<br>the alveolar ridge<br><br>-Elevation of maxillary sinus<br>floor |
Prescription Use
AND/OR
Over-The-Counter Use _________________________________________________________________________________________________________________________________________________________
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
Concurrence of CDRH, Office of Device Evaluation (ODE)
{4}------------------------------------------------
# 2.0 510(k) Summary
### Sponsor/Applicant
NIBEC Co., Ltd. lwol electricity-electronic Agro-industrial Complex, 116, Bamdi-gil, Iwol-myeon, Jincheon-gun, Chungcheongbuk-do, 27816, Korea Phone: 82-10-2889-8590 Fax: 82-2-744-8732 Contact: Dr. Park, Yoon-Jeong
Date Prepared : November 30, 2015
Prior Submission : There were no prior submissions.
#### Device Name and Identification
| Proprietary Name: | Regenomer® Syringe |
|----------------------|------------------------------------|
| | Regenomer® Plug |
| | Regenomer® Block |
| Common/Usual Name: | Bone Filling Augmentation Material |
| Classification Name: | Bone grafting material |
#### Predicate devices
#### Primary predicate
FOUNDATION Bone Filling Augmentation Material (K040783) Manufactured by: Terumo Corporation 44-1, 2 chome Hatagaya, Shibuya-ku Tokyo 151-0072 Japan
#### Reference predicates
Bio-Oss® Collagen (K092428, K033815, K974399) Manufactured by: Geistlich Pharma AG Bahnhofstrasse 40 CH-6110 Wolhusen Switzerland
### Device Category/Class
{5}------------------------------------------------
| Device Class: | Class II |
|--------------------|--------------------|
| Regulation Number: | 21 C.F.R. 872.3930 |
| Product Code: | NPM |
### Indication for use
The Regenomer® is in general recommended for the filling of extraction sockets and periodontal defect. The indication for use depending on the device format is further classified in following table.
| Format | Syringe | Plug | Block |
|-----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|
| Indication<br>for use | - Filling of periodontal<br>defects in conjunction with<br>products intended for<br>Guided Tissue<br>Regeneration (GTR)<br><br>- Filling of peri-implant<br>defects in conjunction with<br>products intended for<br>Guided Bone Regeneration<br>(GBR) | - Filling of extraction sockets<br><br>- Augmentation or<br>reconstructive treatment of the<br>alveolar ridge<br><br>-Elevation of maxillary sinus<br>floor | - Augmentation or<br>reconstructive treatment of<br>the alveolar ridge<br><br>-Elevation of maxillary sinus<br>floor |
# Device Description
Regenomer® is a sponge-like absorbable and porous collagen designed to be used as bone filling augmentation material. The device is manufactured from purified type I collagen derived from porcine skin sources in South Korea. The porcine skin is certified by veterinarian and is obtained by a standardized controlled manufacturing process. Regenomer® is three types, Regenomer® Syringe (sheet shape in syringe and blister), Regenomer® Plug (bullet shape in blister), and Regenome® Block (block shape in blister). Regenomer® are supplied sterile, non-pyrogenic, and for single use only.
# Basis for Substantial Equivalence
Regenomer® is substantially equivalent for purposes of the FDA's medical device regulations to FOUNDATION®, which is cleared for the filling of extraction sockets (K040783) and Bio-Oss® Collagen (K974399), which was cleared for the filling of extraction sockets and other bone augmentation procedures. The differences in indications for Regenome® compared to the predicates do not change the intended use because all are intended for bone augmentation procedures. While Regenomer® is indicated for more specific indications based on device shape and size, this is substantially equivalent to the predicates and does not change the intended use because all are intended for bone augmentation procedures. Bio-Oss® Collagen has block type product, the size range of 6X6X6, 7X8X9, 9X10X11 mm. The size dimension of Regenomer Block type is 6X5X7 and 8X7X9mm, which is 10% of size difference with the dimension of Bio-Oss® Collagen. Intended use, principles of operation, and technological characteristics are substantially equivalent to the corresponding characteristics of the predicate devices. Although minor technological differences exist, Regenomer® and the predicate device, these minor differences raise no new issues of equivalence of Regenomer® with predicate devices.
{6}------------------------------------------------
The following is a table comparing Regenomer®, FOUNDATION®, and Bio-Oss® Collagen cleared for the filling of extraction sockets.
| | Regenomer® | FOUNDATION®<br>(K040783) | Bio-Oss® Collagen<br>(K092428, K033815, K974399) |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ITEM | | | |
| | Syringe type<br>- Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR)<br>- Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR) | The FOUNDATION device is a collagen-based bone filling augmentation material for use in the filling of extraction sockets. | - Augmentation or reconstructive treatment of the alveolar ridge.<br>- Filling of intrabony periodontal defects.<br>- Filling of defects after root resection, apicoectomy, and cystectomy<br>- Filling of extraction sockets<br>- Elevation of maxillary sinus floor<br>- Filling of periodontal defects in conjunction with products intended for Guided Tissue Regeneration (GTR) and Guided Bone Regeneration (GBR)<br>- Filling of peri-implant defects in conjunction with products intended for Guided Bone Regeneration (GBR) |
| Intended Use | Plug type<br>- Filling of extraction sockets<br>- Augmentation or reconstructive treatment of the alveolar ridge<br>-Elevation of maxillary sinus floor | | |
| | Block type<br>- Augmentation or reconstructive treatment of the alveolar ridge<br>- Elevation of maxillary sinus floor | | |
| Target<br>population | Human oral, periodontal | Human oral, periodontal | Human oral, periodontal |
| | | | |
| | | | |
| Dosage<br>form | Syringe (sheet shape in syringe and blister, 25, 50mg) | Bullet shape (8X25, 15X25mm) | a block form in a blister |
| | Plug (Bullet shape in blister, 10X20, 12X25mm) | sheet type (25X25, 50X25mm) | |
| | Block (Block shape in blister, 6X5X7, 8X7X9mm) | Heat-sealed aluminum package | |
| | | | |
# Table 1: Substantial Equivalence Comparison
{7}------------------------------------------------
| | Form<br>at | | Sizes (mm) | Weight<br>(mg/unit) | | Format | | Sizes (mm) | | 0.25mm to 1.0mm or 1.0mm<br>to 2.0mm granules | | | | |
|----------|------------|-----------------------------------------------|---------------|---------------------|--|-------------------------------------|-------------------------|------------|---------------------------------|-----------------------------------------------|-----------------------------|---------------|---------------------------------|--|
| | | Small | 20X10X1<br>.8 | 25 | | sheet | Small | 25X25 | | | | | | |
| | Syringe | Medi<br>um | 25X15X1<br>.8 | 50 | | | Medium | 50X25 | | | | | | |
| Dime | | small | 10 x 20 | 40 | | Plug | small | 8 x 25 | Shape | Dimensio<br>n (mm) | | Weight<br>(g) | | |
| nsion | Plug | medi<br>um | 12 X 25 | 100 | | | medium | 15 X 25 | | Block | 6X6X6 | | 100mg | |
| | | small | 6 X 5 X 7 | 10 | | | | | | Block | 7X8X9 | | 250mg | |
| | Block | medi<br>um | 8 X 7 X 9 | 20 | | | | | | Block | 9X10X11 | | 500mg | |
| | | | | | | | | | | | | | | |
| Materi | | Type I collagen | | | | Type I collagen 85-95% | | | | Anorganic derived<br>osteoconductive | | | | |
| al | | | | | | | Type III collagen 5-15% | | | hydroxyapatite, Collagen | | | | |
| Sourc | | NA | | | | NA | | | | Bovine | | | | |
| e | | | | | | | | | | | | | | |
| bone | | | | | | | | | | | | | | |
| Sourc | Porcine | | | | | Bovine | | | | | Porcine | | | |
| e of | | | | | | | | | | | | | | |
| collag | | | | | | | | | | | | | | |
| en | | | | | | | | | | | | | | |
| Physi | Sponge | | | | | Sponge | | | | | Trabecular, interconnecting | | | |
| cal | | | | | | | | | | | macro and micro pores | | | |
| Morp | | | | | | | | | | | | | | |
| holog | | | | | | | | | | | | | | |
| y | | | | | | | | | | | | | | |
| | | Biocompatible,<br>demonstrated by : | | as | | Biocompatible<br>demonstrated<br>in | | | (as<br>published<br>literature) | Biocompatible<br>demonstrated<br>in | | | (as<br>published<br>literature) | |
| | | - Acute systemic injection test | | | | | | | | | | | | |
| | | - AMES test | | | | | | | | | | | | |
| Bioco | | - Cytotoxicity test | | | | | | | | | | | | |
| mpatible | | - Implantation test | | | | | | | | | | | | |
| | | - Intracutaneous reactivity test | | | | | | | | | | | | |
| | | - Micronucleus Test for Genetic<br>Toxicology | | | | | | | | | | | | |
| | | - Maximization and | | | | | | | | | | | | |
{8}------------------------------------------------
| | sensitization test<br>- Oral Mucosa Irritation test<br>-Sterility test<br>-Pyrogen test<br>- Preclinical safety and efficacy<br>test<br>- Clinical case series | | |
|-----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|----------------------------------|
| Perfor<br>manc<br>e | Bone filling | Bone filling | Bone formation |
| Comp<br>atibilit<br>y<br>w/oth<br>er<br>devic<br>es | | | Can be used with GTR<br>membrane |
| Sterili<br>zation<br>Proce<br>ss | Sterile by Gamma Irradiation | Sterile by Gamma Irradiation | Sterile by Gamma Irradiation |
| Chem<br>ical<br>safety | Biocompatible | Biocompatible | Biocompatible |
| Anato<br>mical<br>sites | Oral, Periodontal | Oral, Periodontal | Oral, Periodontal |
| Non-<br>Pyrog<br>enic | Yes | Yes | Yes |
| Shelf-<br>Life | 36 Months | 36 Months | 36 Months |
{9}------------------------------------------------
# Brief Summary of Data Submitted
The Sponsor evaluated the performance characteristics of Regenomer® and FOUNDATION® with a thorough chemical and physical characterization including pH, loss on drying, amino acid contents, FT-IR, SDS-PAGE, sterility, and bacterial endotoxin. The physical and chemical characteristics of Regenomer were found to be comparable with FOUNDATION®. Both products were found to grow new bone and be subsequently resorbed at similar rates in canine alveolar bone defects model. Finally, in a clinical case series, use of Regenomer® resulted in defect healing and formation of new bone without inflammation.
Regenomer® was the subject of the full range of biocompatibility tests recommended in the FDA's "Class II Special Controls Guidance Document: Dental Bone Grafting Devices" and in accordance with ISO 10993. Test results confirmed product biocompatibility. Regenomer® is made from pure type I collagen obtained by a standardized controlled manufacturing process. The type I collagen has been purified from veterinary certified porcine skin. Further, the product is sterilized by y-irradiation to achieve a sterility assurance level SAL 1 x 10-6.
Based on the information presented herein, it has been demonstrated that Regenomer® is substantially equivalent to FOUNDATION® for the proposed indications for use.
## Conclusion
The Regenomer® presents the same types of potential risks to consumers as the predicate device FOUNDATION®, and has controlled these risks in a similar manner. Comparison with the predicate device shows that the device has similar specification, physico-chemical properties, and performance. In addition, biocompatibility tests show that Regenome® meets the requirements of those standards. Therefore, it is concluded that Regenomer® is substantially equivalent to the predicate device FOUNDATION®.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.