← Product Code [LLZ](/productcode/LLZ) · K150069

# Fine Structure Analysis(fineSA) (K150069)

_Acuitas Medical , Ltd. · LLZ · May 22, 2015 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K150069

## Device Facts

- **Applicant:** Acuitas Medical , Ltd.
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** May 22, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

fineSA software is intended for use as a non-invasive device that processes designated MR images using the 2-D pulse sequence incorporated on Siemens Medical 3T MAGNETOM systems running syngo MR software versions B17 and D13. This includes the Aera, Skyra, Avanto and Verio models. fineSA software generates reports that provide information about the structural features in bone. The MR data, in the form of spectra or spectral images, reflect the spacing of structural features at a scale of approximately 500 microns or larger.

## Device Story

fineSA is a software-only image processing system for characterizing trabecular bone microarchitecture. It processes 1D MR data acquired via standard 2D pulse sequences on Siemens 3T MAGNETOM scanners. Input data is transferred via DICOM to a networked PC running the software. The device uses Fourier and inverse Fourier transformations to analyze spectral data, identifying repetitive structural features in bone at a scale of 500 microns or larger. It generates a PDF report, which is sent back to a DICOM destination for clinician review. Used in radiology environments or remote locations with secure network access. The output assists clinicians in evaluating bone structural features; it does not provide chemical information. The device benefits patients by providing non-invasive, quantitative insights into bone microarchitecture.

## Clinical Evidence

Bench testing only. Software verification and validation were performed according to FDA guidance for a 'moderate' level of concern. Testing utilized phantom and clinical data to confirm the software meets design specifications and intended use requirements.

## Technological Characteristics

Software-only image processing system; operates on networked PCs. Inputs: DICOM MR data (spin echo/gradient echo). Processing: Fourier and inverse Fourier transformations, windowing. Connectivity: DICOM standard. Deployment: Radiology internal network or remote secure access. Software level of concern: Moderate.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- syngo.MR General ([K130749](/device/K130749.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 22, 2015

Acuitas Medical Ltd. % John J. Smith, M.D., J.D. Partner Hogan Lovells US LLP 555 Thirteenth Street NW WASHINGTON DC 20004

Re: K150069

Trade/Device Name: Fine Structure Analysis (fineSA) Software Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ, LNH Dated: April 22, 2015 Received: April 22, 2015

Dear Dr. Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael D.'Hara

For

Robert Ochs, Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

510(k) Number (if known)

#### K150069

Device Name

## Fine Structure Analysis (fineSA) Software

Indications for Use (Describe)

fineSA software is intended for use as a non-invasive device that processes designated MR images using the 2-D pulse sequence incorporated on Siemens Medical 3T MAGNETOM systems running syngo MR software versions B17 and D13. This includes the Aera, Skyra, Avanto and Verio models.

fineSA software generates reports that provide information about the structural features in bone. The MR data, in the form of spectra or spectral images, reflect the spacing of structural features at a scale of approximately 500 microns or larger.

Type of Use (Select one or both, as applicable)Z Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Acuitas Medical. The word "ACUITAS" is written in purple, with a small green square on top of the "I". The word "medical" is written in green and is smaller than the word "ACUITAS". The logo is simple and modern.

Fine Structure Analysis (fineSA)

#### 510(K) SUMMARY (K150069) AS REQUIRED BY 21 CFR 807.92(c)

#### 1. Submitter

Acuitas Medical Ltd. 12 Technium One, Kings Road Swansea Waterfront SA1 8PH Wales UK

Phone: (970) 472-0613 Fax: (970) 416-8896

Contact Person: John Heinrich, Ph.D. Chief Executive Officer Phone: (970) 472-0613 Fax: (970) 416-8896 Email: jheinrich@acuitasmedical.com

# 2. Date Prepared

May 20, 2015

## 3. Device

- . Trade Name: Fine Structure Analysis (fineSA)
- Common Name: System, Image Processing, Radiological ●
- Classification Name: Picture Archiving and Communication System (21 CFR 892.2050) ●
- Regulatory Class: II ●
- . Product Code: LLZ, LNH

## 4. Predicate Device

The predicate device for the Fine Structure Analysis (fineSA) software is the syngo.MR General, which was cleared under 510(k) K130749.

{4}------------------------------------------------

# 5. Device Description

fineSA is a tool for the characterization of trabecular bone microarchitecture using magnetic resonance data. fineSA analyzes data acquired from a region of interest with a 2-D pulse sequence. The result is a spectrum that shows not chemical information, but the average spacing of the structural elements within the region of interest. fineSA provides the ability to identify repetitive structural features as small as 500 um.

Compatible MRI scanners, acquire one-dimensional data along the length of a defined region of interest in the anatomical region using a standard pulse sequence available on the OEM scanner. The additional acquisition time required to collect the fineSA compatible data is less than 2 minutes. The resulting DICOM data object is transferred to a network personal computer loaded with fineSA software for processing and report generation.

The DICOM input data is received from the MRI scanner via standard DICOM protocols. The processing performed is driven by the received data content, and at the completion of data processing and analysis, a PDF formatted report is generated and transferred to a designated DICOM destination as a DICOM encapsulated PDF document for viewing and interpretation.

fineSA is a software only image processing system that can be installed on any network computer that meets the specified minimum hardware and operating system requirements. The system can be deployed on the radiology internal network or at a remote location with secure access to radiology internal network.

## 6. Indications for Use

fineSA software is intended for use as a non-invasive device that processes designated MR images using the 2-D pulse sequence incorporated on Siemens Medical 3T MAGNETOM systems running syngo MR software versions B17 and D13. This includes the Aera, Skyra, Avanto and Verio models.

fineSA software generates reports that provide information about the structural features in bone. The MR data, in the form of spectral images, reflect the spacing of structural features at a scale of approximately 500 microns or larger.

## 7. Predicate Device Comparison

The fineSA software is substantially equivalent to the syngo.MR General post-processing application (K130749).

A summary comparison of characteristics is provided in Table 1.

{5}------------------------------------------------

Fine Structure Analysis (fineSA)

| Characteristic                    | Fine Structure Analysis (fineSA)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | PREDICATE<br>syngo.MR General                                                                                                                                                                                                                                                   |
|-----------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Clearance                  | K150069                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | K130749                                                                                                                                                                                                                                                                         |
| Imaging Modality                  | MRI                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | MRI                                                                                                                                                                                                                                                                             |
| Product Code                      | LLZ                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | LLZ                                                                                                                                                                                                                                                                             |
| Intended Use                      | fineSA software is intended for use as a non-<br>invasive device that processes designated MR<br>images using the 2-D pulse sequence incorporated<br>on Siemens Medical 3T MAGNETOM systems<br>running syngo MR software versions B17 and D13.<br>This includes the Aera, Skyra, Avanto and Verio<br>models.<br><br>fineSA software generates reports that provide<br>information about the structural features in bone.<br>The MR data, in the form of spectra or spectral<br>images, reflect the spacing of structural features at a<br>scale of approximately 500 microns or larger. | Post processing software / applications to be used<br>for viewing and evaluating the designated images<br>provided by a magnetic resonance device.<br><br>syngo.MR General is a syngo based post-processing<br>software for viewing, manipulating, and evaluating<br>MR images. |
| Software Device                   | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     | Yes                                                                                                                                                                                                                                                                             |
| Patient Contact                   | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | No                                                                                                                                                                                                                                                                              |
| Graphical User<br>Interface (GUI) | No                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Yes                                                                                                                                                                                                                                                                             |
| Data Interchange                  | DICOM                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | DICOM                                                                                                                                                                                                                                                                           |
| Data Acquisition                  | Data acquired for fineSA processing uses spin echo<br>and gradient echo sequences.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Data acquired uses spin echo and gradient echo<br>sequences.                                                                                                                                                                                                                    |
| Data Analysis                     | Data analyzed using standard data analysis<br>techniques including Fourier and inverse Fourier<br>transformations and windowing.                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Data analyzed using standard data analysis<br>techniques including Fourier and inverse Fourier<br>transformations and windowing.                                                                                                                                                |
| Data Presentation                 | Data presented in conjunction with MR images.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Data presented as or in conjunction with MR<br>images.                                                                                                                                                                                                                          |
| Data Types                        | Spacings of anatomical elements in bone.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Various, including anatomical, relaxation time,<br>diffusion rate, spectroscopic. and chemical shift.                                                                                                                                                                           |

#### Table 1 - fineSA & Predicates Comparative Analysis

## 8. Performance Data

#### Software Verification and Validation Testing

Software verification and validation testing were conducted and documentation was provided as recommended by the FDA's Guidance for Industry and FDA Staff, "Guidance for Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered a "moderate" level of concern, since a malfunction or latent flaw in the software could delay delivery of appropriate medical care that would likely lead to minor injury.

The fineSA processing software was fully tested, verified and validated by Acuitas as part of its internal design control requirements to ensure it meets defined specifications using phantom and clinical data. Formal test plans were executed to confirm that fineSA meets its formal requirements. This submission includes the verification and validation plan, test results, and verification report confirming the fineSA meets its intended use and performance requirements.

{6}------------------------------------------------

# 9. Risk Management

Product risk management activities were performed in accordance with ISO 14971:2007 throughout the product development process. Risk management verification consisted of both a desk audit and software testing to ensure the implementation of all risk mitigations for the device.

# 10. Substantial Equivalence Conclusion

As noted previously, the fineSA and its predicate devices have similar intended use for viewing and evaluating the designated images provided by a magnetic resonance device. While there are some differences between fineSA and its predicates, these differences are minor and do not raise different questions of safety and effectiveness. Performance data and software verification and validation demonstrate that the fineSA software should perform as intended in the specified use conditions.

Therefore, Acuitas Medical believes that the fineSA software is substantially equivalent to the predicate devices identified in this submission.

---

**Source:** [https://fda.innolitics.com/device/K150069](https://fda.innolitics.com/device/K150069)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
