Published clinical literature on manual TKA complication rates
A literature review of published occurrence rates for standardized TKA complications in manual instrumentation was used to establish a safety benchmark for the investigational device.
Literature review; Manual TKA; Safety benchmark; Complication rates
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
RIO Total Knee Arthroplasty (TKA) IDE study; Prospective, non-randomized, multi-center study; Follow-up/Duration: 3 months
89 subjects implanted with the RIO System; Sample Size: 89; Number of Sites: 3
Published literature on manual TKA complication rates
The Robotic Arm Interactive Orthopedic System (RIO®) is intended to assist the surgeon in providing software defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures. The RIO® is indicated for use in surgical knee procedures in which the use of stereotactic surgery may be appropriate, and where reference to rigid anatomical bony structures can be identified relative to a CT based model of the anatomy. These procedures include: • Total Knee Arthroplasty (TKA) The Implants systems with which the system is compatible: • Kinetis Total Knee System (CR/UC) • Triathlon Total Knee System (CR/CS/PS cemented Primary)
Device Story
RIO is a stereotactic instrument system comprising a robotic arm, optical detector, computer, planning laptop, and dedicated instrumentation. It utilizes patient CT data for pre-surgical implant planning and intraoperative tracking of the femur and tibia. During orthopedic surgery, the robotic arm acts as an intelligent tool guide, providing stereotactic guidance by passively constraining tool movement within software-defined spatial boundaries. The system is operated by a surgeon in an OR setting. By providing real-time navigation and haptic constraints based on the pre-operative plan, the device assists the surgeon in achieving accurate implant placement and bone resection, potentially improving clinical outcomes and limb alignment compared to manual instrumentation.
Clinical Evidence
Prospective, non-randomized, multi-center study (N=89 implanted subjects). Primary endpoint: safety (incidence of device-related adverse events) compared to manual TKA literature; 0/89 subjects experienced primary composite safety events. Secondary endpoint: post-operative limb alignment; mean deviation from plan was 1.53° (± 1.05°) and 1.56° (± 1.04°) by two independent reviewers. Supporting endpoint: WOMAC score improvement; mean reduction of 29.6 points at 3 months. One UADE (retained foreign body) and three possibly related AEs reported. Results demonstrate non-inferiority to manual TKA.
Technological Characteristics
Stereotactic instrument system. Components: robotic arm, optical detector, computer, planning laptop, dedicated instrumentation. Sensing: optical tracking of rigid bony structures. Imaging: CT-based. Connectivity: integrated system platform. Software: application-specific guidance software for spatial boundary definition. Sterilization: reusable and disposable instruments.
Indications for Use
Indicated for patients undergoing surgical knee procedures, specifically Total Knee Arthroplasty (TKA), where stereotactic surgery is appropriate and rigid anatomical bony structures can be identified relative to a CT-based model.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing right, with flowing lines beneath them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 6, 2015
MA Surgical Corporation Jonathan Reeves Principal Senior Regulatory Specialist 2555 Davie Road Ft. Lauderdale, Florida 33317
Re: K143752 Trade/Device Name: Total Knee Application (TKA) Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: July 1, 2015 Received: Julv 7, 2015
Dear Jonathan Reeves,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Mark N. Melkerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
| DEPARTMENT OF HEALTH AND HUMAN SERVICES<br>Food and Drug Administration<br><b>Indications for Use</b> | Form Approved: OMB No. 0910-0120<br>Expiration Date: January 31, 2017<br>See PRA Statement below. |
|-------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number ( <i>if known</i> ) | K143572 |
| Device Name | Total Knee Application (TKA) |
| Indications for Use ( <i>Describe</i> ) | |
| Indications for Use: | |
| | The Robotic Arm Interactive Orthopedic System (RIO®) is intended to assist the surgeon in providing software defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures. |
| | The RIO® is indicated for use in surgical knee procedures in which the use of stereotactic surgery may be appropriate, and where reference to rigid anatomical bony structures can be identified relative to a CT based model of the anatomy. These procedures include:<br>• Total Knee Arthroplasty (TKA) |
| | The Implants systems with which the system is compatible:<br>• Kinetis Total Knee System (CR/UC)<br>• Triathlon Total Knee System (CR/CS/PS cemented Primary) |
| Type of Use ( <i>Select one or both, as applicable</i> ) | |
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the
time to review instructions, search existing data sources, gather and maintain the data needed and complete
and review the collection of information. Send comments regarding this burden estimate or any other aspect
of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services
Food and Drug Administration
Office of Chief Information Officer
Paperwork Reduction Act (PRA) Staff
[PRAStaff@fda.hhs.gov](mailto:PRAStaff@fda.hhs.gov)
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of
information unless it displays a currently valid OMB number."
FORM FDA 3881 (8/14) Page 1 of 1 PSC Publishing Services (301) 443-6740 EF
.
1
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image is a logo for MAKO Surgical Corp. The logo features the word "MAKO" in large, bold, black letters. Below "MAKO" is the text "SURGICAL CORP" in smaller, black letters. Above the word "MAKO" is a blue, curved shape that resembles a wave or an abstract mountain range.
555 Davie Road • Ft. Lauderdale, FL 33317 one 954.927.2044 • Fax 954.927.04 ww.makosurgical.com
510(K) SUMMARY
Submitter: Address: Phone number: Fax number: Contact Person: Date Prepared: Device Trade Name: Regulation Name: Regulation Number: Device Classification: Product Code:
MAKO Surgical Corp. 2555 Davie Road, Fort Lauderdale, FL 33317 954-628-0655 954-927-0446 Jonathan Reeves December 30, 2014 Total Knee Application (TKA) Stereotaxic Instrument 21 CFR 882.4560 Class II OLO
## Substantial Equivalence Claimed To:
Total Knee Application is substantially equivalent to MAKO Surgical Corp.'s (Robotic Arm Interactive Orthopedic System) RIO® - (Partial Knee Application) PKA cleared via K112507.
#### Description:
The Robotic Arm Interactive Orthopedic System (RIO) with Total Knee Arthroplasty Application is a stereotactic instrument that includes a robotic arm, an integrated cutting system, an optical detector, a computer, dedicated instrumentation, operating software, a planning laptop, and tools and accessories. RIO uses patient CT data to assist the physician with pre-surgical implant placement planning and intraoperative tracking of the patient's femur and tibia. RIO's robotic arm serves as an "intelligent" tool holder or tool guide used by a surgeon for stereotactic guidance during orthopedic surgical procedures.
The main RIO platform includes an optical detector, computer, dedicated instrumentation, operating software, tools and accessories, cutting system, and a robotic arm. The system's architecture is designed to support total knee procedures. With application specific hardware and software, the system provides stereotactic guidance during orthopedic surgical procedures by using patient CT data to assist a surgeon with pre-surgical planning and interpretive/intraoperative navigation.
RIO's robotic arm, once configured for a specific application, can serve as surgeon's "intelligent" tool holder or tool guide by passively constraining the preparation of an anatomical site for an orthopedic implant with software-defined spatial boundaries.
{4}------------------------------------------------
## Summary of Technological Characteristics Compared to Predicate Devices:
The technological characteristics of Total Knee Application compared to the predicate device are listed below:
| Technological<br>Characteristics | Total Knee Application | RIO-PKA<br>(K112507) |
|----------------------------------|-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------|
| Major<br>Components | Guidance Module, robotic arm,<br>camera stand, cutting system,<br>preoperative planning laptop. | Guidance Module, robotic arm,<br>camera stand, cutting system,<br>preoperative planning laptop. |
| Tools/accessories | Various reusable and disposable<br>instruments | Various reusable and disposable<br>instruments |
| Images Use | CT | CT |
## Intended Use/Indications for Use:
The Robotic Arm Interactive Orthopedic System (RIO®) is intended to assist the surgeon in providing software defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures.
The RIO® is indicated for use in surgical knee procedures in which the use of stereotactic surgery may be appropriate, and where reference to rigid anatomical bony structures can be identified relative to a CT based model of the anatomy. These procedures include:
- Total Knee Arthroplasty (TKA) ●
The Implant systems with which the system is compatible:
- Kinetis Total Knee System ( CR/UC) .
- Triathlon Total Knee System ( CR/CS/PS cemented Primary)
## Performance Data:
| Validation / Verification<br>Method | Purpose | Validation / Verification<br>Results |
|-------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------|
| Full System Run-Through<br>Test | Verify that the integration<br>of the Robotic Arm<br>Interactive Orthopedic<br>System (RIO), with the<br>Total Knee Application<br>Software, and supporting<br>instrumentation provides<br>adequate functionality to be<br>able to successfully<br>complete a MAKOplasty<br>Total Knee Arthroplasty<br>procedure | Pass |
| System Accuracy Test | Verify the overall system<br>accuracy by combining<br>bone registration and bone | Pass |
{5}------------------------------------------------
| | resection accuracy. | |
|--------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| Bone Registration Accuracy<br>Test | Verify the accuracy of bone<br>registration for the TKA 1.0<br>Application satisfies the<br>specified requirements. | Pass |
| RIO 3.0 System Platform<br>Cutting Accuracy Test | Verify the RIO resection<br>accuracy for the Total Knee<br>Arthroplasty Application<br>satisfies the specified<br>requirements. | Pass |
| TKA Perimeter Retention<br>Accuracy Verification | Verify the RIO perimeter<br>retention accuracy for the<br>Total Knee Arthroplasty<br>Application satisfies the<br>specified requirements. | Pass |
| TKA Validation | Verify in a simulated-use<br>environment that the<br>integration of the Robotic<br>Arm Interactive Orthopedic<br>System (RIO) with the<br>Total Knee Application<br>Software and supporting<br>instrumentation provides<br>adequate functionality to<br>successfully complete a<br>MAKOplasty Total Knee<br>Arthroplasty procedure and<br>satisfies the customer<br>requirements | Pass |
## Clinical Data:
The Robotic Arm Interactive Orthopedic System (RIO) Total Knee Arthroplasty (TKA) study was a prospective, non-randomized, multi-center study conducted at three sites in the US. The purpose of the RIO TKA IDE study was to demonstrate safety and effectiveness of the RIO system in total knee arthroplasty at 3 months. A total of 100 subjects were consented and enrolled. A total of 89 subjects were implanted using the RIO System and all have successfully completed pre-operative, intra-operative, and 3 month post-operative visits.
There were 49 female (55.1%) subjects and 40 (44.9%) male subjects. Refer to table below for age, height and weight distribution for this population
{6}------------------------------------------------
| Label | N | Mean | Std Dev | Median | Lower Quartile | Upper Quartile | Minimum | Maximum |
|-------------|----|--------|---------|--------|----------------|----------------|---------|---------|
| Age | 89 | 65.78 | 8.49 | 65.00 | 60.00 | 72.00 | 45.00 | 83.00 |
| Height (in) | 89 | 67.34 | 3.84 | 67.00 | 65.00 | 70.00 | 60.00 | 75.00 |
| Weight (lb) | 89 | 202.61 | 38.96 | 197.00 | 175.00 | 226.00 | 122.00 | 318.00 |
| BMI | 89 | 31.43 | 5.70 | 31.07 | 27.45 | 33.78 | 21.61 | 52.90 |
The primary study endpoint was to demonstrate the safety of the RIO Total Knee Arthroplasty Application based on surgeon assessment of complications intra-operatively and at a short term follow up. The safety analysis focused on the complications identified by the Knee Society TKA Complications Workgroup in 2012 for total knee arthroplasty. A review of the literature was performed to identify currently published occurrence rates in manually instrumented TKA for each standardized TKA complication. The analysis of the safety profile was designed to provide clinical data necessary to test the hypothesis that there is no clinically significant increase in the incidence of selected device-related adverse events when using the investigational device relative to manual TKA. No subject experienced any of the rare adverse events that comprise the primary composite safety events (0/89). Therefore, the study meets its primary safety study success criterion.
The secondary endpoint was post-operative radiographic limb alignment of the operative knee assessed at the 3 month post-operative by two independent reviewers. The measured post-operative limb alignment was compared to the derived subject-specific value for the intended limb alignment established prior to bone resection. Two subjects were excluded from the assessment due to issues with the methods of collecting the images used to make the assessment. Excluding these two subjects, independent surgeon reviewer one measured an estimated mean limb alignment deviation from plan of 1.53° (± 1.05°) and independent surgeon reviewer two measured an estimated mean limb alignment deviation from plan of 1.56° (± 1.04°).
In addition to reporting accuracy in achieving post-operative limb alignment, malalignment was defined as a radiographically determined alignment angular deformity in the coronal plane greater than 10° from the mechanical axis. Alignment was assessed in descriptive analysis as a categorical variable using the categories appearing in the Knee Society Score, Doctor Portion. These categories are 0-4°, 5-10°, 11-15°, and Other, however, these ranges do not account for the surgeon's subject-specific intended postoperative limb alignment. Forty (40) of 89 subjects (44.9%) reported post-operative KSS limb alignment of 0-4°, which is associated with a 3-point deduction in KSS score per degree, while 49/89 (55.1%) reported KSS limb alignment of 5-10° which received no deduction according to the Knee Society Score. No subject reported malalignment as defined above.
The supporting study endpoint was to demonstrate improvement in post-operative subject function at the three month interval as compared to the subject's pre-operative function as measured by the WOMAC. In order to meet the study success criterion, a reduction in the WOMAC score from pre-op to 3 months was necessary. Subjects exhibited a mean 29.6 points reduction, indicating improvement over pre-operative score.
{7}------------------------------------------------
There were 33 reported Adverse Events (AEs) occurring in 32 of the study subjects. Study investigators classified relatedness of all AEs as definitely related, possibly related, or not related to the study system or procedure. There was 1 (1/33, 3.0%) AE that was considered related to the system or procedure. The AE was for an Unanticipated Adverse Device Effect (UADE) for retained foreign body (saw tooth). There were 3 (3/33, 9.1%) that were considered possibly related to the device. The SAEs possibly related to the device included: UADEs was for retained foreign body (saw tooth) (1) and swelling and pain (2). There were a total of 9 (9/33, 27.2%) reported serious AEs (SAEs) that were not related to the device or procedure. These included (8) manipulation under anesthesia (MUA) and (1) pulmonary embolism.
A second surgery was required for two subjects to remove a retained foreign body. For one of the subjects, a revision of the tibial insert was required to locate the foreign body and irrigate the joint space.
The safety and effectiveness of the Robot Arm Interactive Orthopaedic System (RIO®) used for Total Knee Replacement (TKR) has been demonstrated. No subject experienced any rare adverse events that comprised the composite safety events. Secondary endpoints of postoperative limb alignment and clinical outcomes scores at three months postoperatively demonstrated non-inferiority to TKR used with manual instruments as reported in the literature.
#### Conclusions of Clinical and Non-Clinical Data:
The results of the clinical and non-clinical data indicate the device performed within the intended use and did not raise any new safety and efficacy issues. The device was found to be substantially equivalent to the predicate device.
Image /page/7/Picture/5 description: The image is an abstract design featuring a light blue shape against a white background. The shape is curved and tapers at both ends, resembling a stylized wing or a swooping line. The overall impression is minimalist and airy, with the light blue color providing a soft contrast to the white backdrop.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.