Rhondium One-Visit-Crown is intended for restoration of permanent teeth with a single unit crown.
Device Story
Rhondium One-Visit-Crown (OVC) is a preformed hybrid ceramic crown for single-visit restoration of structurally compromised posterior teeth. Device consists of a fully-cured anatomical occlusal layer and an uncured sub-layer; clinician selects size based on tooth location; no impressions required. Accessories include OVC Replica (vinyl polysiloxane) for checking occlusal clearance; OVC Wedges (polyacetal resin) for interproximal working surface; Selector Keys (polycarbonate) for sizing and stabilization during curing. Clinician places OVC on prepared tooth; uncured layer allows customization to preparation; light-cured to finish. Output is a customized, bonded crown. Benefits include single-visit restoration and simplified procedure compared to traditional onlays or partial crowns.
Clinical Evidence
No clinical data were included in this submission. Substantial equivalence is supported by non-clinical bench testing, including product shelf life, biocompatibility (ISO 10993-1, ISO 10993-5), flexural strength, static compressive strength, and resin depth of cure (ISO 4049).
Technological Characteristics
Hybrid ceramic crown; fully-cured occlusal layer and uncured sub-layer. Materials: ceramic tooth shade resin, vinyl polysiloxane (Replica), polyacetal resin (Wedges), polycarbonate (Selector Keys). Dimensions: 8 designs in 6 sizes (XS-XXL). Energy source: light-cured. Nonsterile, single-use. Standards: ASTM F1980 (accelerated aging), ISO 10993-1/5 (biocompatibility), ISO 4049 (polymer-based restorative materials).
Indications for Use
Indicated for restoration of permanent teeth with a single unit crown in patients requiring posterior tooth restoration.
Regulatory Classification
Identification
Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized image of an eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 15, 2015
Rhondium Ltd. c/o Mr. Kevin A. Thomas PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego California, 92130
Re: K143694
Trade/Device Name: Rhondium One-Visit-Crown (OVC) Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: II Product Code: EBF Dated: April 14, 2015 Received: April 15, 2015
Dear Mr. Thomas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office
of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Tina
Kiang -S
for Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
K143694
Device Name
Rhondium One-Visit-Crown (OVC)
Indications for Use (Describe)
Rhondium One-Visit-Crown is intended for restoration of permanent teeth with a single unit crown.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|
| <div> <span> <b> ☑ </b> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW!*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Druq Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
# Rhondium, Ltd
# Rhondium One-Visit-Crown (OVC)
# K143694
# May 11, 2015
# ADMINISTRATIVE INFORMATION
| Manufacturer Name | Rhondium, Ltd<br>4 Sheffield Street<br>Katikati, 3129, New Zealand<br>Telephone: +64 7 549 5566<br>Fax: +64 7 549 1461 | |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------|------------------------------------------|
| Official Contact | Greer Fricker<br>R&D Projects and Regulatory Manager | |
| Representative/Consultant | Kevin A. Thomas, PhD<br>Floyd G. Larson<br>PaxMed International, LLC<br>12264 El Camino Real, Suite 400<br>San Diego, CA 92130 | |
| | Telephone: | +1 (858) 792-1235 |
| | Fax: | +1 (858) 792-1236 |
| | Email: | kthomas@paxmed.com<br>flarson@paxmed.com |
### DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | Rhondium One-Visit-Crown (OVC) |
|----------------------------|-----------------------------------------------------|
| Common Name | Dental material, filling/restorative, polymer based |
| Classification Name | Tooth Shade Resin Material |
| Classification Regulations | 21 CFR 872.3690, Class II |
| Product Code | EBF |
| Classification Panel | Dental Products Panel |
| Reviewing Branch | Dental Device Branch |
{4}------------------------------------------------
510(k) Summary K143694
### INTENDED USE
Rhondium One-Visit-Crown is intended for restoration of permanent teeth with a single unit crown.
### DEVICE DESCRIPTION
Rhondium One-Visit-Crown (OVC) is a hybrid ceramic crown consisting of a fully-cured anatomical occlusal layer and an uncured sub-layer that allows customization of the OVC to the tooth preparation. When finished, the customized OVC is bonded to a prepared natural tooth. The OVC is for restoring structurally compromised posterior teeth in cases normally restored with an onlay or a partial crown in a single dental visit. The OVC is selected based on tooth location and size and therefore no impressions are necessary. The OVC is provided in designs for the upper and lower first and second premolars and the first and second molars, and each design is provided in six (6) sizes (extra-small (XS), small (S), medium (M), large (L), extra-large (XL), and extra-extra-large (XXL)). The OVC is made from a ceramic tooth shade resin material, and is provided nonsterile for single use only.
Accessories to the OVC Device include the OVC Replica, an exact replica of the cured (occlusal layer) portion of the OVC. The OVC Replica is used to confirm that sufficient occlusal clearance was obtained during tooth preparation and to provide direct visualization of the overall 3D position of the OVC on the prepared tooth and the relationship with adjacent and opposing dentition. The OVC Replica is manufactured from vinyl polysiloxane impression material. OVC Wedges are accessories provided in four configurations (small right, large left and large right), and are placed interproximally to create a working surface and to block out undercuts of the preparation. The Wedges are manufactured from polyacetal resin. The Selector Keys are accessories that allow the clinician to measure the mesial-distal distance in order to choose the appropriate size OVC for the restoration, and are used during the curing of the uncured resin to prevent disturbing the placement of the crown. The Selector Keys are manufactured from polycarbonate resin. All accessories are provided nonsterile for single use only.
# PERFORMANCE DATA
Performance testing to demonstrate substantial equivalence included methods described in the following standards: ASTM F1980 Standard Guide for Accelerated Aging of Sterile Barrier Systems for Medical Devices; ISO 10993-1 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process; ISO 10993-5 Biological evaluation of medical devices – Part 5: Tests for in vitro cytotoxicity; and ISO 4049 Dentistry – Polymer-Based Restorative Materials. Non-clinical testing data submitted, referenced, or relied upon to demonstrate substantial equivalence included: product shelf life testing, biocompatibility testing, flexural strength testing, static compressive strength testing, and resin depth of cure testing. No clinical data were included in this submission.
{5}------------------------------------------------
## EQUIVALENCE TO MARKETED DEVICE
Rhondium, Ltd submits the following information in this Premarket Notification to demonstrate, for the purposes of FDA's regulation of medical devices, the subject device is substantially equivalent in indications and design principles to the following legally marketed predicate devices:
K072733, 3M™ ESPE™ Adult Crown, 3M ESPE Dental Products;
K050401, Ceramage, ShoFu Dental Corporation;
K052501, Resin Tooth Bonding Agent, Cosmedent, Incorporated; and
K092319, Star VPS, First Half, First Quarter, Danville Materials, Incorporated.
The primary predicate device is K072733. A comparison of the technological characteristics of the subject device and the predicate device K072733 is provided in the following table.
| | Subject Device | Predicate Device |
|----------------------------------|----------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| Comparison | Rhondium Ltd<br>One-Visit-Crown (OVC)<br>K143694 | 3M ESPE Dental Products<br>3M™ ESPE™ Adult Crown<br>K072733 |
| Crown Description | Preformed single tooth crown for one visit<br>restoration | Preformed single tooth crown for one visit<br>restoration |
| Teeth to be Restored | Upper and lower premolars and molars | Upper and lower canines, premolars, and molars |
| Specific Crown Designs | N=8 designs :<br>Upper Premolar, Upper Molar, Lower Premolar,<br>Lower Molar<br>Each in Left and Right designs | N=5 designs<br>Canine, Upper Premolar, Upper Molar, Lower<br>Premolar; Lower Molar |
| Crown Sizes | Each design provided in 6 sizes:<br>XS, S, M, L, XL, XXL | Each design provided in 2 sizes, except Lower<br>Premolar (1 size) |
| Crown Material | Light-cured polymer composite resin | Light-cured composite |
| Crown Try-in (Trial)<br>Material | OVC Replica<br>Vinyl polysiloxane | None |
| Sizing Device Material | OVC Selector Key<br>Polycarbonate | 3M ESPE Crown Size Tool<br>Material not stated |
| Other Accessory | OVC Wedges<br>polyacetal resin | None |
The subject device and the predicate device K072733 have the same intended use and very similar indications for use (restoration of permanent teeth with a single unit crown). The subject device and the predicate K072733 are provided in a range of designs and sizes to restore the premolars and molars of the upper and lower jaw. The subject device and the predicate K072733 encompass similar ranges of physical dimensions and characteristics, including design and dimensions, and both are made of light-cured resin materials. The subject device and the predicate devices are provided nonsterile for single use only.
{6}------------------------------------------------
The subject device consists of a fully-cured layer of tooth shade resin, a bonding layer, and a layer of uncured tooth shade resin to allow for customization and to assist with a cohesive fit of the crown to the prepared tooth.
Any differences in the technological characteristics between the subject device and the predicate devices do not affect substantial equivalence.
## CONCLUSION
The data included in this submission demonstrates substantial equivalence to the predicate devices listed above.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.