ASFORA ANTERIOR CERVICAL PLATE SYSTEM

K143688 · Medical Designs, LLC · KWQ · Jul 1, 2015 · Orthopedic

Device Facts

Record IDK143688
Device NameASFORA ANTERIOR CERVICAL PLATE SYSTEM
ApplicantMedical Designs, LLC
Product CodeKWQ · Orthopedic
Decision DateJul 1, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3060
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Asfora Anterior Cervical Plate System is intended to provide temporary stabilization to the anterior spine during the development of cervical spine fusions (C2-T1) for the following indications: degenerative disc disease (DDD) (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), tumor, pseudoathrosis, and failed previous fusion.

Device Story

Anterior cervical plate system providing temporary stabilization during spinal fusion (C2-T1). System consists of metallic plates of varying lengths and locking screws; plates feature locking tabs to prevent screw back-out. Used by surgeons in clinical settings to fixate vertebral bodies. Device acts as internal orthosis to maintain alignment and stability during bone healing. Benefits include structural support for fusion process. System includes associated surgical instruments (drill guides, plate benders, drivers) for implantation.

Clinical Evidence

No clinical data. Bench testing only. Mechanical testing performed in accordance with ASTM F1717-14, including dynamic compression bending, static compression bending, static torsion, screw insertion torque, and screw retention force evaluation.

Technological Characteristics

Spinal intervertebral body fixation orthosis. Components include metallic plates and screws. Features include locking tabs, pre-lordosed plates (190mm radius), and square drive interfaces. Testing conducted per ASTM F1717-14. System is non-sterile; sterilization tray provided for instrument/implant processing.

Indications for Use

Indicated for patients requiring temporary stabilization of the anterior cervical spine (C2-T1) during fusion for DDD, spondylolisthesis, trauma, spinal stenosis, deformities (scoliosis, kyphosis, lordosis), tumor, pseudoarthrosis, or failed previous fusion.

Regulatory Classification

Identification

A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized representation of three human profiles facing right, stacked on top of each other. The profiles are rendered in a dark color, creating a sense of depth and unity. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the profiles. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 July 1, 2015 Medical Designs, LLC Ms. Kristi Vondra Vice President of Operations 6709 South Minnesota Avenue, Suite 204 Sioux Falls, South Dakota 57108 Re: K143688 Trade/Device Name: Asfora Anterior Cervical Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: KWQ Dated: June 3, 2015 Received: June 4, 2015 Dear Ms. Vondra: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set {1}------------------------------------------------ forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours. # Lori A. Wiggins -S for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ #### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below. K143688 Page 1 of 1 510(k) Number (if known) K143688 Device Name Asfora Anterior Cervical Plate (AACP) System #### Indications for Use (Describe) The Asfora Anterior Cervical Plate System is intended to provide temporary stabilization to the anterior spine during the development of cervical spine fusions (C2-T1) for the following indications: degenerative disc disease (DDD) (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), tumor, pseudoathrosis, and failed previous fusion. Type of Use (Select one or both, as applicable) | <span style="font-size:14px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:14px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | |------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------| |------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------| #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary #### 1. Submission Sponsor Medical Designs, LLC 6709 S. Minnesota Ave., Suite 204 Sioux Falls, South Dakota 57108 #### 2. Submission Correspondent Kristi Vondra, Vice President of Operations Tel: (605) 275-1032 FAX: (605) 335-3734 #### 3. Date Prepared June 3, 2015 #### 4. Device Identification | Trade/Proprietary Name: | Asfora Anterior Cervical Plate System | |----------------------------|----------------------------------------------| | Common/Usual Name: | Anterior Cervical Plate System | | Classification Name: | Spinal intervertebral body fixation orthosis | | Classification Regulation: | 21 CFR §888.3060 | | Product Code: | KWQ | | Device Class: | Class II | | Classification Panel: | Orthopaedic and Rehabilitation Devices Panel | #### 5. Predicate Devices | Primary: K122497 | Asfora Anterior Cervical Plate System | |---------------------|---------------------------------------| | Additional: K030866 | Synthes Cervical Spine Locking Plate | #### 6. Device Description The Asfora Anterior Cervical Plate System is composed of plates of varying lengths to accommodate surgical procedures from one to four levels. The plates are designed for application to the anterior aspect of the cervical spine and have three locking tabs to reduce the potential for screws to back out of the vertebral body. The screws are provided in diameters of 4.0mm (Variable Angle Implant Screw and Fixed Angle Implant Screw) and 4.4mm (Rescue Screw) and lengths of 12mm, 14mm and 16mm. {4}------------------------------------------------ ## 7. Indication for Use The Asfora Anterior Cervical Plate System is intended to provide temporary stabilization to the anterior spine during the development of cervical spine fusions (C2-T1) for the following indications: degenerative disc disease (DDD) (defined as neck pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies), spondylolisthesis, trauma (i.e., fracture or dislocation), spinal stenosis, deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), tumor, pseudoarthrosis, and failed previous fusion. # 8. Comparison to Predicates No changes have been made to the intended use. Indications for Use, principle of operation or device materials of the cleared primary predicate Asfora Anterior Cervical Plate System. The subject device is identical to the primary predicate device except for the following differences. The additional plate and screw dimensions are substantially equivalent to the additional predicate. | Characteristic | Difference in Subject Device to Primary Predicate | |----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Plate locking<br>mechanism | Locking tabs instead of locking rings | | Plate dimensions | maximum plate width decreased from 19 to 17.5mm (in range of reference predicate)additional plate lengths at all 4 levels (in range of reference predicate)lateral hole to hole distance decreased (from 9 to 8.5mm) | | Plate pre-bent angle | All plates pre-lordosed with a radius of 190mm | | Screw Drive Feature | Square instead of hexalobe | | Variable Angle<br>Implant Screws | Added 4.0 x 12mm screw (in range of reference predicate); spherical head to fit plate recess | | Variable Angle<br>Rescue Screw | Diameter decreased (5.0 to 4.4mm) and added 4.4 x 12mm screw (in range of reference predicate) | | Fixed Angle Screws | Added fixed angle screws (in range of reference predicate); spherical head to fit plate recess and shank to match plate exit hole | | Temporary Fixation<br>Screws | Removed from system; Casper Pin Style recommended for ease of use and functionality, but not provided | | Extractor Assembly | Square interface instead of hexalobe | | Drill Guide | Change in tip geometry to interface with spherical cavity of plate; addition of fixed angle drill guide configuration | | Plate Bender | Simplified, off the shelf French bender | | Retention Assembly | Nemcomed design; one-piece square driver with thermoplastic insert. | | Sterilization Tray | Minor modification to hold redesigned cervical plates, screws and instruments | {5}------------------------------------------------ ## 9. Functional and Safety Testing Mechanical testing was conducted, in accordance with Guidance for Industry and FDA Staff: Spinal System 510(k)s and ASTM F 1717-14 to ensure the device performs according to specification, to verify that the device is able to withstand clinical loading and maintain mechanical integrity, and is suited for its intended purpose. Testing included dynamic compression bending, static compression bending, static torsion, screw insertion torque evaluation and screw retention force evaluation. ### 10. Conclusion Medical Designs. LLC considers the Asfora Anterior Cervical Plate System to be equivalent to the predicate Asfora Anterior Cervical Plate System listed above. This conclusion is based upon the device similarities in intended use, Indications for Use, principle of operation, materials and performance testing.
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%