← Product Code [EOB](/productcode/EOB) · K143673

# CMOS Video Nasopharyngoscope System (K143673)

_Schoelly Fiberoptic GmbH · EOB · Mar 4, 2015 · Ear, Nose, Throat · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K143673

## Device Facts

- **Applicant:** Schoelly Fiberoptic GmbH
- **Product Code:** [EOB](/productcode/EOB.md)
- **Decision Date:** Mar 4, 2015
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 874.4760
- **Device Class:** Class 2
- **Review Panel:** Ear, Nose, Throat
- **Attributes:** Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K143673 · Mar 4, 2015 | CMOS Video Nasopharyngoscope System | Schoelly Fiberoptic GmbH | Devices previously in clinical use | Devices previously used in clinical settings were subjected to high-level disinfection (HLD) validation testing to ensure the efficacy of the disinfection process under 'real life' conditions. | High-level disinfection; Reprocessing validation; Clinical use devices |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| HLD Validation Study; Validation study using devices previously in clinical use; Follow-up/Duration: 12 minute immersion | Endoscopes previously in clinical use; Sample Size: 5 test results for microbial reduction | Control devices (inoculated but not disinfected) | Log-reduction of Mycobacterium terrae |

## Indications for Use

The Schoelly CMOS Video Nasopharyngoscope System may only be used by persons with an appropriate medical qualification and who are acquainted with the rhino/laryngoscopic technique. The endoscope is used for endoscopic diagnosis within the nasal lumens and airway anatomy, and is intended to provide visualization via a video monitor.

## Device Story

Flexible, steerable endoscope with integrated CMOS imaging sensor and LED light source; camera control unit (CCU) regulates light and connects to external monitors/recorders. Used by physicians for rhino/laryngoscopic diagnosis; provides real-time visualization of nasal and airway anatomy. Device delivered non-sterile; requires high-level disinfection (HLD) between uses. Modification involves updated labeling for validated HLD process using 0.55% orthophthaldehyde.

## Clinical Evidence

Bench testing only. HLD efficacy validated per ASTM E 1837 and ANSI AAMI ST 58 using Mycobacterium terrae as indicator organism. Process achieved 7.5-log reduction, exceeding 6-log requirement. Cytotoxicity testing confirmed no residual disinfectant interference. Functional testing confirmed no material degradation after multiple HLD cycles.

## Technological Characteristics

Flexible steerable endoscope; plastic outer surfaces; CMOS imaging sensor; LED illumination; fiberoptic light transmission. CCU for signal processing and display. HLD process: 12-minute immersion in 0.55% orthophthaldehyde (Cidex OPA) followed by water rinses and air drying. Complies with ASTM E 1837 and ANSI AAMI ST 58.

## Regulatory Identification

A nasopharyngoscope (flexible or rigid) and accessories is a tubular endoscopic device with any of a group of accessory devices which attach to the nasopharyngoscope and is intended to examine or treat the nasal cavity and nasal pharynx. It is typically used with a fiberoptic light source and carrier to provide illumination. The device is made of materials such as stainless steel and flexible plastic. This generic type of device includes the antroscope, nasopharyngolaryngoscope, nasosinuscope, nasoscope, postrhinoscope, rhinoscope, salpingoscope, flexible foreign body claw, flexible biopsy forceps, rigid biopsy curette, flexible biospy brush, rigid biopsy forceps and flexible biopsy curette, but excludes the fiberoptic light source and carrier.

## Predicate Devices

- Schoelly CMOS Video Nasopharyngoscope System ([K132009](/device/K132009.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines connecting them.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

March 4, 2015

Schoelly Fiberoptic Gmbh % Ms. Pamela Papineau President Delphi Medical Device Consulting, Inc. 5 Whitcomb Avenue Ayer, MA 01432

Re: K143673

Trade/Device Name: CMOS Video Nasopharyngoscope System Regulation Number: 21 CFR 874.4760 Regulation Name: Nasopharyngoscope (Flexible Or Rigid) And Accessories Regulatory Class: Class II Product Code: EOB Dated: January 29, 2015 Received: February 5, 2015

Dear Ms. Papineau,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

# Eric A. Mann -S

for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K143673

#### Device Name

Schoelly CMOS Video Nasopharyngoscope System

#### Indications for Use (Describe)

The Schoelly CMOS Video Nasopharyngoscope System may only be used by persons with an appropriate medical qualification and who are acquainted with the rhino/laryngoscopic technique. The endoscopic diagnosis within the nasal lumens and airway and is intended to provide visualization via a video monitor.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

# FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# Special 510(k) Summary

| Preparation date:                                                                   | 27 February 2015                                                                                                                                                                                           |
|-------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number:                                                                      | K143673                                                                                                                                                                                                    |
| Owner's Name:<br>Address:                                                           | Schoelly Fiberoptic GmbH (Registration: 8043903)<br>Robert-Bosch-Str. 1 - 3<br>79211 Denzlingen<br>Germany                                                                                                 |
| Telephone Number:<br>Fax Number:<br>Contact Person:                                 | +49-7666-980-0<br>+49-7666-908-380<br>Dr. Sandra Baumann                                                                                                                                                   |
| Subject Device Name:<br>Trade Name:<br>Common/Usual Name:<br>Classification Name:   | Schoelly CMOS Video Nasopharyngoscope System<br>Schoelly CMOS Video Nasopharyngoscope System<br>Video Nasopharyngoscope System<br>EOB - Nasopharyngoscope (flexible or rigid)<br>21 CFR 874.4760; Class II |
| Predicate Device Name:<br>Trade Name:<br>Common/Usual Name:<br>Classification Name: | Schoelly CMOS Video Nasopharyngoscope System<br>Schoelly CMOS Video Nasopharyngoscope System<br>Video Nasopharyngoscope System<br>EOB - Nasopharyngoscope (flexible or rigid)<br>21 CFR 874.4760; Class II |
| Premarket Notification:                                                             | K132009 (Schoelly Fiberoptic GmbH), SE date April 09, 2014                                                                                                                                                 |

#### Device Description

The Schoelly CMOS Video Nasopharyngoscope System consists of a flexible and steerable endoscope and a camera control unit (CCU) for regulation of light intensity and connection to a monitor, PC, medical video recorder or printer for image display or image documentation.

The endoscope has outer surfaces mainly made from plastic. The endoscope handle incorporates a control lever to bend the distal tip and an integrated LED light source. Light is transmitted through fiberoptic bundles illuminating the anatomy under investigation. The video signal is captured by a CMOS imaging sensor located at the tip of the endoscope shaft and transferred to the CCU.

The endoscope further incorporates a ventilation system to protect the shaust valve at the endoscope handle can further be used for leakage testing. For this purpose the system is accompanied by a leakage tester and accessories.

The Schoelly CMOS Video Nasopharyngoscope System is delivered in a non-sterile condition.

# Indications for Use

The Schoelly CMOS Video Nasopharyngoscope System may only be used by persons with an appropriate medical qualification and who are acquainted with the rhino/laryngoscopic technique. The endoscope is used for endoscopic diagnosis within the nasal lumens and airway anatomy, and is intended to provide visualization via a video monitor.

# Predicate Device

The predicate device is identical to the proposed device with the same device specifications and design, the same indications for use, and the same fundamental scientific technology. The changes described in this submission are limited to a change in the device Instructions for Use to add a validated method for high level 

{4}------------------------------------------------

disinfection to the Schoelly CMOS Video Nasopharyngoscope System that was cleared for marketing by FDA in K132009. This labeling change is the only change to the device cleared in K132009.

# Performance Testing

#### High Level Disinfection Efficacy Testing

This 510(k) contains a summary of the high level disinfection (HLD) validation conducted in accordance with ASTM E 1837 and ANSI AAMI ST 58. This validated HLD process consists of a 12 minute immersion in 0.55% orthophthaldehyde (Cidex OPA) followed by three water rinses, then air drying. The validation was conducted on devices that had previously been in clinical use to ensure "real life" conditions. Test devices and controls were inoculated in several locations with a high-titer (> 10° CFU/mL) suspension of an indicator organism (Mycobacterium terrae). The test devices were disinfected and rinsed according to the device Instructions for Use; the control devices were not disinfected after inoculation. Both the test (inoculated and disinfected) and control (inoculated only) devices were rinsed with a saline buffer solution and the rinse fluid was cultured to determine the number of viable indicator organisms on each device. These data were used to derive the log-reduction in M. terrae achieved through the HLD process. In total, five results for microbial reduction by HLD and three inoculation control results were obtained as part of the efficacy study.

The final recovery portion of this validation demonstrated that the HLD process successfully achieved a 7.5-log reduction in the indicator organism; all of the control devices were found to have the indicator organism present at a minimum concentration of 106 CFU at recovery. Therefore, the HLD process was successfully validated to achieve a minimum 6-log reduction of an appropriate indicator organism.

# Cytotoxicity Testing

Cytotoxicity testing performed on the test devices immediately after HLD confirmed that there was no residual disinfectant material that could adversely affect the test results by obscuring residual indicator organisms remaining after HLD.

#### Functional Testing

To confirm that the HLD process caused no adverse effects on the endoscope, several devices were subjected to multiple, extended HLD cycles followed by visual inspection; no damage or material degradation was observed in any of these test samples.

#### Conclusion

The proposed modification to the Schoelly CMOS Video Nasopharyngoscope System labeling has met all predetermined acceptance criteria as specified by applicable standards, test protocols, and/or customer inputs and does not introduce new patient or user risks. The device with the modified labeling is substantially equivalent to the device cleared in K132009.

---

**Source:** [https://fda.innolitics.com/device/K143673](https://fda.innolitics.com/device/K143673)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
