Noblex Long Pulse Alexandrite Laser

K143666 · Finemec Co., Ltd. · GEX · Apr 23, 2015 · General, Plastic Surgery

Device Facts

Record IDK143666
Device NameNoblex Long Pulse Alexandrite Laser
ApplicantFinemec Co., Ltd.
Product CodeGEX · General, Plastic Surgery
Decision DateApr 23, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4810
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Noblex Long Pulse Alexandrite Laser is indicated for stable long-term or permanent hair reduction. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime. It is used for all skin types (Fitzpatrick I - VI) including tanned skin. It is also indicated for the treatment of benign vascular lesions, benign pigmented lesions and wrinkles.

Device Story

Noblex Long Pulse Alexandrite Laser emits coherent 755nm near-infrared light; energy delivered via fiber-optic coupled hand-piece to target chromophores in human tissue. Rapid localized heating causes thermal destruction of targets for hair removal, vascular/pigmented lesion treatment, and wrinkle reduction. Operated by physicians in clinical settings; system controlled by microprocessor interfaced with LCD touch screen. Physician adjusts laser pulse energy and spot size to optimize treatment based on application. Device consists of main body, hand-piece, and protective goggles. Benefits include non-invasive, targeted thermal therapy for dermatological conditions.

Clinical Evidence

Bench testing only. Device conforms to EN 60825-1:2007 (laser safety), EN 60601-2-22:1996 (therapeutic laser safety), EN 60601-1:2006 (general safety), and EN 60601-1-2:2007 (electromagnetic compatibility).

Technological Characteristics

Near-infrared 755nm pulsed laser; fiber-optic delivery; microprocessor-controlled with LCD touch screen interface. Conforms to EN 60825-1, EN 60601-2-22, EN 60601-1, and EN 60601-1-2 standards. System includes main body and hand-piece.

Indications for Use

Indicated for stable long-term or permanent hair reduction in all skin types (Fitzpatrick I-VI), including tanned skin. Also indicated for treatment of benign vascular lesions, benign pigmented lesions, and wrinkles.

Regulatory Classification

Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with medicine and healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the symbol. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 April 23, 2015 FineMEC Company, Ltd. % Shin Kuk Yoo LSK Biopartners Incorporated 8 East Broadway, Suite 611 Salt Lake City, Utah 84111 Re: K143666 Trade/Device Name: Noblex Long Pulse Alexandrite Laser Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: March 12, 2015 Received: March 20, 2015 Dear Shin Kuk Yoo: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {1}------------------------------------------------ device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Jennifer R. Stevenson -S For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K143666 Device Name Noblex Long Pulse Alexandrite Laser ## Indications for Use (Describe) The Noblex Long Pulse Alexandrite Laser is indicated for stable long-term or permanent hair reduction. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime. It is used for all skin types (Fitzpatrick I - VI) including tanned skin. It is also indicated for the treatment of benign vascular lesions, benign pigmented lesions and wrinkles. Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 20px;">☑</span> | |----------------------------------------------|-----------------------------------------| | Over-The-Counter Use (21 CFR 801 Subpart C) | ☐ | ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and comblete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### 510(k) Summary Prepared date: Mar. 5. 2015 New Device: Noblex Long Pulse Alexandrite Laser - 1. Submitter and US Official Correspondent Submitter: FineMEC Co., Ltd. Address: #230, 7, Galhyeon-ro 11-gil, Eunpyeong-gu Seoul, 122-902, Korea Tel: +82-2-309-1882 Fax: +82-2-376-1884 Contact: Mr. Ghun Hee Lee (General Manager) E-mail: ghlee@finemec.co.kr | US Official Correspondent: | Shin Kuk Yoo, Consultant<br>Telephone No.: 714-313-7442<br>Fax No.: 801-303-7455<br>Email: skyone@LSKBioPartners.com | |----------------------------|----------------------------------------------------------------------------------------------------------------------| |----------------------------|----------------------------------------------------------------------------------------------------------------------| #### 2. Device Information | Proprietary/Trade Name : | Noblex Long Pulse Alexandrite Laser | |--------------------------|--------------------------------------| | Common/Usual Name: | Medical Laser System | | Classification Name: | Instrument, Surgical, Powered, Laser | | Device Class: | 21 CFR 878.4810 GEX Class II | #### 3. Description of Device The Nobelx Long Pulsed Laser system produces a pulsed beam of coherent near infrared (755nm) light. This beam is directed to the treatment zone by fiber optic coupled to hand-piece. When the beam contacts human tissue, the energy in the beam is absorbed, resulting in a very rapid highly localized temperature increase to the target chromospheres such as Hair removal and Pigment treatment. This increases localized temperature of the chromospheres to smaller particles. By directing the beam onto specific tissue locations, using different hand-pieces, and controlling the treatment fluence, the intensity of the temperature of the target can be varied. The physician can optimize the effect for different applications by controlling the energy of the laser pulse and the spot size of the treatment beam. This equipment is composed of the main body and a hand-piece which is an irradiation device and as an accessory part, protective goggles for protection of the worker. Operation principle: This equipment is controlled by a micro processor interfaced to a LCD touch screen control panel. The computer controls start and stop of the treatment. When the key switch of the system is turned clockwise, the main power will be inputted, which will be conveyed to the hand piece through the control board. Meanwhile, the control board connected to the touch screen is connected to the lamp of the hand-piece and controls the same, and controls the whole system through the data connected to the touch screen control panel. When the switch of the hand piece is pressed, the lamp will laser. {4}------------------------------------------------ #### 4. Indications for Use The Noblex Long Pulse Alexandrite Laser is indicated for stable long-term or permanent hair reduction. Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs regrowing when measured at 6, 9 and 12 months after the completion of a treatment regime. It is used for all skin types (Fitzpatrick I -VI) including tanned skin. It is also indicated for the treatment of benign vascular lesions, benign pigmented lesions and wrinkles. #### 5. Performance Testing The Noblex Long Pulse Alexandrite Laser is required to conform and does conform to the laser performance standards as follows; - 1. EN 60825-1:2007, Safety of laser products-Part 1: Equipment classification and requirements - 2. EN 60601-2-22: 1996 Medical electrical equipment- part 2: particular requirements for the safety of diagnostic and therapeutic laser equipment. - 3. EN 60601-1: 2006 Medical electrical equipment Part 1 : General requirements for basic safety and essential performance - 4. EN 60601-1-2: 2007 Medical electrical equipment Part 1-2 : General requirements for basic safety and essential performance- Collateral standard : Electromagnetic compatibility- requirements and tests #### 6. Substantial Equivalence Substantial Equivalence for the Noblex Long Pulse Alexandrite Laser is based in its similarities in indication for use, design features, operational principles and material composition when compared to the predicate devices cleared under the following submissions: - K034030 - Cynosure Apogee Elite Laser, Manufactured by Cynosure, Inc. - K024335 - GentleLASE Family of Laser Systems, Manufactured by Candela Corporation #### 7. Conclusions Based on the technical comparisons between the Noblex Long Pulse Alexandrite Laser and the predicate lasers above, these devices are Substantially Equivalent.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...