DirectInject is a self-setting, calcium phosphate cement intended to repair neurosurgical burr holes, contiguous craniotomy cuts and other cranial defects not intrinsic to the stability of the bony structure. It is also intended for augmentation or restoration of bony contour in the craniofacial skeleton to include the cranial and zygomatic bones. DirectInject is intended to repair cranial defects with a surface area of 4 cm2 or less. DirectInject is indicated for patients in whom skeletal growth is complete. It can be used in patients with surgically created bone defects.
Device Story
DirectInject is a sterile, calcium phosphate-based bone cement supplied as a dual-paste system in a double-barrel delivery syringe. During surgery, the pastes are expressed through a Mixer-Cannula, forming an injectable, moldable cement that hardens in situ to hydroxyapatite, the primary mineral constituent of bone. Used by neurosurgeons or craniofacial surgeons in clinical settings to fill burr holes, craniotomy cuts, or cranial/zygomatic defects. The device provides a biocompatible, osteoconductive material for contour restoration. It is intended for single-patient use. By hardening to match bone mineral composition, it facilitates integration with surrounding tissue, aiding in the repair of cranial defects.
Clinical Evidence
No clinical data. Evidence consists of bench testing and in-vivo animal studies. Bench testing included chemical characterization (XRD, FTIR, XRF), physical properties (porosity per ASTM D4404-10, D4284-12), and performance metrics (pH, setting time, injectability, compressive strength, shelf life). In-vivo studies in sheep and rabbit models demonstrated stability, osteoconductivity, and integration with surrounding bone tissue via macroscopic and histological evaluation.
Technological Characteristics
Calcium phosphate cement forming hydroxyapatite. Supplied as dual-paste in double-barrel syringe with Mixer-Cannula. Materials evaluated per ISO 10993-1. Bench testing standards: ASTM D4404-10, ASTM D4284-12 (porosity), USP <791> (pH), ICH Q1A(R2) (shelf life). Permanent implant (>30 days). Sterile, single-use.
Indications for Use
Indicated for patients with complete skeletal growth requiring repair of neurosurgical burr holes, contiguous craniotomy cuts, or other cranial defects (≤ 4 cm²) not intrinsic to bony stability, and for augmentation or restoration of craniofacial/zygomatic bone contour. Applicable to surgically created bone defects.
Regulatory Classification
Identification
Methyl methacrylate for cranioplasty (skull repair) is a self-curing acrylic that a surgeon uses to repair a skull defect in a patient. At the time of surgery, the surgeon initiates polymerization of the material and forms it into a plate or other appropriate shape to repair the defect.
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 2, 2015
Stryker Craniomaxillofacial Mr. Nathan Van Sweden Senior Manager, Regulatory Affairs 750 Trade Centre Wav. Suite 200 Portage, Michigan 49002
Re: K143661 Trade/Device Name: DirectInject® Regulation Number: 21 CFR 882.5300 Regulation Name: Methyl Methacrylate for Cranioplasty Regulatory Class: Class II Product Code: GXP Dated: July 31, 2015 Received: August 3, 2015
Dear Mr. Van Sweden,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in
{1}------------------------------------------------
the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Michael J.Hoffmann -S
for Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K143661
Device Name DirectInject
#### Indications for Use (Describe)
DirectInject is a self-setting, calcium phosphate cement intended to repair neurosurgical burr holes, contiguous craniotomy cuts and other cranial defects not intrinsic to the stability of the bony structure. It is also intended for augmentation or restoration of bony contour in the craniofacial skeleton to include the cranial and zygomatic bones. DirectIniect is intended to repair cranial defects with a surface area of 4 cm2 or less.
DirectInject is indicated for patients in whom skeletal growth is complete. It can be used in patients with surgically created bone defects.
Type of Use (Select one or both, as applicable)
| <span></span> | <input checked="True" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------|--------------------------------------------------------------------------------------|
| <span></span> | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
This section provides a summary of 510(k) safety and effectiveness information in accordance with the requirements of 21 CFR 807.92.
- I. SUBMITTER
| 510(k) Owner: | Stryker Leibinger GmbH& Co. KG<br>Boetzinger Strasse 41<br>D-79111 Freiburg, Germany |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter/ Contact<br>Person: | Nathan Van Sweden<br>Stryker Craniomaxillofacial<br>750 Trade Centre Way<br>Portage, MI 49002<br>Phone: 269-389-3419<br>Fax: 877-648-7114 |
| Date prepared: | August 31, 2015 |
| II. DEVICE | |
| Trade Name: | DirectInject® |
| Common or Usual<br>name: | Methyl methacrylate for cranioplasty |
| Classification<br>name: | Methyl methacrylate for cranioplasty (882.5300) |
| Regulatory Class: | Class II |
| Product Code: | GXP |
## III. PREDICATE DEVICE
Primary Predicate: Stryker Injectable Cement – K060763 This predicate has not been subject to a design-related recall.
Secondary Predicate: BoneSource HAC - K021440 This predicate has not been subject to a design-related recall.
{4}------------------------------------------------
#### DEVICE DESCRIPTION IV.
Stryker DirectInject consists of a sterile dual paste system, provided pre-filled in a double barrel delivery syringe system, which is calcium phosphate based. Upon injection through the Mixer-Cannula, the two pastes form a cement paste which is injectable, moldable and biocompatible. The injected cement paste will harden under normal body conditions to form hydroxyapatite, which is the principle mineral constituent of bone. The contents are supplied sterile for single patient use in sizes of 3 cc, 5 cc, and 10 cc.
#### V. INTENDED USE/INDICATIONS FOR USE
DirectInject is a self-setting, calcium phosphate cement intended to repair neurosurgical burr holes, contiguous craniotomy cuts and other cranial defects not intrinsic to the stability of the bony structure. It is also intended for augmentation or restoration of bony contour in the craniofacial skeleton to include the cranial and zygomatic bones. DirectInject is intended to repair cranial defects with a surface area of 4 cm2 or less.
DirectInject is indicated for patients in whom skeletal growth is complete. It can be used in patients with surgically created bone defects.
| | Subject Device | Primary Predicate –<br>K060763 | Secondary Predicate<br>– K021440 |
|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended<br>Use/Indications<br>for Use | DirectInject is a self-<br>setting, calcium<br>phosphate cement<br>intended to repair<br>neurosurgical burr<br>holes, contiguous<br>craniotomy cuts and<br>other cranial defects<br>not intrinsic to the<br>stability of the bony<br>structure. It is also<br>intended for<br>augmentation or<br>restoration of bony<br>contour in the | Stryker® Injectable<br>Cement is a self-setting<br>calcium phosphate<br>cement indicated to fill<br>bony voids or gaps of<br>the skeletal system (i.e.<br>extremities,<br>craniofacial, spine, and<br>pelvic).<br>These defects may be<br>surgically created or<br>osseous defects created<br>from traumatic injury to<br>the bone. The Stryker<br>Injectable Cement is<br>indicated only for bony<br>voids or gaps that are | BoneSource® is a self-<br>setting, calcium<br>phosphate cement<br>intended for use in the<br>repair of neurosurgical<br>burr holes, contiguous<br>craniotomy cuts and<br>other cranial defects<br>as well as in the<br>augmentation or<br>restoration of bony<br>contour in the<br>craniofacial skeleton |
### TABLE 5-1: COMPARISON OF INDICATIONS FOR USE
Traditional 510(k)
Stryker DirectInject
{5}------------------------------------------------
## stryker®
| craniofacial skeleton to<br>include the cranial and<br>zygomatic bones.<br>DirectInject is<br>intended to repair<br>cranial defects with a<br>surface area of 4 cm² or<br>less.<br>DirectInject is<br>indicated for patients<br>in whom skeletal<br>growth is complete. It<br>can be used in patients<br>with surgically created<br>bone defects. | not intrinsic to the<br>stability of the bony<br>structure.<br>Stryker® Injectable<br>Cement cured in situ<br>provides an open<br>void/gap filler that can<br>augment provisional<br>hardware (e.g., K-<br>Wires, plates, screws)<br>to help support bone<br>fragments during the<br>surgical procedure. The<br>cured cement acts only<br>as a temporary support<br>media and is not<br>intended to provide<br>structural support<br>during the healing<br>process.<br>Stryker® Injectable<br>Cement is a self-setting,<br>calcium phosphate<br>cement intended for use<br>in the repair of<br>neurosurgical burr<br>holes, contiguous<br>craniotomy cuts and<br>other cranial defects as<br>well as in the<br>augmentation or<br>restoration of bony<br>contour in the<br>craniofacial<br>skeleton. |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
The Indications for Use statement for the Subject device is not identical to the Primary predicate device Stryker Injectable Cement; the differences do not alter the intended therapeutic use of the device nor do they affect the safety and effectiveness of the device relative to the predicate. Both devices have intended use. The Primary
{6}------------------------------------------------
predicate is indicated for a wider anatomical range of the skeletal system (i.e. extremities, craniofacial, spine, and pelvic). The Indications for use of the Stryker DirectInject falls within the scope of the broader Indications for Use statement of the Primary predicate device.
The Indications for Use statement for the Subject device is not identical to the Secondary predicate device Bonesource HAC; however, the differences do not alter the intended therapeutic use of the device nor do they affect the safety and effectiveness of the device relative to the predicate. Both devices have the same intended use.
#### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE VI. PREDICATE DEVICE
The Stryker DirectInject is compared to its predicate devices for substantial equivalence based on the following criteria:
- A. Principle of Operation
- B. Technological Characteristics
## A. Principle of Operation
The operating principle of the Stryker DirectIniect is to repair neurosurgical burr holes, contiguous craniotomy cuts and other cranial defects. It is also intended for the augmentation or restoration of bony contour in the craniofacial skeleton. The Subject device and Predicate devices have the same principle of operation.
## B. Technological and Operational Characteristics
At a high level, the Subject device and Predicate devices are based on the following technological elements:
- Same operating principle: repair neurosurgical burr holes, contiguous craniotomy cuts and other cranial defects.
- -Same area of application: Craniofacial (Cranial and Zygomatic Bones)
- Same duration of implantation: implant is osteointegrated via normal bone physiological conditions.
- -Similar control mechanism: The Subject device uses a dual syringe with pastes and a Mixer-Cannula, while the predicate devices use powder and liquid bases, a mixing apparatus and syringe (primary predicate).
- -Similar material formulation: Both the Primary and Secondary predicates and the Subject device form Hydroxyapatite cement, however, due to the paste form of the Subject device, a proprietary material formulation is used.
{7}------------------------------------------------
## VII. PERFORMANCE DATA
The following performance data are provided in support of the substantial equivalence determination.
## Biocompatibility Testing
The biocompatibility evaluation for the Stryker DirectInject was conducted in accordance with
- The FDA Blue Book Memorandum #G95-1 "Use of International Standard ISO-10993, 'Biological Evaluation of Medical Devices Part 1: Evaluation and Testing,"" May 1, 1995,
- ISO 10993-1 "Biological Evaluation of Medical Devices Part 1: Evaluation and . Testing Within a Risk Management Process," as recognized by FDA.
- FDA draft guidance "Use of International Standard ISO- 10993, "Biological . Evaluation of Medical Devices Part 1: Evaluation and Testing", issued April 23, 2013
Stryker DirectInject is considered a permanent implant, tissue/bone contacting greater than 30 days. The implant is Hydroxyapatite cement, the same as the Predicate devices.
| Test | Standard |
|------------------------------------------------------------|----------------------------|
| Cytotoxicity | ISO-10993-5 |
| Irritation | ISO-10993-10 & ISO-10993-2 |
| Sensitization | ISO-10993-10 & ISO-10993-2 |
| Acute systemic toxicity | ISO-10993-11 & ISO-10993-2 |
| Genotoxicity: Bacterial reverse mutation | ISO-10993-3 |
| Genotoxicity: Mouse lymphoma Assay | ISO-10993-3 |
| Genotoxicity: Mouse peripheral blood micronucleus<br>study | ISO-10993-3 & ISO-10993-2 |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the word "stryker" in a bold, sans-serif font. The word is all lowercase, and the letters are closely spaced together. There is a registered trademark symbol to the right of the letter "r". The overall impression is one of a strong, modern brand.
| Haemocompatibility | ISO-10993-4 & ASTM F756 |
|-------------------------|----------------------------|
| Sub-Chronic Toxicity | ISO-10993-11 & ISO-10993-2 |
| Chronic toxicity | ISO-10993-11 & ISO-10993-2 |
| Neurotoxicity (8 weeks) | ISO-10993-6 & ISO-10993-2 |
### Performance Bench Testing
The following performance bench tests were completed per the requirements of 'Class II Special Controls Guidance Document: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA '.
| Characteristic | Test | Standard |
|---------------------------|--------------------------------------------|-----------------------------------------------------|
| Chemical Characterization | X-ray diffraction | International Centre for<br>Diffraction Data (ICDD) |
| | Fourier Transform Infrared<br>Spectroscopy | N/A |
| | X-ray Fluorescence | N/A |
| Physical Properties | Porosity | ASTM D4404-10, ASTM<br>D4284-12 |
| Performance | pH testing | USP <791> |
| | Setting time | N/A |
| | Dimensional stability | N/A |
| | Setting reaction temperature | N/A |
| | Injectability force | N/A |
#### TABLE 5-3 PERFORMANCE TESTING FOR STRYKER DIRECTINJECT
In addition to the test methods identified above, the following tests were performed to address risk identified or design requirements.
| Characteristic | Test | Standard |
|----------------|-------------------------------|-------------|
| Performance | Compressive Strength | N/A |
| Performance | Max supported defect strength | N/A |
| Performance | Shelf life Assessment | ICH Q1A(R2) |
| TABLE 5-4 PERFORMANCE TESTING FOR STRYKER DIRECTINJECT | |
|--------------------------------------------------------|--|
| | |
{9}------------------------------------------------
## In-Vivo Testing
The following local effects in-vivo tests were performed as per the requirements of ISO-10993-2 and ISO-10993-6. The in-vivo local effects of Stryker DirectInject were evaluated in a sheep and rabbit model, by macroscopic and histological evaluation of the tissue in the treatment areas. These studies demonstrated that Stryker DirectInject is stable, osteoconductive, and integrates with the bone tissue surrounding the defect site.
## Clinical Testing
Clinical testing was not required as a basis for substantial equivalence.
## VIII. CONCLUSIONS
The results of the non-clinical data demonstrate the Stryker DirectInject will perform as intended in the specified use conditions. According to the comparison based on the requirements of 21 CFR 807.87 and the information provided herein, it is concluded that the information included in this submission supports substantial equivalence.
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.