Vector PTA Balloon Dilatation Catheter

K143652 · Vector Corporation, LLC · LIT · Feb 26, 2015 · Cardiovascular

Device Facts

Record IDK143652
Device NameVector PTA Balloon Dilatation Catheter
ApplicantVector Corporation, LLC
Product CodeLIT · Cardiovascular
Decision DateFeb 26, 2015
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1250
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Vector™ PTA Balloon Dilatation Catheter is recommended for Percutaneous Transluminal Angioplasty (PTA) of the femoral, iliac, popliteal, and renal arteries and for the treatment of obstructive lesions of native or synthetic arterio-venous dialysis fistulae. The Vector Catheter is not indicated for use in the coronary arteries or for placement of or post dilatation of stents.

Device Story

High-pressure, non-compliant, coaxial dual-lumen over-the-wire balloon dilatation catheter; used for peripheral artery angioplasty and dialysis fistula treatment. Input: guide wires up to 0.035 inches. Operation: physician advances catheter to stenosis; inflation via outer lumen expands balloon to dilate lesion; radio-opaque marker bands facilitate fluoroscopic visualization. Output: mechanical dilation of vessel/fistula. Used in clinical settings by physicians. Benefits: restores vessel patency; improves blood flow in peripheral vasculature and dialysis access sites.

Clinical Evidence

Bench testing only. Performance testing included rated burst pressure, balloon compliance, diameter sizing, inflation/deflation times, repeat inflation, removal/refold/reuse, flexibility/kink, crossing profile, protector removal, guide wire/introducer compatibility, tensile strength, fluoroscopic visibility, torque strength, and catheter leak. All devices met performance specifications. Biocompatibility confirmed per ISO 10993-1.

Technological Characteristics

Coaxial dual-lumen over-the-wire catheter; Parylene-coated high-pressure composite balloon; 30 atm rated burst pressure; 5-10 mm balloon diameters; 50-135 cm lengths; 6-7F introducer compatibility. Features radio-opaque marker bands. Sterilized via ethylene oxide (SAL 10-6). Biocompatible materials per ISO 10993-1. No natural rubber latex.

Indications for Use

Indicated for patients requiring percutaneous transluminal angioplasty (PTA) of femoral, iliac, popliteal, and renal arteries, or treatment of obstructive lesions in native or synthetic arterio-venous dialysis fistulae. Contraindicated for coronary artery use and stent placement or post-dilation.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health and Human Services. The logo features a stylized image of an eagle with three human profiles incorporated into its design. The profiles are facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 February 26, 2015 Vector Corporation LLC c/o Ms. Laurie Lewandowski Honkanen Consulting, Inc. 738 Saddle Wood Drive Eagan, MN 55123 Re: K143652 Trade/Device Name: Vector PTA Balloon Dilatation Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: LIT Dated: January 30, 2015 Received: February 2, 2015 Dear Ms. Lewandowski: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {2}------------------------------------------------ device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, # Kenneth J. Cavanaugh -S for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ # Indications for Use 510(k) Number (if known) K143652 #### Device Name Vector™ PTA Balloon Dilatation Catheter #### Indications for Use (Describe) The Vector™ PTA Balloon Dilatation Catheter is recommended for Percutaneous Transluminal Angioplasty (PTA) of the femoral, iliac, popliteal, and renal arteries and for the teatment of obstructive or synthetic arterio-venous dialysis fistulae. The Vector Catheter is not indicated for use in the coronary arteries or for placement of or post dilation of stents. Type of Use (Select one or both, as applicable) X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. #### FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) l Health (CDRH) (*Signature*) {4}------------------------------------------------ This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {5}------------------------------------------------ # 510(k) Summary Vector Corporation ## Primary Contact Person Laurie Lewandowski, Consultant Honkanen Consulting, Inc. 738 Saddle Wood Drive Eagan, MN 55123 Phone: 612-770-4038 Email: lalew207@gmail.com ## Sponsor's Contact Jay Sturm Vector Corporation LLC 1270 Ave of Americas. Suite 302 New York, NY 10020 Phone: 646-833-3700 Email: jsturm(a)volcanocap.com # Device Name Trade Name: Vector™ Common Name: Percutaneous Catheter Classification Name: Percutaneous Catheter Classification: II Product Code: LIT Regulation Number: 21 CFR 870.1250 ## Predicate Device The device is substantially equivalent to the r4 Vascular, Inc. Vector4™ PTA Balloon Dilatation Catheter (K131329). # Indications for Use/Intended Use The Vector™ PTA Balloon Dilatation Catheter is recommended for Percutaneous Transluminal Angioplasty (PTA) of the femoral, iliac, popliteal, and renal arteries and for the treatment of obstructive lesions of native or synthetic arterio-venous dialysis fistulae. The Vector" Catheter is not indicated for use in the coronary arteries or for placement of or post dilatation of stents. {6}------------------------------------------------ # Device Description The Vector™ PTA Balloon Dilatation Catheter is a high pressure, non-compliant balloon catheter. It is a coaxial dual lumen over-the-wire balloon dilatation catheter. The central lumen accommodates guide wires up to the 0.035 inches in diameter while the outer lumen is the inflation lumen for the balloon. The balloon is a high pressure composite balloon with radio-opaque marker bands to help visualize the balloon placement within the vasculature during both placement and inflation. The balloon's rated burst pressure is 30 atm. # Technological Characteristics The Vector™ PTA Balloon Dilatation Catheter was cleared under K131329. The fundamental scientific technology and technological characteristics for the Vector™ PTA Balloon Catheter remain the same as Vector4™ PTA Balloon Dilatation Catheter. Vector™ is a coaxial dual lumen over-the-wire balloon dilatation catheter. The central lumen accommodates guide wires up to 0.035 inches in diameter while the outer lumen is the inflation lumen for the balloon. The catheter includes an atraumatic tip to ease advancement of the catheter to and through the stenosis. The balloon is a high pressure, Parylene coated, composite balloon with marker bands located at the proximal and distal end of the balloon working length and at the distal tip to help visualize placement within the lesion or stenosis. The balloon's rated burst pressure is 30 atm. The catheter is supplied sterile and non-pyrogenic. Different balloon sizes (5 mm to 10 mm) and catheter lengths (50 to 135 cm) are available. The proximal end of the catheter contains a hub with two female luer locks. The device is compatible with 6-7F introducers. The introducer compatibility for each device size is listed on the device label. The materials used in the construction are biocompatible and not made with natural rubber latex. The modifications to the Vector™ PTA Balloon Dilatation Catheter are as follows. - 1) Removal of radiopaque coating, BaSO4 tip doping and addition of marker bands - 2) Angled the balloon fibers and modified the fiber denier - 3) Updates to the IFU for company name, product name, use of contrast - 4) Increased flow for reduced inflation / deflation times - 5) Increased product robustness for balloon bonds and reduction in introducer withdrawal force - 6) The 6 Fr balloon protector ID was increased slightly. There are no changes in the balloon catheter dimensions or significant changes in weight with these modifications. {7}------------------------------------------------ # Performance Data Based on the risk analysis of the modified device, and testing was conducted on the modified Vector™ PTA Balloon Dilatation Catheter. The testing performed on the modified devices was: Rated Burst Pressure, Balloon Compliance, Balloon Diameter Sizing, Inflation/deflation times. Repeat Inflation. Removal/Refold & Reuse. Flexibility / Kink, Crossing Profile, Balloon Protector Removal, Guide wire/Introducer Compatibility, Tensile Strength, Visibility under Fluoro-imaging, Torque Strength and Catheter Leak. All devices met the performance specifications. The device is sterilized by ethylene oxide to an SAL 10-6 level and the device is biocompatible meeting the requirements of ISO 10993-1. The bench testing demonstrated that the device meets specifications indicating that the device is as safe and effective as the predicate device. # Substantial Equivalence The modified Vector" PTA Balloon Catheter is substantially equivalent to the Vector4™ PTA catheter (K131329). They have the identical intended use, and treat the same target population. Both devices are intended to treat peripheral arteries. The manner in accessing and treating lesions in all these arteries is identical. The devices are similar in design with changes as described. # Conclusion The modified Vector" PTA Balloon Catheters and the predicate Vector4™ PTA Balloon Catheters have identical intended use and similar technological characteristics and are therefore substantially equivalent. Bench testing of the Vector™ PTA Balloon Dilatation Catheters demonstrated that all product specifications were met. These tests demonstrate that the Vector" PTA Balloon Dilatation Catheters are as safe and effective as the predicate device.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...