Each non-absorbable suture is used as a surgical suture to support wound closure/soft tissue approximation, with the following indications: Except Steel, each non-absorbable suture is indicated for use in general soft tissue approximation and/or ligation including use in cardiovascular, ophthalmic and neurological procedures*. *MOPYLEN is also indicated for microsurgery * POLYESTER is also indicated for orthopaedic surgery *RESOLON is also indicated for microsurgery Steel wire is indicated for use in abdominal wound closure, hernia repair, sternal closure and orthopaedic procedures including cerclage and tendon repair.
Device Story
Non-absorbable surgical sutures; used for soft tissue approximation/ligation; available in monofilament, twisted, or braided forms; supplied sterile with or without needles. Materials include polypropylene, PVDF, silk, polyester, polyamide, and stainless steel. Operated by surgeons in clinical settings (OR). Sutures provide mechanical support to wound sites; facilitate healing by holding tissue edges together. Benefits include secure wound closure and reduced risk of dehiscence.
Clinical Evidence
Bench testing only. Compliance with USP monographs for non-absorbable sutures (diameter, tensile strength, needle attachment). Biocompatibility testing per ISO 10993-1, 10993-10, and 10993-6. Rabbit pyrogen and LAL testing performed. Real-time stability testing supports shelf life.
Technological Characteristics
Materials: Polypropylene, PVDF, Silk (Bombyx Mori L.), Polyester (PET), Polyamide 6-6.6, Stainless Steel (AISI 316L/ASTM F138-13). Forms: Monofilament, twisted, or braided. Sizes: USP 11-0 to 7. Sterilization: Ethylene Oxide. Meets USP requirements for synthetic non-absorbable sutures.
Indications for Use
Indicated for general soft tissue approximation/ligation in cardiovascular, ophthalmic, and neurological procedures. MOPYLEN/RESOLON also for microsurgery; POLYESTER also for orthopaedic surgery. Steel indicated for abdominal wound closure, hernia repair, sternal closure, and orthopaedic procedures (cerclage/tendon repair).
Regulatory Classification
Identification
Nonabsorbable polypropylene surgical suture is a monofilament, nonabsorbable, sterile, flexible thread prepared from long-chain polyolefin polymer known as polypropylene and is indicated for use in soft tissue approximation. The polypropylene surgical suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be undyed or dyed with an FDA approved color additive; and the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image contains the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol is a stylized representation of a human figure, with three faces overlapping to create a sense of depth and community.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
October 29, 2015
Resorba Medical GmbH Mr. Karl-Josef Beck Quality Assurance/Regulatory Affairs Manager Am Flachmoor 16 Nuremberg, 90475 Germany
Re: K143582
Trade/Device Name: MOPYLEN, RESOPREN, SILK, POLYESTER, NYLON, RESOLON, POLY AMIDE PSEUDO, STEEL Regulation Number: 21 CFR 878.5010 Regulation Name: Nonabsorbable polypropylene surgical suture Regulatory Class: Class II Product Code: GAW, MXW, GAP, GAT, GAR, GAQ Dated: September 2, 2015 Received: September 8, 2015
Dear Mr. Beck:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{1}------------------------------------------------
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## David Krause -S
- for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a PDF portfolio icon with the Adobe logo on the cover. Below the icon, there is a message that reads, "For the best experience, open this PDF portfolio in Acrobat X or Adobe Reader X, or later." A button labeled "Get Adobe Reader Now!" is located beneath the message. The image is designed to encourage users to open the PDF portfolio using the recommended software for optimal viewing.
{3}------------------------------------------------
## 5.0 TRADITIONAL 510(K) SUMMARY
| Submitted by: | Resorba Medical GmbH<br>Am Flachmoor 16<br>90475 Nuremberg<br>Germany<br>Tel: +49 9128-9115-0<br>Fax: +49 9128-9115-10 |
|---------------|------------------------------------------------------------------------------------------------------------------------|
|---------------|------------------------------------------------------------------------------------------------------------------------|
Contact Person: Karl-Josef Beck
Date of Summary: 01 September 2015
| Trade Name | Common Name | Classification<br>Name | Regulation<br>Classification | Product<br>Code | Class<br>of<br>device | Predicate<br>device |
|------------|-----------------------------------------------------------|---------------------------------------------------------------------|------------------------------|-----------------|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| MOPYLEN® | Non absorbable<br>suture:<br>Polypropylene | Nonabsorbable<br>polypropylene<br>surgical suture | 21 CFR §878.5010 | GAW | II | Premilene,<br>Aesculap,<br>K980703 |
| RESOPREN® | Non absorbable<br>suture:<br>Polyvinylidene<br>diflouride | Nonabsorbable<br>polypropylene<br>surgical suture | 21 CFR §878.5010 | MXW | II | Pronova,<br>Ethicon,<br>K001625 |
| SILK | Non absorbable<br>suture: Silk | Natural<br>nonabsorbable<br>silk surgical<br>suture | 21 CFR §878.5030 | GAP | II | Silkam,<br>Aesculap,<br>K990089<br>Secondary<br>predicate1:<br>Riversilk<br>Riverpoint<br>Medical<br>K100006<br>Secondary<br>predicate2:<br>WG-Surgical<br>Foosin<br>Medical<br>supplies Inc.<br>K080684 |
| POLYESTER | Non absorbable<br>suture: Polyester | Nonabsorbable<br>poly(ethylene<br>terephthalate)<br>surgical suture | 21 CFR §878.5000 | GAT | II | Cottony II,<br>Tevdek<br>K021019 |
| NYLON | Non absorbable<br>suture: Polyamide | Nonabsorbable<br>polyamide<br>surgical suture | 21 CFR §878.5020 | GAR | II | Dermalon,<br>Davis&Geck,<br>K081582 |
{4}------------------------------------------------
| RESOLON® | Non absorbable<br>suture: Polyamide | Nonabsorbable<br>polyamide<br>surgical suture | 21 CFR §878.5020 | GAR | II | Cardionyl,<br>Promedica<br>K913102 |
|---------------------|-------------------------------------|-----------------------------------------------|------------------|-----|----|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| POLYAMIDE<br>PSEUDO | Non absorbable<br>suture: Polyamide | Nonabsorbable<br>polyamide<br>surgical suture | 21 CFR §878.5020 | GAR | II | Supramid,<br>S Jackson,<br>K904052<br>Secondary<br>predicate:<br>Dermalon,<br>Davis&Geck,<br>K981582 |
| STAINLESS<br>STEEL | Non absorbable<br>suture: Steel | Stainless Steel<br>suture | 21 CFR §878.4495 | GAQ | II | Flexon,<br>Davis&Geck,<br>K955723<br>Secondary<br>predicate1:<br>FerroFibre<br>Pontis<br>Orthopaedics<br>LLC<br>K140127<br>Secondary<br>predicate 2:<br>Stainless steel<br>medical<br>suture<br>CP Medical<br>inc.<br>K030351 |
{5}------------------------------------------------
## Device Description: The subject devices are non-absorbable surgical sutures. They are available undyed and dyed. The sutures are supplied sterile, in monofilament, twisted and braided forms in sizes USP 11-0 to 7 (depending on suture type), with or without needles in a variety of cut lengths.
| Description | Primary<br>material | Form | Colour and Colourant<br>for primary material | Coating | Sizes (USP) |
|---------------------|------------------------------------------------------|-----------------------------------------------------------|----------------------------------------------|----------|-------------|
| MOPYLEN | Polypropylene<br>(PP) | Monofilament | Blue<br>(Phthalocyaninato{2})<br>copper | None | 10-0 to 2 |
| RESOPREN | Polyvinylidene<br>diflouride<br>(PVDF) | Monofilament | Blue<br>(Phthalocyaninato{2})<br>copper | None | 8-0 to1 |
| SILK | Silk<br>Bombyx Mori L. | Braided | Black<br>Logwood Extract<br>White<br>Undyed | Silicone | 10-0 to 6 |
| POLYESTER | Polyester (PET) | Braided | White<br>Undyed<br>Green<br>D&C Green | None | 7-0 to 7 |
| NYLON | Polyamide<br>6-6.6(PA) | Monofilament | White<br>Undyed<br>Black<br>Logwood Extract | None | 11.0 to 2 |
| RESOLON | Polyamide<br>6-6.6(PA) | Monofilament | Blue<br>(Phthalocyaninato{2})<br>copper | None | 11-0 to 2 |
| POLYAMIDE<br>PSEUDO | Polyamide<br>6-6.6(PA)<br>With polyamide<br>6 sheath | Monofilament<br>or twisted<br>with<br>polyamide<br>sheath | White<br>Undyed | None | 7-0 to 7 |
| STEEL | Stainless Steel<br>1.4404 - AISI<br>316 L | Monofilament<br>or twisted | NA | None | 6.0 to 7 |
{6}------------------------------------------------
| Indication for Use: | Each non-absorbable suture is used as a surgical suture to support wound<br>closure/soft tissue approximation, with the following indications: |
|----------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Except Steel, each non-absorbable suture is indicated for use in general soft<br>tissue approximation and/or ligation including use in cardiovascular,<br>ophthalmic and neurological procedures*. |
| | *MOPYLEN is also indicated for microsurgery<br>* POLYESTER is also indicated for orthopaedic surgery<br>*RESOLON is also indicated for microsurgery |
| | Steel wire is indicated for use in abdominal wound closure, hernia repair,<br>sternal closure and orthopaedic procedures including cerclage and tendon<br>repair. |
| Substantial Equivalence: | Each Non-absorbable suture has the same intended use and similar design,<br>materials, labeling, performance characteristics to their predicate device. |
| Technological<br>characteristics | Each Non-absorbable suture is substantially equivalent to the predicate<br>device listed when compared to the technological characteristics and are<br>supplied sterile for single use. All meet USP requirements. |
Comparison of technological characteristics to predicate device:
| | Intended use | Material | Design | Performance<br>Diameter,<br>Needle<br>attachment<br>and Tensile<br>strength | Sterilisation<br>method |
|-----------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|-------------------------|
| MOPYLEN | Intended for use<br>in general soft<br>tissue<br>approximation<br>and/or ligation<br>including use in<br>cardiovascular,<br>microsurgery,<br>ophthalmic and<br>neurological<br>procedures. | Material:<br>Polypropylene<br>Coating: None<br>Dye:<br>(Phthalocyani<br>nato{2})<br>copper | Non-absorbable,<br>monofilament,<br>dyed, uncoated,<br>provided sterile<br>with or without<br>needles | Meets U.S.P | Ethylene<br>Oxide |
| Predicate | Same | Same | Same | Same | Same |
{7}------------------------------------------------
| RESOPREN | Intended for use<br>in general soft<br>tissue<br>approximation<br>and/or ligation<br>including use in<br>cardiovascular,<br>ophthalmic and<br>neurological<br>procedures. | Material:<br>Polyvinylidene<br>diflouride<br>Coating: None<br>Dye:<br>(Phthalocyani<br>nato{2})<br>copper | Non-absorbable,<br>monofilament,<br>dyed, uncoated,<br>provided sterile<br>with or without<br>needles | Meets U.S.P | Ethylene<br>Oxide |
|--------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|-------------|-------------------|
| Predicate | Same | Same | Same | Same | Same |
| SILK | Intended for use<br>in general soft<br>tissue<br>approximation<br>and/or ligation<br>including use in<br>cardiovascular,<br>ophthalmic and<br>neurological<br>procedures. | Material:<br>Silk Bombyx<br>Mori<br>Coating:<br>Silicone<br>Dye:<br>Logwood<br>Extract | Non-absorbable,<br>multifilament,<br>dyed and undyed,<br>coated, provided<br>sterile with or<br>without needles | Meets U.S.P | Ethylene<br>Oxide |
| Predicate | Same | Same | Same | Same | Gamma |
| Each<br>Secondary<br>predicate | Same | Same | Same | Same | Same |
| POLYESTER | Intended for use<br>in general soft<br>tissue<br>approximation<br>and/or ligation<br>including use in<br>cardiovascular,<br>orthopaedic,<br>ophthalmic and<br>neurological<br>procedures. | Material:<br>Polyester<br>Coating: None<br>Dye:<br>D&C Green | Non-absorbable,<br>multifilament,<br>dyed and undyed,<br>uncoated,<br>provided sterile<br>with or without<br>needles | Meets U.S.P | Ethylene<br>Oxide |
| Predicate | Same | Same | Same | Same | Same |
{8}------------------------------------------------
| NYLON | Intended for use<br>in general soft<br>tissue<br>approximation<br>and/or ligation<br>including use in<br>cardiovascular,<br>ophthalmic and<br>neurological<br>procedures | Material:<br>Polyamide<br>Coating: None<br>Dye:<br>Logwood<br>Extract | Non-absorbable,<br>monofilament,<br>dyed and undyed,<br>uncoated,<br>provided sterile<br>with or without<br>needles | Meets U.S.P | Ethylene<br>Oxide |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|-------------|-------------------|
| Predicate | Same | Same | Same | Same | Same |
| RESOLON | Intended for use<br>in general soft<br>tissue<br>approximation<br>and/or ligation<br>including use in<br>cardiovascular,<br>ophthalmic,<br>microsurgery and<br>neurological<br>procedures. | Material:<br>Polyamide<br>Coating: None<br>Dye:<br>(Phthalocyani<br>nato{2})<br>copper | Non-absorbable,<br>monofilament,<br>dyed, uncoated,<br>provided sterile<br>with or without<br>needles | Meets U.S.P | Ethylene<br>Oxide |
| Predicate | Same | Same | Same | Same | Same |
| POLYAMIDE<br>PSEUDO | Intended for use<br>in general soft<br>tissue<br>approximation<br>and/or ligation<br>including use in<br>cardiovascular,<br>ophthalmic and<br>neurological<br>procedures. | Material:<br>Polyamide<br>with<br>polyamide<br>sheath<br>Coating: None<br>Dye: Undyed | Non-absorbable,<br>multifilament<br>with sheath or<br>monofilament<br>undyed,<br>uncoated,<br>provided sterile<br>with or without<br>needles | Meets U.S.P | Ethylene<br>Oxide |
| Predicate | Intended for use<br>in general soft<br>tissue<br>approximation<br>and/or ligation<br>including use in<br>ophthalmic<br>procedures | Same | Same | Same | Same |
| Secondary<br>Predicate | Same | Same | Similar. | Same | Same |
{9}------------------------------------------------
| | | | sheath. | | |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|-----------------------------------------------------------------------------------------------------------------------|-------------|-------------------|
| STAINLESS<br>STEEL | Indicated for use<br>in abdominal<br>wound closure,<br>hernia repair,<br>sternal closure<br>and orthopaedic<br>procedures<br>including cerclage<br>and tendon repair. | Steel 316L | Non-absorbable,<br>monofilament or<br>twisted, undyed,<br>uncoated,<br>provided sterile<br>with or without<br>needles | Meets U.S.P | Ethylene<br>Oxide |
| Predicate | Same | Same | Same | Same | Gamma |
| Each<br>Secondary<br>predicate | Same | Same | Same | Same | Same |
Performance Testing Summary:
As per the FDA's Class II Special Control Guidance Document for Surgical Sutures, the devices were subjected to the requirements of the United States Pharmacopeia (U.S.P) monograph for Synthetic Non-absorbable Sutures. Testing included:
- . Diameter <861>
- Tensile strength <881>
- Needle attachment <871>
Biocompatibility evaluation has been performed in accordance with ISO 10993-1, sensitization according to ISO 10993-10, implantation according to ISO 10993-6, data has also been leveraged from the supplier for Mopylen and Resopren. Rabbit pyrogen and LAL testing has successfully been performed on each subject device. Steel suture complies with ASTM F138-13. Real-time stability testing has been performed to support shelf life. With this data, an established history of use in non-US markets and results of performance testing it is demonstrated that each Non-absorbable suture meets the current performance requirements for Non-absorbable sutures and that it is substantially equivalent to each corresponding predicate device listed.
Conclusion Based on the information provided within this 510(k) submission, Resorba Medical GmbH concludes that the proposed suture products are substantially equivalent to each corresponding predicate device listed.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.