← Product Code [IYE](/productcode/IYE) · K143560

# CygneX I System (K143560)

_Accuthera, Inc. · IYE · Oct 5, 2015 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K143560

## Device Facts

- **Applicant:** Accuthera, Inc.
- **Product Code:** [IYE](/productcode/IYE.md)
- **Decision Date:** Oct 5, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.5050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Therapeutic

## Indications for Use

The CygneX I System is a medical charged-particle radiation therapy system accelerating electrons intended to provide photon radiation for treatment.

## Device Story

CygneX I System provides stereotactic radiosurgery and precision radiotherapy. Inputs: CT data for treatment planning. Operation: Linear accelerator (6 MV) mounted on 6-axis robotic manipulator delivers focused X-ray beams from multiple nodes to target lesions/tumors; minimizes healthy tissue exposure. Surface-guided monitoring (e.g., AlignRT PLUS) and mobile C-arm/fluoroscopy (e.g., Philips Veradius Neo) confirm target position. Output: Controlled photon radiation dose. Used in clinical settings by trained professionals. Physician approves treatment plan generated by CygPlan I subsystem. Benefits: Non-invasive treatment of brain/BOS conditions with high precision.

## Clinical Evidence

Bench testing only. Performance verified via non-clinical testing per FDA Recognized Consensus Standards, including essential performance, basic safety, electromagnetic compatibility, functional performance, and software verification/validation. No clinical data presented.

## Technological Characteristics

6 MV standing wave linear accelerator; 6-axis robotic manipulator; variable secondary collimator (hexagonal fields 5-30mm at 60 SAD); 300 cGy/min max dose rate; software-based treatment planning (CygPlan I); non-isocentric design; safety interlocks and emergency stops included.

## Regulatory Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

## Predicate Devices

- Accuray CyberKnife System ([K984563](/device/K984563.md))

## Reference Devices

- AlignRT PLUS
- Philips Veradius Neo System ([K133819](/device/K133819.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

October 5, 2015

Accuthera, Inc. % Mr. Robert Morton CEO Ouality and Regulatory Services, Inc. 1244 Fairway Valley Court LINCOLN CA 95648

Re: K143560 Trade/Device Name: CygneX I System Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: September 3, 2015 Received: September 8, 2015

Dear Mr. Morton:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{1}------------------------------------------------

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

Michael D'Hara

For

Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K143560

Device Name CygneX I System

Indications for Use (Describe)

The CygneX I System is intended to provide treatment planning and surface-guided stereotactic radiosurgery and precision radiotherapy for lesions (e.g. atteriovenous malformations), tumors and conditions of the brain and base of skull (BOS).

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## ACCUTHERA

### 510(k) Summary

The following information is provided following the format of 21 CFR 807.92.

Submitter: Accuthera Inc. 2-8-22 Kurigi, Asao-ku, Kawasaki-shi Kanagawa, 215-0033 Japan

|        | Contact Name: Eiji Tanabe, Ph.D. |
|--------|----------------------------------|
|        | President & CEO                  |
| Phone: | +81-44-980-1511                  |
| Fax:   | +81-44-980-1522                  |
| Email: | etanabe@accuthera.com            |

Date Summary was prepared: 10 December 2014

#### Name of the Device:

| Trade/Proprietary Name: | CygneX I System                                   |
|-------------------------|---------------------------------------------------|
| Common or Usual Name:   | Medical Linear Accelerator                        |
| Classification Name:    | Medical charged-particle radiation therapy system |
|                         | 21 CFR 892.5050, Class II                         |
|                         | Product Code: IYE                                 |

Predicate Device: Accuray CyberKnife System, K984563

### Description of Device:

The CygneX I System consists of therapeutic CygneX I Treatment Delivery Subsystem and the dedicated CygPlan I Treatment Planning Subsystem. The CygneX I Treatment Delivery Subsystem consists mainly of a Therapeutic Xray Generation Part, 6-Axis Manipulator Part, and X-ray Delivery Control Console Part. A linear accelerator with a variable collimator is incorporated in the X-ray head, the main portion of the Therapeutic X-ray Generation Part, which generates a very fine, highly- energized (6 MV) and controlled dose of X-ray beams. The variable secondary collimator is capable of delivering regular hexagonal fields, with the distances between the two opposite sides

{4}------------------------------------------------

# ACCUTHERA

Accuthera Inc. Head Office 2-8-22 Kurigi, Asao-ku, Kawasaki-Shi, 215-0033 Japa TFI : +81-44-980-1511 FAX: +81-44-980-1522 E-mail: info@accuthera.com URL: http://www.accuthera.com

from 5mm to 30 mm, at 60 SAD. The CygPlan I Treatment Planning Subsystem provides the capability for forward and inverse planning from CT data. It provides a plan after physician approval, to the CygneX I Treatment Delivery Subsystem to deliver the treatment from the selected treatment nodes.

### Statement of Intended Use:

The CygneX I System is a medical charged-particle radiation therapy system accelerating electrons intended to provide photon radiation for treatment.

## Statement of Indications For Use:

The CygneX I System is intended to provide treatment planning and surfaceguided stereotactic radiosurgery and precision radiotherapy for lesions (e.g. arteriovenous malformations), tumors and conditions of the brain and base of skull (BOS).

## Summary of the Technological Characteristics:

The CygneX I System has same Intended Use, principles of operation, and major technological characteristics and a similar Indication For Use as the predicate device, the CyberKnife System For Stereotactic Radiosurgery/Radiotherapy, K984563, manufactured by Accuray, Inc.

A comparison table summarizing the similarities and differences between the CygneX I System and the predicate device is provided in Table ES-1.

{5}------------------------------------------------

| Feature                                                  | Accuray CyberKnife System For<br>Stereotactic<br>Radiosurgery/Radiotherapy<br>K984563                                                                                                                                                                                                                                                                                                                                                                | Accuthera CygneX I System<br>(CygneX I with CygPlan I)                                                                                                                                                                                                                                                                                                                                                                                          | Consideration of the<br>difference(s)                                                                                                                                 |
|----------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use                                             | The CyberKnife System (K984563) is a medical charged-particle radiation<br>therapy system accelerating<br>electrons intended to provide photon<br>radiation for treatment.                                                                                                                                                                                                                                                                           | The CygneX I System is a medical<br>charged-particle radiation therapy<br>system accelerating electrons intended<br>to provide photon radiation for treatment.                                                                                                                                                                                                                                                                                  | Same                                                                                                                                                                  |
| Indication for use                                       | To provide treatment planning and<br>image-guided stereotactic<br>radiosurgery and precision<br>radiotherapy for lesions (e.g.<br>arteriovenous malformations), tumors<br>and conditions of the brain, base of<br>skull (BOS), cervico-thoracic spine<br>(CTS), head and neck.                                                                                                                                                                       | The System is intended to provide<br>treatment planning and surface-guided<br>stereotactic radiosurgery and precision<br>radiotherapy for lesions (e.g.<br>arteriovenous malformations), tumors<br>and conditions of the brain and base of<br>skull (BOS).                                                                                                                                                                                      | No, but while it is less, it is<br>included by the predicate.                                                                                                         |
| Principles of<br>Radiation<br>delivery                   | The System is a type of stereotactic<br>radiotherapy, a non-invasive<br>treatment in which high doses of<br>focused radiation beams are<br>delivered from multiple locations<br>outside of the body to destroy a<br>tumor or lesion within the body. This<br>minimizes radiation exposure to<br>healthy tissue surrounding the tumor.<br>The system uses a linear accelerator<br>(Linac), producing the radiation while<br>mounted on a robotic arm. | The System is a type of stereotactic<br>radiotherapy, a non-invasive treatment in<br>which high doses of focused radiation<br>beams are delivered from multiple<br>locations outside of the body to treat a<br>tumor or lesion within the body. This<br>minimizes radiation exposure to healthy<br>tissue surrounding the tumor.<br>The system uses a linear accelerator<br>(Linac), producing the radiation while<br>mounted on a robotic arm. | Same                                                                                                                                                                  |
| Patient<br>position<br>monitoring                        | The predicate incorporates a<br>fluoroscopic imaging device.                                                                                                                                                                                                                                                                                                                                                                                         | CygneX I uses commercially available<br>(510K cleared) positioning device for<br>monitoring of the patient's body surface<br>movement, such as AlignRT PLUS and a<br>mobile C-Arm imaging or fluoroscopic<br>system be used to confirm the target<br>position, such as the Philips Veradius<br>Neo System, K133819, or equivalent.                                                                                                              | No, but same safety can be<br>achieved by confirmation of the<br>patient position using a device<br>which can image the position of<br>the target inside the patient. |
| Technical characteristic                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                       |
| Single dose<br>and<br>fractionated<br>treatment          | YES                                                                                                                                                                                                                                                                                                                                                                                                                                                  | YES                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                  |
| X-band<br>Accelerator                                    | YES                                                                                                                                                                                                                                                                                                                                                                                                                                                  | YES                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                  |
| X-ray energy                                             | 6MV(Standing wave Linac)                                                                                                                                                                                                                                                                                                                                                                                                                             | 6MV(Standing wave Linac)                                                                                                                                                                                                                                                                                                                                                                                                                        | Same                                                                                                                                                                  |
| Dose rate                                                | 300 cGy/min (maximum BUILD UP)                                                                                                                                                                                                                                                                                                                                                                                                                       | 300 cGy/min (maximum BUILD UP)                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                                                                  |
| Isocenter<br>floor height<br>SAD                         | 127cm (nominal isocenter, system is<br>not isocentric)<br>80cm                                                                                                                                                                                                                                                                                                                                                                                       | 130cm (nominal isocenter, system is not<br>isocentric)<br>60cm                                                                                                                                                                                                                                                                                                                                                                                  | No, but it is only 3 cm higher.<br>No, but it still treats the target to<br>the prescribed dose.                                                                      |
| Source Target<br>positioning                             | Six-axis manipulator                                                                                                                                                                                                                                                                                                                                                                                                                                 | Six-axis manipulator                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                                  |
| Patient<br>Positioning<br>device<br>(Treatment<br>Couch) | Stationary/Adjustable                                                                                                                                                                                                                                                                                                                                                                                                                                | CygneX I System does not include a<br>Treatment Couch. It uses commercially<br>available Adjustable Treatment Couch<br>which complies with conditions specified<br>by Accuthera Inc.                                                                                                                                                                                                                                                            | No, but same level of safety &<br>effectiveness can be achieved<br>by using a specified 510K<br>cleared treatment couch.                                              |
| Mechanical<br>Isocenter<br>accuracy                      | Less than 0.05cm radius                                                                                                                                                                                                                                                                                                                                                                                                                              | Sphere less than 0.05cm radius                                                                                                                                                                                                                                                                                                                                                                                                                  | Same                                                                                                                                                                  |
| Dosimetry<br>system<br>reproducibilit<br>y<br>position   | ±3% or 3MU whichever is greater at<br>any fixed treatment node                                                                                                                                                                                                                                                                                                                                                                                       | ±2% or 1 MU whichever is greater at any<br>fixed treatment node                                                                                                                                                                                                                                                                                                                                                                                 | No, but has improved<br>specification.                                                                                                                                |
| Beam<br>collimation                                      | Fixed secondary collimators<br>delivering circular field sizes of 5, 7.5,<br>10, 12.5, 15, 20, 25, 30, 35, 40, 50<br>and 60 mm diameter at SAD                                                                                                                                                                                                                                                                                                       | Variable secondary collimator delivering<br>regular hexagonal fields sizes, the<br>distances between the two opposite<br>sides are 5, 7.5, 10, 12.5, 15, 20, 25 and<br>30 mm at SAD                                                                                                                                                                                                                                                             | Comparable field sizes up to 30<br>mm.                                                                                                                                |
| Target<br>location<br>reference                          | Patient's skull                                                                                                                                                                                                                                                                                                                                                                                                                                      | Patient's skull                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same                                                                                                                                                                  |
| Treatment<br>Planning<br>System                          | YES                                                                                                                                                                                                                                                                                                                                                                                                                                                  | YES                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                  |
| Safety<br>Interlocks                                     | YES                                                                                                                                                                                                                                                                                                                                                                                                                                                  | YES                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                  |
| Emergency<br>Stop                                        | YES                                                                                                                                                                                                                                                                                                                                                                                                                                                  | YES                                                                                                                                                                                                                                                                                                                                                                                                                                             | Same                                                                                                                                                                  |

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

Table ES-1: Comparison of major feature of CygneX I System and the Predicate Device

{9}------------------------------------------------

#### Summary of Performance Testing:

There are some minor differences. However, comprehensive nonclinical testing in accordance with FDA Recognized Consensus Standards as shown in the TAB 8-2 and in the TAB-14, and testing in accordance with our internal performance specifications as shown in the TAB-15 demonstrates that the minor differences do not adversely impact performance of the device for its intended use nor do the differences raise new safety concerns. The nonclinical testing performed includes essential performance and basic safety testing, electromagnetic compatibility testing, functional performance characteristics testing and software verification and validation testing. All of these testing confirmed that the CygneX I System performs as intended.

Thus the CygneX I System is substantially equivalent to the Accuray CyberKnife System For Stereotactic Radiosurgery/Radiotherapy (K984563).

---

**Source:** [https://fda.innolitics.com/device/K143560](https://fda.innolitics.com/device/K143560)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
