K-Y BRAND TOUCH All-in-one Warming Caressing Creme and Personal Lubricant
Applicant
Reckitt Benckiser
Product Code
NUC · Obstetrics/Gynecology
Decision Date
May 20, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.5300
Device Class
Class 2
Indications for Use
K-Y® BRAND TOUCH® All-in-One Warming Caressing Crème and Personal Lubricant is a personal lubricant, for penile and/or vaginal application, intended to lubricate and enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex, polyisoprene, or polyurethane condoms.
Device Story
K-Y® Brand Touch® is a non-sterile, anhydrous, fragranced, glycol-in-silicone emulsion personal lubricant. It provides a warming sensation upon contact with ambient skin moisture. The device is intended for over-the-counter use to supplement natural lubrication and enhance comfort during intimate sexual activity. It is supplied in 3.0 ounce HDPE bottles with pump closures. The user applies the crème topically to the penis or vagina. The device is not compatible with natural rubber latex, polyisoprene, or polyurethane condoms.
Clinical Evidence
No clinical data. Bench testing included biocompatibility (ISO 10993-5:2009 cytotoxicity, ISO 10993-10:2010 sensitization and vaginal irritation, ISO 10993-11:2006 acute systemic toxicity) and shelf-life stability testing. Condom compatibility was tested per ASTM D7661-10; results indicated incompatibility with natural rubber latex, polyisoprene, and polyurethane condoms.
Technological Characteristics
Non-sterile, anhydrous, glycol-in-silicone emulsion. Ingredients include lubricants, emollients, antioxidant, gelling agent, thickener, film former, stabilizer, surfactant, and fragrance. Form factor: 3.0 oz HDPE bottle with pump. Biocompatibility tested per ISO 10993 standards. Condom compatibility tested per ASTM D7661-10.
Indications for Use
Indicated for adults requiring personal lubrication for penile and/or vaginal application to enhance comfort during intimate sexual activity. Not compatible with natural rubber latex, polyisoprene, or polyurethane condoms.
Regulatory Classification
Identification
A condom is a sheath which completely covers the penis with a closely fitting membrane. The condom is used for contraceptive and for prophylactic purposes (preventing transmission of sexually transmitted infections). The device may also be used to collect semen to aid in the diagnosis of infertility.
Special Controls
*Classification.* (1) Class II (special controls) for condoms made of materials other than natural rubber latex, including natural membrane (skin) or synthetic.(2) Class II (special controls) for natural rubber latex condoms. The guidance document entitled “Class II Special Controls Guidance Document: Labeling for Natural Rubber Latex Condoms Classified Under 21 CFR 884.5300” will serve as the special control. See § 884.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes entwined around it. The logo also includes the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular pattern around the caduceus symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 20, 2015
Reckitt Benckiser Elizabeth Torre Regulatory Operations Lead Morris Corporate Center IV 399 Interpace Parkway Parsippany, NJ 07054
Re: K143511
> Trade/Device Name: K-Y® Brand Touch® All-in-One Warming Caressing Crème and Personal Lubricant Regulation Number: 21 CFR 884.5300 Regulation Name: Condom Regulatory Class: II Product Code: NUC Dated: April 21, 2015 Received: April 27, 2015
Dear Elizabeth Torre,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA mav publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR
{1}------------------------------------------------
Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Joyce M. Whang -S
for Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K143511
#### Device Name
K-Y® BRAND TOUCH® All-in-One Warming Caressing Crème and Personal Lubricant
### Indications for Use (Describe)
K-Y® BRAND TOUCH® All-in-One Warming Caressing Crème and Personal Lubricant is a personal lubricant, for penile and/or vaginal application, intended to lubricate and enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex, polyisoprene, or polyurethane condoms.
| Type of Use (Select one or both, as applicable) | <table border="1"><tr><td>Prescription Use (Part 21 CFR 801 Subpart D)</td><td>Over-The-Counter Use (21 CFR 801 Subpart C)</td></tr></table> | Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
|-------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|---------------------------------------------|
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) | | |
_ | Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Submitted by: | Reckitt Benckiser LLC<br>Morris Corporate Center IV<br>399 Interpace Parkway<br>Parsippany, NJ 07054<br>Phone: (973) 404-2715 |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Elizabeth Torre, MPH<br>Regulatory Operations Lead,<br>Reckitt Benckiser LLC |
| Date Prepared: | December 14, 2014 |
| Proprietary Name(s): | K-Y® BRAND TOUCH® All-in-One Warming<br>Creme and Personal Lubricant |
| 510(k) Number: | K143511 |
| Classification Name: | Condom (21 CFR §884.5300); Class II |
| Product Code: | NUC (lubricant, personal) |
| Cleared Device/Predicate: | K081236<br>K-Y® Brand Intrigue™ 2-in-1 Massage Creme™ |
## Indication for Use:
K-Y® BRAND TOUCH® All-in-One Warming Caressing Crème and Personal Lubricant is a personal lubricant, for penile and/or vaginal application, intended to lubricate and enhance the ease and comfort of intimate sexual activity and supplement the body's natural lubrication. This product is not compatible with natural rubber latex, polyisoprene, or polyurethane condoms.
## Description of the Device:
K-Y® BRAND TOUCH® All-in-One Warming Caressing Creme and Personal Lubricant, is a nonsterile, anhydrous, fragranced, glycol-in-silicone emulsion that provides a warming sensation upon contact with ambient moisture on the skin. It is composed of a combination of lubricants and emollients, an antioxidant, a gelling agent, a thickener, a film former, a stabilizer, a surfactant, and a fragrance. The emulsion is off-white to beige in color with a slight hint of vanilla and/or floral scent. It has a viscosity, specific gravity, and lubricity typical for emulsions of this type.
The product will be filled in printed black 3.0 ounce high density polyethylene (HDPE) heart shaped bottles, sealed with black pump closures, and packaged in printed cartons.
Like the cleared device, the modified device, K-Y® BRAND TOUCH® All-in-One Warming Caressing Creme and Personal Lubricant, will be marketed as a non-prescription medical device for over-the-counter (OTC) use.
{4}------------------------------------------------
# Technological Characteristics
The modified device, K-Y® BRAND TOUCH® All-in-One Warming Caressing Creme and Personal Lubricant, is substantially equivalent to the cleared device. The intended use is identical and the technical characteristics vary only slightly due to the addition of small amount of fragrance. Aside from the fragrance, all of the ingredients in the modified device are identical, and they are present at almost identical concentrations to those in the cleared device.
# Performance Data:
# Biocompatibility
K-Y® BRAND TOUCH® All-in-One Warming Caressing Creme and Personal Lubricant is varied slightly from the cleared predicate device; the only difference in the formula is the addition of a fragrance. All of the ingredients, except the fragrance, have been used in other 510k cleared marketed devices.
Biocompatibility evaluation of the K-Y® BRAND TOUCH® All-in-One Warming Caressing Creme and Personal Lubricant consisted of the following tests:
- Cytotoxicity according to ISO 10993-5:2009
- Guinea Pig Maximization Sensitization according to ISO 10993-10:2010
- Vaginal Irritation according to ISO 10993-10:2010 ●
- Acute Systemic Toxicity according to ISO 10993-11:2006
The results of these tests demonstrate that the K-Y® BRAND TOUCH® All-in-One Warming Caressing Creme and Personal Lubricant is biocompatible.
# Condom Compatibility
Condom compatibility testing was conducted in accordance with ASTM D7661-10. Under current standards, the modified device may have an effect on the mechanical and/or physical integrity of natural rubber latex, synthetic polyisoprene and polyurethane condoms. Therefore, condom compatibility claims are not requested for this modified device.
# Stability:
Shelf life testing was conducted on final devices to support the targeted two-year shelf-life. Based on satisfactory results of this testing, the product will be labeled with a 2 year expiration date.
# Conclusion
Because the minor modifications (i.e., fragrance) to this cleared device do not raise new types of questions of safety or effectiveness, there is no change in intended use or fundamental scientific technology, and the results of additional technological characterization, biocompatibility testing, and performance testing are favorable, we conclude that the modified device, K-Y® BRAND TOUCH® All-in-One Warming Caressing Creme and Personal Lubricant, is substantially equivalent to our legally cleared predicate device, K-Y® Brand Intrigue™ 2-in-1 Massage Creme™.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.