← Product Code [GEX](/productcode/GEX) · K143457

# NeuroBlate System (K143457)

_Monteris Medical, Inc. · GEX · Feb 19, 2015 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K143457

## Device Facts

- **Applicant:** Monteris Medical, Inc.
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** Feb 19, 2015
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

The Monteris Medical NeuroBlate™ System is indicated for use to ablate, necrotize, or coagulate soft tissue through interstitial irradiation or thermal therapy in medicine and surgery in the discipline of neurosurgery with 1064 nm lasers. The Monteris Medical NeuroBlate™ System is intended for planning and monitoring thermal therapies under MRI visualization. It provides MRI based trajectory planning assistance for the stereotaxic placement of MRI compatible (conditional) NeuroBlate™ Laser Delivery Probes. It also provides real time thermographic analysis of selected MRI images. When interpreted by a trained physician, this System provides information that may be useful in the determination or assessment of thermal therapy. Patient management decisions should not be made solely on the basis of the NeuroBlate™ System analysis.

## Device Story

NeuroBlate System provides MRI-guided laser interstitial thermal therapy (LITT) for neurosurgical procedures. System components include gas-cooled laser delivery probes (SideFire/FullFire), probe drivers for stereotaxic positioning, interface platform, electronics rack, and control workstation with M-Vision software. Surgeon uses workstation for trajectory planning and real-time thermographic monitoring of MRI images during ablation. System delivers 1064 nm laser energy to target tissue. Output assists physician in assessing thermal therapy progress; clinical decisions must not rely solely on system analysis. Used in clinical settings with MRI compatibility. This submission introduces 2.2mm reduced-diameter probes to complement existing 3.3mm probes.

## Clinical Evidence

Bench testing only. Verification testing confirmed compliance with electrical, mechanical, and laser emission requirements. Testing demonstrated that the reduced-diameter probes function identically to the predicate larger-diameter probes regarding system interface, cooling, and target tissue placement.

## Technological Characteristics

System utilizes 1064 nm laser energy for thermal ablation. Components include gas-cooled laser delivery probes (2.2mm and 3.3mm diameters), robotic/advanced probe drivers, and MRI-compatible interface platform. Software-based M-Vision workstation provides trajectory planning and real-time thermographic analysis. System is MRI-conditional. Materials are identical to predicate device. Connectivity includes umbilicals and cables for system integration.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- NeuroBlate™ System ([K141983](/device/K141983.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with three overlapping profiles, representing the department's mission to protect the health of all Americans and provide essential human services.

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 19, 2015

Monteris Medical. Inc. % Mr. Craig Coombs President Coombs Medical Device Consulting 1193 Sherman Street Alameda, California 94501

Re: K143457

Trade/Device Name: Neuroblate System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: January 19, 2015 Received: January 21, 2015

Dear Mr. Craig Coombs,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act 

{1}------------------------------------------------

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

# Jennifer R. Stevenson -S

For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K143457

Device Name Monteris Medical NeuroBlate™ System

#### Indications for Use (Describe)

The Monteris Medical NeuroBlate™ System is indicated for use to ablate, necrotize, or coagulate soft tissue through interstitial irradiation or thermal therapy in medicine and surgery in the discipline of neurosurgery with 1064 nm lasers.

The Monteris Medical NeuroBlate™ System is intended for planning and monitoring thermal therapies under MRI visualization. It provides MRI based trajectory planning assistance for the stereotaxic placement of MRI compatible (conditional) NeuroBlate™ Laser Delivery Probes. It also provides real time thermographic analysis of selected MRI images.

When interpreted by a trained physician, this System provides information that may be useful in the determination or assessment of thermal therapy. Patient management decisions should not be made solely on the basis of the NeuroBlate™ System analysis.

| Type of Use ( <i>Select one or both, as applicable</i> ) |
|----------------------------------------------------------|
|----------------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows two identical arrangements of teal-colored circles of varying sizes. Each arrangement consists of seven circles, with one larger circle at the top and smaller circles arranged around it. The circles have shadows, giving them a three-dimensional appearance. The background is plain white.

## Section 5: 510(k) Summary K143457

| Category                                 | Comments                                                                                                                                                   |
|------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor:                                 | Monteris Medical Corp.<br>16305 36th Ave. North, Suite 200<br>Plymouth, MN 55446<br>763-253-4710<br>Fax: 763-746-0084<br>www.monteris.com                  |
| Correspondent Contact<br>Information:    | Craig Coombs<br>Coombs Medical Device Consulting<br>1193 Sherman Street<br>Alameda, CA 94501<br>Tel: 510-337-0140<br>Fax: 510-337-0416                     |
| Device Common Name:                      | Magnetic Resonance Image Guided Laser<br>Thermal Therapy System                                                                                            |
| Device Classification Number:            | 21 CFR 878.4810<br>Laser surgical instrument for use in general and<br>plastic surgery and in dermatology<br><br>21 CFR 882.4560<br>Stereotaxic instrument |
| Device Classification &<br>Product Code: | Class II, GEX<br>Class II, HAW                                                                                                                             |
| Device Proprietary Name:                 | Monteris Medical NeuroBlate™ System                                                                                                                        |

# a. Device Information:

### Predicate Device Information:

| Predicate Device:                                  | NeuroBlate™ System                                                                                                                                         |
|----------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Predicate Device Manufacturer:                     | Monteris Medical                                                                                                                                           |
| Predicate Device Common Name:                      | Monteris NeuroBlate™ System                                                                                                                                |
| Predicate Device Premarket Notification #:         | K141983                                                                                                                                                    |
| Predicate Device Regulation:                       | 21 CFR 878.4810<br>Laser surgical instrument for use in general<br>and plastic surgery and in dermatology<br><br>21 CFR 882.4560<br>Stereotaxic instrument |
| Predicate Device Classification &<br>Product Code: | Class II, GEX<br>Class II, HAW                                                                                                                             |

### b. Date Summary Prepared

19 February 2015

{4}------------------------------------------------

Image /page/4/Figure/0 description: The image shows two identical arrangements of five turquoise circles each, positioned against a white background. Each arrangement consists of one larger circle at the top, two medium-sized circles in the middle, and two smaller circles at the bottom. Each circle has a shadow effect, giving them a three-dimensional appearance. 

### c. Description of Device

The Monteris NeuroBlate™ System is a collection of MRI-compatible laser devices and accessories that create an MRI guided delivery of precision thermal therapy in the practice of neurosurgery.

The NeuroBlate System components consist of:

- . Families of gas-cooled Laser Delivery Probes (Probe) (SideFire & FullFire) to deliver controlled energy to a target zone. This application includes the smaller diameter SideFire Select and FullFire Select families of Probes;
- Probe Drivers (Advanced Probe Driver, Robotic Probe Driver) which allow the . surgeon to precisely position, stabilize and manipulate a probe, endoscope or other device within the target zone.
- An Interface Platform, which attaches to the MRI system patient table and . provides supporting electronics for the Probe Drivers and interconnections for the Laser Delivery Probes;
- A System Electronics Rack and Components, which includes necessary . umbilicals, cables, penetration panels, and small hardware for system mechanical, electrical, and electronic operation, and
- A Control Workstation including the M-Vision™ software, which includes a user . interface for procedure planning, interactive monitoring of NeuroBlateTM procedures, and interfaces to the MRI and hardware subsystems.

This submission adds a line of 2.2mm Reduced Diameter Probes to the existing 3.3mm diameter probes. They will be known as the SideFire Select™ and FullFire Select™ Probes.

### d. Indications for Use

The Monteris Medical NeuroBlate™ System is indicated for use to ablate, necrotize, or coagulate soft tissue through interstitial irradiation or thermal therapy in medicine and surgery in the discipline of neurosurgery with 1064 nm lasers.

The Monteris Medical NeuroBlate™ System is intended for planning and monitoring thermal therapies under MRI visualization. It provides MRI based trajectory planning assistance for the stereotaxic placement of MRI compatible (conditional) NeuroBlate™ Laser Delivery Probes. It also provides real-time thermographic analysis of selected MRI images.

When interpreted by a trained physician. this System provides information that may be useful in the determination or assessment of thermal therapy. Patient management decisions should not be made solely on the basis of the NeuroBlate™ System analysis.

{5}------------------------------------------------

Image /page/5/Figure/0 description: The image shows a pattern of teal-colored circles arranged in two similar clusters. Each cluster consists of a larger circle at the top and several smaller circles below, creating a triangular shape. The circles have soft shadows, giving them a slightly raised or three-dimensional appearance. The background is plain white, which makes the teal circles stand out.

### e. Comparison to Predicate Device

The application Monteris Medical NeuroBlate™ System with the Reduced Diameter Laser Delivery Probes is substantially equivalent to the predicate Monteris NeuroBlate™ System (K141983) in intended use, technology, design and physician use.

The Indications for Use for the modified NeuroBlate System are unchanged from the predicate NeuroBlate System. The fundamental technology is also unchanged.

All patient contacting materials are identical in composition, source, and use with respect to the predicate device.

The technical modes of action and technical principles are materially the same as the predicate devices.

The application System with the Reduced Diameter Laser Delivery Probes has similar laser emission patterns and laser ablation effect as the predicate Laser Delivery Probes. The expected use of the Reduced Diameter Probe for either its SideFire or FullFire is no different than the expected use of the predicate laser delivery probes.

Bench testing has demonstrated that the System is in compliance with the medical community's expectations and the product labeling and specifications. It demonstrates that NeuroBlate system works as well with the Reduced Diameter Probe as it does with the larger diameter versions.

As the modifications presented in the current device do not change the intended use, operating principles, or raise any unaddressed safety concerns, it can be concluded the application NeuroBlate™ System with the added Reduced Diameter Laser Delivery Probes is substantially equivalent to the predicate NeuroBlate™ System.

### f. Summary of Supporting Data

Bench verification testing has demonstrated that the NeuroBlate System in general, and the new sizes of Laser Delivery Probes in particular, are in compliance with the medical community's expectations and the product labeling and product specifications.

In particular, the bench testing demonstrated that the new sizes of the Laser Delivery Probes exhibited compliance to the same electrical requirements, mechanical integrity and operation (e.g., cooling) requirements, and laser emission requirements as the predicate sizes; that they interfaced with the NeuroBlate System in an identical manner; and that adequate labeling information was provided to ensure that they could be accurately placed within the target tissue.

---

**Source:** [https://fda.innolitics.com/device/K143457](https://fda.innolitics.com/device/K143457)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
