UTAH CVX-RIPE

K143424 · Utah Medical Products, Inc. · PFJ · Aug 28, 2015 · Obstetrics/Gynecology

Device Facts

Record IDK143424
Device NameUTAH CVX-RIPE
ApplicantUtah Medical Products, Inc.
Product CodePFJ · Obstetrics/Gynecology
Decision DateAug 28, 2015
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 884.4260
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The UTAH CVX-RIPE™ is intended to mechanically improve the favorability of the cervix of pregnant patients at term gestation, in which induction of labor is medically indicated.

Device Story

UTAH CVX-RIPE is a double-balloon silicone catheter for mechanical cervical ripening. Device consists of 18 Fr., 42 cm silicone catheter with two independent balloons and a malleable stainless steel stylet for placement. Physician inserts catheter through vagina into cervical canal; distal balloon positioned against internal cervical os; proximal balloon against external cervical os. Both balloons inflated with saline via separate color-coded check valves ('U' for uterine, 'V' for vaginal). Simultaneous pressure on internal and external os provides gradual mechanical dilation. Used in clinical setting by healthcare providers to facilitate labor induction. Benefits include mechanical cervical ripening for patients at term gestation.

Clinical Evidence

No clinical data. Substantial equivalence demonstrated via bench testing: balloon deflation reliability, integrity, overuse testing, volume maintenance, response to traction, bond strength, and stylet puncture testing. Biocompatibility confirmed per ISO 10993-1. Shelf life established via accelerated aging study.

Technological Characteristics

Double-balloon catheter; 18 Fr., 42 cm length. Materials: medical grade silicone. Includes malleable stainless steel stylet. Independent inflation lumens with color-coded check valves. Biocompatible per ISO 10993-1.

Indications for Use

Indicated for pregnant patients at term gestation requiring induction of labor to mechanically improve cervical favorability.

Regulatory Classification

Identification

A hygroscopic Laminaria cervical dilator is a device designed to dilate (stretch open) the cervical os by cervical insertion of a conical and expansible material made from the root of a seaweed (Laminaria digitata or Laminaria japonica). The device is used to induce abortion.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling an eagle or a stylized human profile, composed of three interconnected shapes. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 August 28, 2015 Utah Medical Products, Inc. Ben Shirley VP of Quality Assurance & Product Development 7043 South 300 West Midvale, UT 84047 K143424 Re: Trade/Device Name: UTAH CVX-RIPE Regulation Number: 21 CFR 884.4260 Regulation Name: Hygroscopic Laminaria cervical dilator Regulatory Class: II Product Code: PFJ Dated: July 27, 2015 Received: July 29, 2015 Dear Ben Shirley, We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in {2}------------------------------------------------ the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, **Herbert P. Lerner -S** Benjamin R. Fisher, Ph.D. for Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ #### Indications for Use Statement 6.0 | 510(k) Number: | K143424 | |----------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Device Name: | UTAH CVX-RIPE™; UTMD Cervical Ripening Balloon Catheter | | | Indications for Use: The UTAH CVX-RIPE™ is intended to mechanically improve the<br>favorability of the cervix of pregnant patients at term gestation, in which<br>induction of labor is medically indicated. | > Prescription Use (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE—CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) {4}------------------------------------------------ # 510(k) Summary #### Submitter Information | 510(k) Owner: | Utah Medical Products, Inc | |--------------------|----------------------------| | Address: | 7043 South 300 West | | | Midvale, UT 84047 | | | USA | | Telephone No.: | (801) 566-1200 | | Fax No.: | (801) 566-2062 | | Contact: | Ben Shirley | | Contact Phone No.: | (801) 569-4039 | | Contact E-mail: | bshirley@utahmed.com | | Date Prepared: | August 27, 2015 | #### Device Information | Trade Name: | UTAH CVX-RIPE™, UTMD Cervical Ripening Balloon<br>Catheter | |-----------------|------------------------------------------------------------| | Common Name: | Catheter, Balloon, Dilation of Cervix | | Classification: | Class II | | Regulation: | 21 CFR § 884.4260 | | Description: | Hygroscopic Laminaria cervical dilator | | Product Code: | PFJ (catheter, balloon, dilation of cervical canal) | #### Predicate Device The predicate device is the Cook® Cervical Ripening Balloon cleared September 27, 2013 under 510(k) number K131206. #### Device Description The UTAH CVX-RIPE™ is a double balloon catheter. The tip is placed through the vagina and into the cervical canal. The distal balloon is positioned against the internal cervical os and inflated with saline. The proximal balloon is positioned against the external cervical os and inflated with saline, providing gradual mechanical cervical dilation by simultaneously applying pressure on the internal and external cervical os. The materials used to manufacture the device, and which have patient contact, are all medical grade silicone. {5}------------------------------------------------ The UTAH CVX-RIPE™ consists of an 18 Fr., 42 cm silicone catheter with a silicone tip and two silicone balloons. Each balloon is independently inflated via a separate inflation lumen through color coded check valves located at the proximal end of the device. The check valves are marked accordingly, "U" for the distal uterine balloon and "V" for the proximal vaginal balloon. A malleable stainless steel stylet is included for ease of placement. The stylet is placed through a third separate lumen that is closed at the distal tip. The stylet is preloaded in the catheter. ### Intended Use The UTAH CVX-RIPE™ is intended to mechanically improve the favorability of the cervix of pregnant patients at term gestation, in which induction of labor is medically indicated. ## Predicate Device Comparison The UTAH CVX-RIPE™ has the same intended use and technological characteristics as the Cook® Cervical Ripening Balloon cleared under K131206. ## Performance Testing To demonstrate substantial equivalence, the following mechanical performance tests were conducted: - Balloon Deflation Reliability Testing to demonstrate that the balloons ● deflate properly after use. - Balloon Integrity Testing to demonstrate that the balloons do not burst at ● the maximum recommended inflation volume and duration under simulated use conditions. - Balloon Overuse - Testing to demonstrate that balloons do not leak and properly deflate when using an inflation volume and duration greater than maximum recommendations. - Balloon Volume Maintenance Testing to demonstrate that the balloons ● do not leak. - Balloon Response to Traction - Testing to demonstrate that the distal balloon does not pull out from the patient under simulated use conditions. - Bond Strength - Testing to demonstrate that bonded connections do not fail during simulated use conditions. - Stylet Puncture - Testing to demonstrate that the stylet does not puncture through the catheter during simulated use. The UTAH CVX-RIPE™ is biocompatible per the requirements of ISO 10993-1. The shelf life of the UTAH CVX-RIPE™ was established based on the results of an accelerated aging study. {6}------------------------------------------------ # Conclusion The UTAH CVX-RIPE™ is substantially equivalent to the Cook® Cervical Ripening Balloon cleared under K131206.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...