← Product Code [DSI](/productcode/DSI) · K143359

# CG-6108 ACT-3L Continuous ECG Monitor and Arrhythmia Detector (K143359)

_Lifewatch Technologies , Ltd. · DSI · Sep 11, 2015 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K143359

## Device Facts

- **Applicant:** Lifewatch Technologies , Ltd.
- **Product Code:** [DSI](/productcode/DSI.md)
- **Decision Date:** Sep 11, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1025
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

The CG-6108 ACT-3L Continuous ECG Monitor and Arrhythmia Detector is intended for use by patients who experience transient symptoms that may suggest cardiac arrhythmia. The device continuously monitors patient ECG, automatically generates an alert triggered by an arrhythmia detection algorithm, or generates an alert manually triggered by the patient, and transmits the recorded data trans-telephonically to a monitoring center provides the ECG data to the medical practitioner for evaluation. The CG-6108 ACT-3L Continuous ECG Monitor and Arrhythmia Detector is intended to be prescribed for patients who have demonstrated a need for cardiac monitoring and are at low risk of developing life-threatening arrhythmias. Conditions where the system should not be used include patients likely to experience primary Ventricular Fibrillation or Ventricular Tachycardia and patients who have other co-morbid cardiovascular conditions where an arrhythmia could be potentially life threatening. This device has not been tested for and it is not intended for pediatric use.

## Device Story

Chest-worn sensor with 4 electrodes acquires/records ECG signals; transmits via Bluetooth to handheld mobile device (Android-based). Proprietary application on mobile device processes ECG; detects arrhythmias (Atrial Fibrillation, Tachycardia, Bradycardia, Pause). Upon detection or manual patient trigger, handheld device transmits ECG data via cellular link or landline to remote Monitoring Center. Medical professionals at center analyze data; provide findings to prescribing physician. Device enables home-based cardiac monitoring for low-risk patients; facilitates timely clinical intervention for transient symptoms.

## Clinical Evidence

No clinical data. Bench testing only, including wireless co-existence testing, software validation, and algorithm performance testing per ANSI/AAMI EC57:2012.

## Technological Characteristics

Chest-worn sensor with 4 electrodes, 3.6V AA battery, Bluetooth transceiver, patient alert buzzer. Mobile platform (Android). Standards: IEC 60601-1, ISO 14971, IEC 62304, ISO 10993-1, ANSI/AAMI EC57, IEC 60601-1-2, IEC 60601-1-8.

## Regulatory Identification

The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.

## Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.

## Predicate Devices

- CG-6108 ACT-3L Continuous ECG Monitor and Arrhythmia Detector ([K110499](/device/K110499.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Public Health Service

Image /page/0/Picture/2 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized caduceus-like symbol with three intertwined figures, representing health and well-being. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" encircles the symbol, indicating the department's name and national affiliation. The seal is presented in black and white, emphasizing its official and formal nature.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 11, 2015

LifeWatch Technologies Ltd. Stephen Slavens Head of RA and QA 10255 W Higgins Road, Suite 100 Chicago, Illinois 60018

Re: K143359

Trade/Device Name: Cg-6108 Act-31 Continuous ECG Monitor and Arrhythmia Detector Regulation Number: 21 CFR 870.1025 Regulation Name: Arrhythmia Detector and Alarm (Including ST-Segment Measurement and Alarm) Regulatory Class: Class II Product Code: DSI, DXH

Dated: August 23, 2015 Received: August 25, 2015

Dear Stephen Slavens,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

sincerely yours,

Mitchell Stein

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known) K143359

#### Device Name

CG-6108 ACT-3L Continuous ECG Monitor and Arrhythmia Detector

#### Indications for Use (Describe)

The CG-6108 ACT-3L Continuous ECG Monitor and Arrhythmia Detector is intended for use by patients who experience transient symptoms that may suggest cardiac arthythmia. The device continuously monitors patient ECG, automatically generates an alert triggered by an arrhythmia detection algorithm, or generates an alert manually triggered by the patient, and transmits the recorded data trans-telephonically to a monitoring center provides the ECG data to the medical practitioner for evaluation.

The CG-6108 ACT-3L Continuous ECG Monitor and Arrhythmia Detector is intended to be prescribed for patients who have demonstrated a need for cardiac monitoring and are at low risk of developing life-threatening arrhythmias. Conditions where the system should not be used include patients likely to experience primary Ventricular Fibrillation or Ventricular Tachycardia and patients who have other co-morbid cardiovascular conditions where an arrhythmia could be potentially life threatening.

This device has not been tested for and it is not intended for pediatric use.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for LifeWatch Technologies. The logo consists of a pattern of dots on the left, followed by a curved line and the words "LifeWatch" in a blue, sans-serif font. Below "LifeWatch" is the word "Technologies" in a smaller font. There is a small star to the right of the word "LifeWatch".

K143359

# Section 005: 510(k) Summary

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for LifeWatch Technologies. The logo consists of a pattern of dots on the left, followed by the company name in a stylized font. Below the company name is the word "Technologies" in a smaller font.

## 510(k) Summary: CG-6108 ACT-3L Continuous ECG Monitor and Arrhythmia Detector with SW Version 6.

## Introduction

This document contains the 510(k) summary for the CG-6108 ACT-3L Continuous ECG Monitor and Arrhythmia Detector with Software (SW) Ver. 6.1.17. The content of this summary is based on the requirements of 21 CFR Section 807.92(c).

## l. Submitter

| Submitter                               | LifeWatch Technologies Ltd.                                    |                    |
|-----------------------------------------|----------------------------------------------------------------|--------------------|
| Establishment<br>Registration<br>Number | 9681879                                                        |                    |
| Address                                 | 2 Pekeris St., P.O.B. 527, Rehovot, 7610303, Israel            |                    |
| Submitter:                              | Asher Kassel, Director of RA & QA, LifeWatch Technologies Ltd. |                    |
| Phone:                                  | 972-8-9484010 (direct)                                         | Fax: 972-8-9484044 |
| E-mail:                                 | akassel@lifewatch.com                                          |                    |
| Application<br>Correspondent            | Stephen Slavens, Acting Head of Corporate RA & QA, LifeWatch   |                    |
| Phone:                                  | (847) 8134-625                                                 | Fax: (847)720-2111 |
| E-mail                                  | CT-SSlavens@lifewatch.com                                      |                    |
| Date Prepared:                          | Nov 20, 2014                                                   |                    |

## II. Device

| Trade Name:     | CG-6108 ACT-3L Continuous ECG Monitor and Arrhythmia Detector                               |
|-----------------|---------------------------------------------------------------------------------------------|
| Classification: | Detector and alarm, arrhythmia<br>Transmitters and receivers, electrocardiograph, telephone |
| Product Code:   | DSI,<br>DXH                                                                                 |
| Regulation No:  | 870.1025<br>870.2920                                                                        |
| Class:          | II                                                                                          |

## III. Predicate Device

| Predicate device | K110499 cleared on 6th April 2011                             |
|------------------|---------------------------------------------------------------|
|                  | CG-6108 ACT-3L Continuous ECG Monitor and Arrhythmia Detector |

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for LifeWatch Technologies. The logo consists of a pattern of dots on the left, followed by the company name "LifeWatch" in a stylized font. Below the company name is the word "Technologies" in a smaller font. There is a star to the right of the word "LifeWatch".

## IV. Device Description

The CG-6108 ACT-3L Continuous ECG Monitor and Arrhythmia Detector is designed for self-testing by patients at home and for analysis by medical professionals at a remote Monitoring Center.

The chest-worn sensor is used for the acquisition, recording, and transmission of the ECG signal. The device is equipped with 4 electrodes on a harness and it houses a 3.6V AA battery, a Bluetooth transceiver and a patient alert buzzer.

The ECG signals are transmitted via Bluetooth to a handheld device with a proprietary interactive application, configured to process and transmit the ECG recordings. The handheld device is a mobile computing device with a display and a touch input such as a cell phone. It has sufficient memory and processing capability to run the proprietary application.

When an arrhythmia event is detected, the handheld device transmits the recorded ECG information automatically via cellular link, to the Monitoring Center for professional analysis. When cellular service is unavailable the patient has an option to transmit via a landline telephone.

The modification that led to this submission was the change from a mobile platform with a Windows-based OS to one with an Android-based OS.

#### V. Indications for Use

The CG-6108 ACT-3L Continuous ECG Monitor and Arrhythmia Detector is intended for use by patients who experience transient symptoms that may suggest cardiac arrhythmia. The device continuously monitors patient ECG, automatically generates an alert triggered by an arrhythmia detection algorithm, or generates an alert manually triggered by the patient, and transmits the recorded data transtelephonically to a monitoring center. The monitoring center provides the ECG data to the medical practitioner for evaluation.

The CG-6108 ACT-3L Continuous ECG Monitor and Arrhythmia Detector is intended to be prescribed for patients who have demonstrated a need for cardiac monitoring and are at low risk of developing life-threatening arrhythmias. Conditions where the system should not be used include patients likely to experience primary Ventricular Fibrillation or Ventricular Tachycardia and patients who have other co-morbid cardiovascular conditions where an arrhythmia could be potentially life threatening.

The device has not been tested for and it is not intended for pediatric use.

## VI. Summary of the Technological Characteristics / Principles of Operation

The technological characteristics and principles of operation of the ACT device with SW Ver. 6.1.17 are the same as the predicate device. The chest-worn ECG sensor transmits signals via Bluetooth to the handheld device equipped with the Medical Application, which incorporates an algorithm for detection of cardiac events: Atrial Fibrillation, Tachycardia, Bradycardia and Pause.

A detected artifact triggers transmission of the signal to the Monitoring Center for analysis.

## VII. Performance data - Non Clinical for the CG-6108 ACT-3L with SW Ver. 6.1.17

In order to support the new SW Version, the following testing has been performed and passed:

- . Wireless co-existence testing of the ACT-3L system, LifeWatch Technologies, VATR-0163
- . SW Validation Test Report, LifeWatch Technologies, STR-0112-61170
- Testing of the ACT-3L algorithm, LifeWatch Technologies, VATR-0164 o

## Performance data - Standards:

This submission was written according to and in conformance with FDA Guidance "Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm, October 28, 2003." This submission was also prepared per "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, May 11 2005".

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for LifeWatch Technologies. The logo consists of a pattern of dots on the left, followed by the word "LifeWatch" in a serif font. Below "LifeWatch" is the word "Technologies" in a smaller, lighter font. There is a small star to the right of the word "LifeWatch".

- The design of the CG-6108 ACT-3L conforms to the following voluntary standards:
	- i.IEC 60601-1:2005 3rd edition / Cor. 1:2006. Cor. 2:2007. Medical electrical equipment - General requirements for safety
	- ii.ISO 14971:2007, Medical devices Application of risk management to medical devices
	- iii.IEC 62304:2006, Medical Device Software Software Lifecycle Processes
	- iv.ISO 10993-1:2009, Biological evaluation of medical devices -- Part 1: Evaluation and testing within a risk management process.
	- v.ANSI/AAMI EC57:2012: Testing and Reporting Performance Results of Cardiac Rhythm and ST Seqment Measurement Algorithms
	- vi.IEC 60601-1-2: 2014 Ed. 4, Part 1: Medical electrical equipment, Part 1-2; Electromagnetic compatibility - Requirements and tests
	- vii.IEC 60601-1-8:2012. General Requirements for basic safety and essential performance - Collateral standard: General Requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems.

#### Performance data - Clinical

The FDA Guidance document "Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm. October 28. 2003" states the Agency will rely on well-designed bench and/or animal testing rather than requiring clinical studies so long as the studies do not have dissimilar designs, new technology, or dissimilar indications for use. The modifications in this submission do not make changes that require clinical studies.

#### VIII. PeConclusion of Substantial Equivalence:

The CG-6108 ACT-3L device with SW Ver. 6.1.17 is substantially equivalent with respect to the indications for use, technological characteristics and performance characteristics to the identified legally marketed predicate device.

---

**Source:** [https://fda.innolitics.com/device/K143359](https://fda.innolitics.com/device/K143359)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
