← Product Code [PNR](/productcode/PNR) · K143344

# NeoMed NeoConnect Enteral Syringes with ENFit Connector (K143344)

_Neomed, Inc. · PNR · Apr 9, 2015 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K143344

## Device Facts

- **Applicant:** Neomed, Inc.
- **Product Code:** [PNR](/productcode/PNR.md)
- **Decision Date:** Apr 9, 2015
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.5980
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic, Pediatric

## Indications for Use

The device is indicated for use as a dispenser, a measuring device. It is used to deliver fluids into the body via extension sets and feeding tubes in neonatal and small pediatric patients.

## Device Story

NeoConnect™ Enteral Syringes are piston-style syringes (0.5ml to 100ml) used for dispensing and measuring fluids. The device features an integral female ENFit connector tip, designed to interface exclusively with male ENFit feeding tubes and extension sets. This dedicated connection system prevents wrong-route administration of fluids. Used in hospital settings, the device is operated by clinicians to deliver enteral nutrition or medication. The translucent barrel allows for visual verification of fluid volume. By ensuring a secure, non-interchangeable connection, the device reduces the risk of misconnection errors, thereby improving patient safety during enteral feeding or medication administration.

## Clinical Evidence

Bench testing only. Testing included biocompatibility (ISO 10993-5, -10, -11), chemical extractables/leachables, and finished device verification (capacity, scale, piston fit, leakage). ENFit-specific testing included misconnection assessment, stress cracking, axial load resistance, unscrewing resistance, and overriding resistance. Human factors and usability validation were performed to confirm safe operation.

## Technological Characteristics

Piston-style syringe; materials: Polypropylene (Class VI) barrel/plunger, Silicone (Class VI) gasket, Polydimethylsiloxane lubricant. ENFit connector compliant with AAMI/CN3:2014 (PS) Part 3 Table B.2. Sterilization: Ethylene Oxide (EO) to 10-6 SAL. Sizes: 0.5ml to 100ml.

## Regulatory Identification

A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).

## Special Controls

*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.

## Predicate Devices

- NeoMed Oral / Enteral Syringe ([K122373](/device/K122373.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

September 29, 2016

NeoMed, Inc. Melinda Harrison Smith, RAC, ASQ-CBA Director, Quality and Regulatory Affairs 100 Londonderry Court, Suite 112 Woodstock, GA 30188

K143344 Re:

Trade/Device Name: NeoMed NeoConnect™ Enteral Syringes with ENFit Connector Regulation Number: 21 CFR §876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: II Product Code: PNR Dated: March 11, 2015 Received: March 13, 2015

Dear Melinda Harrison Smith,

This letter corrects our substantially equivalent letter of April 7, 2015.

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use bared in the May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of I he goneral ocharing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

## Page 2 - Melinda Harrison Smith, RAC, ASQ-CBA

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely,

For Division

Douglas Silverstein -S 2016.09. -04'00'

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K143344

### Device Name

NeoMed NeoConnect™ Enteral Syringes with ENFit Connector

Indications for Use (Describe)

The device is indicated for use as a dispenser, a measuring device. It is used to deliver fluids into the body via extension sets and feeding tubes in neonatal and small pediatric patients.

| Type of Use (Select one or both, as applicable) |  |
|-------------------------------------------------|--|
|-------------------------------------------------|--|

X Prescription Use (Part 21 CFR 801 Subpart D)

| Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# TRADITIONAL 510(K) SUMMARY (21 CFR § 807.92)

| I. SUBMITTER                          | NeoMed, Inc.<br>100 Londonderry Court<br>Suite 112<br>Woodstock, GA 30188<br>Tel: 770-516-2225<br>Fax: 770-516-2448 |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------|
| Contact:                              | Melinda Harrison Smith, RAC, CBA<br>mharrison@neomedinc.com                                                         |
| Date Prepared:                        | 18 November 2014                                                                                                    |
| Establishment<br>Registration Number: | 3006520777                                                                                                          |

#### ll. DEVICE

| Trade Name:          | NeoConnect™ Enteral Syringes                              |
|----------------------|-----------------------------------------------------------|
| Common Name:         | Enteral Syringe                                           |
| Classification Name: | Gastrointestinal tube and accessories (21 CFR § 876.5980) |
| Regulatory Class:    | II                                                        |
| Product Code :       | PIF                                                       |

#### lll. PREDICATE DEVICE

NeoMed Oral / Enteral Syringe (0.5ml to 100ml) (K122373) This predicate had not been subject to a design-related recall.

#### IV. DEVICE DESCRIPTION

The NeoMed NeoConnect™ Enteral Syringes are standard piston style syringes consisting of a syringe barrel with integral ENFit syringe tip, syringe plunger, syringe gasket, and supplied with a syringe tip cap. They are provided in varying sizes ranging from 0.5ml to 100ml nominal capacity. The integral syringe tip is a female ENFit connector which is compatible only with feeding tubes and extension sets that have ENFit male connectors to form a dedicated system that prevents wrong-route administration of fluids. They possess translucent barrels to provide visualization of fluid contents and volume.

#### V. INDICATIONS FOR USE

The device is indicated for use as a dispenser, a measuring device and a fluid transfer device. It is used to deliver fluids into the body via extension sets and feeding tubes in neonatal and small pediatric patients.

{4}------------------------------------------------

### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE

| Feature                                                               | PREDICATE DEVICE                                                                                                                                                                                                                                 | SUBJECT DEVICE                                                                                                                                                                                                                              |
|-----------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use /<br>Intended Use<br>Statement                    | The device is indicated for use as<br>a dispenser, a measuring device<br>and an oral fluid transfer device. It<br>is used to inject fluids into the body<br>via extension sets and feeding<br>tubes in neonatal and small<br>pediatric patients. | The device is indicated for use as a<br>dispenser, a measuring device and<br>a fluid transfer device. It is used to<br>deliver fluids into the body via<br>extension sets and feeding tubes in<br>neonatal and small pediatric<br>patients. |
| Patient Population /<br>Environment of Use                            | Neonates and Small Pediatrics/<br>Hospital, Disposable and for single<br>patient use only                                                                                                                                                        | Neonates and Small Pediatrics/<br>Hospital, Disposable and for single<br>patient use only                                                                                                                                                   |
| Description of Device                                                 | An oral / enteral syringe consisting<br>of a syringe barrel with integral tip<br>(tapered), plunger, gasket, barrel<br>lubricant and supplied with a<br>syringe tip cap.                                                                         | An enteral syringe consisting of a<br>syringe barrel with integral tip<br>(ENFit), plunger, gasket, barrel<br>lubricant and supplied with a<br>syringe tip cap.                                                                             |
| Syringe Type                                                          | Oral / Enteral Syringe                                                                                                                                                                                                                           | Enteral Syringe                                                                                                                                                                                                                             |
| Principle of Operation                                                | Piston Style Syringe                                                                                                                                                                                                                             | Piston Style Syringe                                                                                                                                                                                                                        |
| Syringe Tip Type                                                      | Tapered                                                                                                                                                                                                                                          | ENFit<br>(Dimensional compliance to<br>AAMI/CN3:2014 (PS) Part 3 Table<br>B.2 Female Enteral Small-Bore<br>Connector).                                                                                                                      |
| Syringe Sizes<br>(nominal volumes)                                    | 0.5ml to 100ml sizes                                                                                                                                                                                                                             | 0.5ml to 100ml sizes                                                                                                                                                                                                                        |
| Syringe Barrel with<br>integral syringe tip and<br>volume graduations | Barrel - Polypropylene, class VI<br>Graduation Ink - Orange                                                                                                                                                                                      | Barrel - Polypropylene, class VI<br>Graduation Ink - Orange or Purple                                                                                                                                                                       |
| Integral Syringe Tip<br>Flexural Modulus                              | 9500 Kg/cm² (932 MPa)                                                                                                                                                                                                                            | 9500 Kg/cm² (932 MPa)                                                                                                                                                                                                                       |
| Syringe Plunger                                                       | Inert Polypropylene, class VI<br>White Colorant                                                                                                                                                                                                  | Inert Polypropylene, class VI<br>White Colorant                                                                                                                                                                                             |
| Syringe Gasket                                                        | Silicone, Class VI, Black Colorant                                                                                                                                                                                                               | Silicone, class VI, Black Colorant                                                                                                                                                                                                          |
| Syringe Tip Cap                                                       | Inert Polypropylene, class VI<br>Orange Colorant                                                                                                                                                                                                 | Inert Polypropylene, class VI<br>Orange or Purple Colorant                                                                                                                                                                                  |
| Barrel Lubricant                                                      | Polydimethylsiloxane class VI                                                                                                                                                                                                                    | Polydimethylsiloxane class VI                                                                                                                                                                                                               |
| Packaging Material                                                    | Tyvek or medical grade paper /<br>film primary pouch, Double Flute<br>paper corrugate case box                                                                                                                                                   | Tyvek or medical grade paper / film<br>primary pouch, Double Flute paper<br>corrugate case box                                                                                                                                              |
| Sterilization Method,<br>SAL                                          | Ethylene Oxide (EO), 10-6 SAL                                                                                                                                                                                                                    | Ethylene Oxide (EO), 10-6 SAL                                                                                                                                                                                                               |
| Human Factors and<br>Usability                                        | Design Validation                                                                                                                                                                                                                                | Design Validation<br>Device Master File #MAF2258                                                                                                                                                                                            |
| Risk Analysis                                                         | Device Risk Assessment including<br>Design and User FMEA                                                                                                                                                                                         | Device Risk Assessment including<br>Design and User FMEA<br>Device Master File #MAF2258                                                                                                                                                     |

{5}------------------------------------------------

#### VII. PERFORMACE DATA (BENCH)

The following performance testing was conducted on the NeoConnect™ Enteral Syringe:

- . Finished Device
O

- Risk Analysis including design, user and process FMEA (Failure Modes and o Effects Analysis) in accordance with EN ISO 14971:2012
- o Human Factors and Usability Validation
- Biocompatibility o
	- · ISO 10993-5: Cytotoxicity
	- ISO 10993-10: Irritation and sensitization .
	- . ISO 10993-11: Acute Toxicity
- Chemical Testing o
	- Extractables and Leachables .
	- Finished Device Verification Testing
		- Critical Dimension verification .
		- . Ink Adhesion
		- =
			- Capacity Tolerance ●
			- Graduated Scale ●
			- Piston Fit in Barrel .
			- Air and Liquid Leakage Testing ●
- Syringe Tip (ENFit connector) ●
	- Enteral Connector Misconnection Assessment o
	- ENFit Connector Risk Management Report (including misconnections FMEA) O
	- Human Factors Validation Study o
	- Dimensional verification O
	- Liquid Leakage Testing O
	- O Stress Cracking
	- Resistance to separation from axial load O
	- Resistance to separation form unscrewing O
	- Resistance to overriding O
	- Disconnection by unscrewing o

#### CONCLUSIONS VIII.

The NeoMed NeoConnect™ Enteral Syringes are substantially equivalent to the NeoMed Oral / Enteral Syringe (K122373).

---

**Source:** [https://fda.innolitics.com/device/K143344](https://fda.innolitics.com/device/K143344)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
