K143298 · Bk Medical Aps · IYN · Jan 16, 2015 · Radiology
Device Facts
Record ID
K143298
Device Name
Ultrasound Scanner System bk2300
Applicant
Bk Medical Aps
Product Code
IYN · Radiology
Decision Date
Jan 16, 2015
Decision
SESE
Submission Type
Abbreviated
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
Device Story
The bk2300 is a diagnostic ultrasound scanner system used in clinical settings by healthcare professionals. It processes ultrasonic signals from transducers to generate B-mode, M-mode, PWD, CWD, Color Doppler, Tissue/Contrast Harmonic Imaging, and Elastography images. It features an optional 3-D unit for volume reconstruction and a Vector Flow Imaging (VFI) module for visualizing axial and transverse velocities. The device supports biopsy needle guidance and geometric parameter calculations. Clinicians use the displayed images and measurements to inform diagnostic decisions and patient management. The system provides non-invasive imaging, aiding in the assessment of various anatomical structures and blood flow, potentially improving diagnostic accuracy and patient outcomes.
Clinical Evidence
Bench testing only. The device was evaluated for acoustic output, thermal, electrical, electromagnetic, and mechanical safety, and biocompatibility. It conforms to standards including AIUM/NEMA UD-2, AIUM/NEMA UD-3, IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 14971, EN ISO 10993-1, IEC 62304, and IEC 62359. No clinical data was required for this submission.
Technological Characteristics
Diagnostic ultrasound system utilizing pulsed echo and Doppler imaging. Materials are biocompatible. Connectivity includes standard ultrasound interfaces. Software follows IEC 62304 lifecycle processes. Acoustic output complies with Track 3 requirements (Ispta < 720 mW/cm², MI < 1.9). Thermal Index (TI) max 6.0. Sterilization is not applicable to the console; transducers are compatible with standard clinical cleaning/disinfection protocols.
Indications for Use
Indicated for fetal (obstetrics), abdominal, intra-operative (including neuro), pediatric, small organ, neonatal cephalic, adult cephalic, trans-rectal, trans-vaginal, trans-urethral, musculo-skeletal (conventional and superficial), cardiac adult, and peripheral vessel clinical applications.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{1}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol that resembles a stylized caduceus, with three human profiles facing to the right.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 16, 2015
BK Medical ApS % Mrs. Randi Hauerberg Regulatory Affairs Lead Manager Mileparken 34 Herlev DK-2730 DENMARK
Re: K143298 Trade/Device Name: Ultrasound Scanner System bk2300 Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO, ITX Dated: November 11, 2014 Received: November 26, 2014
Dear Mrs. Hauerberg:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{2}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert A Ochs
Robert Ochs, Ph.D. Acting Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
CEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
Form Annroved: OMB No. 0910-0120
sweck fir rember iif known
K143298
Device Name
Ultrasound Scanner System bk2300
#### Indications for Use (Describe)
Ultrasound scanner and transducers for B, Tissue and Contrast Harmonic Imaging, M, PWD, CWD, Color Doppler, Vector Flow Imaging and combined mode imaging and Elastography.
Signal Analysis.
Guidance of biopsy needles, geometrical measurements and calculation of parameters. And optimal 3-D unit can reconstruct a series of 2-D imaqes into a single 3-D volume and display this on the screen.
An optional Vector Flow Imaging (VFI) module: Color Flow Mapping (CFM) imaging mode with the ability to visualize both the axial and the transverse velocity.
Clinical Applications:
- Fetal (Sometimes called Obstetrics) .
- Abdominal ●
- Intra-operative ●
- Intra-operative (Neuro) (sometimes called Neuro Surgical) ●
- Pediatrics .
- Small Organ (Sometimes called Small Parts) .
- . Neonatal Cephalic (Cephalic is sometimes called trans-cranial)
- Adult Cephalic (Cephalic is sometimes called trans-cranial) ●
- Trans-rectal
- . Trans-vaqinal
- . Trans-urethral
- Musculo-skeletal (Conventional)
- Musculo-skeletal (Superficial)
- Cardiac Adult
- Peripheral Vessel (Sometimes called Peripheral Vascular) .
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescliption Use (Part 21 CFR 801 Subpart D)
Over-The Counter Use (21 CFR 801 Suppart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FOA USE ONLY
Concurrence of Certer for Devices and Radiclogical Health (CDRH) (Signature)
{4}------------------------------------------------
#### This section applies only to requirements of the Paperwork Recuction Act of 1995. "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response including the time to review instructions, search existing data sources, gather and maintain the dista needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burder, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Intornation Officer Paperwork Reduction Act (PRA) Staff PRAStaff@ida.hhs.gov
Ar agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number
{5}------------------------------------------------
#### Fill out one form for each ultrasound system and each transducer.
#### System: bk2300
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | Modes | | | | | | | | | | |
|---------------------------------|-------|---|-----|------------|-------------------------------|----------------------------------------|------------------|----------------------|------------------------------|----------------|--------------|
| Specific<br>(Tracks I & III) | B | M | PWD | CWD<br>25) | Tissue<br>Harmonic<br>Imaging | Contrast<br>Harmonic<br>Imaging<br>26) | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>1) | Other |
| Ophthalmic | | | | | | | | | | | |
| Fetal 2) | P | P | P | N | P | N | P | P | P | P | N24) |
| Abdominal 3) | P | P | P | N | P | N | P | P | P | P | N24) |
| Intra-operative (Specify 4) | P | P | P | N | P | N | P | P | P | P | N24) |
| Intra-operative (Neuro 5) | P | P | P | N | P | N | P | P | P | P | N24) |
| Laparoscopic 6) | | | | | | | | | | | |
| Pediatric 7) | P | P | P | N | P | N | P | P | P | P | N24) |
| Small Organ (Specify 8) | P | P | P | N | P | N | P | P | P | P | N24) |
| Neonatal Cephalic 9) | P | P | P | N | P | N | P | P | P | P | N24) |
| Adult Cephalic 10) | P | P | P | N | P | N | P | P | P | P | N24) |
| Trans-rectal 11) | P | P | P | N | P | N | P | P | P | P | N24) |
| Trans-vaginal 12) | P | P | P | N | P | N | P | P | P | P | N24) |
| Trans-urethral 13) | P | P | P | N | P | N | P | P | P | P | N24) |
| Trans-esoph. (non-Card.) 14) | | | | | | | | | | | |
| Musculo-skel. (Conventional 15) | P | P | P | N | P | N | P | P | P | P | N24) |
| Musculo-skel. (Superficial 16) | P | P | P | N | P | N | P | P | P | P | N24) |
| Intra-luminal 17) | | | | | | | | | | | |
| Other (Specify 18) | | | | | | | | | | | |
| Cardiac Adult 19) | P | P | P | N | P | N | P | P | P | P | |
| Cardiac Pediatric 20) | | | | | | | | | | | |
| Trans-esoph. (Cardiac 21) | | | | | | | | | | | |
| Other (Specify 22) | | | | | | | | | | | |
| Peripheral vessel 23) | P | P | P | N | P | N | P | P | P | P | N23)<br>N24) |
| Other (Specify 24) | P | P | P | | | | | | | | N |
N= new indication; P= previously cleared by FDA; E= added under Appendix E *Examples may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, Color Velocity Imaging. Tissue Harmonic Imaging.
The numbering in the table above refers to the comments provided below and, in case it is considered relevant, further comments and history in the Attachment 006 Instructions For Use - Attachment. Please do observe that the numbering from 1 to 24 is fixed to make the document more consistent meaning that comments are only provided in the table below if relevant.
Additional Comments:
| 1. | B+M,<br>B+D,<br>B+C,<br>B+D+C.<br>B includes Tissue Harmonic Imaging<br>D:PWD.<br>C: Color Flow mapping Doppler incl.<br>Amplitude(power)Doppler |
|----|--------------------------------------------------------------------------------------------------------------------------------------------------|
| | Mode combinations: |
Vector Flow Imaging (VFI) 23)
24) Elastography.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH; Office of In Vitro Diagnostics and Radiological Health (OlR)
{6}------------------------------------------------
### Fill out one form for each ultrasound system and each transducer.
#### System: bk2300
#### Transducer: 9018
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | | | | | Modes | | | | | |
|---------------------------------|---|---|-----|-----|-------------------------------|---------------------------------|------------------|----------------------|------------------------------|-------------|-------|
| Specific<br>(Tracks I & III) | B | M | PWD | CWD | Tissue<br>Harmonic<br>Imaging | Contrast<br>Harmonic<br>Imaging | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined 1) | Other |
| Ophthalmic | | | | | | | | | | | |
| Fetal 2) | | | | | | | | | | | |
| Abdominal 3) | | | | | | | | | | | |
| Intra-operative (Specify 4) | | | | | | | | | | | |
| Intra-operative (Neuro 5) | | | | | | | | | | | |
| Laparoscopic 6) | | | | | | | | | | | |
| Pediatric 7) | | | | | | | | | | | |
| Small Organ (Specify 8) | | | | | | | | | | | |
| Neonatal Cephalic 9) | | | | | | | | | | | |
| Adult Cephalic 10) | | | | | | | | | | | |
| Trans-rectal 11) | P | P | P | P | P | | P | P | | P | N24) |
| Trans-vaginal 12) | P | P | P | P | P | | P | P | | P | N24) |
| Trans-urethral 13) | | | | | | | | | | | |
| Trans-esoph. (non-Card.) 14) | | | | | | | | | | | |
| Musculo-skel. (Conventional 15) | | | | | | | | | | | |
| Musculo-skel. (Superficial 16) | | | | | | | | | | | |
| Intra-luminal 17) | | | | | | | | | | | |
| Other (Specify 18) | | | | | | | | | | | |
| Cardiac Adult 19) | | | | | | | | | | | |
| Cardiac Pediatric 20) | | | | | | | | | | | |
| Trans-esoph. (Cardiac 21) | | | | | | | | | | | |
| Other (Specify 22) | | | | | | | | | | | |
| Peripheral vessel 23) | | | | | | | | | | | |
| Other (Specify 24) | | | | | | | | | | | N |
N= new indication; P= previously cleared by FDA; E= added under Appendix E *Examples may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, Color Velocity Imaging. Tissue Harmonic Imaging.
The numbering in the table above refers to the comments provided below and, in case it is considered relevant, further comments and history in the Attachment 006_Instructions For Use - Attachment. Please do observe that the numbering from 1 to 24 is fixed to make the document more consistent meaning that comments are only provided in the table below if relevant.
Additional Comments:
| 1. | |
|----|-------------------------------------------------------------|
| | Mode combinations: |
| | B+M, |
| | B+D, |
| | B+C, |
| | B+D+C. |
| | B includes Tissue Harmonic Imaging |
| | D:PWD. |
| | C: Color Flow mapping Doppler incl. Amplitude(power)Doppler |
24) Elastography.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH; Office of In Vitro Diagnostics and Radiological Health (OlR)
{7}------------------------------------------------
### Fill out one form for each ultrasound system and each transducer.
#### System: bk2300
#### Transducer: 9051
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | Modes | | | | | | | | | |
|---------------------------------|---|-------|-----|-----|-------------------------------|---------------------------------|------------------|----------------------|------------------------------|-------------|--------------|
| Specific<br>(Tracks I & III) | B | M | PWD | CWD | Tissue<br>Harmonic<br>Imaging | Contrast<br>Harmonic<br>Imaging | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined 1) | Other |
| Ophthalmic | | | | | | | | | | | |
| Fetal 2) | | | | | | | | | | | |
| Abdominal 3) | | | | | | | | | | | |
| Intra-operative (Specify 4) | | | | | | | | | | | |
| Intra-operative (Neuro 5) | | | | | | | | | | | |
| Laparoscopic 6) | | | | | | | | | | | |
| Pediatric 7) | P | P | P | P | P | | P | P | | P | N24) |
| Small Organ (Specify 8) | P | P | P | P | P | | P | P | | P | N24) |
| Neonatal Cephalic 9) | | | | | | | | | | | |
| Adult Cephalic 10) | | | | | | | | | | | |
| Trans-rectal 11) | | | | | | | | | | | |
| Trans-vaginal 12) | | | | | | | | | | | |
| Trans-urethral 13) | | | | | | | | | | | |
| Trans-esoph. (non-Card.) 14) | | | | | | | | | | | |
| Musculo-skel. (Conventional 15) | P | P | P | P | P | | P | P | | P | N24) |
| Musculo-skel. (Superficial 16) | P | P | P | P | P | | P | P | | P | N24) |
| Intra-luminal 17) | | | | | | | | | | | |
| Other (Specify 18) | | | | | | | | | | | |
| Cardiac Adult 19) | | | | | | | | | | | |
| Cardiac Pediatric 20) | | | | | | | | | | | |
| Trans-esoph. (Cardiac 21) | | | | | | | | | | | |
| Other (Specify 22) | | | | | | | | | | | |
| Peripheral vessel 23) | P | P | P | P | P | | P | P | | P | N23)<br>N24) |
| Other (Specify 24) | | | | | | | | | | | N |
N= new indication; P= previously cleared by FDA; E= added under Appendix E *Examples may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, Color Velocity Imaging. Tissue Harmonic Imaging.
The numbering in the table above refers to the comments provided below and, in case it is considered relevant, further comments and history in the Attachment 006_Instructions For Use - Attachment. Please do observe that the numbering from 1 to 24 is fixed to make the document more consistent meaning that comments are only provided in the table below if relevant.
Additional Comments:
| 1. | Mode combinations: | B+M,<br>B+D,<br>B+C,<br>B+D+C.<br>B includes Tissue Harmonic Imaging<br>D:PWD.<br>C: Color Flow mapping Doppler incl.<br>Amplitude(power)Doppler |
|----|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
|----|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
Vector Flow Imaging (VFI) 23)
24) Elastography.
### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH; Office of In Vitro Diagnostics and Radiological Health (OIR)
{8}------------------------------------------------
### Fill out one form for each ultrasound system and each transducer.
#### System: bk2300
#### Transducer: 9062
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | | | | | Modes | | | | | |
|---------------------------------|---|---|-----|------------|-------------------------------|----------------------------------------|------------------|----------------------|------------------------------|-------------|-------|
| Specific<br>(Tracks I & III) | B | M | PWD | CWD<br>25) | Tissue<br>Harmonic<br>Imaging | Contrast<br>Harmonic<br>Imaging<br>26) | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined 1) | Other |
| Ophthalmic | | | | | | | | | | | |
| Fetal 2) | | | | | | | | | | | |
| Abdominal 3) | | | | | | | | | | | |
| Intra-operative (Specify 4) | P | P | P | | P | | P | P | | P | N24) |
| Intra-operative (Neuro 5) | P | P | P | | P | | P | P | | P | N24) |
| Laparoscopic 6) | | | | | | | | | | | |
| Pediatric 7) | P | P | P | | P | | P | P | | P | N24) |
| Small Organ (Specify 8) | | | | | | | | | | | |
| Neonatal Cephalic 9) | P | P | P | | P | | P | P | | P | N24) |
| Adult Cephalic 10) | | | | | | | | | | | |
| Trans-rectal 11) | | | | | | | | | | | |
| Trans-vaginal 12) | | | | | | | | | | | |
| Trans-urethral 13) | | | | | | | | | | | |
| Trans-esoph. (non-Card.) 14) | | | | | | | | | | | |
| Musculo-skel. (Conventional 15) | | | | | | | | | | | |
| Musculo-skel. (Superficial 16) | | | | | | | | | | | |
| Intra-luminal 17) | | | | | | | | | | | |
| Other (Specify 18) | | | | | | | | | | | |
| Cardiac Adult 19) | | | | | | | | | | | |
| Cardiac Pediatric 20) | | | | | | | | | | | |
| Trans-esoph. (Cardiac 21) | | | | | | | | | | | |
| Other (Specify 22) | | | | | | | | | | | |
| Peripheral vessel 23) | | | | | | | | | | | |
| Other (Specify 24) | | | | | | | | | | | N |
N= new indication; P= previously cleared by FDA; E= added under Appendix E *Examples may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, Color Velocity Imaging. Tissue Harmonic Imaging.
The numbering in the table above refers to the comments provided below and, in case it is considered relevant, further comments and history in the Attachment 006_Instructions For Use - Attachment. Please do observe that the numbering from 1 to 24 is fixed to make the document more consistent meaning that comments are only provided in the table below if relevant.
Additional Comments:
| 1. | Mode combinations: | B+M,<br>B+D,<br>B+C,<br>B+D+C.<br>B includes Tissue Harmonic Imaging<br>D:PWD.<br>C: Color Flow mapping Doppler incl.<br>Amplitude(power)Doppler |
|----|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
|----|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
24) Elastography.
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH; Office of In Vitro Diagnostics and Radiological Health (OlR)
{9}------------------------------------------------
### Fill out one form for each ultrasound system and each transducer.
#### System: bk2300
#### Transducer: 9077
Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:
| Clinical Application | | | | | | Modes | | | | | |
|---------------------------------|---|---|-----|------------|-------------------------------|----------------------------------------|------------------|----------------------|------------------------------|-------------|-------|
| Specific<br>(Tracks I & III) | B | M | PWD | CWD<br>25) | Tissue<br>Harmonic<br>Imaging | Contrast<br>Harmonic<br>Imaging<br>26) | Color<br>Doppler | Amplitude<br>Doppler | Color<br>Velocity<br>Imaging | Combined 1) | Other |
| Ophthalmic | | | | | | | | | | | |
| Fetal 2) | | | | | | | | | | | |
| Abdominal 3) | P | P | P | N | P | | P | P | | P | |
| Intra-operative (Specify 4) | | | | | | | | | | | |
| Intra-operative (Neuro 5) | | | | | | | | | | | |
| Laparoscopic 6) | | | | | | | | | | | |
| Pediatric 7) | | | | | | | | | | | |
| Small Organ (Specify 8) | | | | | | | | | | | |
| Neonatal Cephalic 9) | | | | | | | | | | | |
| Adult Cephalic 10) | P | P | P | N | P | | P | P | | P | |
| Trans-rectal 11) | | | | | | | | | | | |
| Trans-vaginal 12) | | | | | | | | | | | |
| Trans-urethral 13) | | | | | | | | | | | |
| Trans-esoph. (non-Card.) 14) | | | | | | | | | | | |
| Musculo-skel. (Conventional 15) | | | | | | | | | | | |
| Musculo-skel. (Superficial 16) | | | | | | | | | | | |
| Intra-luminal 17) | | | | | | | | | | | |
| Other (Specify 18) | | | | | | | | | | | |
| Cardiac Adult 19) | P | P | P | N | P | | P | P | | P | |
| Cardiac Pediatric 20) | | | | | | | | | | | |
| Trans-esoph. (Cardiac 21) | | | | | | | | | | | |
| Other (Specify 22) | | | | | | | | | | | |
| Peripheral vessel 23) | | | | | | | | | | | |
| Other (Specify 24) | | | | | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under Appendix E *Examples may include: A-mode, Amplitude Doppler, 3-D Imaging, Harmonic Imaging, Tissue Motion Doppler, Color Velocity Imaging. Tissue Harmonic Imaging.
The numbering in the table above refers to the comments provided below and, in case it is considered relevant, further comments and history in the Attachment 006_Instructions For Use - Attachment. Please do observe that the numbering from 1 to 24 is fixed to make the document more consistent meaning that comments are only provided in the table below if relevant.
Additional Comments:
| 1. | Mode combinations: | B+M,<br>B+D,<br>B+C,<br>B+D+C.<br>B includes Tissue Harmonic Imaging<br>D:PWD.<br>C: Color Flow mapping Doppler incl.<br>Amplitude(power)Doppler |
|----|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
|----|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------|
25) Continuous Wave Doppler (CWD/CW)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH; Office of In Vitro Diagnostics and Radiological Health (OIR)
{10}------------------------------------------------
# 510(k) Summary:
This summary is provided as part of this Premarket Notification in compliance with 21CRF, Section 807.92.
Submitters name: BK Medical Address: Mileparken 34, DK-2730 Herlev, Denmark Phone: +45 44528100 Fax: +45 44528199 Contact person: Randi Hauerberg, Regulatory Affairs Lead Manager Date prepared: November 6, 2014
Trade name: Ultrasound Scanner System bk 2300 Common name: Diagnostic Ultrasound System Classification names: Ultrasonic Pulsed Echo Imaging System (90 IYO, CFR 892.1560) Ultrasonic Pulsed Doppler Imaging System (90 IYN, CFR 892.1550) Diagnostic Ultrasonic Transducer (90 ITX, CFR 892.1570)
Identification of predicate, legally marketed device: B-K Medical Ultrasound Scanner Pro Focus 2202, K043524 B-K Medical Ultrasound Scanner Pro Focus 2202, K100919 BK Medical Ultrasound Scanner bk2300, K140428 Siemens Ultrasound Scanner System Acuson S1000, K130619
### Device description:
Ultrasound scanner and transducers for B, Tissue and Contrast Harmonic Imaging, M, PWD, CWD, Color Doppler, Vector Flow Imaging and combined mode imaging and Elastography.
Signal Analysis.
Guidance of biopsy needles, geometrical measurements and calculation of parameters. And optimal 3-D unit can reconstruct a series of 2-D images into a single 3-D volume and display this on the screen.
An optional Vector Flow Imaging (VFI) module: Color Flow Mapping (CFM) imaging mode with the ability to visualize both the axial and the transverse velocity.
{11}------------------------------------------------
## Transducers
All the transducers in this submission have previous been cleared.
## Only the new modes
- Σ Elastography
- Σ Vector Flow Imaging (VFI).
- Continuous Wave Doppler (CWD/CW)
has been added to the functionality of relevant transducers as indicated in the Instructions For Use Section in VOL_004
There are no changes to the physical design of the transducers There are no changes to the patient contact materials. All patient contact materials are biocompatible.
All transducers used together with bk2300 are Track 3 transducers.
### Acoustic output
The system controlling the Acoustic Output in the modified bk2300 is the same as the system in the predicate device Pro Focus 2202. The system will assure that the acoustic output always will stay below the pre-amendments upper limits i.e. Ispta [] 720 mW/cm² and MI [] 1.9 (Track 3, non ophthalmic).
The Thermal Index values are maximum 6.0, i.e. TI [] 6.0
### Clinical measurement accuracy
Clinical measurements and calculations are described and accuracies are provided in the User Information.
Thermal, mechanical and electrical safety. The ultrasound scanner system bk2300 has been tested by a recognized Certified Body.
### Acoustic Output Reporting
The Acoustic Output Reporting is made according to the standards required by "Information for Manufacturers Seeking Clearance of Diagnostic Ultrasound Systems and Transducers, FDA, CDRH, September 9, 2008"
Intended use: Diagnostic ultrasound imaging or fluid flow analysis of the human body.
{12}------------------------------------------------
| | | Summary of Technological Characteristics - Predicate Device Compared to Modified Device | | | |
|--|--|-----------------------------------------------------------------------------------------|--|--|--|
| | | | | | |
| | | | | | |
| Devices and<br>Predicates | Predicate device<br>K140428,<br>Ultrasound<br>scanner bk2300 | Predicate device<br>K043524,<br>Ultrasound<br>scanner Pro<br>Focus 2202 | Predicate device<br>K100919,<br>Ultrasound<br>scanner Pro<br>Focus 2202 | Predicate device<br>K130619,<br>Ultrasound<br>scanner system<br>Acuson S1000 | Modified device<br>(this<br>application),<br>Ultrasound<br>scanner bk2300 |
|--------------------------------------------------|---------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|
| Modes of<br>operation<br>Ref.: [1]<br>Appendix G | B | B | B | B | B |
| | THI<br>M,<br>PWD,<br>CFM 1)<br>and combinations | THI<br>M,<br>PWD,<br>CFM 1)<br>and combinations<br>CHI<br>CWD, | THI<br>M,<br>PWD,<br>CFM 1)<br>and combinations<br>VFI | THI<br>M,<br>PWD,<br>CFM 1)<br>and combinations<br>Elastography | THI<br>M,<br>PWD,<br>CFM 1)<br>and combinations<br>CHI<br>CWD,<br>VFI<br>Elastography |
| Intended Use: | Diagnostic<br>ultrasound<br>imaging or fluid<br>flow analysis of<br>the human body<br>as follows: | Diagnostic<br>ultrasound<br>imaging or fluid<br>flow analysis of<br>the human body<br>as follows: | Diagnostic<br>ultrasound<br>imaging or fluid<br>flow analysis of<br>the human body<br>as follows: | Diagnostic<br>ultrasound<br>imaging or fluid<br>flow analysis of<br>the human body<br>as follows: | Diagnostic<br>ultrasound<br>imaging or fluid<br>flow analysis of<br>the human body<br>as follows: |
| Indications For<br>Use: | Fetal (incl<br>Obstetrics)<br>Abdominal<br>Intraoperative<br>Neurosurgery | Fetal (incl<br>Obstetrics)<br>Abdominal<br>Intraoperative<br>Neurosurgery | Fetal (incl<br>Obstetrics)<br>Abdominal<br>Intraoperative<br>Neurosurgery | Fetal (incl<br>Obstetrics)<br>Abdominal<br>Intraoperative<br>Neurosurgery | Fetal (incl<br>Obstetrics)<br>Abdominal<br>Intraoperative<br>Neurosurgery |
| | Pediatrics | Pediatrics | Pediatrics | Pediatrics | Pediatrics |
| | Small Parts<br>(organs) | Small Parts<br>(organs) | Small Parts<br>(organs) | Small Parts<br>(organs) | Small Parts<br>(organs) |
| | Adult Cephalic | Adult Cephalic | Adult Cephalic | Adult Cephalic | Adult Cephalic |
| | Neonatal<br>Cephalic | | | Neonatal<br>Cephalic | Neonatal<br>Cephalic |
| | Transrectal | Transrectal | Transrectal | Transrectal | Transrectal |
| | Transvaginal | Transvaginal | Transvaginal | Transvaginal | Transvaginal |
| | Transurethral | Transurethral | Transurethral | Transurethral | Transurethral |
| | Muskulo-skeletal<br>(conventional and<br>superficial) | Muskulo-skeletal<br>(conventional and<br>superficial) | Muskulo-skeletal<br>(conventional and<br>superficial) | Muskulo-skeletal<br>(conventional and<br>superficial) | Muskulo-skeletal<br>(conventional and<br>superficial) |
| | Cardiac | Cardiac | Cardiac | Cardiac | Cardiac |
| | Peripheral Vessel | Peripheral Vessel | Peripheral Vessel | Peripheral Vessel | Peripheral Vessel |
| Features | | ECG (not<br>monitoring) | ECG (Not<br>monitoring) | | |
{13}------------------------------------------------
1) CFM= Color Flow Mapping=Color Doppler and Amplitude (Power) Doppler.
{14}------------------------------------------------
## A brief discussion of non-clinical tests submitted, referenced, or relied on in the 510(k) for a determination of substantial equivalence.
The device has been evaluated for acoustic output, thermal, electrical, electromagnetic and mechanical safety, biocompatibility and has been found to conform with applicable medical device safety standards. The system complies with the following voluntary standards:
- Σ AIUM/NEMA UD-2, Acoustic Output Measurement Standard For Diagnostic Ultrasound Equipment
- 2 AIUM/NEMA UD-3, Real-time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment
- A IEC 60601-1, Medical Electrical Equipment, Part 1: General requirements for safety
- IEC 60601-1-2, General requirements for safety, Collateral Standard, Electromagnetic Compatibility - Requirements and tests
- バ IEC 60601-2-37, Particular requirements for the safety of ultrasonic diagnostic medical and monitoring equipment
- ISO 14971, Application of Risk Management of Medical Devices
- EN ISO 10993-1. Biocompatibility
- IEC 62304, Medical Device Software Software lifecycle processes
- IEC 62359, Ultrasonics Field characterization Test methods for the determination of thermal and mechanical indices related to medical diagnostic ultrasonic fields
Thermal, electrical, electromagnetic, mechanical and biocompatibility safety is unchanged.
## Technological characteristics compared to the predicate device
The predicate devices have the same major technological characteristics as the subject device described above.
Minor differences consist with respect to the Ultrasound Scanner System: Addition of new modes.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.